October 1971

Halberg TonometerAn Evaluation

Author Affiliations

Ghent, Belgium
From the Ophthalmological Clinic, University of Ghent, Ghent, Belgium.

Arch Ophthalmol. 1971;86(4):376-384. doi:10.1001/archopht.1971.01000010378004
Abstract

The Halberg tonometer, a portable, gravitational, optical, direct-reading applanation tonometer, based on the Maklakoff principle, has been evaluated. Comparative studies were performed with the Goldmann, Draeger, and Filatov-Kalfa instruments. The Halberg tonometer compared favorably to the Goldmann type and the Maklakoff type instruments, when both simple and differential tonometry were practiced. General principles of evaluation of a tonometer are discussed. The data obtained are subjected to mathematical treatment and statistical analysis. The study validates the Halberg tonometer as a reliable clinical instrument.

References
1.
Halberg GP:  Hand applanation tonometer . Trans Amer Acad Ophthal Otolaryng 72:112-114, 1968.
2.
Gloster J, Martin B:  Evaluation of Posner Inglima tonometer . Brit J Ophthal 49:617-625, 1965.Article
3.
Goldmann H, Schmidt T:  Ueber applanations tonometrie . Ophthalmologica 134:221-242, 1957.Article
4.
Goldmann H, Schmidt T:  Weiterer Beitrag zur Applanationstonometrie . Ophthalmologica 141:441-456, 1961.Article
5.
Gloster J, Perkins ES:  The validity of the Imbert-Fick law as applied to applanation tonometry . Exp Eye Res 2:274-283, 1963.Article
6.
Ourgaud AG, Etienne R: L'exploration fonctionnelle de l'oeil glaucomateux . Paris, Masson et Cie, 1961, vol 1.
7.
Dasvevskij A: Vestn Oftalm , 1941, p 531
8.
1944, pp 3-21
9.
1945, pp 14-21
10.
1949, pp 1-14
11.
1961, pp 5-10.
12.
Kruse W:  Eine neue Messmethode des Rigiditätskoeffizienten . Graefe Arch Klin Exp Ophthal 162:78-96, 1960.Article
13.
Moses RA, Tarkkanen A:  The pressure-volume relationship in the intact human eye at low pressures . Amer J Ophthal 47:557-564, 1959.
14.
Friedenwald JS:  Contribution to the theory and practice of tonometry: An analysis of the work of Professor Kalfa with the applanation tonometer . Amer J Ophthal 22:375-383, 1939.
15.
Goldmann H, Schmidt T:  Ueber applanations-tonometrie . Ophthalmologica 131:335-341, 1956.Article
16.
Quintana M:  La relation Pt-W (presión-peso) en tonometria . Rev Brasil Oftalm 25:515, 1966.
17.
Schargel K, Grom E, de Uzcategui CE:  Tonometrias por aplanacion con los tonometros de Halberg y Goldmann . Rev Oftal Ven 24:196-209, 1970.
×