[Skip to Content]
[Skip to Content Landing]
Article
April 1986

Electrodes for ElectroretinographyA Comparison of Four Different Types

Author Affiliations

From the Department of Ophthalmology, Karolinska Institute, Huddinge (Sweden) University Hospital.

Arch Ophthalmol. 1986;104(4):569-570. doi:10.1001/archopht.1986.01050160125027
Abstract

• Electroretinograms of 11 healthy eyes were recorded on two occasions. Four different electrodes were used, and the B-wave amplitudes were compared. The Lovac contact lens electrode produced higher amplitudes than did the ERG-Jet lens electrode or the Burian-Allen lens electrode. The gold foil electrode gave the lowest amplitudes, but was superior in comfort.

×