Comparison of MacKay-Marg, Goldmann, and Perkins Tonometers in Abnormal Corneas

Author Affiliations

From the Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami (Fla) School of Medicine. Dr. McMillan is now located in Jackson, Miss.

Arch Ophthalmol. 1975;93(6):420-424. doi:10.1001/archopht.1975.01010020434007
Abstract

Intraocular pressure was determined manometrically in owl monkey eyes with normal and edematous corneas. The pressure was measured with Perkins hand-held, Goldmann, and MacKay-Marg tonometers to compare relative accuracy. At manometric levels of 20, 30, and 40 mm Hg, a statistically significant difference in readings was found between normal and edematous corneas with use of the Perkins and Goldmann applanators, whereas no such difference was found with the MacKay-Marg tonometer. The MacKay-Marg tonometer most closely approximated the true intraocular pressure in edematous corneas.

References
1.
Kaufman HE, Wind CA, Waltman SR:  Validity of MacKay-Marg electronic applanation tonometer in patients with scarred irregular corneas . Am J Ophthalmol 69:1003-1007, 1970.
2.
Wind CA, Kaufman HE:  Validity of MacKay-Marg applanation tonometry following penetrating keratoplasty in man . Am J Ophthalmol 72:117-118, 1971.
3.
Wind CA, Irvine AR:  Electronic applanation tonometry in corneal edema and keratoplasty . Invest Ophthalmol 8:620-624, 1969.
4.
Goldmann H:  Applanation tonometry , in Newell FW (ed): Second Josiah Macy Jr. Foundation Conference on Glaucoma, 1956 . New York, Josiah Macy Jr. Foundation, 1957, pp 186-187.
5.
Moses RA:  The Goldmann applanation tonometer . Am J Ophthalmol 46:865-869, 1958.
6.
Moses RA, Marg E, Oechsli R:  Evaluation of the basic validity and clinical usefulness of the MacKay-Marg tonometer . Invest Ophthalmol 1:78-85, 1962.
7.
Moses RA:  The MacKay-Marg tonometer . Trans Am Acad Ophthalmol Otolaryngol , 1962, pp 88-95
8.
Hilton GF, Shaffer RN:  Electronic applanation tonometry . Am J Ophthalmol 62:838-843, 1966.
9.
Tierney JP, Rubin ML:  A clinical evaluation of the electronic applanation tonometer . Am J Ophthalmol 62:263-272, 1966.
10.
Irvine AR, Kaufman HF:  Intraocular pressure following penetrating keratoplasty . Am J Ophthalmol 68:835-844, 1969.
11.
Mishima S, Hedbys BO:  Measurement of corneal thickness with the Haag-Streit pachometer : Arch Ophthalmol 80:710-713, 1968.Article
12.
Perkins ES:  Hand-held applanation tonometer . Br J Ophthalmol 49:591-593, 1965.Article
13.
Kaufman HE:  Pressure measurement: Which tonometer? Invest Ophthalmol 11:80-85, 1972.
×