Figure 1.
Noninvasive Vogele-Bale-Hohner head holder with the individual dental cast, headrest, and adjustable fixation elements in place. The arm-based Viewing Wand is also set up.

Noninvasive Vogele-Bale-Hohner head holder with the individual dental cast, headrest, and adjustable fixation elements in place. The arm-based Viewing Wand is also set up.

Figure 2.
Sandstrøm headset in place. Note the fixation of the device along the nasion and the outer ear canal and the counterfixation at the crown. The infrared emitter is mounted to the holder; the patient is wearing a Vogele-Bale-Hohner registration element.

Sandstrøm headset in place. Note the fixation of the device along the nasion and the outer ear canal and the counterfixation at the crown. The infrared emitter is mounted to the holder; the patient is wearing a Vogele-Bale-Hohner registration element.

Figure 3.
Endonasal 3-dimensional–assisted opening of a pyocele of the left sphenoid sinus. Top, Macroscopic view. Bottom, A light-emitting diode stylus is inserted into the nose and into the opened sphenoid sinus under videoendoscopic control.

Endonasal 3-dimensional–assisted opening of a pyocele of the left sphenoid sinus. Top, Macroscopic view. Bottom, A light-emitting diode stylus is inserted into the nose and into the opened sphenoid sinus under videoendoscopic control.

Figure 4.
Microscopic view of an ARTMA videoendoscopic procedure: the patient is wearing the registration element and is tracked, as is the suction-irrigation endoscope. Note that the sensor is mounted on top of the hand piece.

Microscopic view of an ARTMA videoendoscopic procedure: the patient is wearing the registration element and is tracked, as is the suction-irrigation endoscope. Note that the sensor is mounted on top of the hand piece.

Figure 5.
A 15-year-old girl with a teratoma of the retromaxillary space. A screen shot of the crosshairs indicates the stylus position inside the tumor. If the line of sight from the light-emitting diode (LED) to the camera is obstructed, as in this case, a warning message appears.

A 15-year-old girl with a teratoma of the retromaxillary space. A screen shot of the crosshairs indicates the stylus position inside the tumor. If the line of sight from the light-emitting diode (LED) to the camera is obstructed, as in this case, a warning message appears.

Figure 6.
Endonasal endoscopic decompression of the orbit and the optic nerve with the ARTMA System. The graphic structure floating over the video and the computed tomographic scan indicates the path to the orbit and the optic nerve.

Endonasal endoscopic decompression of the orbit and the optic nerve with the ARTMA System. The graphic structure floating over the video and the computed tomographic scan indicates the path to the orbit and the optic nerve.

1.
Schlöndorff  GMösges  RMeyer-Ebrecht  DKrybus  WAdams  L CAS (computer assisted surgery: a new procedure in head and neck surgery [in German]. HNO. 1989;37187- 190
2.
Gunkel  ARFreysinger  WThumfart  WF Computer-assisted surgery in the frontal and maxillary sinus. Laryngoscope. 1997;107631- 633Article
3.
Klimek  LMösges  R Computer-assisted surgery in the ENT specialty: developments and experiences from the first decade [in German]. Laryngorhinootologie. 1998;77275- 282Article
4.
Gunkel  ARFreysinger  WMartin  A  et al.  Three-dimensional image-guided endonasal surgery with a microdebrider. Laryngoscope. 1997;107834- 838Article
5.
Anon  JBLipman  SPOppenheim  D  et al.  Computer-assisted endoscopic sinus surgery. Laryngoscope. 1994;104901- 905Article
6.
Freysinger  WGunkel  ARBale  R  et al.  Three-dimensional navigation in otorhinolaryngological surgery with the Viewing Wand. Ann Otol Rhinol Laryngol. 1998;107953- 958
7.
Bale  RJVogele  MFreysinger  W  et al.  Minimally invasive head holder to improve the performance of frameless stereotactic surgery. Laryngoscope. 1997;107373- 377Article
8.
Gunkel  AR Computer-assisted 3D-brachytherapy in metastases of the head and neck region: International Symposium on Metastases in Head and Neck Cancer, Kiel, Germany, 15-18 January 1998. Br J Cancer. 1998;77 (suppl 1) 35Article
9.
Martin  ABale  RJVogele  MGunkel  ARThumfart  WFFreysinger  W Vogele-Bale-Hohner mouthpiece: registration device for frameless stereotactic surgery. Radiology. 1998;208261- 265
10.
Auer  THensler  EEichberger  P  et al.  3D-navigation for interstitial stereotaxic brachytherapy. Strahlenther Onkol. 1998;17482- 87Article
11.
Freysinger  WGunkel  ARThumfart  WF Image-guided endoscopic ENT surgery. Eur Arch Otorhinolaryngol. 1997;254343- 346Article
12.
Laitinen  LV Noninvasive multipurpose stereoadapter. Neurol Res. 1987;9137- 141
13.
Sandeman  DRPatel  NChandler  C  et al.  Advances in image-directed neurosurgery: preliminary experience with the ISG Viewing Wand compared with the Leksell G frame. Br J Neurosurg. 1994;8529- 544Article
14.
Drake  JMRutka  JTHofmann  HJ ISG Viewing Wand system. Neurosurgery. 1994;341094- 1097Article
15.
Freysinger  WGunkel  ARMartin  ABale  RJVogele  MThumfart  WF Advancing ear, nose, and throat computer-assisted surgery with the arm-based ISG Viewing Wand: the stereotactic suction tube. Laryngoscope. 1997;107690- 693Article
16.
Gunkel  ARFreysinger  WThumfart  WFPototschnig  C Complete sphenoethmoidectomy and computer-assisted surgery. Acta Otorhinolaryngol Belg. 1995;49257- 261
17.
Thumfart  WFGunkel  AR The latest developments in intraoperative 3D-navigation in the ENT area [in German]. Laryngorhinootologie. 1997;76700- 703Article
18.
Gunkel  ARVogele  MFreysinger  WMartin  ABale  RJThumfart  WF Computer aided surgery in the petrous bone. Laryngoscope. 1999;1091793- 1799Article
19.
Truppe  MJFreysinger  WGunkel  ARThumfart  WF Teleassisted stereotactic endoscopic surgery. Lemke HUVannier MWInamura KFarman AGeds.Proceedings of the CAR ‘96 10th International Symposium on Computer and Communication Systems for Image Guided Diagnosis and Therapy, Paris, June 1996. Amsterdam, the Netherlands Elsevier Science BV1996;686- 692
20.
Gunkel  ARFreysinger  WThumfart  WF Practical experience with the ARTMA Virtual Patient System in ENT surgery. Holl WWurm Geds. Proceedings of the 1st International Congress on Computer Integrated Surgery in the Areas of Head and Spine 1-5, September 1997 Linz, AustriaComput Aided Surg. 1997; (suppl) S06- S33
21.
Ossoff  RHReinisch  L Computer assisted surgical techniques: a vision for the future of otolaryngonology–head and neck surgery. J Otolaryngol. 1994;23354- 359
22.
Korves  BKlimek  LMösges  R Surgical decompression in endocrine orbitopathy: a three dimensional localizing device ensures greater safety. ORL J Otorhinolaryngol Relat Spec. 1996;5846- 50Article
23.
Zinreich  SJTebo  SLong  DL  et al.  Frameless stereotaxic integration of CT imaging data: accuracy and initial applications. Radiology. 1993;188735- 742
Original Article
March 2000

Experience With Various 3-Dimensional Navigation Systems in Head and Neck Surgery

Author Affiliations

From the Ear, Nose and Throat Department, University of Innsbruck, Innsbruck, Austria.

Arch Otolaryngol Head Neck Surg. 2000;126(3):390-395. doi:10.1001/archotol.126.3.390
Abstract

Objective  To evaluate the benefits and difficulties encountered when using various 3-dimensional (3-D) navigation systems in head and neck procedures.

Design  Five different navigation systems were used for preoperative planning and intraoperative 3-D navigation in procedures at the paranasal sinuses, the frontal and lateral skull bases, and the petrous bone.

Intervention  Intraoperative 3-D localizing systems (position-sensitive mechanical arms, infrared cameras, etc) demand reliable patient fixation on the operating table. We achieved this by developing a noninvasive head holder. Other systems allow patient movements by using magnetic digitizing technology (ARTMA System) and sophisticated programming.

Results  Having surpassed an initial learning curve, we now achieve an accuracy of 1 to 2 mm regularly. Especially in paranasal and frontal basal surgery, all navigation systems used provide valuable positioning information during surgery. In particular for revision or tumor surgery, decisive benefits resulted from use of these systems: shorter overall operation time; safer manipulation near delicate structures; and reliable identification of the skull base even in patients with bleeding, scarring, or missing anatomical landmarks.

Conclusions  We performed approximately 250 operations with different systems and introduced navigation at the lateral skull base and the petrous bone with mechanical, optic, and magnetic digitizers. In these anatomical areas, navigation was used successfully; the technical challenge is greatest at the lateral skull base, however.

AMONG OTHERS, Schlöndorff et al1 introduced 3-dimensional (3-D) navigation to otolaryngology in 1987 with a position-sensitive articulated arm. A successor thereof was the ISG Viewing Wand, which was for sometime considered the criterion standard in 3-D navigation. After initial scepticism of otolaryngologists, 3-D navigation emerged as an accepted and useful technique in nasal, paranasal, and skull base surgery. Minimally invasive surgery has gained enormous popularity because delicate structures at risk can be identified safely even when reliable landmarks are missing (tumors, revision, bleeding, etc).15 We have gained experience with different 3-D navigation systems in approximately 250 operations mainly at the paranasal sinuses and the frontal and lateral skull bases. We report some aspects of our work with 3-D navigation systems, emphasizing those that are crucial to obtaining optimal intraoperative results.

MATERIALS AND METHODS
IMAGING

We use 3-D computed tomographic and magnetic resonance (CT/MR) imaging as described by Freysinger et al.6

PATIENT FIXATION

Various options are available for intraoperative patient fixation to the operating table for navigation: tape fixation, the Vogele-Bale-Hohner (VBH) head holder (Medical Intelligence, Schäwbmünchen, Germany), and the Sandstrøm headset (Carl Zeiss, Oberkochen, Germany). The most important factor affecting patient-image registration is reliable patient fixation that still provides full access to the operating field and is noninvasive. Classic stereotactic frames are not desirable, and therefore we developed a noninvasive head holder, the VBH head holder.7 It consists of a headrest, hydraulic fixation elements, and an individual upper dental cast. The patient rests on a head plate and the dental cast is inserted, which is secured by underpressure and mounted to the fixation elements. The mouthpiece can carry registration elements. A counterfixation at the patient's occiput completes immobilization: it is firm and reliable but not invasive. Its variability supports all operations8,9 as well as computer-assisted interstitial brachytherapy.10 Preoperative 3-D imaging is done with the dental cast registration rods assembly in place, permitting creation of a "virtual patient" for preoperative planning and simulation.11Figure 1 illustrates the VBH head holder.

THE SANDSTRØM HEAD FIXATION DEVICE

Another option for patient fixation is the Sandstrøm head fixation device. This is also noninvasive and consists of an adjustable U-shaped carbon fixation element running across the head; its ends fit into the outer ear canals with plastic "olives." A V-shaped aluminum alloy element locks into the first element and provides counterfixation by resting on the nasion.12 The patient is fixated by tightening screws on both sides of the frame and locking the holder to the headrest plate (Figure 2). Counterfixation is provided by a strap running across the back of the head.

PATIENT-IMAGE REGISTRATION

A set of well-defined landmarks, either anatomical or fiducial, is matched with identical points in the images. Accuracy is affected by a variety of factors tentatively known as theoretical resolution of the 3-D digitizer; actual touching and correlating of landmarks with the stylus; skin shift; motion artifacts during imaging; movement of fiducial markers, the patient, or the operating table; and combinations thereof.

3-D NAVIGATION SYSTEMS
ISG Viewing Wand

The Viewing Wand (ISG Technologies, Mississauga, Ontario) mainly consists of a position-sensitive mechanical measuring arm with a localizing stylus of varying shapes at the end. The arm has 6 degrees of freedom and can be moved inside the operating field. The joints' positions are converted to position and orientation in space, which are shown as crosshairs in the preoperative CT/MR images. This allows searching of preplanned paths or identification of anatomical structures during surgery.13,14 To improve functionality we added tools,4,15 thus providing straight and bent probes, suction devices, power instrumentation, needles, cannulae, etc. Other than for ear, nose, and throat surgery, we have successfully used this system for interstitial brachytherapy to place hollow radiation in optimal preplanned positions inside inoperable tumor masses of the head and neck region.10

Philips Easy Guide, Zeiss STN, and ISG/ELEKTA Free-hand

These systems (Philips Easy Guide [Philips, Eindhoven, the Netherlands], Zeiss STN [Carl Zeiss], and ISG/ELEKTA free-hand [ELEKTA, Stockholm, Sweden]) use pointers that are equipped with infrared light-emitting diodes. The position and orientation of the light-emitting diodes are detected with 2 or 3 infrared cameras close to the operative field, and the position of the probe is indicated as crosshairs in the 3-D data sets. These systems require stable intraoperative patient fixation, and the registration procedures are nearly identical. We have used these systems in procedures at the paranasal sinuses, the frontal skull base, and the petrous bone (Figure 3).

ARTMA System

This virtual patient system (ARTMA, Vienna, Austria) is considerably different from the latter, apart from the fact that it uses magnetic digitizing. Sensors fixed to the instrument, the endoscope (microscope), and the patient provide 3-D data for navigation. In addition—and this is the unique feature of this navigation system—the positions, access paths, and additional graphical structures can be visualized on the live video. This allows us to directly superimpose 3-D information into the video sequence. For example, a predefined path can be shown by a sequence of colored frames, similar to a tunnel, floating as permanent information over the video and the data set at the same time, guiding the surgeon to the planned target. To that end, digital photographs of the patient (lateral and frontal views), the medical image data sets (2-dimensional and 3-D) are registered to the patient. The optical function of the endoscope (microscope) has to be determined, and thereafter navigation is possible: pathways are faded into the live video, and navigation in CT/MR is possible.16 The intraoperative setup with the tracked endoscope and the patient is shown in Figure 4.

Zeiss MKM System

The Zeiss MKM system (Carl Zeiss) is mainly designed for neurosurgical and otologic operations. The microscope is mounted on a high-precision robotic arm. The patient is rigidly mounted to the operating table, where the reference emitter of the digitizer is mounted too. This emitter is referenced with the robot. The position of the focal point inside the patient can be displayed relative to the patient's 3-D image data sets after a standard patient-image registration procedure. The Zeiss MKM system can superimpose contours, targets, or diagnostic images to the microscopic view, showing the actual position of the autofocal point inside the patient.

RESULTS AND COMMENT
VBH AND SANDSTRØM HEAD HOLDER

The cardinal factor for image-guided surgery is reliable intraoperative patient fixation in those systems requiring it (eg, ISG Viewing Wand, ISG/ELEKTA free-hand, Zeiss STN, Zeiss MKM, and Philips Easy Guide). The VBH head holder has provided superior patient fixation and flexibility in ear, nose, and throat surgery and for computer-assisted interstitial brachytherapy. The Sandstrøm head fixation device12 could not achieve the same accuracy in terms of repositioning; its construction is inappropriate for external approaches to the paranasal sinuses and the frontal and lateral skull bases, and the fixation can lead to pressure ulceration in the region of the outer ear or the nasion and cause considerable pain. Therefore, we use the VBH head holder almost exclusively.

The ISG Viewing Wand has given us reliable results for a long time. However, technological advances have provided other newer 3-D digitizing systems that are used. The arm must be mounted properly to the operating table to provide access to the whole surgical field. In addition, "working around corners," which is frequent in sinus surgery, is awkward.17 We have reached an accuracy of less than 2 mm regularly in all operations, and the intraoperative preparation and registration time—including patient fixation—has been shortened to 15 minutes on average. In operations at the lateral skull base, the Viewing Wand's probes are too bulky and working with the arm under the microscope unwieldy, although we were able to achieve superior accuracy, even down to CT resolution.18Figure 5 shows the application in a girl with a tumor in the retromaxillary region and the pterygopalatine fossa.

Another important aspect for working with a 3-D localizer is that the surgeon has to change from the monitor (binoculars) that he or she is working with to the computer monitor. This is especially critical during navigated microscopic ear surgery in which the surgeon touches an anatomical structure with the stylus and then looks away to the monitor for verification with the crosshairs.2,3,11,18 This is obviously inconvenient and risky because inadvertent movements might violate delicate structures unintentionally.

PHILIPS EASY GUIDE, ZEISS STN AND MKM, AND ISG/ELEKTA FREE-HAND

These systems provide patient-image registrations similar to those of the Viewing Wand (<2 mm) but are more convenient by providing navigation with infrared-tracked tools. It is crucial to position the infrared camera during surgery so that an unobstructed view is established throughout surgery. One must ensure that the suction-irrigation endoscope with videocamera, video rack, irrigation unit, navigation unit, sterile draping, instruments, and surgeon and nurse not be in the line of sight. We found the ideal positioning for endoscopic sinus and frontal skull base procedures as follows: the camera is opposite the surgeon, who looks in the face of the patient, at the patient's head. Right-handed surgeons move the stylus with the right hand under video endoscopic control (left hand) so that the camera can detect the diodes.

In the lateral skull base and the petrous bone, sufficient accuracy down to 1 mm was possible with the microscope. However, current probes are still too bulky to provide convenient navigation. Similar precautions as those mentioned for the Viewing Wand apply here too; moreover, the length of the stylus might be problematic when a focal distance of 200 mm is used. Sometimes the stylus simply cannot be introduced into the operating field, eg, deep in the petrous bone or the cerebellopontine angle.

The user interface is important because the surgeon needs a software structure that simplifies the reading and matching of landmarks to the imaging data and immediately informs and warns him or her of possible errors and system failure. In this respect, the Zeiss STN system currently seems most advanced. The pointer of the Philips system is equipped with many light-emitting diodes so that it can be detected by the camera from practically all working positions and, despite its linear form, perform real-time multiplanar reformatting.

Zeiss MKM is suitable for operations at the petrous bone and the lateral skull base, for open procedures at the frontal skull base, and, of course, for neurosurgical procedures. It is not useful for endonasal procedures because it is difficult to work under microscopic view and navigate inside the nose. Because the angle of view often has to be changed in these procedures, the robotic arm is too slow and immobile to allow proper surgical handling. The MKM has its domain in cases in which microscopic preparation is performed and the approach is minimized as much as possible. The surgeon gets relevant spatial information and the position of the autofocal point relative to the CT image faded into his or her view. Therefore, the surgeon need not look away at the monitor. In the meantime, the quality of the graphic overlay has improved in sharpness and contrast. The surgeon can now perform the registration without technical assistance simply by touching keys on a sterilely draped microscope, easing registration. Accuracy with the MKM reaches 1 mm and therefore approaches the resolution of the CT scan.

ARTMA SYSTEM

This system uses magnetic digitizing, and the patient need not be fixed on the operating table, which is advantageous. Because magnetic fields penetrate the whole operating field, a direct view from light-emitting diodes to a camera is not necessary. So the surgeon never interferes with the "measuring beam" of systems with optical detection. This system has another feature: it connects the live video with the patient's images and provides graphical structures (lines, pathways, etc) to guide the surgeon in the live video. These structures show the spatial position of the region of interest or a preplanned path. In addition, these structures can be changed, modified, and readapted on the basis of identified landmarks in the video or the CT or from outside the operating theater by a supervisor or "master surgeon" in the sense of telepresence.19 The system offers many options for the field of telemedicine and telesurgery. We have already successfully transferred and modified surgical 3-D navigation data via the Internet and ISDN.20 This system is powerful and has several promising features. Permanent system development has decisively improved handling, accuracy, and convenience. Currently, accuracy lies in the range of 1 to 4 mm for endoscopic and microscopic operations of the head and neck region. Having previously used the ARTMA System mainly in endoscopic sinus operations, we now have successfully used it in operations at the petrous bone. In Figure 6, the system is used in a case of endocrine ophthalmopathy in which an endoscopic endonasal decompression of the orbit and the optic nerve was performed.

APPLICATION AT THE FRONTAL AND LATERAL SKULL BASES

For navigation at the midface, the paranasal sinuses, and the frontal skull base, one can use a variety of anatomical landmarks, fiducial markers, or markers of the head fixation system, which mostly yield satisfactory registration results. In the petrous bone or the lateral skull base, the patient's face and the available landmarks are covered and not available for intraoperative registration. On the other hand, well-defined landmarks are rare at the lateral aspect of the skull. This complicates the situation, and registration is prone to errors. We solved this problem by introducing landmarks in this region using the head holder to avoid implantation of screws.

GENERAL ASPECTS

Minimally invasive procedures in the region of the paranasal sinuses, the frontal skull base, the petrous bone, and toward the brain bear a distinctive risk for violating delicate structures. The advantage of 3-D navigation is obvious provided that some key requirements are satisfied.21,22 Only then can these techniques successfully support the surgeon and help minimize the potential risks. Successful system use requires intensive cooperation with the radiology department to achieve 3-D data sets that comply with all requirements of navigation, especially for otolaryngological operations.23 Every 3-D navigation system available today cannot simply be "switched on and used for exact navigation" like an ordinary surgical tool. The systems are still too complicated, and the possibilities for pitfalls and consecutive inaccuracy are numerous. Our practical experience has clearly shown that "good results" (ie, high accuracy) and ease of use are only possible if the user faces the problem of navigation as a whole, especially failures. This starts with adapting head fixation for imaging, registration, and, finally, navigation itself. All systems that we have used up to now are still "2-person systems," that is, intraoperative technical assistance is still necessary.

FAILURE DETECTION

Although the software of all systems we have used so far has been steadily improved, there still is demand for an immediate failure detection, eg, inadvertent patient movement. The surgeon has to be aware of this possibility and has to repeatedly check the registration's validity during surgery by touching well-defined landmarks in the operating field to ensure that they are still identical to the corresponding location in the images. Whenever in doubt, the registration needs to be verified or redone. If this is not possible, the operation must be continued "classically."

CONCLUSIONS

We have used various 3-D navigation systems in about 250 operations. We believe that computer-assisted navigation significantly improves safety and orientation in minimally invasive procedures at the paranasal sinuses, the frontal skull base, the petrous bone, and the lateral skull base. Different optical 3-D digitizing systems regularly achieved an accuracy of 1 to 2 mm provided that a reliable patient fixation is used if necessary. We now routinely use these systems during surgery for endonasal procedures. For microscopic operations at the frontal skull base, the petrous bone, and the brain, the Zeiss MKM system offers valuable intraoperative support by superimposing spatial information into the microscopic images. The ARTMA System, using magnetic digitizing technology, offers a range of additional measuring options for endoscopic and microscopic procedures, including the possibility of teleteaching and telesurgery. In conclusion, all systems must become simpler to use to become efficient tools for everyday surgery.

Back to top
Article Information

Accepted for publication November 12, 1999.

Presented ay the annual meeting of the American Head and Neck Society, Palm Desert, Calif, April 27, 1999.

Reprints: Andreas Rüdiger Gunkel, MD, Ear, Nose, and Throat Department, University of Innsbruck, Anichstrasse 35, A-6020 Innsbruck, Austria.

References
1.
Schlöndorff  GMösges  RMeyer-Ebrecht  DKrybus  WAdams  L CAS (computer assisted surgery: a new procedure in head and neck surgery [in German]. HNO. 1989;37187- 190
2.
Gunkel  ARFreysinger  WThumfart  WF Computer-assisted surgery in the frontal and maxillary sinus. Laryngoscope. 1997;107631- 633Article
3.
Klimek  LMösges  R Computer-assisted surgery in the ENT specialty: developments and experiences from the first decade [in German]. Laryngorhinootologie. 1998;77275- 282Article
4.
Gunkel  ARFreysinger  WMartin  A  et al.  Three-dimensional image-guided endonasal surgery with a microdebrider. Laryngoscope. 1997;107834- 838Article
5.
Anon  JBLipman  SPOppenheim  D  et al.  Computer-assisted endoscopic sinus surgery. Laryngoscope. 1994;104901- 905Article
6.
Freysinger  WGunkel  ARBale  R  et al.  Three-dimensional navigation in otorhinolaryngological surgery with the Viewing Wand. Ann Otol Rhinol Laryngol. 1998;107953- 958
7.
Bale  RJVogele  MFreysinger  W  et al.  Minimally invasive head holder to improve the performance of frameless stereotactic surgery. Laryngoscope. 1997;107373- 377Article
8.
Gunkel  AR Computer-assisted 3D-brachytherapy in metastases of the head and neck region: International Symposium on Metastases in Head and Neck Cancer, Kiel, Germany, 15-18 January 1998. Br J Cancer. 1998;77 (suppl 1) 35Article
9.
Martin  ABale  RJVogele  MGunkel  ARThumfart  WFFreysinger  W Vogele-Bale-Hohner mouthpiece: registration device for frameless stereotactic surgery. Radiology. 1998;208261- 265
10.
Auer  THensler  EEichberger  P  et al.  3D-navigation for interstitial stereotaxic brachytherapy. Strahlenther Onkol. 1998;17482- 87Article
11.
Freysinger  WGunkel  ARThumfart  WF Image-guided endoscopic ENT surgery. Eur Arch Otorhinolaryngol. 1997;254343- 346Article
12.
Laitinen  LV Noninvasive multipurpose stereoadapter. Neurol Res. 1987;9137- 141
13.
Sandeman  DRPatel  NChandler  C  et al.  Advances in image-directed neurosurgery: preliminary experience with the ISG Viewing Wand compared with the Leksell G frame. Br J Neurosurg. 1994;8529- 544Article
14.
Drake  JMRutka  JTHofmann  HJ ISG Viewing Wand system. Neurosurgery. 1994;341094- 1097Article
15.
Freysinger  WGunkel  ARMartin  ABale  RJVogele  MThumfart  WF Advancing ear, nose, and throat computer-assisted surgery with the arm-based ISG Viewing Wand: the stereotactic suction tube. Laryngoscope. 1997;107690- 693Article
16.
Gunkel  ARFreysinger  WThumfart  WFPototschnig  C Complete sphenoethmoidectomy and computer-assisted surgery. Acta Otorhinolaryngol Belg. 1995;49257- 261
17.
Thumfart  WFGunkel  AR The latest developments in intraoperative 3D-navigation in the ENT area [in German]. Laryngorhinootologie. 1997;76700- 703Article
18.
Gunkel  ARVogele  MFreysinger  WMartin  ABale  RJThumfart  WF Computer aided surgery in the petrous bone. Laryngoscope. 1999;1091793- 1799Article
19.
Truppe  MJFreysinger  WGunkel  ARThumfart  WF Teleassisted stereotactic endoscopic surgery. Lemke HUVannier MWInamura KFarman AGeds.Proceedings of the CAR ‘96 10th International Symposium on Computer and Communication Systems for Image Guided Diagnosis and Therapy, Paris, June 1996. Amsterdam, the Netherlands Elsevier Science BV1996;686- 692
20.
Gunkel  ARFreysinger  WThumfart  WF Practical experience with the ARTMA Virtual Patient System in ENT surgery. Holl WWurm Geds. Proceedings of the 1st International Congress on Computer Integrated Surgery in the Areas of Head and Spine 1-5, September 1997 Linz, AustriaComput Aided Surg. 1997; (suppl) S06- S33
21.
Ossoff  RHReinisch  L Computer assisted surgical techniques: a vision for the future of otolaryngonology–head and neck surgery. J Otolaryngol. 1994;23354- 359
22.
Korves  BKlimek  LMösges  R Surgical decompression in endocrine orbitopathy: a three dimensional localizing device ensures greater safety. ORL J Otorhinolaryngol Relat Spec. 1996;5846- 50Article
23.
Zinreich  SJTebo  SLong  DL  et al.  Frameless stereotaxic integration of CT imaging data: accuracy and initial applications. Radiology. 1993;188735- 742
×