Table 1. 
Detecting Angle Closure Using SPAC and SL-OCT (153 Eyes)
Detecting Angle Closure Using SPAC and SL-OCT (153 Eyes)
Table 2. 
Sensitivity, Specificity, and Area Under the Curve for SL-OCT and SPAC
Sensitivity, Specificity, and Area Under the Curve for SL-OCT and SPAC
Table 3. 
Quadrant-by-Quadrant Comparison of Gonioscopy and SL-OCT
Quadrant-by-Quadrant Comparison of Gonioscopy and SL-OCT
1.
Wong  TYLoon  SCSaw  SM The epidemiology of age related eye diseases in Asia. Br J Ophthalmol 2006;90 (4) 506- 511
PubMedArticle
2.
Foster  PJOen  FTMachin  D  et al.  The prevalence of glaucoma in Chinese residents of Singapore: a cross-sectional population survey of the Tanjong Pagar District. Arch Ophthalmol 2000;118 (8) 1105- 1111
PubMedArticle
3.
Foster  PJBaasanhu  JAlsbirk  PHMunkhbayar  DUranchimeg  DJohnson  GJ Glaucoma in Mongolia: a population-based survey in Hovsgol Province, Northern Mongolia. Arch Ophthalmol 1996;114 (10) 1235- 1241
PubMedArticle
4.
Dandona  LDandona  RMandal  P  et al.  Angle-closure glaucoma in an urban population in Southern India: the Andhra Pradesh Eye Disease Study. Ophthalmology 2000;107 (9) 1710- 1716
PubMedArticle
5.
Foster  PJJohnson  GJ Glaucoma in China: how big is the problem? Br J Ophthalmol 2001;85 (11) 1277- 1282
PubMedArticle
6.
Nolan  WPAung  TMachin  D  et al.  Detection of narrow angles and established angle closure in Chinese residents of Singapore: potential screening tests. Am J Ophthalmol 2006;141 (5) 896- 901
PubMedArticle
7.
Devereux  JGFoster  PJBaasanhu  J  et al.  Anterior chamber depth measurement as a screening tool for primary angle-closure glaucoma in an East Asian population. Arch Ophthalmol 2000;118 (2) 257- 263
PubMedArticle
8.
Kashiwagi  KTokunaga  TIwase  AYamamoto  TTsukahara  S Usefulness of peripheral anterior chamber depth assessment in glaucoma screening. Eye 2005;19 (9) 990- 994
PubMedArticle
9.
Foster  PJDevereux  JGAlsbirk  PH  et al.  Detection of gonioscopically occludable angles and primary angle closure glaucoma by estimation of limbal chamber depth in Asians: modified grading scheme. Br J Ophthalmol 2000;84 (2) 186- 192
PubMedArticle
10.
Kashiwagi  KKashiwagi  FToda  YOsada  KTsumura  TTsukahara  S A newly developed peripheral anterior chamber depth analysis system: principle, accuracy and reproducibility. Br J Ophthalmol 2004;88 (8) 1030- 1035
PubMedArticle
11.
Wirbelauer  CKarandish  AHaberle  HPham  DT Noncontact goniometry with optical coherence tomography. Arch Ophthalmol 2005;123 (2) 179- 185
PubMedArticle
12.
Müller  MDahmen  GPorksen  E  et al.  Anterior chamber angle measurement with optical coherence tomography: intraobserver and interobserver variability. J Cataract Refract Surg 2006;32 (11) 1803- 1808
PubMedArticle
13.
Radhakrishnan  SRollins  AMRoth  JE  et al.  Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol 2001;119 (8) 1179- 1185
PubMedArticle
14.
Radhakrishnan  SGoldsmith  JHuang  D  et al.  Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol 2005;123 (8) 1053- 1059
PubMedArticle
15.
Baskaran  MOen  FTChan  YH  et al.  Comparison of the scanning peripheral anterior chamber depth analyzer and the modified van Herick grading system in the assessment of angle closure. Ophthalmology 2007;114 (3) 501- 506
PubMedArticle
16.
Nolan  WPSee  JLChew  PT  et al.  Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology 2007;114 (1) 33- 39
PubMedArticle
17.
Sakata  LMLavanya  RFriedman  DS  et al.  Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology 2008;115 (5) 769- 774
PubMedArticle
18.
Scheie  HG Width and pigmentation of the angle of the anterior chamber: a system of grading by gonioscopy. Arch Ophthalmol 1957;58 (4) 510- 512
PubMedArticle
Clinical Sciences
May 11, 2009

Comparison of Slitlamp Optical Coherence Tomography and Scanning Peripheral Anterior Chamber Depth Analyzer to Evaluate Angle Closure in Asian Eyes

Author Affiliations

Author Affiliations: Tan Tock Seng Hospital (Dr Wong), Singapore National Eye Centre and Singapore Eye Research Institute (Drs Chua, M. H. Y. Wong, and T. Aung), and Yong Loo Lin School of Medicine, National University of Singapore (Dr T. Aung), Singapore.

Arch Ophthalmol. 2009;127(5):599-603. doi:10.1001/archophthalmol.2009.41
Abstract

Objective  To evaluate the effectiveness of slitlamp optical coherence tomography (SL-OCT) and Scanning Peripheral Anterior Chamber depth analyzer (SPAC) in detecting angle closure, using gonioscopy as the reference standard.

Methods  A total of 153 subjects underwent gonioscopy, SL-OCT, and SPAC. The anterior chamber angle (ACA) was classified as closed on gonioscopy if the posterior trabecular meshwork could not be seen; with SL-OCT, closure was determined by contact between the iris and angle wall anterior to the scleral spur; and with SPAC by a numerical grade of 5 or fewer and/or a categorical grade of suspect or potential.

Results  A closed ACA was identified in 51 eyes with gonioscopy, 86 eyes with SL-OCT, and 61 eyes with SPAC (gonioscopy vs SL-OCT, P < .001; gonioscopy vs SPAC, P = .10; SL-OCT vs SPAC, P < .001; McNemar test). Of the 51 eyes with a closed ACA on gonioscopy, SL-OCT detected a closed ACA in 43, whereas SPAC identified 41 (P = .79). An open angle in all 4 quadrants was observed in 102 eyes with gonioscopy, but SL-OCT and SPAC identified 43 and 20 of these eyes, respectively, as having angle closure. The overall sensitivity and specificity for SL-OCT were 84% and 58% vs 80% and 80% for SPAC.

Conclusion  Using gonioscopy as the reference, SL-OCT and SPAC showed good sensitivity for detecting eyes at risk of angle closure.

Glaucoma is an important cause of blindness and an increasing economic burden on health care systems worldwide. The prevalence of glaucoma across Asia has been estimated to range from 2% to 5% in persons 50 years or older.1 Primary angle-closure glaucoma is an important form of the disease and more visually debilitating than primary open-angle glaucoma in population surveys conducted in Singapore, Mongolia, India, and China.25 The identification of eyes at risk of angle closure may reduce the morbidity of this disease because early treatment may halt the process and, hence, reduce the risk of developing primary angle-closure glaucoma.

Previous studies have evaluated various screening methods for angle closure that assess the central and peripheral anterior chamber depth (ACD).69 Scanning peripheral ACD analyzer (SPAC) (model SM-70; Takagi Seiko Co, Nagano, Japan)10 and anterior segment optical coherence tomography (AS-OCT)1114 are new noncontact devices that enable evaluation of the peripheral ACD and anterior chamber angle (ACA), respectively. The SPAC grading correlates well with limbal ACD assessed using the modified van Herick system.15 In addition, AS-OCT allows dynamic imaging of the ACA, and previous studies have been performed to evaluate the use of the Visante-OCT (Carl Zeiss Meditec, Dublin, California) in detecting angle closure.16,17 Both SPAC and AS-OCT can be performed quickly and are easily tolerated by patients.

The aim of this study was to evaluate the effectiveness of a new AS-OCT method, slitlamp optical coherence tomography (SL-OCT) (Heidelberg Engineering, Heidelberg, Germany) in detecting angle closure, using gonioscopy as the reference standard. We also compared the performance of this instrument with SPAC.

METHODS

This prospective hospital-based comparative study was conducted with the approval of the ethics review board and performed in accordance with the tenets of the Declaration of Helsinki. Informed consent was obtained from every participant.

Consecutive patients who fulfilled study eligibility criteria were recruited from a glaucoma clinic at a Singapore hospital from January 1 to July 31, 2007. Patients who had undergone any prior intraocular procedures or had any penetrating eye injuries and corneal disorders, such as corneal endothelial dystrophy, pterygium, or a corneal scar, that may preclude satisfactory imaging were excluded from the study. Patients who had undergone laser iridotomy were not excluded. A comprehensive ophthalmic examination was performed on the day of recruitment and included visual acuity examination, slitlamp biomicroscopy, Goldmann applanation tonometry, gonioscopy, and fundus examination followed by imaging with the SPAC and SL-OCT.

GONIOSCOPY

Gonioscopy was performed in the dark by one of us with glaucoma subspecialty training (H.T.W.) who was masked to findings from SPAC and SL-OCT. Static and dynamic gonioscopy was performed using the Sussman 4-mirror lens under high magnification (×16) with the eye in primary gaze position. During gonioscopic examination, care was taken to avoid any light falling onto the pupil and any accidental indentation on the globe. The ACA in each quadrant was graded using the Scheie grading system, which is based on the anatomical structures observed during nonindentation gonioscopy (grade I, visible ciliary body; grade II, visible scleral spur; grade III, visible anterior trabecular meshwork; grade IV, angle structures not visible).18 The ACA was considered closed in that quadrant if the posterior trabecular meshwork could not be seen in the primary position without indentation (Scheie grade 3 or 4).

SCANNING PERIPHERAL ANTERIOR CHAMBER DEPTH ANALYZER

The SPAC imaging was performed by one of us (H.T.A.) who was masked to gonioscopic and clinical findings. Study participants were required to place their chin and forehead against the headrest of the slitlamp and remain fixated on the red light-emitting diode in front while the examiner aligned the reflection of the red light with the pupil center. In SPAC, optical methods are used to measure the peripheral ACD, and the ACD is scanned from the optical axis to the temporal limbus at a rate of 12 mm/s. The instrument acquired 21 images at 0.4-mm intervals with a magnification of 0.4. Three measurements were automatically performed at each point, and the mean of the ACD values were used for analysis. These measurements were compared with those in a normative database included with the device (derived from a sample of Japanese subjects), and SPAC software provides numerical and categorical grades to classify the ACD. The numerical scale ranged from 1 to 12, with 12 representing the deepest ACD. The categorical grading indicates the risk for angle closure: S, suspect angle closure; P, potential angle closure; and no suffix (for normal results).

SLITLAMP OPTICAL COHERENCE TOMOGRAPHY

The SL-OCT was performed by one of us (H.T.A.) who was masked to gonioscopic and clinical findings. All participants were required to place their chin and forehead against the headrest of the slitlamp while the beam was positioned onto the area of interest. Of note, a low-intensity narrow light beam from the slitlamp was kept on during SL-OCT scan acquisition to facilitate gross adjustment of the scanning beam location because that beam scans the eye's structures along the same line of the slitlamp light beam.

The SL-OCT is incorporated into a slitlamp microscope, with a scanner unit attached to the slitlamp's illumination system. Using coherence interferometry technology with a 1310-nm laser, the SL-OCT enables dynamic imaging of the anterior segment, with an image acquisition rate of 200 A-scans per second, using 670 points per A-scan, to a depth of 7 mm. The SL-OCT images are composed horizontally of 200 A-scans over 15 mm. These images are then resized to the actual scale, where the horizontal data are stretched to more than twice their size to represent the actual anterior chamber dimensions. The final SL-OCT image has a maximum transverse resolution of 75 μm and an axial resolution of 10 to 25 μm.

Image acquisition with the SL-OCT required imaging of the entire cross-section of the anterior segment in 1 single-image frame. Moving the upper and lower eyelids away was necessary to image the ACA at the 12- and 6-o’clock positions, and care was taken to avoid inadvertent pressure on the globe. Two scans of the ACA of each eye were obtained: one at the 3- and 9-o’clock meridian and another at the 12- and 6-o’clock meridian.

The ACA status in each of the 4 quadrants of the eye was assessed by one of us with glaucoma subspecialty training (L.M.S.) who was masked to other test results. The ACA was considered closed on SL-OCT imaging if there was contact between the iris and angle wall anterior to the scleral spur in that quadrant.

STATISTICAL ANALYSIS

If both eyes were eligible for inclusion, 1 eye from each patient was randomly selected for analysis. Parametric and nonparametric tests were used to compare continuous variables. The χ2 test was used to compare categorical data, whereas the McNemar test was used to compare differences in distribution of a categorical variable between 2 dependent samples.

For the purposes of this study, an eye was defined as having angle closure on gonioscopy or SL-OCT if 1 or more of the quadrants were closed. For SPAC, an eye was considered at risk of angle closure if the numerical grade was 5 or fewer and/or the categorical grade was S or P.15 Receiver operating characteristic curves as well as calculations of area under the curve and 95% confidence intervals were used as an index of each instrument's performance. Values for specificity and sensitivity were also calculated with gonioscopy as the reference standard.

P < .05 was considered statistically significant. Statistical analyses were performed using JMP version 5 (SAS Institute Inc, Cary, North Carolina) and MedCalc (Mariakerke, Belgium) software.

RESULTS

A total of 188 participants (188 eyes) were recruited for the study. Their mean (SD) age was 63.3 (10.5) years (range, 37-99 years), and 57.0% were women. Most were Chinese (162 [86.2%]) and the rest were Malay (8 [4.3%]), Indian (12 [6.4%]), and other races (6 [3.2%]). The mean (SD) Scheie grade was 1.92 (0.96).

Fourteen participants (7.5%) were excluded from the analysis because of failure in obtaining SL-OCT images the examiner was unable to move the eyelids out of the way in 12 [6.4%], and motion artifacts during imaging in 2 [1.1%] resulted in poor image quality). Another 21 participants (11.2%) were also excluded because their SL-OCT images could not be graded owing to poor definition of the scleral spur. Therefore, further analysis was only conducted for 153 participants (81.4%) with satisfactory SL-OCT and SPAC assessments. Of note, there were no significant differences in the mean age, sex, race distribution, and number of gonioscopically closed quadrants for excluded vs included cases (P = .44). Of 153 included participants, 102 (66.7%) had an open ACA in all 4 quadrants of the eye on gonioscopic examination, whereas 51 (33.3%), all of whom had previously undergone laser iridotomy, had at least 1 closed-quadrant ACA.

ANALYSIS BY EYE

Overall, angle closure in 1 or more quadrants was detected by gonioscopy in 51 eyes and by SL-OCT in 86 eyes; SPAC identified 61 eyes at risk of angle closure (numerical grade <5 and/or categorical grade S or P). Therefore, SL-OCT identified significantly more eyes with angle closure than gonioscopy or SPAC (gonioscopy vs SL-OCT, P < .001; SL-OCT vs SPAC, P < .001; gonioscopy vs SPAC, P = .10). Of 51 eyes identified by gonioscopy as having a closed ACA in at least 1 quadrant, SL-OCT identified 43 (84%) and SPAC identified 41 (80%) (P = .79) (Table 1).

Of 102 eyes with an open ACA in all 4 quadrants on gonioscopic evaluation, SL-OCT detected a closed ACA in at least 1 quadrant in 43 (42.2%), whereas SPAC identified 20 (19.6%) as at risk for angle closure. Table 2 shows the area under the receiver operating characteristic curve, sensitivity, and specificity values for SL-OCT and SPAC. Using the definition of angle closure as an eye with 1 quadrant of closed ACA on gonioscopic evaluation as the reference standard, the overall sensitivity and specificity for SL-OCT was 84% and 58% vs 80% and 80% for SPAC. With a definition of 2 quadrants of closed ACA, the overall sensitivity and specificity for SL-OCT was 65% and 72% vs 83% and 78% for SPAC.

ANALYSIS BY QUADRANT

Overall, SL-OCT detected more closed ACA quadrants than gonioscopy (P = .006) (Table 3), particularly in the superior, inferior, and temporal quadrants (superior and inferior quadrants, P < .001; temporal quadrant, P = .003). There was no difference in the proportion of closed nasal quadrants detected using gonioscopy vs SL-OCT.

Intraobserver agreement for qualitatively grading the SL-OCT images of 83 randomly selected eyes by the examiner (L.M.S.) was substantial (κ, 0.71 [95% confidence interval, 0.63-0.79]). No comparisons were made with the SPAC because it does not analyze the anterior segment in a quadrant-by-quadrant fashion.

COMMENT

To our knowledge, this is the first study to evaluate the use of SL-OCT for angle imaging and to compare this device with SPAC, another noncontact instrument for angle-closure detection. We found interesting differences between SL-OCT, SPAC, and gonioscopy in the assessment of angle closure. Both SL-OCT and SPAC detected most of the eyes with a closed ACA in at least 1 quadrant on gonioscopy. However, SL-OCT identified more eyes as having angle closure compared with gonioscopy and SPAC. In fact, SL-OCT detected a closed ACA in at least 1 quadrant in 30 eyes that were determined to be open by gonioscopy and SPAC. It is important to note that SL-OCT actually produces images of all 4 quadrants of the ACA, and these images must be interpreted by an examiner, whereas SPAC evaluates the peripheral ACD of the temporal quadrant only and uses a normative database to interpret the results of each examination.

Previous studies have shown that AS-OCT imaging performed with Visante-OCT also detected more closed ACAs than gonioscopy, particularly in the superior and inferior quadrants.17 The discrepancies between gonioscopy and AS-OCT may be owing to differences in the methods of assessing and interpreting the ACA configuration with each technique, including the influence of visible light on the pupil during gonioscopy (AS-OCT is performed in the dark), inadvertent indentation during gonioscopy, and the level of irido-angle contact required to define a closed ACA on AS-OCT imaging.16,17 Of interest, a low-intensity narrow slit of light from the slitlamp was kept on during image acquisition with SL-OCT to facilitate the positioning of the scan beam at the apex of the cornea, as suggested in the SL-OCT manual. In spite of this, SL-OCT imaging still detected more closed ACAs than gonioscopy in this study.

Compared with the SL-OCT, our study showed that the SPAC had a similar sensitivity but a higher specificity (80% vs 55%) in detecting eyes with angle closure using the definition of 1 quadrant of closed ACA on gonioscopy. When the definition was 2 quadrants of closed ACA, SPAC performed better than SL-OCT, with an overall sensitivity and specificity of 83% and 78% for SPAC compared with 65% and 72% for SL-OCT. However, SPAC diagnosed more eyes as at risk of angle closure relative to gonioscopy, although this difference was not statistically significant. A comparison between SPAC and the modified van Herick grading system relative to gonioscopy15 also found fairly similar sensitivity and specificity results with SPAC as in this study. This overestimation in detecting eyes at risk of angle closure with SPAC could be attributed to the influence of inadvertent indentation during gonioscopy and the use of a normative database derived from a different (Japanese) population.

Using gonioscopy as the reference standard, SL-OCT and SPAC showed reasonably good sensitivity in detecting eyes at risk for angle closure. However, the relatively low specificity values demonstrated by both devices, even when more stringent criteria were adopted, make them currently unsuitable for community-based screening. In general, a very high specificity would be desirable to reduce the number of false-positive cases and, thus, improve the cost-effectiveness of the screening method. Of interest, using a stricter definition of angle closure (ie, 2 quadrants of closed ACA on gonioscopy) did not markedly change the performance of SPAC, but the sensitivity and specificity of SL-OCT were worse.

Our study has some limitations. Almost one-third of participants were excluded, mainly because of the poor quality of their SL-OCT images. The use of a single examiner could result in a systematic bias for the gonioscopy findings. The sample was not population based but instead was drawn from patients at a glaucoma clinic. We did not investigate the performance of these instruments in participants with different stages of angle-closure disease (suspects, primary angle closure, and primary angle-closure glaucoma). The interobserver reproducibility of detecting a closed ACA in AS-OCT images was not assessed in this study. Future studies should evaluate the reproducibility of the qualitative assessment of SL-OCT images, including variations related to image acquisition and those related to the interobserver grading of the ACA.

In conclusion, there were differences in angle-closure detection between SL-OCT and SPAC compared with gonioscopy. In particular, more persons were found to have closed angles with SL-OCT than with gonioscopy. Although SPAC and SL-OCT devices have high sensitivity in detecting eyes with angle closure, these methods may have limited use in population-based screening programs because of their low specificity. Longitudinal prospective studies may help clarify the clinical value of each instrument's findings and determine whether eyes classified as closed by SL-OCT and SPAC, but not gonioscopy, are at risk of developing angle-closure glaucoma.

Back to top
Article Information

Corresponding Author: Tin Aung, FRCS(Ed), PhD, Glaucoma Department, Singapore National Eye Centre, 11 Third Hospital Ave, Singapore 168751 (tin11@pacific.net.sg).

Submitted for Publication: August 31, 2008; final revision received November 3, 2008; accepted November 24, 2008.

Financial Disclosure: Dr T. Aung has received grant funding as well as financial support and honoraria for travel to conferences from Carl Zeiss Meditec, the manufacturer of Visante-OCT.

Funding/Support: The study was supported by a grant from the National Medical Research Council, Singapore. The SL-OCT was loaned by Heidelberg Engineering for the study.

References
1.
Wong  TYLoon  SCSaw  SM The epidemiology of age related eye diseases in Asia. Br J Ophthalmol 2006;90 (4) 506- 511
PubMedArticle
2.
Foster  PJOen  FTMachin  D  et al.  The prevalence of glaucoma in Chinese residents of Singapore: a cross-sectional population survey of the Tanjong Pagar District. Arch Ophthalmol 2000;118 (8) 1105- 1111
PubMedArticle
3.
Foster  PJBaasanhu  JAlsbirk  PHMunkhbayar  DUranchimeg  DJohnson  GJ Glaucoma in Mongolia: a population-based survey in Hovsgol Province, Northern Mongolia. Arch Ophthalmol 1996;114 (10) 1235- 1241
PubMedArticle
4.
Dandona  LDandona  RMandal  P  et al.  Angle-closure glaucoma in an urban population in Southern India: the Andhra Pradesh Eye Disease Study. Ophthalmology 2000;107 (9) 1710- 1716
PubMedArticle
5.
Foster  PJJohnson  GJ Glaucoma in China: how big is the problem? Br J Ophthalmol 2001;85 (11) 1277- 1282
PubMedArticle
6.
Nolan  WPAung  TMachin  D  et al.  Detection of narrow angles and established angle closure in Chinese residents of Singapore: potential screening tests. Am J Ophthalmol 2006;141 (5) 896- 901
PubMedArticle
7.
Devereux  JGFoster  PJBaasanhu  J  et al.  Anterior chamber depth measurement as a screening tool for primary angle-closure glaucoma in an East Asian population. Arch Ophthalmol 2000;118 (2) 257- 263
PubMedArticle
8.
Kashiwagi  KTokunaga  TIwase  AYamamoto  TTsukahara  S Usefulness of peripheral anterior chamber depth assessment in glaucoma screening. Eye 2005;19 (9) 990- 994
PubMedArticle
9.
Foster  PJDevereux  JGAlsbirk  PH  et al.  Detection of gonioscopically occludable angles and primary angle closure glaucoma by estimation of limbal chamber depth in Asians: modified grading scheme. Br J Ophthalmol 2000;84 (2) 186- 192
PubMedArticle
10.
Kashiwagi  KKashiwagi  FToda  YOsada  KTsumura  TTsukahara  S A newly developed peripheral anterior chamber depth analysis system: principle, accuracy and reproducibility. Br J Ophthalmol 2004;88 (8) 1030- 1035
PubMedArticle
11.
Wirbelauer  CKarandish  AHaberle  HPham  DT Noncontact goniometry with optical coherence tomography. Arch Ophthalmol 2005;123 (2) 179- 185
PubMedArticle
12.
Müller  MDahmen  GPorksen  E  et al.  Anterior chamber angle measurement with optical coherence tomography: intraobserver and interobserver variability. J Cataract Refract Surg 2006;32 (11) 1803- 1808
PubMedArticle
13.
Radhakrishnan  SRollins  AMRoth  JE  et al.  Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol 2001;119 (8) 1179- 1185
PubMedArticle
14.
Radhakrishnan  SGoldsmith  JHuang  D  et al.  Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol 2005;123 (8) 1053- 1059
PubMedArticle
15.
Baskaran  MOen  FTChan  YH  et al.  Comparison of the scanning peripheral anterior chamber depth analyzer and the modified van Herick grading system in the assessment of angle closure. Ophthalmology 2007;114 (3) 501- 506
PubMedArticle
16.
Nolan  WPSee  JLChew  PT  et al.  Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology 2007;114 (1) 33- 39
PubMedArticle
17.
Sakata  LMLavanya  RFriedman  DS  et al.  Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology 2008;115 (5) 769- 774
PubMedArticle
18.
Scheie  HG Width and pigmentation of the angle of the anterior chamber: a system of grading by gonioscopy. Arch Ophthalmol 1957;58 (4) 510- 512
PubMedArticle
×