Figure 1.
Schematic representation of odor identification and analysis. A, The headspace over a liquid sample is aspirated into an analyzer (electronic nose). B, Headspace gas (containing the sample aroma) is allowed to interact with an array of 32 conducting polymers with differing sensitivities to specific chemical types (eg, alcohols, ketones). C, The electrical resistance of each of the conducting polymers changes reproducibly after exposure to an aroma, allowing the aroma to be represented as a point in a 32-dimensional space. D, Principle component analysis (PCA) is used to compare individual odors on a 2-dimensional map. CSF indicates cerebrospinal fluid; Std, standard (reference sample [water]).

Schematic representation of odor identification and analysis. A, The headspace over a liquid sample is aspirated into an analyzer (electronic nose). B, Headspace gas (containing the sample aroma) is allowed to interact with an array of 32 conducting polymers with differing sensitivities to specific chemical types (eg, alcohols, ketones). C, The electrical resistance of each of the conducting polymers changes reproducibly after exposure to an aroma, allowing the aroma to be represented as a point in a 32-dimensional space. D, Principle component analysis (PCA) is used to compare individual odors on a 2-dimensional map. CSF indicates cerebrospinal fluid; Std, standard (reference sample [water]).

Figure 2.
First set of data, run manually. Each symbol corresponds to an individual patient. Ovals encircle cerebrospinal fluid (CSF) samples and serum samples as labeled. PCA indicates principle component analysis.

First set of data, run manually. Each symbol corresponds to an individual patient. Ovals encircle cerebrospinal fluid (CSF) samples and serum samples as labeled. PCA indicates principle component analysis.

Figure 3.
Second set of data, run on an automated device. The code for each patient's data points is in the accompanying chart. Cerebrospinal fluid (CSF) and serum samples are identified respectively with triangles and squares. Water (diamonds) is used as a control. PCA indicates principle component analysis.

Second set of data, run on an automated device. The code for each patient's data points is in the accompanying chart. Cerebrospinal fluid (CSF) and serum samples are identified respectively with triangles and squares. Water (diamonds) is used as a control. PCA indicates principle component analysis.

Table 1. 
Cerebrospinal Fluid (CSF) and Serum Data for Patients in Figure 2 (Manual)
Cerebrospinal Fluid (CSF) and Serum Data for Patients in Figure 2 (Manual)
Table 2. 
Cerebrospinal Fluid (CSF) and Serum Data for Patients in Figure 3 (Automated)
Cerebrospinal Fluid (CSF) and Serum Data for Patients in Figure 3 (Automated)
1.
Hanson  CWSteinberger  HA The use of a novel electronic nose to diagnose the presence of intrapulmonary infection. Anesthesiology. 1997;87A269Article
2.
Skedros  DGCass  SPHirsch  BEKelly  RH Sources of error in use of beta-2 transferrin analysis for diagnosing perilymphatic and cerebral spinal fluid leaks. Otolaryngol Head Neck Surg. 1993;109861- 864
Original Article
January 2000

Use of an Electronic Nose to Distinguish Cerebrospinal Fluid From Serum

Author Affiliations

From the Departments of Otorhinolaryngology (Drs Thaler and Kennedy and Ms Bruney) and Anesthesia (Dr Hanson), University of Pennsylvania, Philadelphia.

Arch Otolaryngol Head Neck Surg. 2000;126(1):71-74. doi:10.1001/archotol.126.1.71
Abstract

Background  Efforts to mimic the biologic olfactory system have resulted in the development of an electronic nose, whereby volatile gases may be identified by means of organic semiconductors. Such devices have been used in the food and beverage industry for quality-control purposes, but to date have not been used in the field of medicine.

Objective  To present the application of an electronic nose for clinical decision making by assessing the ability of an electronic nose to distinguish cerebrospinal fluid (CSF) from serum.

Design  Randomized, prospective, masked study.

Subjects  Nineteen matched sets of CSF and serum from inpatients at a university hospital.

Results  The electronic nose was able to distinguish CSF from serum in 18 of 19 patients. The data points for 18 of 19 CSF and 18 of 19 serum samples were within statistically distinct cluster groups, suggesting that the device is able to identify an unknown sample as CSF or serum.

Conclusions  This new technology is able to distinguish CSF from serum with a high degree of accuracy and speed, and with small sample quantity, potentially allowing the physician to identify reliably CSF otorrhea or rhinorrhea. This revolutionary diagnostic approach may have further, widespread application in the field of otorhinolaryngology and in medicine as a whole.

THE HUMAN sense of smell has been little utilized in modern medicine because of its lack of precision. However, other species such as dogs, with highly adapted olfactory systems, have legendary abilities to distinguish odor, and have even been used for diagnostic purposes in medicine. Efforts to artificially reproduce biologic olfactory systems have resulted in several technologies that are capable of analyzing odor with remarkable precision.

Several technologies are used in electronic noses. The first kind of sensors that were developed were the metal-oxide semiconductors, which measure the change in conductivity at a catalytic surface when gas molecules decompose at high temperatures. There are several types of mass-dependent electronic noses, such as quartz-crystal microbalances and surface acoustic wave devices, which measure a change in sensor mass as gas molecules are adsorbed. Optical-based sensors, eg, glass fibers coated with solvatochromic dyes, measure the change in fluorescence as gas molecules interact with the dye. In other sensors, conductivity changes as a gas diffuses into a polymer. Conducting polymers are organic semiconductors with a backbone in which single and double bonds alternate, giving rise to a large delocalized electron cloud. The polymers interact with volatile molecules (in odors) by changing conductance. Different polymers can be designed to respond preferentially to specific stereochemical characteristics of the odorant (eg, alcohols, ketones). Neural networks and principle component analysis are commonly used to identify and distinguish odors from one another with each of these technologies.

Electronic nose technology has been used in the food and beverage industries, to identify spoilage, and by NASA (National Aeronautics and Space Administration), to identify noxious gases. One application in the field of medicine has been to distinguish patients with pneumonia from control subjects by analyzing exhaled gases.1

The differentiation of cerebrospinal fluid (CSF) from mucous or serum is of critical importance in managing patients with potential CSF otorrhea or rhinorrhea. The most widely used method of distinguishing CSF from mucous or serum is the β2 transferrin assay. While a positive assay reliably identifies CSF, results are not available immediately (in many institutions taking longer than 24 hours), and may involve delays in management or unnecessary treatments.2 In addition, a reliable assay requires 0.5 mL of fluid, an amount that may not be readily attainable. This study was designed to determine whether the electronic nose technology could be used to correctly differentiate CSF from serum, rapidly and with small amounts of fluid.

MATERIALS AND METHODS

After obtaining institutional review board approval, matched sets of CSF and serum were obtained from the Pepper Laboratory of the University of Pennsylvania Medical Center in Philadelphia. The study was prospective, randomized, and masked in that no information regarding the patients was known prior to analyzing the samples, and samples were obtained based only on availability of matched sets on several different test days. The samples used were excess fluid to be otherwise discarded after all necessary clinical testing had been accomplished.

Samples of CSF and serum of 0.1 to 0.2 mL each were vaporized and analyzed by an organic semiconductor-based electronic nose (Osmetech, Hollis, NH). This device consists of a 32-element array of conducting polymers with differing responsiveness to various chemical species. Odor analysis was performed by passing the vaporized gas sample over the conductor array (Figure 1). The response of the sensor array to a sample gas is represented as a point in a 32-dimensional space (the relative amplitude of response of each of the 32 sensors to a sample odor). Principle component analysis was then used to depict odors on a 2-dimensional graph. Statistical analysis was assessed using the Mahalanobis distance, a measure of class separation between different cluster centroids. A value greater than 8 indicates distinct classes.

RESULTS

A total of 19 matched sets of CSF and serum were analyzed by the electronic nose. Results were available in 90 seconds. The specimens were run in 2 separate batches. The first, with 6 samples, was run manually, and the second, with 13 samples, was run on an automated version of the device once it became clear after the first run that a difference between substances was identifiable (Figure 2 and Figure 3). Eighteen of 19 patients' CSF specimens were graphically distinct and separate from their serum specimens. Furthermore, 18 of 19 CSF specimens and 18 of 19 serum specimens mapped out graphically in clusters, with CSF and serum reliably falling in distinct locations. The Mahalanobis distance for the first batch was 7.5, due to the one overlapping sample. Excluding that data point, the Mahalanobis distance was 20. For the second batch, this distance was 16. These results indicate statistical significance.

Laboratory data available for patients' CSF and serum samples were then retrospectively reviewed, to determine if there were any distinct features discernible through their routine laboratory testing (Table 1 and Table 2). No trends were identified, including in the 1 patient for whom the 2 specimens did not appear to fall within the range of the others.

COMMENT

These early data demonstrate that an electronic nose is able to distinguish CSF from serum rapidly, reliably, and with small sample quantity. Such a testing device, made clinically available, could provide useful real-time analysis for medical decision making in patients with suspected CSF rhinorrhea or otorrhea. The small amount of material necessary for testing and the speed with which it is accomplished represent improvements over the current standard of care, a test for presence of β2 transferrin.

The one patient whose specimens fell outside the representative data clusters for both serum and CSF did not have any identified abnormalities on available testing. There are several possible explanations for this outcome. It is possible that the specimens were contaminated or "overwhelmed" by a stronger scent than those typically picked up by the electronic nose. A systemic disorder productive of a high level of nitrogen might be one such explanatory overwhelming scent. It is also possible that the aberrance simply represents a technical error of the device. More data must be accumulated before this can be interpreted further.

CONCLUSIONS

The electronic nose appears to be able to distinguish CSF from serum rapidly and with small sample quantity. This device may have potential clinical use in the management of patients with suspected CSF rhinorrhea. The current technology simply requires collection of a 0.1- to 0.2-mL aliquot of suspected rhinorrhea as well as a control serum sample. These samples are analyzed with the electronic nose and compared with previous cluster groups for purposes of identification. The device theoretically has the capability of identifying samples of smaller size, collected on swabs or pledgets, although these sorts of specimens have not yet been studied. Further work is in progress to assess the reliability of the device using these other techniques.

The device has the potential for further applications in otorhinolaryngology. These include intraoperative identification of perilymph in exploring for oval or round window fistula, rapid identification of CSF leak during skull base procedures, and following posttraumatic CSF leaks as they heal without operative intervention. The device may also be able to identify the presence of bacteria in a specimen, thereby permitting real-time evaluation for the presence of ongoing infection.

Back to top
Article Information

Accepted for publication November 3, 1999.

Reprints: Erica R. Thaler, MD, Department of Otorhinolaryngology, 5 Silverstein, 3400 Spruce St, Philadelphia, PA 19104 (e-mail: thaler@mail.med.upenn.edu).

References
1.
Hanson  CWSteinberger  HA The use of a novel electronic nose to diagnose the presence of intrapulmonary infection. Anesthesiology. 1997;87A269Article
2.
Skedros  DGCass  SPHirsch  BEKelly  RH Sources of error in use of beta-2 transferrin analysis for diagnosing perilymphatic and cerebral spinal fluid leaks. Otolaryngol Head Neck Surg. 1993;109861- 864
×