Figure 1.
Test pattern of the Humphrey Field Analyzer 24-2 (Carl Zeiss Meditec) paradigm for a right eye (A) and a GDx Nerve Fiber Analyzer with variable corneal compensator (Laser Diagnostic Technologies, Inc, San Diego, Calif) retardation image of a right eye (B), with a measurement circle superimposed (the fixed corneal compensator has the exact same arrangement). In the present study, visual field test points and peripapillary GDx measurements were grouped in corresponding sectors, as suggested by Garway-Heath et al. Corresponding sectors were grayscaled and named after the position of the sector in the GDx variable corneal compensator image in relation to the optic disc.

Test pattern of the Humphrey Field Analyzer 24-2 (Carl Zeiss Meditec) paradigm for a right eye (A) and a GDx Nerve Fiber Analyzer with variable corneal compensator (Laser Diagnostic Technologies, Inc, San Diego, Calif) retardation image of a right eye (B), with a measurement circle superimposed (the fixed corneal compensator has the exact same arrangement). In the present study, visual field test points and peripapillary GDx measurements were grouped in corresponding sectors, as suggested by Garway-Heath et al.18 Corresponding sectors were grayscaled and named after the position of the sector in the GDx variable corneal compensator image in relation to the optic disc.

Figure 2.
Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “number” (affected eyes). r = −0.524.

Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “number” (affected eyes). r = −0.524.

Figure 3.
Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “maximum modulation” (affected eyes). r = 0.526.

Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “maximum modulation” (affected eyes). r = 0.526.

Figure 4.
Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “ellipse modulation” (affected eyes). r = 0.5026.

Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “ellipse modulation” (affected eyes). r = 0.5026.

Figure 5.
Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “maximum modulation” (difference between unaffected and affected eyes). r = 0.625.

Correlation between visual field mean deviation (MD) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) global parameter “maximum modulation” (difference between unaffected and affected eyes). r = 0.625.

Figure 6.
Correlation between superior visual field mean deviation (Garway-Heath et al sectors 2 and 3) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) sectoral parameter “inferior average” (difference between unaffected and affected eyes). r = 0.683.

Correlation between superior visual field mean deviation (Garway-Heath et al18 sectors 2 and 3) and GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) sectoral parameter “inferior average” (difference between unaffected and affected eyes). r = 0.683.

Figure 7.
GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) data for a patient with nonarteritic anterior ischemic optic neuropathy in the right eye and a contralateral unaffected left eye. Note the flattening of the superior retinal nerve fiber layer hump.

GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, Inc, San Diego, Calif) data for a patient with nonarteritic anterior ischemic optic neuropathy in the right eye and a contralateral unaffected left eye. Note the flattening of the superior retinal nerve fiber layer hump.

Table 1. 
Global VF Indexes and RNFL Parameters by SLP Between Affected NAION Eye and Contralateral Normal Eye
Global VF Indexes and RNFL Parameters by SLP Between Affected NAION Eye and Contralateral Normal Eye
Table 2. 
Correlation Between Global VF Index (Mean Deviation) and Global SLP Parameters
Correlation Between Global VF Index (Mean Deviation) and Global SLP Parameters
Table 3. 
Sectoral Humphrey VF Indexes and Sectoral SLP RNFL Parameters Between NAION Involved Eyes and Contralateral Uninvolved Eyes
Sectoral Humphrey VF Indexes and Sectoral SLP RNFL Parameters Between NAION Involved Eyes and Contralateral Uninvolved Eyes
Table 4. 
Correlation Between Sectoral VF Indexes and Sectoral SLP RNFL Parameters by SLP (Affected Eyes)
Correlation Between Sectoral VF Indexes and Sectoral SLP RNFL Parameters by SLP (Affected Eyes)
Table 5. 
Qualitative VF Parameters for Involved Eyes Along With GDx RNFL Measurements*
Qualitative VF Parameters for Involved Eyes Along With GDx RNFL Measurements*
1.
Arnold  AC Pathogenesis of nonarteritic anterior ischemic optic neuropathy. J Neuroophthalmol 2003;23157- 163PubMedArticle
2.
Hayreh  SS Anterior ischemic optic neuropathy: differentiation of arteritic from non-arteritic type and its management. Eye 1990;425- 41PubMedArticle
3.
Henkind  PCharles  NCPearson  J Histopathology of ischemic optic neuropathy. Am J Ophthalmol 1970;6978- 90PubMed
4.
Quigley  HAAddicks  EM Quantitative studies of retinal nerve fibre layer defects. Arch Ophthalmol 1982;100807- 814PubMedArticle
5.
Sommer  AMiller  NRPollack  IMaumenee  AEGeorge  T The nerve fiber layer in the diagnosis of glaucoma. Arch Ophthalmol 1977;952149- 2156PubMedArticle
6.
Sommer  AKatz  JQuigley  HA  et al.  Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol 1991;10977- 83PubMedArticle
7.
Weinreb  RNShakiba  SSample  PA  et al.  Associations between quantitative nerve fiber layer measurement and visual field loss in glaucoma. Am J Ophthalmol 1995;120732- 738PubMed
8.
Reus  NJLemij  HG The relationship between standard automated perimetry and GDx VCC measurements. Invest Ophthalmol Vis Sci 2004;45840- 845PubMedArticle
9.
Weinreb  RNDreher  AWColeman  A  et al.  Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness. Arch Ophthalmol 1990;108557- 560PubMedArticle
10.
Reus  NJLemij  HG Diagnostic accuracy of the GDx VCC for glaucoma. Ophthalmology 2004;1111860- 1865PubMedArticle
11.
Mohammadi  KBowd  CWeinreb  RNMedeiros  FASample  PAZangwill  LM Retinal nerve fiber layer thickness measurements with scanning laser polarimetry predict glaucomatous visual field loss. Am J Ophthalmol 2004;138592- 601PubMedArticle
12.
Foroozan  RBuono  LMSavino  PJSergott  RC Scanning laser polarimetry of the retinal nerve fiber layer in central retinal artery occlusion. Ophthalmology 2003;110715- 718PubMedArticle
13.
Meier  FMBernasconi  PSturmer  JCaubergh  MJLandau  K Axonal loss from acute optic neuropathy documented by scanning laser polarimetry. Br J Ophthalmol 2002;3285- 287Article
14.
Steel  DHWWaldock  A Measurement of the retinal nerve fibre layer with scanning laser polarimetry in patients with previous demyelinating optic neuritis. J Neurol Neurosurg Psychiatry 1998;64505- 509PubMedArticle
15.
Miyahara  TKurimoto  YKurokawa  TKuroda  TYoshimura  N Alterations in retinal nerve fiber layer thickness following indirect traumatic optic neuropathy detected by nerve fiber analyzer, GDx-N. Am J Ophthalmol 2003;136361- 364PubMedArticle
16.
Colen  TPvan Everdingen  JAMLemij  HG Axonal loss in a patient with anterior ischemic optic neuropathy as measured with scanning laser polarimetry. Am J Ophthalmol 2000;130847- 848PubMedArticle
17.
Weinreb  RNShakiba  SZangwill  L Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol 1995;119627- 636PubMed
18.
Garway-Heath  DFPoinoosawmy  DFitzke  FWHitchings  RA Mapping the visual field to the optic disc in normal tension glaucoma eyes. Ophthalmology 2000;1071809- 1815PubMedArticle
19.
Weinreb  RNZangwill  LBerry  CCBathija  RSample  PA Detection of glaucoma with scanning laser polarimetry. Arch Ophthalmol 1998;1161583- 1589PubMedArticle
20.
Bowd  CZangwill  LBerry  CC  et al.  Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci 2001;421993- 2003PubMed
21.
Tjon-Fo-Sang  MJLemij  HG The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol 1997;12362- 69PubMed
22.
Trible  JRSchultz  RORobinson  JCRothe  TL Accuracy of scanning laser polarimetry in the diagnosis of glaucoma. Arch Ophthalmol 1999;1171298- 1304PubMedArticle
23.
Greenfield  DSKnighton  RWFeuer  WJSchiffman  JCZangwill  LWeinreb  RN Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol 2002;13427- 33PubMedArticle
24.
Weinreb  RNBowd  CGreenfield  DSZangwill  LM Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol 2002;120901- 906PubMedArticle
25.
Lan  Y-WHenson  DBKwartz  AJ The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma. Br J Ophthalmol 2003;871135- 1141PubMedArticle
26.
Kwon  YHHong  SHonkanen  HAAlward  WL Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma 2000;9281- 288PubMedArticle
27.
Choplin  NTZhou  QKnighton  RW Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments. Ophthalmology 2003;110719- 725PubMedArticle
28.
Horn  FKJonas  JBMartus  P  et al.  Polarimetric measurement of retinal nerve fiber layer thickness in glaucoma diagnosis. J Glaucoma 1999;8353- 362PubMedArticle
29.
Gardiner  SKJohnson  CACioffi  GA Evaluation of the structure-function relationship in glaucoma. Invest Ophthalmol Vis Sci 2005;463712- 3717PubMedArticle
30.
Greenfield  DSKnighton  RWHuang  XR Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am J Ophthalmol 2000;129715- 722PubMedArticle
31.
Knighton  RWHuang  XRGreenfield  DS Analytical model of scanning laser polarimetry for retinal nerve fiber layer assessment. Invest Ophthalmol Vis Sci 2002;43383- 392PubMed
32.
Reus  NJColen  TPLemij  HG Visualization of localized retinal nerve fiber layer defects with the GDx with individualized and with fixed compensation of anterior segment birefringence. Ophthalmology 2003;1101512- 1516PubMedArticle
33.
Tannenbaum  DPHoffman  DLemij  HGGarway-Heath  DFGreenfield  DSCaprioli  J Variable corneal compensation improves discrimination between normal and glaucomatous eyes with the scanning laser polarimeter. Ophthalmology 2004;111259- 264PubMedArticle
34.
Niessen  AGvan den Berg  TJLangerhorst  CTGreve  EL Retinal nerve fiber layer assessment by scanning laser polarimetry and standardized photography. Am J Ophthalmol 1996;121484- 493PubMed
35.
Schlottmann  PGCilla  SDGreenfield  DSCaprioli  JGarway-Heath  DF Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by scanning laser polarimetry. Invest Ophthalmol Vis Sci 2004;451823- 1829PubMedArticle
Clinical Sciences
December 2006

Correlation of Retinal Nerve Fiber Layer Measured by Scanning Laser Polarimeter to Visual Field in Ischemic Optic Neuropathy

Author Affiliations

Author Affiliations: Departments of Ophthalmology (Drs Danesh-Meyer, Carroll, Ku, Hsiang, and Gaskin) and Medicine (Mr Gamble), Faculty of Medical and Health Sciences, University of Auckland, Auckland, New Zealand; and Wills Eye Hospital, Neuro-Ophthalmology Service, Thomas Jefferson Medical College, Philadelphia, Pa (Dr Savino).

Arch Ophthalmol. 2006;124(12):1720-1726. doi:10.1001/archopht.124.12.1720
Abstract

Objective  To evaluate the relationship between retinal nerve fiber layer (RNFL) measurement with scanning laser polarimetry (SLP) and standard automated perimetry (SAP) in nonarteritic anterior ischemic optic neuropathy (NAION).

Methods  In this prospective observational case series, all subjects (28 eyes with NAION) underwent SAP and SLP. The RNFL retardation measurements and visual field test points were grouped into 6 corresponding sectors. The contralateral uninvolved eye was used as control. The relationship between RNFL retardation and SAP was evaluated with the Spearman nonparametric technique and linear regression analysis. The main outcome measure was correlation of SLP RNFL parameters and SAP.

Results  Global and sectoral SLP parameters showed a significant difference in affected eyes compared with controls. The strongest correlations were seen between mean deviation and number (r = −0.524; P = .004), ellipse modulation (r = 0.5026; P = .006), and maximum modulation (r = 0.526; P = .004). Superior sectoral visual field indexes showed a strong correlation with inferior RNFL changes (r = 0.522; P<.008). Linear regression confirmed a strong relationship between the superior sectoral visual field indexes and the inferior RNFL.

Conclusion  Scanning laser polarimetry was able to identify structural changes of the RNFL globally and in the inferior SLP sector with functional loss in NAION.

Nonarteritic anterior ischemic optic neuropathy (NAION) is an acute generalized or sectoral swelling of the optic disc resulting in visual field (VF) loss with or without reduction in visual acuity. Nonarteritic anterior ischemic optic neuropathy is thought to be due to vascular insufficiency within the optic nerve head microcirculation,1 but the precise mechanism and specific location of vascular compromise remain unproven.1,2 Histological studies have shown that NAION induces a thinning of the retinal nerve fiber layer (RNFL) in the affected sectors.3 Once the acute swelling resolves clinically, visible optic disc pallor and RNFL defects are evident.4 However, using conventional methods of ophthalmoscopy or red-free photography, it has been only possible to qualitatively evaluate the morphological changes seen following an episode of NAION.46 Recently, several studies have investigated the relationship between RNFL damage and visual function in glaucomatous optic neuropathy with various optical imaging techniques.7,8 One such device is the GDx Nerve Fiber Analyzer (Laser Diagnostic Technologies, San Diego, Calif), a scanning laser polarimeter (SLP) that estimates the thickness of the RNFL by measuring the summed retardation of a polarized scanning laser beam, presumably induced by the form-birefringent microtubules that support the retinal ganglion cell axons and then compensated, partially or completely, for the birefringence of the anterior segment.9 Although it has principally been used in glaucoma,10,11 its potential use in central retinal artery occlusion,12 acute retrobulbar optic neuritis,13 demyelinating optic neuritis,14 and traumatic optic neuropathy15 has been reported. A single case report16 has demonstrated loss of retinal nerve fiber bundles after an episode of NAION with corresponding VF defect. The purposes of this study were to investigate the ability of SLP to identify structural changes in the RNFL of eyes following an event of NAION and to examine the relationship between optic nerve structure and function using measurements of the peripapillary RNFL retardation measured with SLP and standard automated perimetry.

METHODS
SUBJECTS

An observational case series was undertaken involving 28 patients with isolated unilateral NAION occurring at least 3 months prior to initial examination with acute optic disc swelling. They were recruited from the Neuro-Ophthalmology Service at Wills Eye Hospital between September 2000 and October 2003. Patients were considered eligible for recruitment if they had a diagnosis of unilateral NAION, with the initial disc swelling and the diagnosis documented and confirmed at Wills Eye Hospital by a neuro-ophthalmologist. The uninvolved eye was required to have normal ophthalmic assessment findings with a visual acuity of 20/40 or better. Patients with glaucoma, a family history of glaucoma, intraocular pressure of greater than 21 mm Hg in either eye, giant cell arteritis, other optic neuropathies, posterior segment diseases, corneal diseases, refractive surgery, or neurological diseases or systemic diseases with possible ocular involvement, such as diabetes mellitus, were excluded from the study. Patients were excluded if they had ametropia greater than 5 diopters (D) or astigmatism greater than 2.5 D. All patients had experience with VF testing and had undergone at least 2 prior tests with reliable results. Because there is a wide intersubject variability in the RNFL thickness, the unaffected eye was used as the control.

All patients underwent a complete neuro-ophthalmic examination performed by a neuro-ophthalmologist at the time of initial examination and on follow-up visits. This was followed by Swedish Interactive Threshold Algorithm (SITA) 24-2 standard automated perimetry (SAP) and SLP (GDx Nerve Fiber Analyzer with fixed corneal compensation [software version 2.0.09]; Laser Diagnostic Technologies), which were performed within 1 week of each other. The research adhered to the tenets of the Declaration of Helsinki. The institutional review committee had approved the research and informed consent had been obtained.

SCANNING LASER POLARIMETRY

The GDx is a confocal scanning laser ophthalmoscope with an integrated polarization modulator, a fixed corneal compensator, and a polarization detection unit. Details of its operation are described elsewhere.17 In all subjects, both eyes were scanned starting with the right eye. The pupils of the patients were undilated, and the room lights were left on. Three high-quality retardation images were acquired through the undilated pupil and the results were averaged for the analysis. The default sectors were defined as temporal (−25° to 25°), superior (25° to 145°), nasal (145° to 215°), and inferior (215° to 335°), giving two 120o sectors, a 70o sector, and a 50o sector. The parameters automatically generated by the SLP software were evaluated. These were divided into parameters that represented a global assessment of the RNFL and parameters that represented the sectoral RNFL status. Global parameters were the number, symmetry, ellipse modulation, maximum modulation, and ellipse average. The number is a global measure derived from more than 200 other parameters that are analyzed by a neural network trained to discriminate between healthy and glaucomatous eyes. The maximum modulation and ellipse modulation are indicators of the difference between the thickest and thinnest parts of the RNFL. Sectoral parameters were superior maximum, inferior maximum, superior average, inferior average, temporal median (superior maximum–superior ratio), nasal median (superior maximum–superior nasal ratio), superior integral, superior (superior temporal) ratio, inferior (inferior temporal) ratio, and superior nasal ratio.

VF TESTING

Standard automated perimetry was performed with the SITA standard 24-2 program. All VF test results were reliable with less than 33% fixation losses, false-positive responses, and false-negative responses. Each VF was divided into 6 zones based on the optic disc–VF map described by Garway-Heath et al18 (Figure 1). Garway-Heath et al produced the first high-resolution map from structure to function. Using this map allowed a detailed comparison of structure (GDx image) and function (SAP) measures. This mapping allowed correlation of sectors of disc RNFL with sectors of the VF. Because sectoral SLP parameters are divided into 4 fixed sectors, the 2 superior (sectors 2 and 3) and 2 inferior (sectors 4 and 5) VF zones were combined to permit a direct 4-sector comparison between VFs and SLP. Thus, sector 1 (temporal VF, nasal optic disc 121° to 230°) corresponds to the nasal SLP measurements. Sectors 2 and 3 (superior and superior nasal VF, inferior optic disc sectors 231° to 270° and 271° to 310°) correspond to the inferior nasal and inferior temporal SLP measurements, respectively. Sectors 4 and 5 (inferior paracentral and inferior VF, superior disc sectors 41° to 80° and 81° to 120°) correspond to superior temporal and superior nasal SLP measurements, respectively. Sector 6 (macular VF, temporal disc 311° to 340°) corresponds to temporal parapapillary SLP measurements.

The mean deviation (MD) was calculated for each of the sectors from the raw retinal sensitivity on the VF printout. We then correlated SLP measurements with the global and sectoral MDs of the appropriate VF zones using linear and logarithmic regression based on the assumption that the RNFL is a predictor of VF sensitivity.

STATISTICAL ANALYSIS

The unaffected contralateral eyes were used as the control group. A Spearman ranked correlation coefficient (r) was calculated for each sector to measure the degree of association between SAP and SLP. A t test for related groups was used to compare affected with unaffected eyes. “Number” and “symmetry” are not normally distributed parameters; therefore, the sign test was used for these parameters. All tests were 2-tailed and a P value of <.05 was considered statistically significant because all comparisons were preplanned. However, a more conservative critical significance level could be adopted based on correction for multiple comparisons (Bonferroni correction P<.002 or otherwise as stated with relevant table) at the risk of an increased type II error rate. All analyses were performed by a medical statistician using SAS (version 8.12; SAS Institute Inc, Cary, NC).

RESULTS

A total of 28 patients (13 men, 15 women) were evaluated. The mean ± SD age was 61 ± 11 years (range, 39-79 years). The mean time between initial examination and study enrollment was 4.1 months, ranging from 97 days to 159 days. At the final visit, the mean visual acuity was 0.06 logarithm of the minimum angle of resolution (logMAR) (approximately 20/30 Snellen equivalent) in the control eyes and 0.44 logMAR (20/60 Snellen equivalent) in the affected eyes (P = .01).

GLOBAL PARAMETERS

The MD of the affected group ranged from −26 to 0 dB with a mean ± SD of −11.74 ± 6.65 dB, while the unaffected MD ranged from −4.36 to 3.74 dB with a mean ± SD of −0.63 ± 2.03 dB. The difference was statistically significant (P<.001). All global SLP parameters with the exception of symmetry were significantly different between the affected and unaffected eyes (Table 1).

The relationship between SAP and SLP measurements is graphically presented in Figures 2, 3, 4, and 5 for global parameters in eyes with NAION. The global SLP parameters that significantly correlated with MD in affected eyes were number (r = −0.524; P = .001), maximum modulation (r = 0.526; P = .004), and ellipse modulation (r = 0.5026; P = .006) (Table 2) (Bonferroni correction P<.008). The correlation between RNFL loss and MD only modestly improved when the difference between affected and unaffected eyes was considered for both RNFL and VF loss; that is, the affected RNFL was subtracted from the unaffected RNFL and this value correlated with the difference between affected and unaffected VF sensitivities.

SECTORAL PARAMETERS

The VF indexes in all sectors of affected eyes were significantly depressed compared with the contralateral unaffected eyes (P<.001) (Table 3). The majority of the sectoral SLP parameters of affected eyes were significantly different from unaffected eyes (all P<.001), with the exception of those that represented the temporal and nasal regions. In healthy, unaffected eyes, no statistically significant correlations between SAP and SLP measurements were found in any sector (P = .39).

When the relation between structure and function was evaluated using the sectoral SLP parameters with their corresponding VF sectors, only those representing the inferior RNFL region showed significant association with the corresponding superior VFs (Table 4). This included inferior deviation from normal (r = 0.522; P = .008), inferior average (r = 0.517; P = .005), and inferior maximum (r = 0.564; P = .002). Interestingly, inferior maximum and inferior average also correlated with nasal VF sector (r = 0.624; P< .001 and r = 0.611; P = .001, respectively). The relation of structure and function was not identifiable between any of the other sectors, including superior RNFL and the corresponding inferior VF. Table 5 tabulates the qualitative interpretation of the VF defect along with the quantitative VF raw sensitivities and the GDx RNFL parameters. Figure 6 shows the fit with a least-squares linear regression mode for the relationship between RNFL and decibel–differential light sensitivity with r = 0.683 (P<.001). Figure 7 illustrates results for a patient with NAION studied with SLP and the corresponding VF defect.

COMMENT

In glaucoma, there is a correlation between structural measures of optic neuropathy, which is measured with SLP, and functional loss, which is measured with SAP.8 Scanning laser polarimetry (both with fixed corneal compensation and variable corneal compensation) has also been shown to differentiate patients with glaucoma from normal individuals with sensitivities and specificities of 72% to 78% and 56% to 92%, respectively,11,1922 with the variable corneal compensator showing a stronger relationship.23,24 However, to date, there is a paucity of studies that quantitate RNFL loss in nonglaucomatous optic neuropathies. The appearance of the optic nerve in NAION has not been evaluated quantitatively beyond a case report16; rather it is usually subjectively described by estimating the degree and extent of pallor or qualitatively commenting on defects in the nerve fiber layer. Moreover, no available studies have been performed that quantitatively correlate RNFL loss in nonglaucomatous optic neuropathies with a corresponding decrease in VF sensitivities. To our knowledge, this study is the first to demonstrate that when examined with SLP, eyes affected with NAION show defects in the RNFL that can be identified both with the global and the sectoral SLP parameters. Furthermore, to our knowledge, this study is also the first to quantify a relationship between changes of the RNFL and corresponding changes in VF sensitivity in NAION.

All global SLP parameters (with the exception of symmetry) were significantly different between affected and unaffected eyes. Our findings in NAION are similar to studies using SLP in glaucoma, with “number” and “maximum modulation” demonstrating the strongest correlation.25,26 The modulation parameters are theoretically more reliable, correcting for individual variation in the RNFL by using the patient's own temporal and nasal RNFL as a baseline.27 In our study, the absolute indexes of RNFL thickness, average, and ellipse average were not correlated with VF MD, in keeping with the findings of Lan et al.25 The impact of a fixed corneal compensator improperly compensating for intersubject variability in corneal birefringence is the most likely explanation. Additionally, normal VF findings can be associated with a wide range of RNFL measurements. Thus, for a global parameter such as “average” to be significantly reduced in some subjects, a certain amount of RNFL damage must occur; hence, small focal RNFL or optic nerve head damage might be overlooked.

Scanning laser polarimetry demonstrated attenuation of the RNFL in the superior and inferior sectors in the affected eyes compared with the contralateral unaffected eyes, with a greater loss seen in the superior RNFL. The temporal and nasal SLP and VF sectors were not significantly different between the affected and unaffected eyes. The inferior SLP indexes showed a significant correlation with the MD, reflecting that a significant sectoral RNFL defect can result in both sectoral and global loss of VF sensitivity. Similar correlations between sectoral parameters and MD have been reported in glaucomatous eyes.28 The inferior sectoral SLP parameters were also strongly correlated with the superior VF indexes, reflecting that VF defects in NAION tend to be altitudinal defects.

Surprisingly, no structure-function relationship was found between the superior RNFL segment and the corresponding inferior VF sectors. A similar finding was found in glaucomatous eyes using SLP, where inferior SLP indexes correlated strongly with superior VF indexes and superior SLP indexes did not correlate with inferior VF indexes.25 Recently, others have also identified weaker correlations between the inferior VF indexes and corresponding superior SLP indexes.29

There are several possibilities that may account for this lack of correlation. The strength of the correlation in all the sectors may be affected by the fixed corneal compensator, which has been shown to affect the discriminating power of SLP in mild to moderate glaucoma.24,3032 The fixed corneal compensator introduces an error in measurements by erroneously compensating for corneal birefringence where the axis and magnitude of birefringence differ from population mode values. The introduction of the variable corneal compensator to achieve individualized corneal compensation has improved the relationship between the RNFL measured with SLP and the VF test.23,33 Inadequate (under) corneal compensation may result in a general increase in full-field retardation31 that may translate to spuriously high RNFL retardation, decreasing structure-function associations.27,30,31,34

Another contributing factor may be a “floor effect,” described by Schlottmann et al,35 which is when an RNFL of less than 20 μm is not detected. This would result in the detection of noise from other structures rather than signal from the remaining superior RNFL. In the present study, the superior RNFL was significantly thinner than the inferior (mean ± SD, 14.21 ± 8.41 μm vs 19.11 ± 10.55 μm), with the majority of affected eyes less than 20 μm. The MD was also more depressed in the inferior field (−15 dB) than the superior field (−8 dB) when compared with normal eyes. This loss of VF sensitivity, particularly in the inferior VF, may in fact equate to almost total loss of the superior RNFL.

Finally, the lack of correlation could be partly explained by the arbitrary division of sectors of the SLP and VF map. The division of the VF map is not based on altitudinal pattern. Therefore, measurements of the superior and inferior VF sectors are approximates that exclude half of the nasal and temporal fields that cross over into the respective superior or inferior hemifields. Moreover, the structure-function mapping used in this study attributes only 4 points to the temporal field and thus only 2 points for the superotemporal and inferotemporal fields. The nasal field is not sectorally represented on its own. Therefore, there may not be enough points to detect functional loss in such an analysis.

In summary, to our knowledge, this study is the first to quantitatively measure the attenuation of the RNFL with SLP in a group of patients who had a previous episode of NAION. It demonstrates a partial topographic correlation of optic nerve structure with MD, where magnitude of focal RNFL attenuation was related to the magnitude of the decreased VF sensitivity.

The ability to objectively measure the degree of damage to the optic nerve directly, rather than indirectly by the amount of VF damage, marks a juncture in the study of nonglaucomatous optic neuropathies. This study has shown that there is a quantitative relation between structure and function in an optic neuropathy other than glaucoma. Future research should further investigate the structure-function relationship of neuro-ophthalmic conditions and determine whether these technologies may provide us with diagnostic and prognostic information to improve patient care.

Correspondence: Helen V. Danesh-Meyer, FRANZCO, Department of Ophthalmology, Private Bag 92019, University of Auckland, Auckland Mail Centre, Auckland 1142, New Zealand (h.daneshmeyer@auckland.ac.nz).

Submitted for Publication: September 12, 2005; final revision received March 26, 2006; accepted May 14, 2006.

Author Contributions: Dr Danesh-Meyer had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

Financial Disclosure: None reported.

Funding/Support: Drs Carroll, Ku, and Gaskin were supported by the Alcon Optic Nerve Research Fellowship.

References
1.
Arnold  AC Pathogenesis of nonarteritic anterior ischemic optic neuropathy. J Neuroophthalmol 2003;23157- 163PubMedArticle
2.
Hayreh  SS Anterior ischemic optic neuropathy: differentiation of arteritic from non-arteritic type and its management. Eye 1990;425- 41PubMedArticle
3.
Henkind  PCharles  NCPearson  J Histopathology of ischemic optic neuropathy. Am J Ophthalmol 1970;6978- 90PubMed
4.
Quigley  HAAddicks  EM Quantitative studies of retinal nerve fibre layer defects. Arch Ophthalmol 1982;100807- 814PubMedArticle
5.
Sommer  AMiller  NRPollack  IMaumenee  AEGeorge  T The nerve fiber layer in the diagnosis of glaucoma. Arch Ophthalmol 1977;952149- 2156PubMedArticle
6.
Sommer  AKatz  JQuigley  HA  et al.  Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol 1991;10977- 83PubMedArticle
7.
Weinreb  RNShakiba  SSample  PA  et al.  Associations between quantitative nerve fiber layer measurement and visual field loss in glaucoma. Am J Ophthalmol 1995;120732- 738PubMed
8.
Reus  NJLemij  HG The relationship between standard automated perimetry and GDx VCC measurements. Invest Ophthalmol Vis Sci 2004;45840- 845PubMedArticle
9.
Weinreb  RNDreher  AWColeman  A  et al.  Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness. Arch Ophthalmol 1990;108557- 560PubMedArticle
10.
Reus  NJLemij  HG Diagnostic accuracy of the GDx VCC for glaucoma. Ophthalmology 2004;1111860- 1865PubMedArticle
11.
Mohammadi  KBowd  CWeinreb  RNMedeiros  FASample  PAZangwill  LM Retinal nerve fiber layer thickness measurements with scanning laser polarimetry predict glaucomatous visual field loss. Am J Ophthalmol 2004;138592- 601PubMedArticle
12.
Foroozan  RBuono  LMSavino  PJSergott  RC Scanning laser polarimetry of the retinal nerve fiber layer in central retinal artery occlusion. Ophthalmology 2003;110715- 718PubMedArticle
13.
Meier  FMBernasconi  PSturmer  JCaubergh  MJLandau  K Axonal loss from acute optic neuropathy documented by scanning laser polarimetry. Br J Ophthalmol 2002;3285- 287Article
14.
Steel  DHWWaldock  A Measurement of the retinal nerve fibre layer with scanning laser polarimetry in patients with previous demyelinating optic neuritis. J Neurol Neurosurg Psychiatry 1998;64505- 509PubMedArticle
15.
Miyahara  TKurimoto  YKurokawa  TKuroda  TYoshimura  N Alterations in retinal nerve fiber layer thickness following indirect traumatic optic neuropathy detected by nerve fiber analyzer, GDx-N. Am J Ophthalmol 2003;136361- 364PubMedArticle
16.
Colen  TPvan Everdingen  JAMLemij  HG Axonal loss in a patient with anterior ischemic optic neuropathy as measured with scanning laser polarimetry. Am J Ophthalmol 2000;130847- 848PubMedArticle
17.
Weinreb  RNShakiba  SZangwill  L Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol 1995;119627- 636PubMed
18.
Garway-Heath  DFPoinoosawmy  DFitzke  FWHitchings  RA Mapping the visual field to the optic disc in normal tension glaucoma eyes. Ophthalmology 2000;1071809- 1815PubMedArticle
19.
Weinreb  RNZangwill  LBerry  CCBathija  RSample  PA Detection of glaucoma with scanning laser polarimetry. Arch Ophthalmol 1998;1161583- 1589PubMedArticle
20.
Bowd  CZangwill  LBerry  CC  et al.  Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci 2001;421993- 2003PubMed
21.
Tjon-Fo-Sang  MJLemij  HG The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol 1997;12362- 69PubMed
22.
Trible  JRSchultz  RORobinson  JCRothe  TL Accuracy of scanning laser polarimetry in the diagnosis of glaucoma. Arch Ophthalmol 1999;1171298- 1304PubMedArticle
23.
Greenfield  DSKnighton  RWFeuer  WJSchiffman  JCZangwill  LWeinreb  RN Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol 2002;13427- 33PubMedArticle
24.
Weinreb  RNBowd  CGreenfield  DSZangwill  LM Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol 2002;120901- 906PubMedArticle
25.
Lan  Y-WHenson  DBKwartz  AJ The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma. Br J Ophthalmol 2003;871135- 1141PubMedArticle
26.
Kwon  YHHong  SHonkanen  HAAlward  WL Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma 2000;9281- 288PubMedArticle
27.
Choplin  NTZhou  QKnighton  RW Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments. Ophthalmology 2003;110719- 725PubMedArticle
28.
Horn  FKJonas  JBMartus  P  et al.  Polarimetric measurement of retinal nerve fiber layer thickness in glaucoma diagnosis. J Glaucoma 1999;8353- 362PubMedArticle
29.
Gardiner  SKJohnson  CACioffi  GA Evaluation of the structure-function relationship in glaucoma. Invest Ophthalmol Vis Sci 2005;463712- 3717PubMedArticle
30.
Greenfield  DSKnighton  RWHuang  XR Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am J Ophthalmol 2000;129715- 722PubMedArticle
31.
Knighton  RWHuang  XRGreenfield  DS Analytical model of scanning laser polarimetry for retinal nerve fiber layer assessment. Invest Ophthalmol Vis Sci 2002;43383- 392PubMed
32.
Reus  NJColen  TPLemij  HG Visualization of localized retinal nerve fiber layer defects with the GDx with individualized and with fixed compensation of anterior segment birefringence. Ophthalmology 2003;1101512- 1516PubMedArticle
33.
Tannenbaum  DPHoffman  DLemij  HGGarway-Heath  DFGreenfield  DSCaprioli  J Variable corneal compensation improves discrimination between normal and glaucomatous eyes with the scanning laser polarimeter. Ophthalmology 2004;111259- 264PubMedArticle
34.
Niessen  AGvan den Berg  TJLangerhorst  CTGreve  EL Retinal nerve fiber layer assessment by scanning laser polarimetry and standardized photography. Am J Ophthalmol 1996;121484- 493PubMed
35.
Schlottmann  PGCilla  SDGreenfield  DSCaprioli  JGarway-Heath  DF Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by scanning laser polarimetry. Invest Ophthalmol Vis Sci 2004;451823- 1829PubMedArticle
×