Electro-oculogram Testing in Fundus Flavimaculatus

Author Affiliations

From the Department of Ophthalmology, University of Illinois Eye and Ear Infirmary, Chicago.

Arch Ophthalmol. 1979;97(10):1896-1898. doi:10.1001/archopht.1979.01020020344011
Abstract

• Electro-oculogram (EOG) measurements were obtained on seven patients with moderately extensive lesions of fundus flavimaculatus. Each patient was tested by two different methods of measuring EOG light-peak to dark-trough ratios. The results show that erroneously low ratios can be observed unless a diffusing sphere is employed in the determination of light-peak to dark-trough ratios in fundus flavimaculatus and presumably other diseases that affect primarily the posterior pole.

×