Figure 1.
A simple schematic showing myopic(M), emmetropic (E), and hyperopic (H) retinal objects forming their respective aerial images (M′, E′, and H′) inside the camera. The distance between the camera objective lens and the film plane is altered with the focusing knob to bring the desired image into focus on the film plane.

A simple schematic showing myopic(M), emmetropic (E), and hyperopic (H) retinal objects forming their respective aerial images (M′, E′, and H′) inside the camera. The distance between the camera objective lens and the film plane is altered with the focusing knob to bring the desired image into focus on the film plane.

Figure 2.
A photograph of the knob position measuring tool (vernier scale) as used on the TRC-50X camera (Topcon America Corp, Paramus, NJ) and the CR6-45NM Non-mydriatic Retinal Camera (Canon Inc, Tokyo, Japan).

A photograph of the knob position measuring tool (vernier scale) as used on the TRC-50X camera (Topcon America Corp, Paramus, NJ) and the CR6-45NM Non-mydriatic Retinal Camera (Canon Inc, Tokyo, Japan).

Figure 3.
Linear regression curves (best fit) for the 3 camera types with spectacle refraction in diopters (D) plotted on the y-axis and focusing knob place or position on the x-axis. The TRC-50F(Topcon America Corp, Paramus, NJ) results are plotted on graph A (r = 0.96), the TRC-50X (Topcon America Corp) results on graph B (r = 0.99), and the CR6-45NM Non-mydriatic Retinal Camera (Canon Inc, Tokyo, Japan) results on graph C (r = 0.97).

Linear regression curves (best fit) for the 3 camera types with spectacle refraction in diopters (D) plotted on the y-axis and focusing knob place or position on the x-axis. The TRC-50F(Topcon America Corp, Paramus, NJ) results are plotted on graph A (r = 0.96), the TRC-50X (Topcon America Corp) results on graph B (r = 0.99), and the CR6-45NM Non-mydriatic Retinal Camera (Canon Inc, Tokyo, Japan) results on graph C (r = 0.97).

1.
Jonas  JBBergua  ASchmitz-Valckenberg  PPapastathopoulos  KIBudde  WM Ranking of optic disc variables for detection of glaucomatous optic nerve damage. Invest Ophthalmol Vis Sci. 2000;411764- 1773
2.
Bengtsson  BKrakau  CET Correction of optic disc measurements on fundus photographs. Graefes Arch Clin Exp Ophthalmol. 1992;23024- 28Article
3.
Littman  H Determination of the true size of an object on the fundus of the living eye. Williams TD, trans. Optom Vis Sci. 1992;69717- 720Originally published as: Zur Bestimmung der wahren Grosse eines Objektes auf dem Hintergrund des leben den Auges.Klin Monatsbl Augenheilkd 1982;180286- 289Article
4.
Wong  D The fundus camera. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 8
5.
Colenbrander  A Principles of ophthalmoscopy. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 21
6.
Bengtsson  BKrakau  CET Some essential optical features of the Zeiss fundus camera. Acta Ophthalmol (Copenh). 1977;55123- 131Article
7.
Varma  RTielsch  JMQuigley  HA  et al.  Race-, age-, gender-, and refractive error–related differences in the normal optic disc. Arch Ophthalmol. 1994;1121068- 1076Article
8.
Guyton  DL Automated clinical refraction. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 43
9.
Dubé  PQuigley  M Comparison of optic disc area measurements performed with a new technique of fundus photography and Heidelberg retinal tomograph.  Presented at: Annual Meeting of the American Academy of Ophthalmology October 22-25, 2000 Dallas, Tex
Surgical Technique
May 2003

A New Fundus Camera Technique to Help Calculate Eye-Camera MagnificationA Rapid Means to Measure Disc Size

Author Affiliations

From the Department of Ophthalmology, McGill University (Drs Quigley and Dubé), and the Department of Ophthalmology, University of Montreal(Dr Quigley), Montreal, Quebec. Dr Quigley has applied for a patent on the process described in this article.

Arch Ophthalmol. 2003;121(5):707-709. doi:10.1001/archopht.121.5.707
Abstract

Objective  To find a simple means for calculating eye-camera magnification to permit estimation of true retinal object size from a retinal photograph.

Methods  The position of the focusing knob on 3 different retinal cameras (TRC-50F and TRC-50X; Topcon America Corp, Paramus, NJ; and the CR6-45NM Non-mydriatic Retinal Camera; Canon Inc, Tokyo, Japan) was measured during optic nerve photography and correlated with the refractive error, or spectacle refraction, of the subject (N = 11 for each camera).

Results  A strong correlation was found between focusing knob position and spectacle refraction for each of the 3 cameras tested (r =0.96, r = 0.99, and r =0.97, respectively).

Conclusions  The focusing knob position reflects the spectacle refraction of the eye being photographed, and spectacle refraction is known to correlate well with eye-camera magnification. Therefore, focusing knob position can be used to help calculate eye-camera magnification and, hence, true retinal object size.

Clinical Relevance  The true size of the optic nerve head is important for the diagnosis of glaucoma from a retinal photograph. This technique is a simple means to calculate optic nerve head size, which may be especially useful in mass retinal photographic screening programs.

THE DIAGNOSIS of glaucoma from a fundus photograph may be aided by knowing the true size of the optic nerve head and the true area of the nerve tissue.1 However, determining scale from photographs taken with a fundus camera is a challenge. Currently, the best method for estimating true optic disc size from a retinal photograph involves discovering the camera constant (or camera magnification), obtaining an ultrasonogram of the eye to ascertain its axial length, then using these values to calculate the image magnification, per the 1992 article by Bengtsson and Krakau.2 Their article also described 2 other formulas for calculating image magnification; the first uses the refractive error, or spectacle refraction, of the eye, and the second uses the principal point refraction of the eye in addition to the corneal curvature deviation from normal, a technique originally described by Littman.3 Although none of these methods provide an exact solution to the image magnification problem, each technique does offer a good estimate of image magnification and, hence, optic disc size. All of these procedures are somewhat tedious because supplemental testing of the subject is required to derive the eye-camera magnification factor. A simpler means to elucidating this image magnification calculation would be desirable, especially in view of the potential some newer nonmydriatic digital cameras have for population-based mass screening for glaucoma.

The optical design of the fundus camera is similar to that of the indirect ophthalmoscope.4,5 A light source within the camera illuminates the retina. The luminous retinal structures are projected through the pupil and then through the front lens or ophthalmoscopic lens of the camera where an image (the so-called first aerial image of the fundus) is formed. Because the ophthalmoscopic lens of any given camera is a constant, and the distance from the eye being photographed is fixed, the position of this first aerial image in relation to the lens can be seen as a function of the optical error or spectacle refraction of the eye (Figure 1). This first aerial image is then brought into focus on the film plane by means of the focusing mechanism (often a knob located on the side of the camera which is attached to the camera body to adjust the distance between the camera objective and the film plane6). It is reasonable that the position of this focusing knob will reflect the spectacle refraction of the eye. This study tests this hypothesis using 3 different fundus cameras.

METHODS

The study was approved by an institutional ethics committee. Subjects were recruited from among our friends, colleagues, and the clinical practice of one of us (M.G.Q.), and care was taken to recruit subjects with a range of refractive errors. Subjects were photographed with a fundus camera, and the position of the focusing knob was recorded. For the first camera, the TRC-50F (Topcon America Corp, Paramus, NJ), a 180° protractor was affixed to the focusing knob, and the position of the knob, as measured against a fixed point marked on the camera body, was recorded as the degree number marked on the protractor (N = 11). For the other 2 fundus cameras, the TRC-50X (Topcon American Corp) (N = 11) and the CR6-45NM Non-mydriatic Retinal Camera (Canon Inc, Tokyo, Japan) (N = 11), a photocopy of a 150-mm ruler was taped to the circumference of the focusing knob. A reference mark was then made on the camera body just abovethe focusing knob so that a measure of the position of the knob on the ruler scale could be made in millimeters (Figure 2). This measurement was recorded for each photographed subject.

Subjects photographed with the Topcon cameras had their pupils dilated with 1 drop of 0.8% tropicamide hydrochloride and 5% phenylephrine hydrochloride before being photographed, whereas those photographed with the CR6-45NM did not have their pupils dilated. Additionally, photographs taken with the Topcon cameras were done using a red-free filter and a 20° field (these maneuvers were felt to best minimize depth of field and, hence, minimize errors in focusing knob position). Photographs taken with the CR6-45NM were performed at the standard wide-angle 45° field with no special filter in place. The position of the camera focusing knob was then correlated with each patient's spectacle refraction using standard computer statistical software (Excel; Microsoft Corp, Redmond, Wash).

RESULTS

The position of the focusing knob on all 3 cameras correlated highly with the refractive error of the eye being photographed (r = 0.96for the TRC-50F; r = 0.99 for the TRC-50X; and r = 0.97 for the CR6-45NM) (Figure 3).

COMMENT

The results show a previously unreported, highly significant correlation between the position of the focusing knob and the spectacle refraction of the eye for all 3 fundus cameras. In other words, once the camera is appropriately positioned and focused on the retina, the position of the focusing knob reflects the absolute spherical equivalent refraction of the subject being photographed. Given that spectacle refraction can be incorporated into a formula that provides a good approximation of eye-camera magnification, the strong correlation of thefocusing knob position with the subject's spectacle refraction would permit eye-camera magnification to be calculated for any retinal photograph. From this eye-camera magnification, a good approximation of the true size of the optic nerve, or any retinal structure, can be readily determined.

Varma et al7 used an eye-camera magnification correction factor based solely on the refractive error of the patient to determine optic nerve size in their population-based glaucoma screening study. However, they were obliged to obtain refractive error measurements with supplementary testing. The technique that we are describing no longer requires that this additional measurement of refractive error be done. Hence, in any mass screening programs, such as that of Varma et al, using this technique would obviate the separate step of obtaining the spectacle refraction; instead, the position of the focusing knob could be used to determine the spectacle refraction to help in calculating the eye-camera magnification.

Our results also confirm that there is a linear relationship between focusing knob position and the change in spectacle refraction.6 This is in agreement with the well-known optometer principle that, in short, permits continuous linear variation of power in refracting instruments, including automatic refractometers and lensometers.8 In this way, the fundus camera can be regarded as a refracting instrument. Interestingly, the linear relationship has been previously noted by Bengtsson and Krakau, 6 and they proposed that it be used as a means to help photographers with flexible accommodation to obtain better retinal photographs.

As explained by Littman3 and by Bengtsson and Krakau, 2 a technique employing the spectacle refraction to derive eye-camera magnification is good but inexact. Subjects whose refractive errors do not correlate with their axial length (eg, subjects with lenticular ametropia or refractive ametropia occurring after corneal refractive surgery) will have a less accurate sizing of their optic nerve or retinal structure of interest. However, the current practice of not attempting to use eye-camera magnification correction factors at all does not permit us to increase our diagnostic accuracy for glaucoma by sizing the optic nerve.1

In addition, other retinal objects, including retinal blood vessels, choroidal neovascular membranes, and tumors, can be sized with this technique. The practicality of this technique is evident when considering some of the newer digital cameras, such as the CR6-45NM, because retinal photographs obtained with these types of cameras could have the image magnification or scale adjusted automatically and simultaneously to the retinal photograph; therefore, the true size of the optic nerve and other retinal structures could be measured directly from the digital image.

In summary, the advantage of this technique can readily be seen when contemplating any population-based mass screening programs for glaucoma because all the information required to calculate optic nerve size is collected at the same sitting. Simply recording the focusing knob position (which reflects the refractive error of the patient), knowing the camera constant (representing its magnification), and incorporating this information into a formula (such as one previously derived by Bengtsson and Krakau2 for calculation of eye-camera magnification), will permit a good estimate of optic nerve size. Preliminary results comparing optic nerve head size calculations performed with this technique to optic nerve size calculations obtained with the Heidelberg scanning laser ophthalmoscope show a good correlation between the 2 techniques.9 A larger study comparing the 2 techniques is in progress.

Back to top
Article Information

Submitted for publication May 15, 2002; final revision received December 2, 2002; accepted December 26, 2002.

This study was presented as a poster at the Association for Research in Vision and Ophthalmology annual meeting, April 30 to May 5, 2000, Ft Lauderdale, Fla.

We thank Olga Overbury, PhD, for her assistance in the statistical analysis of the data.

Corresponding author and reprints: Michael G. Quigley, MD, 396 Roslyn Ave, Montreal, Quebec, Canada H3Z-2L6 (e-mail: quigley.wilson@sympatico.ca).

References
1.
Jonas  JBBergua  ASchmitz-Valckenberg  PPapastathopoulos  KIBudde  WM Ranking of optic disc variables for detection of glaucomatous optic nerve damage. Invest Ophthalmol Vis Sci. 2000;411764- 1773
2.
Bengtsson  BKrakau  CET Correction of optic disc measurements on fundus photographs. Graefes Arch Clin Exp Ophthalmol. 1992;23024- 28Article
3.
Littman  H Determination of the true size of an object on the fundus of the living eye. Williams TD, trans. Optom Vis Sci. 1992;69717- 720Originally published as: Zur Bestimmung der wahren Grosse eines Objektes auf dem Hintergrund des leben den Auges.Klin Monatsbl Augenheilkd 1982;180286- 289Article
4.
Wong  D The fundus camera. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 8
5.
Colenbrander  A Principles of ophthalmoscopy. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 21
6.
Bengtsson  BKrakau  CET Some essential optical features of the Zeiss fundus camera. Acta Ophthalmol (Copenh). 1977;55123- 131Article
7.
Varma  RTielsch  JMQuigley  HA  et al.  Race-, age-, gender-, and refractive error–related differences in the normal optic disc. Arch Ophthalmol. 1994;1121068- 1076Article
8.
Guyton  DL Automated clinical refraction. In:Duane  Ted.Clinical Ophthalmology 1 Philadelphia, Pa JB Lippincott Co1991;1- 43
9.
Dubé  PQuigley  M Comparison of optic disc area measurements performed with a new technique of fundus photography and Heidelberg retinal tomograph.  Presented at: Annual Meeting of the American Academy of Ophthalmology October 22-25, 2000 Dallas, Tex
×