Hypodermic Stainless Steel Tacks and Companion Inserter Designed for Peripheral Fixation of Retina

Author Affiliations

From the Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami School of Medicine (Drs O'Grady, Flynn, Olsen, Blankenship, and Clarkson and Mr Parel), and the Veterans Administration Medical Center, Miami (Drs O'Grady and Olsen and Mr Lee).

Arch Ophthalmol. 1988;106(2):271-275. doi:10.1001/archopht.1988.01060130285046
Abstract

• Mechanical fixation of the torn or cut edge of retina with 30-gauge hypodermic stainless steel tacks has been performed in 15 patients at the Bascom Palmer Eye Institute, Miami, and Miami Veterans Administration Medical Center with six to 18 months of follow-up. The companion tack inserter allows the surgeon to select an angle for the shaft of the tack in relation to the inserter to allow the tack to be driven perpendicularly into sclera even when the insertion is anterior to the equator. If residual detachment remains following fluid-gas exchange after preliminary placement of the retinal tacks, a given tack may be removed and reinserted to facilitate complete retinal reattachment. This utilitarian feature allows the surgeon to eliminate residual anteroposterior traction following complete membrane peeling by extending relaxing retinotomies and tacking the posterior cut edge of the retina securely between the ora serrata and the equator.

References
1.
Machemer R, McCuen B, deJuan E:  Relaxing retinotomies and retinectomies . Am J Ophthalmol 1986;102:7-12.
2.
Ando F, Kondo J:  A plastic tack for the treatment of retinal detachment with giant tear . Am J Ophthalmol 1983;95:260-261.Article
3.
DeJuan E, Hickingbotham D, Machemer R:  Retinal tacks . Am J Ophthalmol 1985;99:272-274.
4.
Abrams GW, Williams GA, Neuwirth J, et al:  Clinical results of titanium retinal tacks with pneumatic insertion . Am J Ophthalmol 1986;102:13-19.
5.
American National Standards Institute:  Specifications for stainless steel bars and wires for surgical implants , in the 1982 Annual Book of ASTM Standards: Magnetic Properties and Magnetic Materials; Metallic Materials for Thermostats and for Electrical Resistance, Heating, and Contacts; Temperature Measurement . Philadelphia, American Society for Testing and Materials (ASTM) publication (ASTM) F55-82, 1982.
6.
Smik JA:  Specifying metal tubing . Med Device Diagnost Industry 1985;7:70-73.
7.
Morton WG: Toxicology of the Eye . Spring-field, Ill, Charles C Thomas Publishers, 1984, p 941.
8.
Dollfus M-A, Borsotti I:  Tolerance des aciers inoxydables dans l'oeil du lapin . Bull Soc Ophthalmol Fr 1983;50:422-423.
9.
Gronvall H, Hultgren A:  Ocular splinters of steel: Some clinical and metallographic data . Acta Ophthalmol 1982;37:413-466.Article
10.
Synderman MH:  Mechanical and thermal strengthening of hypodermic needle tubing . Medical Device & Diagnostic Industry 1985;7:56-59.
11.
Loertscher H:  Laser-induced breakdown for ophthalmic applications , in Trokel SL (ed): YAG Laser Ophthalmic Microsurgery . Norwalk, Conn, Appleton-Century-Crofts, 1983, pp 45-49.
12.
Flynn HW Jr, Blumenkranz MS, Parel J-M, et al:  Cannulated subretinal fluid aspirator for vitreoretinal microsurgery . Am J Ophthalmol 1987;103:106-108.
13.
The Retina Society Terminology Committee:  The classification of retinal detachment with proliferative vitreoretinopathy . Ophthalmology 1983;90:121-125.
14.
Chang S, Coleman DJ, Lincoff H, et al:  Perfluoropropane gas in the management of proliferative vitreoretinopathy . Am J Ophthalmol 1984;98:180-188.
15.
Parel J-M, Machemer R, O'Grady G, et al:  Intraocular diathermy coagulation . Graefes Arch Clin Exp Ophthalmol 1983;221:31-34.
16.
Schlegel HJ:  Ein einfache Weitwinkeloptik zur spaltlampen mikroscopischen Untersuchung der Augenhintergrundes . Doc Ophthalmol 1969;26:300-308.
17.
Algvere P, Stenkula S, Crafoord S, et al:  Sealing of retinal breaks with metal tacks: Evaluation of a new procedure in retinal re-attachment surgery . Acta Ophthalmol 1986;64:421-424.
18.
DeJuan E, McCuen B, Machemer R:  The use of retinal tacks in the repair of complicated retinal detachment . Am J Ophthalmol 1986;102:20-24.
19.
Burke JM, McDonald HR, Neuwirth J, et al:  Titanium retinal tacks with pneumatic insertion: Histologic evaluation in rabbits . Arch Ophthalmol 1986;105:404-408.
20.