Article
April 1978

Increased Corneal Thickness Simulating Elevated Intraocular Pressure

Author Affiliations

From the Department of Ophthalmology, Washington University School of Medicine, St Louis.

Arch Ophthalmol. 1978;96(4):664-665. doi:10.1001/archopht.1978.03910050360012
Abstract

• A 17-year-old girl had intraocular pressure readings of 30 to 40 mm Hg in both eyes, with normal visual fields and optic nerve heads. Medical treatment was unsuccessful in lowering the IOP measurements substantially. The central corneal thickness was 0.90 mm in each eye in the absence of corneal edema. Cannulation of the left anterior chamber revealed an IOP of 11 mm Hg, while the Perkins' and Schiötz' tonometers gave readings of 35 and 34 mm Hg, respectively. Previously, elevated tonometry readings had been obtained by Goldmann's, Perkins', Schiötz', MacKay and Marg's, pneumatic tonometer and air puff tonometer. This suggested that measurement of central corneal thickness be performed in cases in which IOP recordings do not correspond to other clinical findings.

References
1.
Goldmann H, Schmidt T:  Uber applanations tonomotrie . Ophthalmologica 134:221-242, 1957.Article
2.
Ehlers N, Bramsen T, Sperling S:  Applanation tonometry and central corneal thickness . Acta Ophthalmol 53:34-43, 1975.Article
3.
Hansen FK, Ehlers N:  Elevated tonometer readings caused by a thick cornea . Acta Ophthalmol 49:775-778, 1971.Article
4.
Hansen FK:  A clinical study of the normal human central corneal thickness . Acta Ophthalmol 49:82-89, 1971.
5.
Ehlers N, Hansen FK, Aasved H:  Biometric correlation of corneal thickness . Acta Ophthalmol 53:652-658, 1975.Article
6.
Ehlers N, Hansen FK:  Central corneal thickness in low tension glaucoma . Acta Ophthalmol 52:740-746, 1974.Article
×