Figure 1.
Scanning electron microscopy shows the differences in tube dimensions. A, Poly(styrene-b-isobutylene-b-styrene) tube with smaller inner diameter (65 μm). B, A tab was attached to prevent migration. C, The silicone tube had a larger inner diameter (300 μm). D, A silicone tab was added.

Scanning electron microscopy shows the differences in tube dimensions. A, Poly(styrene-b-isobutylene-b-styrene) tube with smaller inner diameter (65 μm). B, A tab was attached to prevent migration. C, The silicone tube had a larger inner diameter (300 μm). D, A silicone tab was added.

Figure 2.
Intraocular pressure (IOP) comparison using 2 different tonometers. A, Perkins. B, Pneumatonometer. SIBS indicates poly(styrene-b-isobutylene-b-styrene).

Intraocular pressure (IOP) comparison using 2 different tonometers. A, Perkins. B, Pneumatonometer. SIBS indicates poly(styrene-b-isobutylene-b-styrene).

Figure 3.
Slitlamp and gonioscopic examination of the inserted poly(styrene-b-isobutylene-b-styrene) (SIBS) and silicone implants shows the excellent biocompatibility of both materials in the anterior chamber. Silicone tubes migrated almost entirely within 2 days (A), while SIBS tubes migrated within 2 weeks (D). At the end of follow-up, all animals in the silicone group (B) and the SIBS group (E) had deep and quiet anterior chambers. Gonioscopy demonstrated the silicone tubes in a close position or in contact with the iris (C, arrow). Without directly contacting them, SIBS tubes were parallel to the iris or the cornea (F).

Slitlamp and gonioscopic examination of the inserted poly(styrene-b-isobutylene-b-styrene) (SIBS) and silicone implants shows the excellent biocompatibility of both materials in the anterior chamber. Silicone tubes migrated almost entirely within 2 days (A), while SIBS tubes migrated within 2 weeks (D). At the end of follow-up, all animals in the silicone group (B) and the SIBS group (E) had deep and quiet anterior chambers. Gonioscopy demonstrated the silicone tubes in a close position or in contact with the iris (C, arrow). Without directly contacting them, SIBS tubes were parallel to the iris or the cornea (F).

Figure 4.
A, Low diffuse bleb in the poly(styrene-b-isobutylene-b-styrene) (SIBS) group. B, Giant blebs in the silicone group. C, In SIBS, the bleb was diffuse and vascularized. D, Encapsulation with silicone. Two animals in the silicone group had tube extrusion (E, F).

A, Low diffuse bleb in the poly(styrene-b-isobutylene-b-styrene) (SIBS) group. B, Giant blebs in the silicone group. C, In SIBS, the bleb was diffuse and vascularized. D, Encapsulation with silicone. Two animals in the silicone group had tube extrusion (E, F).

Figure 5.
Diffuse bleb demonstrated by optical coherence tomography (OCT) imaging (B) compared with a normal subconjunctival space in a nonoperated control (D). A and C, Location of the OCT scan of the ocular surface.

Diffuse bleb demonstrated by optical coherence tomography (OCT) imaging (B) compared with a normal subconjunctival space in a nonoperated control (D). A and C, Location of the OCT scan of the ocular surface.

Figure 6.
Fluorescein injected into the anterior chamber and drainage into the subconjunctival space to demonstrate patency. A, Immediate preinjection appearance showing position of the implant. Fluorescein exited from the distal tip of the tube into the subconjunctival space (B) and slowly diffused around (C). D, Fluorescein continued to diffuse through the tube into the subconjunctival space to define a 90° to 120° bleb.

Fluorescein injected into the anterior chamber and drainage into the subconjunctival space to demonstrate patency. A, Immediate preinjection appearance showing position of the implant. Fluorescein exited from the distal tip of the tube into the subconjunctival space (B) and slowly diffused around (C). D, Fluorescein continued to diffuse through the tube into the subconjunctival space to define a 90° to 120° bleb.

Figure 7.
Immunofluorescence detection of extracellular matrix deposition and myofibroblasts in the poly(styrene-b-isobutylene-b-styrene) (SIBS) and silicone groups 3 months after surgical implantation. Hematoxylin-eosin staining in the SIBS group showed no excessive collagen deposition, lymphocytes, or inflammatory cell infiltration (A). Collagen IV deposition was weak around the tube and did not surround the entire tube circumference (B). α Smooth muscle actin (α-SMA) did not stain the cells surrounding the tube (C). In the silicone glaucoma drainage implant, the iris stroma seemed quiet, and no stromal inflammatory cell infiltrates were observed at the entrance of the tube into the anterior chamber (D). E, Abundant collagen IV deposition was evident around the tube in all studied sections. In contrast to the SIBS tubes, α-SMA was expressed around the silicone tubes (F). Bar represents 150 μm.

Immunofluorescence detection of extracellular matrix deposition and myofibroblasts in the poly(styrene-b-isobutylene-b-styrene) (SIBS) and silicone groups 3 months after surgical implantation. Hematoxylin-eosin staining in the SIBS group showed no excessive collagen deposition, lymphocytes, or inflammatory cell infiltration (A). Collagen IV deposition was weak around the tube and did not surround the entire tube circumference (B). α Smooth muscle actin (α-SMA) did not stain the cells surrounding the tube (C). In the silicone glaucoma drainage implant, the iris stroma seemed quiet, and no stromal inflammatory cell infiltrates were observed at the entrance of the tube into the anterior chamber (D). E, Abundant collagen IV deposition was evident around the tube in all studied sections. In contrast to the SIBS tubes, α-SMA was expressed around the silicone tubes (F). Bar represents 150 μm.

Figure 8.
Immunofluorescence detection of extracellular matrix deposition and myofibroblasts in the poly(styrene-b-isobutylene-b-styrene) (SIBS) 6 months after surgical implantation. Interlaminar spaces were clearly observed in the subconjunctival space, suggesting the presence of interstitial aqueous humor (A). A light deposition of collagen IV completely surrounded the tube (B, marked by white stars; collagen immunoreactivity was also observed in basement membranes and blood vessels), which was clearly evident at higher magnifications (C). No smooth muscle actin or macrophages reactivity were observed in any of the SIBS tubes (D) in any time frame. Bars represent 150 μm.

Immunofluorescence detection of extracellular matrix deposition and myofibroblasts in the poly(styrene-b-isobutylene-b-styrene) (SIBS) 6 months after surgical implantation. Interlaminar spaces were clearly observed in the subconjunctival space, suggesting the presence of interstitial aqueous humor (A). A light deposition of collagen IV completely surrounded the tube (B, marked by white stars; collagen immunoreactivity was also observed in basement membranes and blood vessels), which was clearly evident at higher magnifications (C). No smooth muscle actin or macrophages reactivity were observed in any of the SIBS tubes (D) in any time frame. Bars represent 150 μm.

Table. 
Summary of Different Variables Studied During the Experiments
Summary of Different Variables Studied During the Experiments
1.
Stanworth  A Conjunctival fibrosis after filtration operations. Trans Opthal Soc U K 1958;7843- 55
PubMed
2.
Tribin-Piedrahita  A Beta therapy in filtering operations for the treatment of glaucoma. Am J Ophthalmol 1965;60140- 141
PubMed
3.
Rollet  M Le drainage au irin de la chamabre anterieure contre l'hypertonie et la douleur. Rev Gen Ophtalmol 1906;25481
4.
Bock  RH Subconjunctival drainage of the anterior chamber by a glass seton. Am J Ophthalmol 1950;33929- 933
PubMed
5.
Steffanson  J An operation for glaucoma. Am J Ophthalmol 1925;8681- 692
6.
Zorab  A The reduction of tension in chronic glaucoma. Ophthalmoscope 1912;10258- 261
7.
Helies  PLegeais  JMSavoldelli  M  et al.  Artificial trabeculum (MESH): clinical and histological study in the rabbit [in French]. J Fr Ophtalmol 1998;21351- 360
PubMed
8.
Kondo  HFantes  FKato  H  et al.  Synthetic meshwork implant for glaucoma filtering surgery: effect of adjunct heparin and sodium hyaluronate in rabbits. Ophthalmic Surg Lasers 1998;29669- 676
PubMed
9.
Nyska  AGlovinsky  YBelkin  MEpstein  Y Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant. J Glaucoma 2003;12275- 280
PubMedArticle
10.
Wilcox  MJBarad  JPWilcox  CC  et al.  Performance of a new, low-volume, high-surface area aqueous shunt in normal rabbit eyes. J Glaucoma 2000;974- 82
PubMedArticle
11.
Molteno  AC New implant for drainage in glaucoma: animal trial. Br J Ophthalmol 1969;53161- 168
PubMedArticle
12.
Molteno  AC New implant for drainage in glaucoma: clinical trial. Br J Ophthalmol 1969;53606- 615
PubMedArticle
13.
Molteno  ACStraughan  JLAncker  E Long tube implants in the management of glaucoma. S Afr Med J 1976;501062- 1066
PubMed
14.
Prata  JA  JrMermoud  ALaBree  LMinckler  DS In vitro and in vivo flow characteristics of glaucoma drainage implants. Ophthalmology 1995;102894- 904
PubMedArticle
15.
Krupin  TPodos  SMBecker  BNewkirk  JB Valve implants in filtering surgery. Am J Ophthalmol 1976;81232- 235
PubMed
16.
Coleman  ALHill  RWilson  MR  et al.  Initial clinical experience with the Ahmed Glaucoma Valve implant. Am J Ophthalmol 1995;12023- 31
PubMed
17.
Molteno  AC The optimal design of drainage implants for glaucoma. Trans Ophthalmol Soc N Z 1981;3339- 41
PubMed
18.
Lloyd  MABaerveldt  GHeuer  DK  et al.  Initial clinical experience with the baerveldt implant in complicated glaucomas. Ophthalmology 1994;101640- 650
PubMedArticle
19.
Siegner  SWNetland  PAUrban  RC  Jr  et al.  Clinical experience with the Baerveldt glaucoma drainage implant. Ophthalmology 1995;1021298- 1307
PubMedArticle
20.
Ayyala  RSMichelini-Norris  BFlores  A  et al.  Comparison of different biomaterials for glaucoma drainage devices: part 2. Arch Ophthalmol 2000;1181081- 1084
PubMedArticle
21.
Jacob  JTBurgoyne  CFMcKinnon  SJ  et al.  Biocompatibility response to modified Baerveldt glaucoma drains. J Biomed Mater Res 1998;4399- 107
PubMedArticle
22.
Pandya  ADRich  CEifrig  DE  et al.  Experimental evaluation of a hydroxylapatite reservoir tube shunt in rabbits. Ophthalmic Surg Lasers 1996;27308- 314
PubMed
23.
Kennedy  JPIvan  B Designed Polymers by Carbocationic Macromolecular Engineering: Theory and Practice.  Munich, Germany Hanser1992;
24.
Pinchuk  LKhan  IJMartin  JB  et al.  A new family of thermoplastic elastomers for ultra-long term implant based upon a backbone of alternating quaternary and secondary carbons.  Paper presented at: 44th Annual Conference ASAIO/CVST April 23-25, 1998 New York, NY
25.
Richard  REBarry  JJKamath  KMiller  SVRanade  SV Polymer technology for the controlled release of paclitaxel from a vascular compatible matrix: the TaxusTM Express paclitaxel-eluting coronary stent.  Paper presented at: Biomaterials in Regenerative Medicine: The Advent of Combination Products October 16-18, 2004 Philadelphia, Pa
26.
Ranade  SVMiller  KMRichard  RE  et al.  Physical characterization of controlled release of paclitaxel from the TAXUS Express2 drug-eluting stent. J Biomed Mater Res A 2004;71625- 634
PubMedArticle
27.
Pinchuk  L Biostable elastomeric polymers having quaternary carbons. US Patent 5 741 331 April21 1998;
28.
Stone  GWEllis  SGCox  DA  et al.  A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med 2004;350221- 231
PubMedArticle
29.
Stone  GWEllis  SGCox  DA  et al.  One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial. Circulation 2004;1091942- 1947
PubMedArticle
30.
Molteno  ACVan Rooyen  MMBartholomew  RS Implants for draining neovascular glaucoma. Br J Ophthalmol 1977;61120- 125
PubMedArticle
31.
Molteno  ACAncker  EVan Biljon  G Surgical technique for advanced juvenile glaucoma. Arch Ophthalmol 1984;10251- 57
PubMedArticle
32.
Da Mata  ABurk  SENetland  PA  et al.  Management of uveitic glaucoma with Ahmed glaucoma valve implantation. Ophthalmology 1999;1062168- 2172
PubMedArticle
33.
Molteno  ACSayawat  NHerbison  P Otago glaucoma surgery outcome study: long-term results of uveitis with secondary glaucoma drained by Molteno implants. Ophthalmology 2001;108605- 613
PubMedArticle
34.
Beebe  WEStarita  RJFellman  RL  et al.  The use of Molteno implant and anterior chamber tube shunt to encircling band for the treatment of glaucoma in keratoplasty patients. Ophthalmology 1990;971414- 1422
PubMedArticle
35.
Arroyave  CPScott  IUFantes  FE  et al.  Corneal graft survival and intraocular pressure control after penetrating keratoplasty and glaucoma drainage device implantation. Ophthalmology 2001;1081978- 1985
PubMedArticle
36.
Lee  RKFantes  F Surgical management of patients with combined glaucoma and corneal transplant surgery. Curr Opin Ophthalmol 2003;1495- 99
PubMedArticle
37.
Heuer  DKLloyd  MAAbrams  DA  et al.  Which is better? One or two? A randomized clinical trial of single-plate versus double-plate Molteno implantation for glaucomas in aphakia and pseudophakia. Ophthalmology 1992;991512- 1519
PubMedArticle
38.
Mills  RPReynolds  AEmond  MJ  et al.  Long-term survival of Molteno glaucoma drainage devices. Ophthalmology 1996;103299- 305
PubMedArticle
39.
Molteno  ACFucik  MDempster  AGBevin  TH Otago Glaucoma Surgery Outcome Study: factors controlling capsule fibrosis around Molteno implants with histopathological correlation. Ophthalmology 2003;1102198- 2206
PubMedArticle
40.
Geankoplis  CJ Transport Processes and Separation Process Principles (Includes Unit Operations). 3rd Upper Saddle River, NJ Prentice Hall1993;
41.
Acosta  ACFernandez  VLamar  PD  et al.  Ocular biocompatibility of QuatromertmTM (polystyrene-polyisobutylene triblock polymers) for glaucoma implants. Invest Ophthalmol Vis Sci 2004;45E-Abstract 2929
42.
Jacob  JTLacour  OJBurgoyne  CF Slow release of the antimetabolite 5-fluorouracil (5-FU) from modified Baerveldt glaucoma drains to prolong drain function. Biomaterials 2001;223329- 3335
PubMedArticle
43.
Morales  JKelleher  PJCampbell  DCrosson  CE Effects of daunomycin implants on filtering surgery outcomes in rabbits. Curr Eye Res 1998;17844- 850
PubMedArticle
Laboratory Sciences
December 2006

A Newly Designed Glaucoma Drainage Implant Made of Poly(styrene-b-isobutylene-b-styrene)Biocompatibility and Function in Normal Rabbit Eyes

Author Affiliations

Author Affiliations: Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine (Drs Acosta, Espana, Yamamoto, Orozco, Dubovy, Fantes, and Parel), and Department of Biomedical Engineering, University of Miami College of Engineering (Drs Orozco and Parel), Miami, Fla; InnFocus LLC, Miami, Fla (Drs Davis, Pinchuk, and Weber); and the University of Liege, CHU Sart-Tilman, Liege, Belgium (Dr Parel).

Arch Ophthalmol. 2006;124(12):1742-1749. doi:10.1001/archopht.124.12.1742
Abstract

Objective  To report clinical evaluation, flow patency, and histopathological findings of a novel glaucoma drainage implant (GDI) made of poly(styrene-b-isobutylene-b-styrene) (SIBS) in rabbits.

Methods  In 16 normal eyes, the proximal end of the SIBS GDI was inserted into the anterior chamber while the distal end was placed in the subconjunctival space. A control group underwent implantation of a similarly designed silicone GDI. Slitlamp follow-up and intraocular pressure measurements were recorded. Flow patency was evaluated by injecting 0.01% fluorescein into the anterior chamber. Immunostaining against collagen IV, macrophages, and α smooth muscle actin was performed.

Results  Slitlamp examination suggested adequate biocompatibility. A low and diffuse bleb was observed in the SIBS group. All SIBS tubes were patent 6 months after insertion. Immunostaining demonstrated noncontinuous collagen deposition. No macrophages or myofibroblasts were visible around the SIBS tubes. In contrast, silicone induced collagen deposition and myofibroblast differentiation.

Conclusion  This new GDI is clinically biocompatible in the rabbit and maintained 100% patency at 6 months. A remarkable difference was the absence of myofibroblasts in the surrounding tissue in the SIBS group.

Clinical Relevance  This novel GDI made of SIBS would prevent the feared complication of hypotony and will decrease the amount of subconjunctival fibrosis.

Drainage surgery to control intraocular pressure (IOP) with different devices has limited success because of a buildup of extracellular matrix.1,2 The initial devices used to drain aqueous were horsehair threads.3 Other materials including glass rods,4 gold,5 silk,6 and even artificial trabecular meshworks7,8 were tried without success because of subconjunctival scarring. New devices including the Ex-PRESS glaucoma implant9 and Wilcox shunt10 are described in the literature; however, long-term effectiveness data are not available.

Modern glaucoma drainage surgery began with the pioneer work of Molteno in rabbits11 and humans,12 where the aqueous humor was drained from the anterior chamber to a plate adjacent to the limbus. He later increased the silicone tube length to allow the drainage of aqueous into a more posterior area.13 Another modification in glaucoma drainage implants (GDIs) was the introduction of unidirectional valve systems to avoid postoperative hypotony.14 Krupin et al15 reported their results with a valved system in 1976, followed by Coleman et al16 with the Ahmed implant. Subsequent design changes by Molteno17 and Baerveldt18,19 were aimed at increasing the plate area.

It is believed that the fibrotic and inflammatory reactions induced by biomaterials are a major determinant of success.2022 Other factors such as shape, flexibility, modulus, and texture could also be associated with erosion, extrusion, inflammation, and scarring.20 Hypothetically, selecting a biomaterial and a design that produces minimal inflammation and fibrosis is a means to increase success.

Poly(styrene-b-isobutylene-b-styrene) (SIBS), a novel synthetic polymer consisting of a triblock of polystyrene-polyisobutylene-polystyrene,23 has been studied in different medical fields.24,25 Multiple investigators26,27 observed that SIBS has excellent biostability. The lack of biodegradation byproducts is believed to be a major factor enhancing its biocompatibility. In addition, SIBS can be loaded with different pharmaceutical agents.25 These unique material properties of SIBS are crucial for the fabrication of life-lasting implants, as demonstrated by the TAXUS coronary stent (Boston Scientific, Natick, Mass). The TAXUS stent uses SIBS as a carrier for the antiproliferative agent paclitaxel, which is released into the vessel wall to prevent the proliferation of smooth muscle cells.28,29

In this article, we report the preliminary results of biocompatibility, flow patency, and histopathology of a novel GDI made of SIBS, where the new features include a new biostable elastomeric polymer, a small and flexible design, and a valveless tube with an inner diameter of 65 μm.

METHODS
ANIMALS AND MATERIALS

The animals used in the study were 25 normal female New Zealand white rabbits from Harlan Laboratories (Indianapolis, Ind), weighing 2.5 to 3.0 kg. All animals were treated in accordance to the Association for Research in Vision and Ophthalmology statement for the use of animals in ophthalmic research. The Committee and Review Board for Animal Research of the University of Miami (Miami, Fla) approved all animal studies. Prior to surgery, all animals were examined to exclude ocular diseases. A Perkins tonometer (Clement Clarke, London, England) and a pneumatonometer (Mentor Inc, Norwell, Mass) were calibrated. Evaluation of the central corneal thickness was performed using an ultrasonic pachymeter (DGH 500 Pachette; DGH Technology, Frazer, Pa). The newly designed GDIs were made of SIBS, with 24 mol% styrene (InnFocus LLC, Miami, Fla).

The polydimethylsiloxane tubes, known as silicone, were purchased (Silastic; Dow Corning, Midland, Mich). The lumens of the SIBS tubes were coated with glycerol to avoid sticking during manipulation. The initial design of the SIBS tube had a mean inner diameter of 65 ± 10 μm, and a mean outer diameter of 250 ± 10 μm (Figure 1A). The silicone rubber tubes had a mean inner diameter of 300 ± 10 μm and a mean outer diameter of 640 ± 15 μm (Figure 1C). This design was later modified by attaching a tab composed of the same materials to the wall in the middle region of the tube in both the SIBS and silicone tubes (Figures 1B and 1D, respectively). A disposable inserter device with a grip, a 27-gauge slotted needle, and a deployment slide were developed to facilitate insertion.

SURGICAL PROCEDURE

The right or left eye of 25 rabbits was randomly assigned for surgery; the other eye was used as a control. Animals were given an intramuscular injection of a ketamine-xylazine-acepromazine mixture (35 mg/kg, 5 mg/kg, and 0.75 mg/kg, respectively). All surgeries were performed by the same surgeon (A.C.A.). The surgical area was exposed with a corneal traction suture. A 90° fornix-based conjunctival peritomy was made in the superotemporal quadrant and the subconjunctival space was dissected posteriorly 10 to 14 mm using Westcott scissors. A 27-gauge needle was inserted 2 mm posterior to the limbus and directed to the anterior chamber. The SIBS GDI was preloaded into the inserter. The preloaded inserter entered the sclera following the needle scleral tract. Once the GDI was 3 mm into the anterior chamber, the GDI was released by retracting the thumb slide. Flow through the tube was confirmed by gently pressing the cornea until aqueous came out of the distal end. The distal end was then positioned under the conjunctiva. The conjunctiva was sutured with 7-0 Vicryl (Ethicon, Somerville, NJ). The silicone GDI was inserted in a similar manner, except that a 23-gauge needle was used for the scleral tract. The animals were treated with a subcutaneous injection of 0.03 mg/kg of buprenorphine after surgery and bacitracin-neomycin-polymyxin ointment and prednisolone acetate 1% (Allergan Inc, Irvine, Calif) twice a day for 3 days.

SLITLAMP BIOMICROSCOPY, IOP MEASUREMENTS, AND CORNEAL PACHYMETRY

Slitlamp follow-up was performed weekly. After topical anesthesia, IOP was measured with the Perkins tonometer and pneumatonometer in both operated and nonoperated eyes at 0, 1, 3, 14, 21, 60, 120, and 150 days after surgery. Animals were examined under general anesthesia at days 7, 28, 90, and 180 after surgery.

FLUORESCEIN PATENCY TEST

To evaluate the patency of the GDI and the passage of aqueous humor from the anterior chamber to the subconjunctival space, a solution of 0.01% sodium fluorescein (AK-Fluor; Akorn, Decatur, Ill) in a balanced salt solution was injected into the anterior chamber of 16 eyes in the SIBS group and 6 eyes in the silicone group. These experiments were performed at 90 and 180 days postoperatively. Briefly, the anterior chamber was entered with a 30-gauge needle, and the aqueous humor was allowed to drain out from the anterior chamber to avoid excessive IOP. A second 30-gauge needle was introduced into the anterior chamber and approximately 0.5 mL of fluorescein was injected for 20 minutes. Intraocular pressure was monitored during the procedure and did not exceed 25 mm Hg. Photographic documentation was acquired using an illuminator paraxial to the operation microscope (OMS-300; Topcon USA, Paramatta, NJ) and blue dichroic excitation and yellow barrier filters (Y54-653 and NT47-247; Edmond Industrial Optics, Barrington, NJ).

LIGHT MICROSCOPY AND IMMUNOFLUORESCENT STAINING

Animals were euthanized with 390 mg/ml Euthasol (Delmarva Laboratories Inc, Midlothian, Va). The GDI and the surrounding tissue were dissected, immersed in optimal cutting temperature component (Tissue-Tek; Sakura, Torrance, Calif), and snap-frozen in liquid nitrogen. Frozen sections, 5-μm-thick, were cut and fixated in acetone for 10 minutes at −20°C and blocked with 1% goat serum for 30 minutes. Sections were incubated overnight with the following antibodies: α smooth muscle actin (α-SMA) (monoclonal, 1:100 dilution; DAKO, Carpintera, Calif), macrophages (monoclonal, 1:50 dilution), and collagen IV (polyclonal that cross-reacts with collagens I, II, and III, 1:100 dilution; Southern Biotechnical, Birmingham, Ala). Rhodamine-B-isothiocyanate–antigoat and fluorescein-5-isothiocyanate–antimouse secondary antibodies (Sigma-Aldrich, St Louis, Mo) were used at 1:100 dilution. The 4',6-diamidino-2-phenylindole mounting media Vectashield was from Vector laboratories (Burlingame, Calif). Some frozen sections were stained with hematoxylin-eosin. Photographs were taken using a confocal laser scanning microscope (LSM-510; Carl Zeiss, Oberkochen, Germany).

RESULTS
SIBS GDIs AND POSTOPERATIVE HYPOTONY

Surgical time varied between 11 and 18 minutes in both groups. The immediate postoperative period was unremarkable in the SIBS group. In contrast, all rabbits in the silicone group had flat or shallow anterior chambers for 2 to 3 days. Hyphema formed in all rabbits in the silicone group during the surgical procedure. One rabbit in this group had blood and fibrin inside the tube and did not form a bleb. Immediate postoperative IOP readings were taken only with the Perkins tonometer to avoid the pressure exerted by the pneumatonometer in a shallow chamber. In the SIBS group, the mean IOP taken immediately after tube insertion and still under the effect of general anesthesia was 5.6 ± 2.1 mm Hg. The lowest IOP recorded was 3.5 mm Hg. One day postoperatively, the mean IOP was 7.9 ± 1.8 mm Hg in all eyes with SIBS implants. In contrast, all controls in the silicone group had flat anterior chambers with IOP of 0 mm Hg immediately after surgery. No statistically significant difference in IOP reduction was observed between the operated and nonoperated eyes in the SIBS or silicone groups or between the SIBS and silicone groups 7 days postoperatively (Figure 2). Taken together, these observations demonstrate the advantage of the SIBS tubes in avoiding immediate postoperative hypotony.

CLINICAL BIOCOMPATIBILITY OF SIBS GDIs

Tubes without a tab (the initial design in both materials) migrated into the anterior chamber. Silicone tubes migrated almost entirely within 2 days (Figure 3A), while SIBS tubes migrated within 2 weeks (Figure 3D). No changes in central corneal thickness were recorded in any group after surgery or during the 6-month follow-up. The mean central corneal thickness was 360 ± 8 μm. No tube migration was observed once a silicone or SIBS tab was attached to the silicone or SIBS tubes, respectively. No other complications were observed in any of the SIBS tubes with an attached tab. In contrast, all silicone tubes produced flat anterior chambers and hyphema during the immediate postoperative period (not shown). At the end of follow-up, all animals in the silicone group (Figure 3B) and the SIBS group (Figure 3E) had deep and quiet anterior chambers. Gonioscopy demonstrated the silicone tubes in a close position or in contact with the iris (Figure 3C). Without directly contacting them, SIBS tubes were parallel to the iris or the cornea (Figure 3F). Small blebs were a typical finding during the first 3 postoperative days in the SIBS group (Figure 4A), compared with very large blebs that slowly flattened in the silicone group (Figure 4B). One week after surgery, only loose subconjunctival space was noticed at the end of the tube, suggesting the presence of a diffuse bleb in both SIBS and silicone groups. In the SIBS group, the bleb was diffuse and vascularized; however, encapsulation and neovascularization were never observed around the tube for the duration of the study (Figure 4C). Fibrous encapsulation and neovascularization around the tube were clearly observed at 1 month (Figure 4D) in eyes that underwent silicone tubes implantation. Two animals in the silicone group were euthanized because of tube extrusion at the limbus area 45 and 109 days postoperatively (Figures 4E and 4F).

SIBS GLAUCOMA DRAINAGE IMPLANTS AFTER 6 MONTHS

Although we observed a low bleb surrounding the tube during slitlamp examination and confirmed it with ancillary imaging (Figure 5), fluorescein flow and tube patency was clearly observed through all tubes in 16 eyes of 16 rabbits evaluated in the SIBS group. Ten of these animals had a follow-up of 6 months at the time of the fluorescein assay (Table). After injection of fluorescein into the chamber, fluorescein began to fill the perilimbal region and the tube lumen within seconds. Fluorescein exited from the distal tip of the tube into the subconjunctival space (Figure 6B) and slowly diffused around (Figure 6C). Fluorescein continued to diffuse through the tube into the subconjunctival space to define a 90° to 120° bleb (Figure 6D). In contrast, only 2 of 6 silicone tubes were patent at 3 months. The blebs were limited to the area adjacent to the silicone tube and did not diffuse to the surrounding subconjunctival space after 30 minutes of injection (not shown).

FIBROTIC RESPONSES OF SIBS GDIs

Sections were obtained at different locations along the tube track in the SIBS group at 3 and 6 months and at 3 months in the silicone group. Hematoxylin-eosin staining in the SIBS group showed no excessive collagen deposition, lymphocytes, or inflammatory cell infiltration (Figure 7A), even in the area where the tab was located (not shown). Collagen IV deposition was weak around the tube and did not surround the entire tube circumference (Figure 7B). α Smooth muscle actin did not stain the cells surrounding the tube (Figure 7C). No staining against macrophages was observed (not shown). Similarly, sparse collagen IV deposition was observed around the tab with the absence of a continuous capsule (not shown). No reactivity against α-SMA was observed around the tab (not shown).

In the silicone GDI, the iris stroma seemed quiet and no stromal inflammatory cell infiltrates were observed at the entrance of the tube into the anterior chamber (Figure 7D) or at a distal region (not shown). Abundant collagen IV deposition was evident around the tube in all studied sections (Figure 7E). In contrast to the SIBS tubes, α-SMA was expressed around the silicone tubes (Figure 7F), and highly deposited areas of collagen were observed around the silicone tab (not shown). At 6 months, hematoxylin-eosin staining of the SIBS tubes did not show collagen deposition or cell infiltration around the tube. Interlaminar spaces were clearly observed in the subconjunctival space suggesting the presence of interstitial aqueous humor (Figure 8A). A light deposition of collagen IV completely surrounded the tube (Figure 8B), which was clearly evident at higher magnifications (Figure 8C). No smooth muscle actin or macrophages reactivity were observed in any of the SIBS tubes (Figure 8D) in any time frame.

COMMENT

Drainage implants are valuable in the management of neovascular30 and juvenile glaucoma,31 as well as glaucoma associated with uveitis,32,33 penetrating keratoplasty,3436 aphakia,37 and failed filtering surgery.38 Despite different modifications to the material, surgical technique, and size and design of implants, the most important cause of failure continues to be the excessive scarring that impedes the exit of aqueous.39

Although the biocompatibility of any device is mainly related to the material, other issues, such as design and flexibility, are critical. This new SIBS GDI is made of an inert, soft, and flexible thermoformable material more conforming to the eye curvature in contrast to silicone controls. The thermoset nature of silicone rubber provides a memory to the tubes, which remain in their original shape. This straightening may be responsible for the 2 extrusions described. The combination of flexibility and conformability, the ability of SIBS to take on a new relaxed shape, the overall design, the chemical inertness, and the lack of molecules that elute from the material all contribute to the reduction in microtrauma and inflammation, therefore decreasing scarring.

The excellent biocompatibility observed clinically with the SIBS implants was confirmed by histological evaluation that showed reduced collagen deposition around the SIBS tubes at 3 and 6 months. Immunofluorescence demonstrated a modest and discontinuous collagen deposition in the SIBS group, whereas a complete circular collagen deposition was observed around the silicone tubes. To further study the amount of reaction induced by the implants, the cellular components responsible for scarring were characterized. Surprisingly, myofibroblasts in the tissue surrounding the SIBS devices—either in areas where SIBS was in contact with the iris, in the area where the tab was located, or in the subconjunctival space where the bleb was formed—were never observed. In contrast, the silicone drainage device always induced the expression of α-SMA. This expression was more evident in the area where the silicone tab was located. These histological results support the concept that SIBS is indeed a material that reduces the capsule formation by decreasing the amount of collagen deposition and the differentiation of myofibroblasts. This lack of scarring allows the outflow of aqueous humor into the subconjunctival space.

It was a challenge to confirm that there was indeed flow in the aqueous shunts in the long-term in the absence of a discernible bleb and without the ability to visualize direct flow in the healthy rabbit model. For this reason, we chose to inject fluorescein into the anterior chamber at 3 and 6 months after surgery. Immediate flow of aqueous was observed with the fluid dispersing into the subconjunctival space within seconds, with subsequent enlargement into a diffuse bleb 20 to 30 minutes later. This experiment confirmed that the tubes that showed fluorescein flow were patent. As reported, all (16 of 16) SIBS tubes (100%) were patent in the fluorescein study, in contrast to the 2 of 6 (33%) silicone tubes.

The SIBS GDI did not demonstrate postoperative hypotony, in contrast to the silicone rubber tubes, which did. The reason for this disparity is that the lumen of the SIBS tube was significantly smaller than the silicone tube. The SIBS lumen was designed to be approximately 60 μm with an 11-mm length, as suggested by the Hagen-Poiseuille equation,40 given a flow rate of approximately 2.5 μL/min and a desired resultant inflow pressure of 5 to 10 mm Hg in the absence of a distal resistance to outflow. However, the effects of a continuous force on the eye or Valsalva maneuvers on eye pressure are unknown because there is no capsule at the distal end that limits the outflow. The 300-μm lumen of the silicone tube extrapolates to negligible pressure by the Hagen-Poiseuille equation at a similar flow rate, as was confirmed by the hypotony observed in vivo.

Although the argument can be made that the hypotony, capsule formation, and other factors relating to the poor performance of the silicone tubes were inherent to their design and size, and were perhaps a result of immediate hypotony and the spewing of inflammatory cytokines into the subconjunctival space, we are convinced that this is not entirely the case, albeit large amounts of cytokines in the subconjunctival space would not be desirable nor would hypotony, and the SIBS GDIs were designed appropriately smaller for this reason. In a previous pilot study,41 0.3-mm-thick discs with a 3-mm diameter made of SIBS showed minimal inflammation without cellular infiltration, neovascularization, infection, or toxic reaction when compared with identically sized cross-linked poly(dimethylsiloxane) discs that demonstrated marked neovascularization and fibrotic reaction. These materials were implanted both in the corneal stroma and sub-Tenon space of healthy rabbits. It is noteworthy that in these cases there was no aqueous humor or other active source of cytokines in contact with the biomaterials, clearly indicating a local response to the material as opposed to the geometry.

Using normal rabbits without established high IOP may explain the similarities in the IOP between the silicone and the SIBS tubes and between operated and nonoperated eyes. Although the SIBS GDIs were patent and formed a flat and extended bleb, no statistically significant difference in IOP between the SIBS and silicone groups was found after 7 days. We hypothesize that the pressure similarities between the SIBS and silicone groups relate to the remaining normal drainage system parallel to the shunt. Finally, a major advantage of SIBS is the possibility of loading the distal end of the GDIs with antiproliferative or anti-inflammatory agents.42,43

Long-term studies in humans will determine whether these small GDIs made of SIBS are clinically useful in long-term IOP reduction. Hopefully, this novel GDI, with its simple and quick method of implantation, will avoid and prevent some of the complications observed with the implantation of current GDIs. The small size of the SIBS GDI, as well as ease of removal, if necessary, may provide a valuable tool to treat patients suffering from glaucoma.

Correspondence: Jean-Marie Parel, PhD, Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, 1638 NW 10 Ave, Miami, FL 33136 (jmparel@med.miami.edu).

Submitted for Publication: October 25, 2005; final revision received April 26, 2006; accepted July 2, 2006.

Financial Disclosure: Drs Acosta, Espana, Yamamoto, Orozco, Dubovy, Fantes, and Parel are employed at the University of Miami; and Drs Davis, Pinchuk, and Weber are employed at InnFocus LLC.

Funding/Support: This study was supported by Florida Lions Eye Bank; InnFocus LLC; Research to Prevent Blindness; Henri and Flore Lesieur Foundation; and National Institutes of Health Center Grant P30-EY014801. InnFocus LLC and the University of Miami have proprietary rights to the devices described herein.

Acknowledgment: Magda Celdran prepared the tissue for light microscopy; Eleut Hernandez, LAT, provided daily animal care; and Izuru Nose, BS, gave technical assistance during the project. John B. Martin and Saul Gottlieb built the molds and inserters used in this study.

References
1.
Stanworth  A Conjunctival fibrosis after filtration operations. Trans Opthal Soc U K 1958;7843- 55
PubMed
2.
Tribin-Piedrahita  A Beta therapy in filtering operations for the treatment of glaucoma. Am J Ophthalmol 1965;60140- 141
PubMed
3.
Rollet  M Le drainage au irin de la chamabre anterieure contre l'hypertonie et la douleur. Rev Gen Ophtalmol 1906;25481
4.
Bock  RH Subconjunctival drainage of the anterior chamber by a glass seton. Am J Ophthalmol 1950;33929- 933
PubMed
5.
Steffanson  J An operation for glaucoma. Am J Ophthalmol 1925;8681- 692
6.
Zorab  A The reduction of tension in chronic glaucoma. Ophthalmoscope 1912;10258- 261
7.
Helies  PLegeais  JMSavoldelli  M  et al.  Artificial trabeculum (MESH): clinical and histological study in the rabbit [in French]. J Fr Ophtalmol 1998;21351- 360
PubMed
8.
Kondo  HFantes  FKato  H  et al.  Synthetic meshwork implant for glaucoma filtering surgery: effect of adjunct heparin and sodium hyaluronate in rabbits. Ophthalmic Surg Lasers 1998;29669- 676
PubMed
9.
Nyska  AGlovinsky  YBelkin  MEpstein  Y Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant. J Glaucoma 2003;12275- 280
PubMedArticle
10.
Wilcox  MJBarad  JPWilcox  CC  et al.  Performance of a new, low-volume, high-surface area aqueous shunt in normal rabbit eyes. J Glaucoma 2000;974- 82
PubMedArticle
11.
Molteno  AC New implant for drainage in glaucoma: animal trial. Br J Ophthalmol 1969;53161- 168
PubMedArticle
12.
Molteno  AC New implant for drainage in glaucoma: clinical trial. Br J Ophthalmol 1969;53606- 615
PubMedArticle
13.
Molteno  ACStraughan  JLAncker  E Long tube implants in the management of glaucoma. S Afr Med J 1976;501062- 1066
PubMed
14.
Prata  JA  JrMermoud  ALaBree  LMinckler  DS In vitro and in vivo flow characteristics of glaucoma drainage implants. Ophthalmology 1995;102894- 904
PubMedArticle
15.
Krupin  TPodos  SMBecker  BNewkirk  JB Valve implants in filtering surgery. Am J Ophthalmol 1976;81232- 235
PubMed
16.
Coleman  ALHill  RWilson  MR  et al.  Initial clinical experience with the Ahmed Glaucoma Valve implant. Am J Ophthalmol 1995;12023- 31
PubMed
17.
Molteno  AC The optimal design of drainage implants for glaucoma. Trans Ophthalmol Soc N Z 1981;3339- 41
PubMed
18.
Lloyd  MABaerveldt  GHeuer  DK  et al.  Initial clinical experience with the baerveldt implant in complicated glaucomas. Ophthalmology 1994;101640- 650
PubMedArticle
19.
Siegner  SWNetland  PAUrban  RC  Jr  et al.  Clinical experience with the Baerveldt glaucoma drainage implant. Ophthalmology 1995;1021298- 1307
PubMedArticle
20.
Ayyala  RSMichelini-Norris  BFlores  A  et al.  Comparison of different biomaterials for glaucoma drainage devices: part 2. Arch Ophthalmol 2000;1181081- 1084
PubMedArticle
21.
Jacob  JTBurgoyne  CFMcKinnon  SJ  et al.  Biocompatibility response to modified Baerveldt glaucoma drains. J Biomed Mater Res 1998;4399- 107
PubMedArticle
22.
Pandya  ADRich  CEifrig  DE  et al.  Experimental evaluation of a hydroxylapatite reservoir tube shunt in rabbits. Ophthalmic Surg Lasers 1996;27308- 314
PubMed
23.
Kennedy  JPIvan  B Designed Polymers by Carbocationic Macromolecular Engineering: Theory and Practice.  Munich, Germany Hanser1992;
24.
Pinchuk  LKhan  IJMartin  JB  et al.  A new family of thermoplastic elastomers for ultra-long term implant based upon a backbone of alternating quaternary and secondary carbons.  Paper presented at: 44th Annual Conference ASAIO/CVST April 23-25, 1998 New York, NY
25.
Richard  REBarry  JJKamath  KMiller  SVRanade  SV Polymer technology for the controlled release of paclitaxel from a vascular compatible matrix: the TaxusTM Express paclitaxel-eluting coronary stent.  Paper presented at: Biomaterials in Regenerative Medicine: The Advent of Combination Products October 16-18, 2004 Philadelphia, Pa
26.
Ranade  SVMiller  KMRichard  RE  et al.  Physical characterization of controlled release of paclitaxel from the TAXUS Express2 drug-eluting stent. J Biomed Mater Res A 2004;71625- 634
PubMedArticle
27.
Pinchuk  L Biostable elastomeric polymers having quaternary carbons. US Patent 5 741 331 April21 1998;
28.
Stone  GWEllis  SGCox  DA  et al.  A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med 2004;350221- 231
PubMedArticle
29.
Stone  GWEllis  SGCox  DA  et al.  One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial. Circulation 2004;1091942- 1947
PubMedArticle
30.
Molteno  ACVan Rooyen  MMBartholomew  RS Implants for draining neovascular glaucoma. Br J Ophthalmol 1977;61120- 125
PubMedArticle
31.
Molteno  ACAncker  EVan Biljon  G Surgical technique for advanced juvenile glaucoma. Arch Ophthalmol 1984;10251- 57
PubMedArticle
32.
Da Mata  ABurk  SENetland  PA  et al.  Management of uveitic glaucoma with Ahmed glaucoma valve implantation. Ophthalmology 1999;1062168- 2172
PubMedArticle
33.
Molteno  ACSayawat  NHerbison  P Otago glaucoma surgery outcome study: long-term results of uveitis with secondary glaucoma drained by Molteno implants. Ophthalmology 2001;108605- 613
PubMedArticle
34.
Beebe  WEStarita  RJFellman  RL  et al.  The use of Molteno implant and anterior chamber tube shunt to encircling band for the treatment of glaucoma in keratoplasty patients. Ophthalmology 1990;971414- 1422
PubMedArticle
35.
Arroyave  CPScott  IUFantes  FE  et al.  Corneal graft survival and intraocular pressure control after penetrating keratoplasty and glaucoma drainage device implantation. Ophthalmology 2001;1081978- 1985
PubMedArticle
36.
Lee  RKFantes  F Surgical management of patients with combined glaucoma and corneal transplant surgery. Curr Opin Ophthalmol 2003;1495- 99
PubMedArticle
37.
Heuer  DKLloyd  MAAbrams  DA  et al.  Which is better? One or two? A randomized clinical trial of single-plate versus double-plate Molteno implantation for glaucomas in aphakia and pseudophakia. Ophthalmology 1992;991512- 1519
PubMedArticle
38.
Mills  RPReynolds  AEmond  MJ  et al.  Long-term survival of Molteno glaucoma drainage devices. Ophthalmology 1996;103299- 305
PubMedArticle
39.
Molteno  ACFucik  MDempster  AGBevin  TH Otago Glaucoma Surgery Outcome Study: factors controlling capsule fibrosis around Molteno implants with histopathological correlation. Ophthalmology 2003;1102198- 2206
PubMedArticle
40.
Geankoplis  CJ Transport Processes and Separation Process Principles (Includes Unit Operations). 3rd Upper Saddle River, NJ Prentice Hall1993;
41.
Acosta  ACFernandez  VLamar  PD  et al.  Ocular biocompatibility of QuatromertmTM (polystyrene-polyisobutylene triblock polymers) for glaucoma implants. Invest Ophthalmol Vis Sci 2004;45E-Abstract 2929
42.
Jacob  JTLacour  OJBurgoyne  CF Slow release of the antimetabolite 5-fluorouracil (5-FU) from modified Baerveldt glaucoma drains to prolong drain function. Biomaterials 2001;223329- 3335
PubMedArticle
43.
Morales  JKelleher  PJCampbell  DCrosson  CE Effects of daunomycin implants on filtering surgery outcomes in rabbits. Curr Eye Res 1998;17844- 850
PubMedArticle
×