For nearly 2 decades, experts and professional society bodies had pressed manufacturers to improve the analytical performance of assays for cardiac troponin (cTn).1-4 This effort engendered a steady increase in analytical and clinical sensitivity and has led to the current generation of high-sensitivity assays for cTn (hsTn) that are used routinely in many regions of the world, but are not available in the United States. The increased sensitivity of current assays for cTn, inclusive of contemporary sensitive assays4 used in the United States and hsTn assays used elsewhere, has presented the practitioner with both advantages and disadvantages. A marked increase in the proportion of patients with elevated cTn results has frustrated physicians; many such patients ultimately receive a diagnosis for a cause of myocardial injury other than myocardial infarction (MI). At the same time, hsTn assays have opened the door to emerging applications, such as for risk stratification in atrial fibrillation, stable ischemic heart disease, and heart failure as well as screening for structural heart disease.5 In addition, use of hsTn for rapid triage of patients presenting to the emergency department (ED) with chest pain is an application likely to be embraced by practitioners. In this issue of JAMA Cardiology, Neumann et al6 and Carlton et al7 report 2 separate prospective studies that move the field forward toward evidence-based strategies incorporating hsTn for this purpose.
Current State of Troponin Testing in the ED
Of more than 8 million ED visits related to chest pain that occur annually in the United States, well less than a third will result in a final diagnosis of an acute coronary syndrome.8 Despite this low pretest probability, physicians are obligated to exclude myocardial ischemia with a high degree of probability and engage in time-consuming and costly testing strategies to do so. To manage costs and the adverse effects of overcrowding in the ED, it is a high priority to rapidly and safely identify patients with a sufficiently low probability for acute coronary syndrome (<0.5%-1%) so that they can be discharged efficiently and avoid unnecessary testing.
Cardiac biomarkers have proven useful for this purpose. Accelerated diagnostic protocols that incorporate the clinical history, electrocardiogram, and cTn concentrations provide a framework to rapidly evaluate and triage patients with chest pain suspicious for ischemia. Introduction of cTn has truncated the historical 3 serial blood samples 6 to 8 hours apart necessary when using creatine kinase. With the use of contemporary sensitive cTn assays and a diagnostic cutoff for MI at the 99th percentile upper reference limit, a strategy of measurement at presentation and 3 to 6 hours later provides a negative predictive value (NPV) greater than 99% in low-risk patients without ischemic electrocardiographic abnormalities or other higher risk features. Therefore, present US practice guidelines recommend (class I) testing at presentation (0 hours) and 3 to 6 hours after symptom onset in low-risk patients, with additional testing beyond 6 hours in patients who have electrocardiographic changes or intermediate/high-risk clinical features.9
In parallel, accumulating data support the proposal that when using hsTn assays, the interval between serial sampling for cTn can be further reduced to 1 to 2 hours while maintaining an NPV of approximately 99% among otherwise low-risk patients. In light of such data, the 2015 European Society of Cardiology10 practice guidelines for management of non–ST-segment elevation acute coronary syndrome recommend (class I) that, if an hsTn assay with a validated 0-hour and 1-hour algorithm is available, testing at presentation and 1 hour, together with the absence of high-risk clinical features, is an alternative to testing at 0 hours and 3 hours. This recommendation establishes a need for well-validated algorithms, tested in robustly sized, geographically diverse, clinically relevant populations to guide hospitals using hsTn now and forecasts the same need for health systems that may adopt hsTn in the future.
Leveraging the Sensitivity in hsTn Assays
As a foundation for this evolution in guidelines,10 the superior analytical performance of hsTn assays has been leveraged as a strength in the ED to detect myocardial injury earlier after onset and in concentrations lower than those determined with previous cTn assays.11 For example, in a multicenter European study of patients presenting to the ED within 12 hours of symptom onset (1439 patients [18.6% with MI]), an initial hsTnT less than the 99th percentile upper reference limit (14 ng/L or 0.014 ng/mL [conversion to micrograms per liter is 1:1]) achieved an NPV of 97.1%. Repeating hsTnT at 1 hour improved the NPV to 99.6%. Therefore, a strategy using baseline and 1-hour hsTnT testing and the 99th percentile upper reference limit missed only 3 patients with acute MI (1.1% of all MIs).12 In that study, among the 25% of patients with an hsTnT value below the limit of detection (<5 ng/L) at presentation, the NPV for MI was already 99.7%, suggesting that serial testing may be unnecessary in such patients. Similarly, Shah and colleagues13 reported that, among 4870 patients presenting with chest symptoms, the optimal threshold for ruling out MI with a single hsTnI concentration determined at presentation was less than 5 ng/L (approximately 50% of the population) with an NPV of 99.6%. In both of these populations, application of a very low cutoff (<99th percentile upper reference limits for these assays) provided an excellent NPV based on a single sample at presentation.
The reports from Neumann et al6 and Carlton et al7 build on these data. Neumann and colleagues conducted a single-center study in Germany enrolling 1040 patients with suspected acute coronary syndrome without ST-elevation (17.7% with MI). In this population, 12.3% of patients had baseline and 1-hour hsTnI values below the limit of detection (2 ng/L), with an NPV of 100% for MI. Using a cutoff level of 6 ng/L, offered as optimal by the authors, the proportion meeting the criterion increased to 39% and the NPV was 99.0% (CI, 97.5%-99.7%). The sensitivity was 97.8%, missing 2.2% of non–ST-segment elevation MIs. The investigators validated the NPV in 2 separate populations with resultant values of 99.7% and 99.2%. Not surprisingly, use of a higher cutoff at the 99th percentile upper reference limit (27 ng/L) at 1 hour had a higher rate of false-negative results (NPV, 94.8%). Moreover, considering only a first hsTn result less than 6 ng/L, the NPV was 97.1%. Data using a single sample with a cutoff at 2 ng/L were not reported.
Carlton et al7 report a pooled analysis of 5 prospective ED cohort studies in Australia, New Zealand, and England (3155 patients [9.2% with MI], 33.5% presenting <2 hours from symptom onset). Among patients with a nonischemic electrocardiogram and using the same assay as Neumann et al,6 Carlton et al found that a single hsTnI result less than 1.2 ng/L (18.8% of patients) delivered an NPV of 99.5% (95% CI, 98.4%-99.9%) and sensitivity of 99.0% (95% CI, 96.8%-99.7%). For comparison with other studies of the same assay, their analysis revealed NPVs at 2 ng/L, 5 ng/L, and 6 ng/L that were 99.3%, 99.2%, and 99.1% with corresponding sensitivities of 98%, 94.5%, and 93%, respectively. Carlton et al emphasize that the clinical sensitivity of cutoffs suggested by prior studies13 and that are higher than the limit of detection did not meet a criterion of 99% or greater sensitivity (ie, missing ≤1% of all MIs). Only a cutoff at 1.2 ng/L met this aim based on the first sample alone. Moreover, in early presenters (ie, time from symptom onset to ED presentation ≤2 hours), the sensitivity, even at 1.2 ng/L, was slightly diminished at 98.6%.
Summary and Lessons Learned
Taken together with prior studies,10,13 the findings from the studies of Neumann et al6 and Carlton et al7 lend strong support to the notion that accelerated diagnostic protocols that incorporate hsTn can facilitate earlier triage while maintaining an acceptable NPV. These studies illustrate key concepts that will continue to shape emerging accelerated diagnostic protocols. Better-performing assays enable a shift to more rapid exclusion strategies. To maintain an acceptable sensitivity using a 1- to 2-hour algorithm to exclude MI, it is necessary to use an hsTn assay with cutoffs lower than the 99th percentile upper reference limit. Assessing for a change in cTn between 2 time points (“Δ value”) improves both the negative (flat Δ) and positive (meaningful Δ) predictive performance.6,14 A single very low hsTn result at presentation may be sufficient to exclude MI with a greater than 99% NPV in otherwise low-risk patients. Present data suggest that the limit of detection is preferable as a cutoff to deliver an acceptable NPV at presentation. This criterion may be met in approximately 20% of patients with chest pain and a nonischemic electrocardiogram. Predictably, higher cutoff levels will include more patients at the cost of more false negatives. Considering the timing of symptom onset is critical. Shorter times to sampling diminish the NPV. It seems prudent to obtain a sample at 3 hours or later in early presenters (or ambiguous timing) and in patients with clinical high-risk features. The proportion of patients with hsTn in the very low range will vary with demographics in the local ED, and the NPV will depend on the prevalence of MI in the local population. The NPVs reported in these studies6,7 will be misleading if the accelerated diagnostic protocol is applied in high-risk patients. It is important to know the performance both of your cTn assay and in your population. Assays for cTn are only tests and must be integrated with other clinical assessments. Finally, it is necessary and possible to use evidence-based algorithms for rapid rule-out protocols specific to individual hsTn assays.
Despite many strengths of the available evidence, there are some reasons that these data may overstate the diagnostic performance of rapid strategies to exclude acute coronary syndromes.15 There is a need for additional studies performed in diverse health care settings, including in the United States, to add to the robustness of the estimated NPV across a variety of populations. At present, because hsTn assays are not available in the United States, serial testing at presentation and 3 to 6 hours with a contemporary sensitive assay remains the US standard of care.
Corresponding Author: David A. Morrow, MD, MPH, TIMI Study Group, Cardiovascular Division, Brigham and Women’s Hospital, Harvard Medical School, 75 Francis St, Boston, MA 02115 (dmorrow@partners.org).
Published Online: June 1, 2016. doi:10.1001/jamacardio.2016.1205.
Conflict of Interest Disclosures: The author has completed and submitted the ICMJE Form for Disclosure of Potential Conflicts of Interest. Dr Morrow reported receiving grants from Abbott Laboratories, Amgen, AstraZeneca, Daiichi Sankyo/Eli Lilly, diaDexus, Eisai, Gilead, GlaxoSmithKline, Merck and Co, Novartis, and Roche Diagnostics and personal fees from Abbott Laboratories, AstraZeneca, Daiichi Sankyo/Eli Lilly, diaDexus, Gilead, GlaxoSmithKline, Instrumentation Laboratory, Merck and Co, Novartis, Provencio, and Roche Diagnostics. No other disclosures were reported.
1.Wu
AH, Apple
FS, Gibler
WB, Jesse
RL, Warshaw
MM, Valdes
R
Jr. National Academy of Clinical Biochemistry Standards of Laboratory Practice: recommendations for the use of cardiac markers in coronary artery diseases.
Clin Chem. 1999;45(7):1104-1121.
PubMedGoogle Scholar 2.Jaffe
AS, Ravkilde
J, Roberts
R,
et al. It’s time for a change to a troponin standard.
Circulation. 2000;102(11):1216-1220.
PubMedGoogle ScholarCrossref 3.Morrow
DA, Cannon
CP, Jesse
RL,
et al. National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines: clinical characteristics and utilization of biochemical markers in acute coronary syndromes.
Clin Chem. 2007;53(4):552-574.
PubMedGoogle ScholarCrossref 6.Neumann
JT, Sörenson
NA, Schwemer
T,
et al. Diagnosis of myocardial infarction using a high-sensitivity troponin I 1-hour algorithm [published online June 1, 2016].
JAMA Cardiol. doi:
10.1001/jamacardio.2016.0695.
Google Scholar 7.Carlton
E, Greenslade
J, Cullen
L,
et al. Evaluation of high-sensitivity cardiac troponin I levels in patients with suspected acute coronary syndrome [published online June 1, 2016].
JAMA Cardiol. doi:
10.1001/jamacardio.2016.1309.
Google Scholar 8.Amsterdam
EA, Kirk
JD, Bluemke
DA,
et al; American Heart Association Exercise, Cardiac Rehabilitation, and Prevention Committee of the Council on Clinical Cardiology, Council on Cardiovascular Nursing, and Interdisciplinary Council on Quality of Care and Outcomes Research. Testing of low-risk patients presenting to the emergency department with chest pain: a scientific statement from the American Heart Association.
Circulation. 2010;122(17):1756-1776.
PubMedGoogle ScholarCrossref 9.Amsterdam
EA, Wenger
NK, Brindis
RG,
et al; ACC/AHA Task Force Members. 2014 AHA/ACC guideline for the management of patients with non–ST-elevation acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines.
Circulation. 2014;130(25):e344-e426.
PubMedGoogle ScholarCrossref 10.Roffi
M, Patrono
C, Collet
JP,
et al. 2015 ESC Guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation: Task Force for the Management of Acute Coronary Syndromes in Patients Presenting without Persistent ST-Segment Elevation of the European Society of Cardiology (ESC).
Eur Heart J. 2016;37(3):267-315.
PubMedGoogle ScholarCrossref 12.Hillinger
P, Twerenbold
R, Jaeger
C,
et al. Optimizing Early rule-out strategies for acute myocardial infarction: utility of 1-hour copeptin.
Clin Chem. 2015;61(12):1466-1474.
PubMedGoogle ScholarCrossref 13.Shah
AS, Anand
A, Sandoval
Y,
et al; High-STEACS investigators. High-sensitivity cardiac troponin I at presentation in patients with suspected acute coronary syndrome: a cohort study.
Lancet. 2015;386(10012):2481-2488.
PubMedGoogle ScholarCrossref 14.Morrow
DA, Bonaca
MP. Real-world application of “delta” troponin: diagnostic and prognostic implications.
J Am Coll Cardiol. 2013;62(14):1239-1241.
PubMedGoogle ScholarCrossref 15.Korley
FK, Jaffe
AS. Preparing the United States for high-sensitivity cardiac troponin assays.
J Am Coll Cardiol. 2013;61(17):1753-1758.
PubMedGoogle ScholarCrossref