Small Case Series
June 2012

Two Cases of Cosmetic Iris Implant Explantation Secondary to Uveitis, Glaucoma, and Corneal Decompensation

Author Affiliations

Author Affiliations: Department of Ophthalmology, Massachusetts Eye and Ear Infirmary, Boston (Drs Veldman, Behlau, and Pineda); Paulista School of Medicine, São Paulo, Brazil (Dr Soriano); and Naval Medical Center, Portsmouth, Virginia (Dr Starling).

Arch Ophthalmol. 2012;130(6):787-789. doi:10.1001/archophthalmol.2011.2594

Cosmetic alteration of the iris color has long intrigued individuals. This has been accomplished with cosmetic contact lenses and most recently with intraocular artificial iris implants. We report 2 cases of uveitis, glaucoma, and corneal decompensation in women who underwent bilateral simultaneous implantation of a non–US Food and Drug Administration–approved cosmetic iris implant (NewColorIris). This device is a plate-styled colored silicone prosthesis designed for placement in the anterior chamber. Both patients had traveled to Panama City, Panama, to undergo the elective procedure. After surgery, both developed variants of an ocular complex comprising uveitis, glaucoma, and corneal decompensation and requiring explantation of the cosmetic iris implants. Various in situ and extirpated imaging techniques demonstrate implant placement, size, and surface characteristics that likely contributed to the ocular complications in these 2 cases.

Report of Cases
Case 1

Three weeks after bilateral NewColorIris implantation, a 28-year-old Brazilian woman had bilateral anterior segment inflammation, intraocular pressure elevation, and discomfort (Figure 1A). Anterior segment optical coherence tomography (OCT) showed close approximation of the inserted iris prosthesis to the corneal endothelium and anatomical angle (Figure 1B). Endothelial specular microscopy performed at that time revealed marked endothelial cell loss, with cell counts of 1533/mm2 OD and 1215/mm2 OS (Figure 1C and D). The iris implants were removed 1 week later after a short course of topical steroids and brimonidine tartrate.

Case 1
Figure 1. Findings in case 1. A, Cosmetic iris implant. B, Anterior segment optical coherence tomographic image demonstrating in situ apposition of the disc to the corneal endothelium peripherally. Decreased endothelial cell counts with irregular cellular morphology in the right (C) and left (D) eyes. Ave indicates average number of cells per cell area; Max, maximum number of cells per cell area; Min, minimum number of cells per cell area; Num, number of areas analyzed; CD, cell density per square millimeter; SD, standard deviation of the cell mean area; CV, coefficient of variation; and 6A, percentage of cells that are hexagonal.

Figure 1. Findings in case 1. A, Cosmetic iris implant. B, Anterior segment optical coherence tomographic image demonstrating in situ apposition of the disc to the corneal endothelium peripherally. Decreased endothelial cell counts with irregular cellular morphology in the right (C) and left (D) eyes. Ave indicates average number of cells per cell area; Max, maximum number of cells per cell area; Min, minimum number of cells per cell area; Num, number of areas analyzed; CD, cell density per square millimeter; SD, standard deviation of the cell mean area; CV, coefficient of variation; and 6A, percentage of cells that are hexagonal.

Case 2

One year after bilateral NewColorIris implantation, a 29-year-old American woman had blurred vision and right eye pain (Figure 2A). At the initial visit, her intraocular pressure was 54 mm Hg OD and 30 mm Hg OS. The right eye demonstrated marked corneal decompensation and an associated visual acuity reduction to 20/400. After treatment with topical glaucoma medications, endothelial cell counts were found to be 2392/mm2 OD and 1357/mm2 OS. Central corneal thickness measured 547 μm OD and 553 μm OS. Anterior segment OCT demonstrated crowding of the anatomical angles bilaterally by the cosmetic iris implant's edge (Figure 2B). Following explantation, gonioscopic examination revealed broad peripheral anterior synechiae, nearly 360° OU. Subsequently, she developed severe, acute, secondary angle-closure glaucoma in both eyes, necessitating urgent trabeculectomies. In addition, her left cornea decompensated and she recently underwent a combined penetrating keratoplasty and cataract extraction. Currently, she is receiving maintenance therapy with multiple topical agents for control of the intraocular pressure in her left eye.

Case 2
Figure 2. Findings in case 2. A, Cosmetic iris implant. B, Anterior segment optical coherence tomographic image demonstrating the cosmetic iris implant's close approximation to the patient's anatomical angle as well as in direct contact with the anterior surface of the iris.

Figure 2. Findings in case 2. A, Cosmetic iris implant. B, Anterior segment optical coherence tomographic image demonstrating the cosmetic iris implant's close approximation to the patient's anatomical angle as well as in direct contact with the anterior surface of the iris.

Analysis of Explanted NewColorIris Implant

Analysis using both light and scanning electron microscopy of one of the explanted NewColorIris devices shows surface and edge characteristics that deviate significantly from the appearance of other currently US Food and Drug Administration–approved anterior chamber implants on scanning electron microscopy.1 Specifically, the posterior surface of the NewColorIris implant has an irregular surface architecture and the edge has a very rough, nearly serrated appearance on close examination (Figure 3). These morphologic features support our contention that the surface characteristics of the NewColorIris contributed to ocular inflammation, secondary glaucoma, and endothelial cell loss in both patients. The edge profile appears to have been particularly deleterious to patient 2 due to direct apposition of the device against the angle structures as demonstrated by OCT.

Analysis of Explanted NewColorIris Implant
Figure 3. Microscopy findings. A, Light microscopy photomicrograph of the explanted device (original magnification ×5). B, Scanning electron microscopy photomicrograph of the anterior surface of the explant with irregular surfaces in the area of color pigmentation (original magnification ×500). C, Scanning electron microscopy photomicrograph of the posterior surface of the explant demonstrating a rough surface (original magnification ×1000). D, Light microscopy photomicrograph demonstrating irregular edge characteristics of the explant (original magnification ×100).

Figure 3. Microscopy findings. A, Light microscopy photomicrograph of the explanted device (original magnification ×5). B, Scanning electron microscopy photomicrograph of the anterior surface of the explant with irregular surfaces in the area of color pigmentation (original magnification ×500). C, Scanning electron microscopy photomicrograph of the posterior surface of the explant demonstrating a rough surface (original magnification ×1000). D, Light microscopy photomicrograph demonstrating irregular edge characteristics of the explant (original magnification ×100).

Comment

These 2 cases add to a growing body of literature regarding profound anterior segment damage, including uveitis, glaucoma, and corneal decompensation, secondary to NewColorIris implantation.2 While not approved by the US Food and Drug Administration for implantation in the United States, the NewColorIris poses a very real risk to patients willing to travel abroad to have it implanted. As we have demonstrated by anterior segment OCT and light and scanning electron microscopy, the NewColorIris device has intraocular positioning and surface characteristics that likely contribute to its poor tolerance in the anterior chamber of some patients. Specifically, if the device is undersized relative to the anterior chamber, it may not sit properly and can then contact the corneal endothelium (as in case 1). Alternatively, it can cause direct damage to angle structures in those with less generously sized anterior segments, which can lead to glaucomatous angle changes such as synechiae. This was particularly evident in case 2, in which profound angle damage was inflicted in the setting of an implant that was shown by OCT to be directly apposing the angle structures prior to its explantation. It is our feeling that the surface characteristics of the implant, particularly its relative surface irregularity compared with other better-tolerated anterior segment implants, were responsible for significant ocular inflammation in both patients in this series. The medical literature demonstrates that inappropriately sized anterior chamber implants deleteriously affect corneal endothelial cell counts and angle structures, as shown in the analysis of phakic intraocular lens tolerance.3 Although implant design and positioning issues strongly influenced these outcomes, the biocompatibility of the material and chemical coloring of the NewColorIris device may have also contributed to these complications; however, we were unable to confirm this.

Scientific data on the safety of this device are lacking in the published literature. Therefore, given its lack of adequate clinical testing and a growing body of literature describing potentially vision-threatening complications related to its implantation, we advise patients to avoid undergoing implantation of the NewColorIris prosthesis. Additionally, eye care professionals should warn patients interested in this procedure of these potential complications. Other iris devices have proven very safe and effective, and patients requiring noncosmetic iris reconstruction may consider using other posterior chamber and secured iris reconstructive devices such as those from Morcher, Ophtec, and HumanOptics (ArtificialIris). Finally, those who have already undergone cosmetic iris implantation should be closely monitored for complications, and explantation of the iris prosthesis should be undertaken at the earliest sign of ocular intolerance.

Back to top
Article Information

Correspondence: Dr Veldman, Department of Ophthalmology, Massachusetts Eye and Ear Infirmary, 243 Charles St, Boston, MA 02114 (peter_veldman@meei.harvard.edu).

Financial Disclosure: None reported.

Disclaimer: The views expressed in this article are those of the authors and do not necessarily reflect the official policy or position of the Department of the Navy, Department of Defense, or US government.

Additional Contributions: Ann S. Tisdale, MS, assisted with scanning electron microscopy.