Volume Displacement of Scleral Buckles

Author Affiliations

From the Wilmer Ophthalmological Institute, The Johns Hopkins University School of Medicine, Baltimore.

Arch Ophthalmol. 1985;103(12):1822-1824. doi:10.1001/archopht.1985.01050120056019
Abstract

• Indentation of the eye wall by a scleral buckle displaces volume from the vitreous cavity. We developed a mathematical formula to calculate the volume displacement caused by a scleral buckle and verified the accuracy of this mathematical model by performing scleral buckles in 21 cadaver eyes and three eyes undergoing retinal reattachment surgery. A single 5-mm radial sponge of moderate height displaces about 0.2 mL of fluid and a circumferential 2.5-mm-wide band of moderate height displaces about 0.5 mL of fluid. Larger circumferential tires of 7to 10-mm width displace 1.1 to 1.8 mL of fluid, depending on the height and configuration of the scleral buckle. A 7- to 10-mm-wide circumferential scleral buckle with a buckle height of 4 mm may displace up to 45% of the volume of the vitreous cavity. This volume displacement should be considered when injecting expansile gases or pharmacologic agents into the vitreous cavity.

×