Article
March 1979

The PneumatonographA Laboratory Study

Author Affiliations

From the Department of Ophthalmology and the Oscar Johnson Institute, Washington University School of Medicine, St Louis.

Arch Ophthalmol. 1979;97(3):547-552. doi:10.1001/archopht.1979.01020010291023
Abstract

• The scale reading of the pneumatonograph (PTG) is dependent on the force of application to the cornea. Since this force is relatively greater at low intraocular pressure, the scale is compressed, yielding an overestimate of intraocular pressure in physiologic range. When the PTG is held vertically, the weight of the plunger is added to the gas propulsion force and the compression of the scale is accentuated. When a 10-g plunger load is added, as is advocated for tonography, the scale compression is further accentuated.

We conclude that the PTG scale is compressed, the PTG is position-sensitive, and that the PTG is totally unsuited to tonography. The PTG is not an applanation tonometer and should not be confused with applanation tonometers.

References
1.
Langham ME:  Developments in the detection of glaucoma by tonometric procedures . Contacto Int Contact Lens J , vol 14, 1970.
2.
Langham ME, McCarthy E:  A rapid pneumatic applanation tonometer . Arch Ophthalmol 79:389-399, 1968.Article
3.
Langham ME:  Pneumatonography: A new methodological approach to the analysis of intraocular pressure and aqueous humor dynamics in human eyes , in Etienne R, Paterson GD (eds): International Glaucoma Symposium Albi (France) 1974 (held in conjunction with the 22d International Congress of Ophthalmology, Paris, 1974). Marseille, France, Diffusion Generale de Librairie, 1974, p 477.
4.
Walker RE, Litovitz TL:  An experimental and theoretical study of the pneumatic tonometer . Exp Eye Res 13:14-23, 1972.Article
5.
Walker RE, Litovitz TL, Langham ME:  Pneumatic applanation tonometer studies: II. Rabbit cornea data . Exp Eye Res 13:187-193, 1972.Article
6.
Tonjum AM:  A new sensor for pneumatic tonometry: Construction and working principle . Acta Ophthalmol 120( (suppl) ):12-19, 1973.
7.
Walker RE, Langham ME:  Pneumatic applanation tonometer studies: III. Analysis of the floating tip sensor . Exp Eye Res 20:167-172, 1975.Article
8.
Stepanik J:  The Mackay-Marg tonometer: Correlation of the tonogram to the corneal applanations induced by the tonometer . Acta Ophthalmol 48:1140-1144, 1970.Article
9.
Armaly MF:  Studies on intraocular effect of the orbital parasympathetic pathway: III. Effect on steady state dynamics . Arch Ophthalmol 62:817-827, 1959.Article
10.
Moses RA, Grodzki WJ Jr:  The Mackay-Marg tonometer: A note on calibration methods . Acta Ophthalmologica 49:800-804, 1971.Article
×