A New Test of Visual Acuity Using a Holographic Phase Grating and a Laser

Author Affiliations

From the Department of Ophthalmology, University of Iowa Hospitals and Clinics, Iowa City (Drs Smith and Kolder and Ms Snyder); Institute of Optics, University of Rochester, Rochester, NY (Dr Remijan); and Randwal Instrument and Machine, Southbridge, Mass (Mr Remijan).

Arch Ophthalmol. 1979;97(4):752-754. doi:10.1001/archopht.1979.01020010404027
Abstract

• A new visual acuity testing device is able to determine a specific visual acuity of the retina even in the presence of opacities of the ocular media, such as a cataract. This is achieved by forming two spatially coherent, quasimonochromatic point sources near the nodal point of the eye. The two spherical waves that emanate from the point sources interact to produce a pattern of stripes on the retina. The intensity of the pattern is easily varied. More importantly, the pattern can be projected over different-sized retinal areas corresponding to visual fields of 0.5 to 15° in diameter. The device may be useful for vision research and preoperative assessments of visual potential.

References
1.
Goldmann H:  Examination of the fundus of the cataractous eye . Am J Ophthalmol 73:309-320, 1972.
2.
Newman M:  Visual acuity , in Moses RA (ed): Adler's Physiology of the Eye , ed 6. St Louis, CV Mosby Co, 1975.
3.
Green DG:  Testing the vision of cataract patients by means of laser-generated interference fringes . Science 168:1240-1242, 1970.Article
4.
Cavonius CR, Hilz R:  A technique for testing visual function in the presence of opacities . Invest Ophthalmol 12:933-936, 1973.
5.
Lotmar W:  Use of Moiré fringes for testing visual acuity of the retina . Appl Optics 11:1266-1268, 1972.Article
×