Figure 1
After flap reflection (black arrows), trephination with a freehand trephine is centered on the stromal bed surface(white arrows), avoiding the cut of the flap hinge. The host stromal bed (white arrowheads) is ready to receive a similar disc from a donor. Asterisk indicates the artificial anterior chamber.

After flap reflection (black arrows), trephination with a freehand trephine is centered on the stromal bed surface(white arrows), avoiding the cut of the flap hinge. The host stromal bed (white arrowheads) is ready to receive a similar disc from a donor. Asterisk indicates the artificial anterior chamber.

Figure 2
The donor button is sutured in the recipient bed with 8 interrupted sutures in the host stromal bed. The flap hinge must be free of sutures (arrows). Asterisk indicates the artificial anterior chamber.

The donor button is sutured in the recipient bed with 8 interrupted sutures in the host stromal bed. The flap hinge must be free of sutures (arrows). Asterisk indicates the artificial anterior chamber.

Figure 3
Once the donor button is in place, the flap is repositioned and sutured with 3 interrupted sutures (arrows) to avoid flap dislodgment. Asterisk indicates the artificial anterior chamber.

Once the donor button is in place, the flap is repositioned and sutured with 3 interrupted sutures (arrows) to avoid flap dislodgment. Asterisk indicates the artificial anterior chamber.

Figure 4
Example of preoperative and postoperative corneal topography. NAS indicates nasal; TMP, temporal.

Example of preoperative and postoperative corneal topography. NAS indicates nasal; TMP, temporal.

Table 1 
Data Collected From Donor and Host Corneas
Data Collected From Donor and Host Corneas
Table 2 
Keratometric Analysis of Host Corneoscleral Rims in the Artificial Anterior Chamber
Keratometric Analysis of Host Corneoscleral Rims in the Artificial Anterior Chamber
1.
von Hippel  A Über die operative Behandlung totaler stationärer Hornhauttrübungen. Albrecht von Graefes Arch Klin Exp Ophthalmol. 1877;2379- 100Article
2.
von Hippel  A Eine neue Methode der Hornhaut Transplantation. Albrecht von Graefes Arch Klin Exp Ophthalmol. 1885;34108- 130Article
3.
Barraquer  JI Queratomileusis para la corrección de la miopía. Arch Soc Am Oftalmol Optom. 1964;527- 48
4.
Barraquer  JI Lamellar keratoplasty (special techniques). Ann Ophthalmol. 1972;4437- 469
5.
Haimovici  RCulbertson  WW Optical lamellar keratoplasty using the Barraquer microkeratome. Refract Corneal Surg. 1991;742- 45
6.
Aiken-O'Neill  PMannis  MJ Summary of corneal transplant activity: Eye Bank Association of America. Cornea. 2002;211- 3Article
7.
Seitz  BMuller  EELangenbucher  AKus  MMNaumann  GOH Endothelial keratopathy in pseudoexfoliation syndrome: quantitative and qualitative morphometry using automated video image analysis [in German]. Klin Monatsbl Augenheilkd. 1995;207167- 175Article
8.
Neumann  GOHSchlotzer-Schrehardt  U Keratopathy in pseudoexfoliation syndrome as a cause of corneal endothelial decompensation: a clinicopathologic study. Ophthalmology. 2000;1071111- 1124Article
9.
Gospodarowicz  DGreenburg  GAlvarado  J Transplantation of cultured bovine corneal endothelial cells to rabbit cornea: clinical implications for human studies. Proc Natl Acad Sci U S A. 1979;76464- 468Article
10.
Mohay  JLange  TMSoltau  JBWood  TOMcLaughlin  BJ Transplantation of corneal endothelial cells using a cell carrier device. Cornea. 1994;13173- 182Article
11.
McCulley  JPMaurice  DMSchwartz  BD Corneal endothelial transplantation. Ophthalmology. 1980;87194- 201Article
12.
Jumblatt  MMMaurice  DMMcCulley  JP Transplantation of tissue-cultured corneal endothelium. Invest Ophthalmol Vis Sci. 1978;171135- 1141
13.
Aboalchamat  BEngelmann  KBohnke  MEggli  PBednarz  J Morphological and functional analysis of immortalized human corneal endothelial cells after transplantation. Exp Eye Res. 1999;69547- 553Article
14.
McDonnell  PJ Emergence of refractive surgery. Arch Ophthalmol. 2000;1181119- 1120Article
15.
Melles  GREggink  FALander  F  et al.  A surgical technique for posterior lamellar keratoplasty. Cornea. 1998;17618- 626Article
16.
Melles  GRLander  FBeekhuis  WHRemeijer  LBinder  PS Posterior lamellar keratoplasty for a case of pseudophakic bullous keratopathy. Am J Ophthalmol. 1999;127340- 341Article
17.
Melles  GRLander  Fvan Dooren  BTPels  EBeekhuis  WH Preliminary clinical results of posterior lamellar keratoplasty through a sclerocorneal pocket incision. Ophthalmology. 2000;1071850- 1856Article
18.
Terry  MA The evolution of lamellar grafting techniques over 25 years. Cornea. 2000;19611- 616Article
19.
Terry  MAOusley  PJ Endothelial replacement without surface corneal incisions or sutures: topography of the deep lamellar endothelial keratoplasty procedure. Cornea. 2001;2014- 18Article
20.
Pallikaris  IGPapatzanaki  MESiganos  DSTsilimbaris  MKPallikaris  I A corneal flap technique for laser in situ keratomileusis: human studies. Arch Ophthalmol. 1991;1091699- 1702Article
21.
Azar  DTJain  SSambursky  RStrauss  L Microkeratome-assisted posterior keratoplasty. J Cataract Refract Surg. 2001;27353- 356Article
22.
Busin  MArffa  RCSebastiani  A Endokeratoplasty as an alternative to penetrating keratoplasty for the surgical treatment of diseased endothelium: initial results. Ophthalmology. 2000;1072077- 2082Article
23.
Ehlers  NEhlers  HHjortdal  JMoller-Pedersen  T Grafting of the posterior cornea: description of a new technique with 12-month clinical results. Acta Ophthalmol (Scand). 2000;78543- 546Article
24.
Behrens  ADolorico  AMTKira  DT  et al.  Precision and accuracy of an artificial anterior chamber system in obtaining corneal lenticules for lamellar keratoplasty. J Cataract Refract Surg. 2001;271679- 1687Article
25.
Li  LEllis  KRBehrens  A  et al.  Laboratory model for microkeratome-assisted posterior lamellar keratoplasty utilizing a running graft suture and a sutureless hinged flap. Cornea. 2002;21192- 195Article
26.
Lim-Bon-Siong  RWilliams  JMSamapunphong  SChuck  RSPepose  JS Screening of myopic photorefractive keratectomy in eye bank eyes by computerized videokeratography. Arch Ophthalmol. 1998;116617623;Article
27.
Seitz  BBehrens  ALangenbucher  AKus  MMNaumann  GOH Experimental 193 nm excimer laser corneal trephination with divergent cut angles for penetrating keratoplasty. Cornea. 1998;17410- 416Article
28.
Behrens  ALangenbucher  AKus  MMRummelt  CSeitz  B Experimental evaluation of two current-generation automated microkeratomes: the Hansatome and the Supratome. Am J Ophthalmol. 2000;12959- 67Article
29.
Behrens  ASeitz  BLangenbucher  A  et al.  Evaluation of corneal flap dimensions and cut quality using a manually guided microkeratome. J Refract Surg. 1999;15118- 123
30.
Perlman  EM An analysis and interpretation of refractive errors after penetrating keratoplasty. Ophthalmology. 1981;8839- 45Article
31.
Gross  RHPoulsen  EJDavitt  S  et al.  Comparison of astigmatism after penetrating keratoplasty by experienced cornea surgeons and cornea fellows. Am J Ophthalmol. 1997;123636- 643
32.
Belmont  SCZimm  JLStorch  RL  et al.  Astigmatism after penetrating keratoplasty using the Krumeich guided trephine system. Refract Corneal Surg. 1993;9250- 254
33.
Seitz  BLangenbucher  AKus  MMKuchle  MNaumann  GO Nonmechanical corneal trephination with the excimer laser improves outcome after penetrating keratoplasty. Ophthalmology. 1999;1061156- 1164Article
Laboratory Sciences
April 2003

Endothelial Lamellar Keratoplasty Using an Artificial Anterior Chamber and a Microkeratome

Author Affiliations

From the Departments of Biomedical Engineering (Dr Chuck) and Ophthalmology(Drs Behrens, Ellis, Li, and Chuck and Ms Sweet), University of California, Irvine. The authors have no relevant financial interest in this article.

Arch Ophthalmol. 2003;121(4):503-508. doi:10.1001/archopht.121.4.503
Abstract

Objective  To compare postoperative astigmatic change and graft stability using 2 different donor button diameters in endothelial lamellar keratoplasty to treat corneal endothelial failure.

Methods  A 200-µm-thick corneal flap keratectomy was performed in human donor corneoscleral rims (n = 20; 10 donors and 10 recipients) using an artificial anterior chamber and a manual microkeratome (ALTK System; Moria USA, Doylestown, Pa). After flap reflection, stromal bed trephination was performed to obtain a disc consisting of posterior stroma, Descemet membrane, and endothelium. Host beds of 7.0 mm and 7.25-mm (n = 5) or 7.50-mm (n = 5) donor buttons were obtained using a freehand trephine. The graft was secured with 8 interrupted sutures (10-0 nylon) in the stromal bed. The flap was sutured with 3 interrupted sutures. Transplanted corneas were submitted to increasing intrachamber pressures to detect graft stability, and preoperative and postoperative videokeratographic data were recorded to assess astigmatic change.

Results  The mean (SD) postoperative astigmatic change was 1.14 (3.17) diopters(D) in the 7.25-mm donor button group and 2.27 (1.77) D in the 7.50-mm donor button group (P = .69). Mean (SD) resisted pressures of 75.4 (44.81) mmHg and 100.4 (46.86) mmHg were observed in the 7.25-mm and 7.50-mm groups, respectively (P = .54).

Conclusion  Both donor button sizes exhibited similar graft stability and astigmatic postoperative change in this experimental model.

Clinical Relevance  As endothelial lamellar keratoplasty becomes further developed as a clinical alternative to penetrating keratoplasty, this laboratory model system should be useful in evaluating different mechanical factors that contribute to graft success.

PENETRATING KERATOPLASTY is a safe and effective method for restoring corneal transparency in patients with disabling corneal opacities.1,2 Endothelial cell dysfunction accounts for most penetrating keratoplasties performed in developed countries.35 Endothelial cell failure has been frequently associated with either ocular trauma after intraocular lens implantation or endothelial corneal dystrophies.6 Other entities associated with corneal endothelial damage, such as pseudoexfoliation keratopathy, may affect these statistics in the future.7,8

Endothelial cell transplantation has been attempted.913 With the development of new biomaterials, this method might be ideal to correct endothelial cell failure.14 However, present clinical applications appear to be feasible only using a posterior stromal lamellar carrier (posterior lamellar keratoplasty). In this approach, replacement of the posterior stroma, Descemet membrane, and endothelial cell layer is accomplished through a scleral pocket.1519 The technique has shown promising results. However, it is laborious and requires a highly skilled surgeon.

Laser-assisted in situ keratomileusis (LASIK) has recently led to further developments in corneal lamellar surgery20;LASIK instruments can be used to perform posterior lamellar transplantation using a corneal flap technique with trephination of the stromal bed, including Descemet membrane and the damaged endothelium. Several recent studies have shown promising results using these techniques in human subjects.2123 Advantages include preservation of the original central corneal surface, avoidance of extensive superficial suturing, and a decrease in the amount of tissue transplanted. Thus, the potential exists for improvement in postoperative visual recovery, fewer problems related to sutures, and a lower rejection rate with endothelial lamellar keratoplasty.

The purpose of this study was to evaluate graft stability and postoperative astigmatic change with endothelial lamellar keratoplasty using 2 different donor button sizes.

METHODS
DONOR TISSUE

After approval by the Institutional Review Board of the University of California, Irvine, donor corneoscleral rims (n = 20) not suitable for transplantation were obtained from the Doheny Eye and Tissue Bank (Los Angeles, Calif). The corneas were preserved under standard eye bank conditions in Optisol-GS medium(Bausch & Lomb Surgical, Inc, San Dimas, Calif) at 4°C until the operation. The procedure was performed no longer than 15 days after death. Upon availability, corneas were sequentially assigned to 1 of 2 groups, the donor group (n = 10) or the host group (n = 10). Demographic data are shown in Table 1.

EXPERIMENTAL SETTING

A manual microkeratome (LSK One; Moria USA, Doylestown, Pa) was used to perform the hinged-flap keratectomy. New stainless steel blades were repeatedly used no more than 4 times to avoid cutting edge deterioration and cut surface irregularity.22 A 200-µm head thickness was used in all corneas. An artificial anterior chamber (ALTK System; Moria USA) was used to support the corneoscleral rims, as described in previous reports.21,24,25 The gearless tracks on the base plate of the artificial anterior chamber are designed to fit into the microkeratome head, so that its pass across the cornea maintains the same plane and direction. All corneal buttons were obtained using a freehand trephine. Hosts' stromal beds were cut with a 7.0-mm trephine, whereas donor stromal buttons were oversized by either 0.25 mm (n = 5) or 0.5 mm (n = 5). As they became available, transplanted corneas were alternately placed into 2 study groups, the 7.25-mm donor button group or the 7.5-mm donor button group, for comparative purposes.

TECHNIQUE

To reduce the number of air bubbles inside the chamber, infusion of isotonic sodium chloride was released before the corneoscleral rims were placed on the base of the anterior chamber to clear the residual air from the infusion line and underneath the cornea. The isotonic sodium chloride solution bottle was raised 1.5 m above the level of the chamber to obtain adequate intrachamber pressures (110 mmHg) for the microkeratome pass. Corneas were centered according to circular guides in the base of the chamber. Mechanical epithelial scraping was performed with a Bard-Parker No. 11 surgical blade (Becton, Dickinson and Co, Hancock, NY) to avoid surface irregularities due to loose epithelium, which may introduce errors in pachymetric and videokeratographic measurements.

Intended Flap Diameter

Applanation lenses were placed on the cornea to determine the diameter to be obtained, as described elsewhere.24 The artificial anterior chamber was set to achieve a 10-mm-diameter flap in all cases. The largest flap diameter was intended to leave as much area in the stromal bed as possible for performing trephination and donor button suturing in the host.

Keratectomy

Several drops of 0.5% proparacaine hydrochloride were applied to the corneal surface prior to the microkeratome pass to resemble clinical conditions. Keratectomy was performed by passing the microkeratome head with its oscillating blade at a relatively constant speed across the plate. The same experienced surgeon (A.B.) performed all keratectomies using the right hand for a right-to-left microkeratome pass.

Flap Thickness Measurements

After epithelium removal, the isotonic sodium chloride infusion was closed, and corneal thickness was measured using an ultrasound pachymeter(Ophthasonic A Scan/Pachymeter; Mentor O & O, Inc, Norwell, Mass) in the center of the cornea. A second measurement was made after the hinged flap was created and reflected from the stromal bed. Central flap thickness was calculated.

Stromal Bed Trephination

Intrachamber pressure was returned to 18 to 20 mmHg by lowering the height of the isotonic sodium chloride solution bottle to 25 cm above the cornea level, and the trephine was centered according to the keratectomy edge. A freehand trephine of 7.0 mm for the host cornea and either 7.25 mm or 7.5 mm for the donor buttons was used to perform a circular cut of the stromal bed. This was manually centered in the recipient bed relative to the flap. The trephine blade was rotated until perforation, and the circular cut was completed with corneal scissors (Figure 1). Donor buttons were sutured in recipient beds with 8 interrupted sutures, the knots were buried, and the flap was replaced in the bed (Figure 2). The suturing technique (suture tightness and length) was the same in both groups to ensure comparable results. Eight sutures were successfully placed in all cases. However, because of the size of the flap and button, shorter than normal suture passes were required. The flap was secured with 3 interrupted sutures (Figure 3).

VIDEOKERATOGRAPHY AND GRAFT STABILITY

For surface curvature analysis, we used a commercial videokeratoscope(Eye Sys Laboratories, Inc, Houston, Tex). The Placido disc was placed in a vertical position and the chamber centered according to the monitor control. This setting to obtain reproducible measurements was adopted from a previous study.26 The orientation was preserved in preoperative and postoperative recordings. Three measurements were performed preoperatively and postoperatively for each recipient cornea (Figure 4).

To assess graft stability, intrachamber pressure was raised progressively by changing stepwise the height of the bottle, as previously reported.27 Under visual control with a surgical microscope at×12 magnification (Ophthalmic 900S; Moeller-Wedel, Hamburg, Germany), presence of leakage was monitored, and the height at which leakage occurred was recorded. Intrachamber pressure was then calculated in millimeters of mercury, according to previous measurements performed using a Tonopen XL (Medtronic Xomed, Inc, Jacksonville, Fla).

STATISTICAL ANALYSIS

Calculations were made using Stats Direct, version 1.9.0, biostatistical software (Cam Code, Ashwell, England). Mean, SD, and minimum and maximum values were described. Comparisons between groups were performed using the nonparametric Mann-Whitney U test for unpaired samples and the Wilcoxon signed rank test for paired samples. A Spearman rank correlation test was performed to assess the dependence of resisted pressure on donor size. P ≤.05 was considered statistically significant.

RESULTS
SURGICAL TECHNIQUE

The surgical procedure was simple and similar to a combination of a corneal flap technique and penetrating keratoplasty. Full-depth trephination of the stromal bed was performed in a similar fashion for both recipient and donor corneas. Some difficulties in suturing were encountered, because less space for suture placement was left under the flap. Likewise, a full-thickness perforating needle pass was more likely to occur because the donor button is considerably thinner than a normal cornea.

The technique is somewhat more laborious than a standard penetrating keratoplasty because of the microkeratome involvement. However, the suturing time is considerably shorter, which decreases the total surgical time.

SURGICAL OUTCOME

The mean (SD) flap thickness was 191.1 (70.55) µm in donors and 240.4 (102.82) µm in hosts.

The mean (SD) preoperative astigmatism in hosts was 1.41 (1.39) diopters(D), compared with 3.32 (1.51) D postoperatively. The mean (SD) postoperative astigmatic change was 1.70 (2.49) D. Keratometric data are shown in Table 2.

Comparing the 2 donor sizes, we found no statistically significant differences in postoperative astigmatic change. The mean (SD) postoperative astigmatic change was 1.14 (3.17) D for the 7.25-mm donor button group and 2.27 (1.76) D for the 7.5-mm donor button group. Despite a lower mean astigmatic change using the smaller donor button, this difference was not significant (P = .69).

There were very large variations in the preoperative and postoperative change in average keratometry values in both groups. The mean (SD) change in average keratometry value for the 7.25-mm donor button group was 1.45 (7.36) D; for the 7.5-mm group, the mean (SD) change was 2.76 (4.00) D.

In terms of stability of the graft, we observed great variability in both groups. A higher resistance with larger donor buttons was observed. The mean (SD) calculated pressure was 100.4 (46.86) mmHg (range, 48-143 mmHg)in the 7.5-mm donor button group, opposed to 75.4 (44.81) mmHg (range, 37-147mmHg) in the 7.25-mm donor button group. However, this difference was not statistically significant (P = .63), and no correlation between size and intrachamber resisted pressure was found (rs = 0.243; P = .49).

COMMENT

New surgical techniques are being developed in parallel with advances in ophthalmic technology. Lamellar corneal surgery has become more popular recently because of the improvement in microkeratome instrumentation. Safety while using these instruments has improved considerably, and a remarkable cut quality has been achieved with the latest systems.28 In lamellar transplants, the smoothness of the cut surface may lead to a better surgical outcome with a clearer interface, which is essential to obtaining a good optical result.5 This has led to the proposal of endothelial lamellar keratoplasty as an alternative to penetrating keratoplasty in patients with endothelial cell dysfunction. Busin et al22 described the procedure in 7 patients; they secured the donor button in its recipient bed with an 8-bite running suture and the flap with an additional 8-bite running suture. In 4 cases, they used a full-thickness corneal donor button oversized by 0.5 mm, and in 3 cases, they used a partial-thickness button obtained by means of an artificial anterior chamber, a rotating blade microkeratome, and a trephine. All donor discs were obtained through a trephination from the endothelial side. In all cases, the corneal flap in the recipient was obtained with an oscillating blade microkeratome.

In our study, the same instruments for flap creation and trephination were used for both recipient and donor corneas. We believe that it is important to preserve the same conditions of the cut in both recipient and donor during corneal flap creation and trephination. Surface cuts have displayed considerable variation using different microkeratomes.28,29 We found a large variation in flap thickness even using the same microkeratome. This could be due to variations in stromal hydration before the microkeratome cut is created. Raising the pressure in the artificial anterior chamber to adequate levels to create the flap might induce significant amounts of stromal hydration. Even donor button orientation might be important to avoid cut surface mismatches in the interface. Moreover, donor button thickness may vary significantly in different sectors according to the quadrant location, which might introduce additional irregularities in overall corneal thickness. In our study, the buttons were excised with full-depth trephination and scissors. This could result in slightly ragged edges, which may compromise wound stability.

The resultant astigmatism after penetrating keratoplasty is variable. With standard trephination methods, the mean astigmatism fluctuates between 3 and 6 D.3032 With excimer laser trephination, the mean keratometric astigmatism fluctuates between 3 and 3.5 D.33 In endokeratoplasty, the mean refractive astigmatism in the small series reported was 2.9 D, 22 whereas in our series it was 3.3 D. Astigmatic change (preoperative astigmatism minus postoperative astigmatism), a more reliable method of comparison between techniques, is not usually presented in the literature. In addition, the terms "refractive, ""keratometric, " and "topographic astigmatism" are mistakenly used interchangeably, adding another source of error for comparative purposes.

In myopic LASIK, a flap is replaced on a stromal bed that has been flattened, and this flattening is transmitted to the surface by the flap. In endothelial lamellar keratoplasty, an irregular stromal bed is created by the sutured button. This irregularity is also transmitted to the surface by the flap, as we have shown in the shifts in average corneal power and astigmatism. In a human subject case report of this procedure, a large change in corneal power was observed in the early postoperative period.21 Thus, the challenge in this form of endothelial lamellar keratoplasty is to create as regular a stromal bed surface as possible before laying down the conforming anterior corneal flap. If interrupted sutures are insufficient, other possibilities include running sutures, glues, and sutureless buttons.

Our initial goal was to place the sutures only in the graft bed, with a sutureless hinged flap, as is performed in LASIK. However, the endothelial pump function is severely altered in a donor button, which may compromise flap adherence and, therefore, promote flap slippage after the procedure. Because of this, we sutured the flap with 3 interrupted sutures in the 3 free quadrants (excluding the hinge) to add more stability. These sutures may be removed in the early postoperative period once the endothelial pump function is recovered and the epithelium has healed.

We found no statistical differences in bursting pressure using different donor sizes. On average, the bursting pressure was above physiologic levels and presumably would result in a watertight procedure. However, there was a wide range in bursting pressure, and the lowest (37 mmHg) could potentially result in a leaking graft clinically. Thus, further refinements of the procedure, possibly more sutures in the flap, are required to obtain consistent results.

Limitations of this study include the effect of confounding variables other than donor size on induced astigmatism and bursting pressure. In particular, variations in the tightness of the flap sutures can have a profound effect on induced astigmatism. Because it is expected that these sutures would be removed early in the postoperative course, the true effect of the differences in astigmatism due to donor size may not be apparent. The tightness of these sutures was carefully controlled to minimize the effect of this variable. Also, the high variability in measured flap thickness may result in a mismatch of donor graft to button removed.

In conclusion, donor grafts oversized by either 0.25 mm or 0.50 mm exhibited similar graft stability and astigmatic changes. The wide variations in average keratometry resulted in significant differences in corneal power postoperatively. This could yield large refractive shifts in the clinical setting and minimize the benefit of reduced astigmatism in this procedure. Endothelial lamellar keratoplasty appears to be a feasible procedure that has several potential advantages over penetrating keratoplasty for endothelial cell layer dysfunction. Further testing is required to optimize this procedure before extensive use in the clinical setting.

Corresponding author and reprints: Roy S. Chuck, MD, PhD, Department of Ophthalmology, University of California, Irvine, 118 Med Surge I, Irvine, CA 92697 (e-mail: rschuck@uci.edu).

Back to top
Article Information

Submitted for publication June 29, 2001; final revision received June 25, 2002; accepted December 20, 2002.

This study was supported by grants from the Pan-American Association of Ophthalmology, Arlington, Tex (Dr Behrens), and the LASIK Institute, Fairfax, Va (Drs Ellis and Chuck), and grant KO8 EY00412-02 from the National Institutes of Health, Bethesda, Md (Dr Chuck).

We thank Bausch & Lomb Surgical, Inc, for providing the Optisol-GS.

References
1.
von Hippel  A Über die operative Behandlung totaler stationärer Hornhauttrübungen. Albrecht von Graefes Arch Klin Exp Ophthalmol. 1877;2379- 100Article
2.
von Hippel  A Eine neue Methode der Hornhaut Transplantation. Albrecht von Graefes Arch Klin Exp Ophthalmol. 1885;34108- 130Article
3.
Barraquer  JI Queratomileusis para la corrección de la miopía. Arch Soc Am Oftalmol Optom. 1964;527- 48
4.
Barraquer  JI Lamellar keratoplasty (special techniques). Ann Ophthalmol. 1972;4437- 469
5.
Haimovici  RCulbertson  WW Optical lamellar keratoplasty using the Barraquer microkeratome. Refract Corneal Surg. 1991;742- 45
6.
Aiken-O'Neill  PMannis  MJ Summary of corneal transplant activity: Eye Bank Association of America. Cornea. 2002;211- 3Article
7.
Seitz  BMuller  EELangenbucher  AKus  MMNaumann  GOH Endothelial keratopathy in pseudoexfoliation syndrome: quantitative and qualitative morphometry using automated video image analysis [in German]. Klin Monatsbl Augenheilkd. 1995;207167- 175Article
8.
Neumann  GOHSchlotzer-Schrehardt  U Keratopathy in pseudoexfoliation syndrome as a cause of corneal endothelial decompensation: a clinicopathologic study. Ophthalmology. 2000;1071111- 1124Article
9.
Gospodarowicz  DGreenburg  GAlvarado  J Transplantation of cultured bovine corneal endothelial cells to rabbit cornea: clinical implications for human studies. Proc Natl Acad Sci U S A. 1979;76464- 468Article
10.
Mohay  JLange  TMSoltau  JBWood  TOMcLaughlin  BJ Transplantation of corneal endothelial cells using a cell carrier device. Cornea. 1994;13173- 182Article
11.
McCulley  JPMaurice  DMSchwartz  BD Corneal endothelial transplantation. Ophthalmology. 1980;87194- 201Article
12.
Jumblatt  MMMaurice  DMMcCulley  JP Transplantation of tissue-cultured corneal endothelium. Invest Ophthalmol Vis Sci. 1978;171135- 1141
13.
Aboalchamat  BEngelmann  KBohnke  MEggli  PBednarz  J Morphological and functional analysis of immortalized human corneal endothelial cells after transplantation. Exp Eye Res. 1999;69547- 553Article
14.
McDonnell  PJ Emergence of refractive surgery. Arch Ophthalmol. 2000;1181119- 1120Article
15.
Melles  GREggink  FALander  F  et al.  A surgical technique for posterior lamellar keratoplasty. Cornea. 1998;17618- 626Article
16.
Melles  GRLander  FBeekhuis  WHRemeijer  LBinder  PS Posterior lamellar keratoplasty for a case of pseudophakic bullous keratopathy. Am J Ophthalmol. 1999;127340- 341Article
17.
Melles  GRLander  Fvan Dooren  BTPels  EBeekhuis  WH Preliminary clinical results of posterior lamellar keratoplasty through a sclerocorneal pocket incision. Ophthalmology. 2000;1071850- 1856Article
18.
Terry  MA The evolution of lamellar grafting techniques over 25 years. Cornea. 2000;19611- 616Article
19.
Terry  MAOusley  PJ Endothelial replacement without surface corneal incisions or sutures: topography of the deep lamellar endothelial keratoplasty procedure. Cornea. 2001;2014- 18Article
20.
Pallikaris  IGPapatzanaki  MESiganos  DSTsilimbaris  MKPallikaris  I A corneal flap technique for laser in situ keratomileusis: human studies. Arch Ophthalmol. 1991;1091699- 1702Article
21.
Azar  DTJain  SSambursky  RStrauss  L Microkeratome-assisted posterior keratoplasty. J Cataract Refract Surg. 2001;27353- 356Article
22.
Busin  MArffa  RCSebastiani  A Endokeratoplasty as an alternative to penetrating keratoplasty for the surgical treatment of diseased endothelium: initial results. Ophthalmology. 2000;1072077- 2082Article
23.
Ehlers  NEhlers  HHjortdal  JMoller-Pedersen  T Grafting of the posterior cornea: description of a new technique with 12-month clinical results. Acta Ophthalmol (Scand). 2000;78543- 546Article
24.
Behrens  ADolorico  AMTKira  DT  et al.  Precision and accuracy of an artificial anterior chamber system in obtaining corneal lenticules for lamellar keratoplasty. J Cataract Refract Surg. 2001;271679- 1687Article
25.
Li  LEllis  KRBehrens  A  et al.  Laboratory model for microkeratome-assisted posterior lamellar keratoplasty utilizing a running graft suture and a sutureless hinged flap. Cornea. 2002;21192- 195Article
26.
Lim-Bon-Siong  RWilliams  JMSamapunphong  SChuck  RSPepose  JS Screening of myopic photorefractive keratectomy in eye bank eyes by computerized videokeratography. Arch Ophthalmol. 1998;116617623;Article
27.
Seitz  BBehrens  ALangenbucher  AKus  MMNaumann  GOH Experimental 193 nm excimer laser corneal trephination with divergent cut angles for penetrating keratoplasty. Cornea. 1998;17410- 416Article
28.
Behrens  ALangenbucher  AKus  MMRummelt  CSeitz  B Experimental evaluation of two current-generation automated microkeratomes: the Hansatome and the Supratome. Am J Ophthalmol. 2000;12959- 67Article
29.
Behrens  ASeitz  BLangenbucher  A  et al.  Evaluation of corneal flap dimensions and cut quality using a manually guided microkeratome. J Refract Surg. 1999;15118- 123
30.
Perlman  EM An analysis and interpretation of refractive errors after penetrating keratoplasty. Ophthalmology. 1981;8839- 45Article
31.
Gross  RHPoulsen  EJDavitt  S  et al.  Comparison of astigmatism after penetrating keratoplasty by experienced cornea surgeons and cornea fellows. Am J Ophthalmol. 1997;123636- 643
32.
Belmont  SCZimm  JLStorch  RL  et al.  Astigmatism after penetrating keratoplasty using the Krumeich guided trephine system. Refract Corneal Surg. 1993;9250- 254
33.
Seitz  BLangenbucher  AKus  MMKuchle  MNaumann  GO Nonmechanical corneal trephination with the excimer laser improves outcome after penetrating keratoplasty. Ophthalmology. 1999;1061156- 1164Article
×