Background Vena cava filters represent an alternative treatment option for patients with contraindications to anticoagulation, or they might serve as adjunctive treatment for continued emboli despite anticoagulation. The fracture of a filter strut with subsequent end-organ embolization is a rarely reported but potentially life-threatening occurrence.
Methods We sought to determine the prevalence of fracture and embolization of the Bard Recovery (first generation) and the Bard G2 (second generation) vena cava filters. A retrospective, single-center, cross-sectional study was conducted by evaluating all patients who received either a Bard Recovery or Bard G2 filter from April 2004 until January 2009. A total of 189 patients had undergone implantation: 1 pregnant woman and 35 patients who died were excluded from our study. In addition, 10 patients who had the filter removed were also excluded. Ultimately, 80 patients participated in the trial. Subjects underwent fluoroscopy to assess the filter's integrity. Embolized struts were localized by fluoroscopy. Echocardiography and cardiac computed tomography were performed in patients with fragment embolization to the heart.
Results Thirteen of 80 patients had at least 1 strut fracture (16%). At least 1 strut in 7 of the 28 Bard Recovery filters fractured and embolized (25%). In 5 of these 7 cases, patients had at least 1 fragment embolize to the heart (71%). Three patients experienced life-threatening symptoms of ventricular tachycardia and/or tamponade, including 1 patient who experienced sudden death at home. Six of 52 Bard G2 filters fractured (12%). In 2 of these 6 cases, the patients had asymptomatic end-organ fragment embolization.
Conclusion The Bard Recovery and Bard G2 filters had high prevalences of fracture and embolization, with potentially life-threatening sequelae.
Venous thromboembolism occurs in more than 200 000 US citizens per year.1 Anticoagulation is the standard therapy for these patients. Vena cava filters have been used as an alternative therapy for patients who have contraindications to anticoagulation or as adjunctive treatment in patients with continued emboli despite anticoagulation.
Widely recognized complications from filter placement have included erosion through the vena cava wall, visceral perforation, and filter thrombosis, obstruction, and migration. Fracturing of a filter strut with subsequent migration to the heart is a rarely reported occurrence.
The Bard Recovery filter (Bard Peripheral Vascular, Tempe, Arizona) was developed as a permanent but potentially retrievable filter and was commercially available from April 2003 until October 2005. The basic design of the filter's construction (Figure 1A) consists of 2 levels of 6 radially distributed nitinol arms and legs that anchor the filter into the inferior vena cava and serve to trap any embolizing clot.
A patient was seen at our institution more than 3 years after filter placement with pleuritic chest pain, cardiac tamponade, and perforation of the right ventricle as a result of filter fragmentation and embolization. The patient underwent emergency open-heart surgery for removal of the fractured fragments and drainage of the pericardial effusion.
In September 2005, Bard modified the design of the Bard Recovery filter to improve fracture resistance by reducing stress concentration at the apex of the filter using longer arms with curved ends to reduce the load to the arms as they exit the apex. This new design was labeled and marketed as the Bard G2 cava filter, replacing the initial Bard Recovery filter. Over 65 000 Bard G2 filters have been implanted since September 2005.
Based on our institution's herald case and the limited case reports in the literature, we undertook a retrospective review of all patients at our institution who received the Bard Recovery filter or Bard G2 filter.
The institutional review board approved this study. Retrospective review of the procedure logs was used to identify all patients who received a Bard Recovery or Bard G2 filter from April 2004 until January 2009. We attempted to contact all nonpregnant living patients at their last known places of residence. Patients were asked to submit to fluoroscopy of the filter to assess its integrity. If a device had fragmented, all embolized nitinol arms or legs were localized on fluoroscopy, and their locations were recorded. Transthoracic echocardiography was performed in all patients who had fragment embolization to the heart. Selected patients with embolization to the heart also underwent cardiac computed tomography to further define the exact location of the fragments.
A total of 189 patients underwent implantation of the Bard Recovery or Bard G2 vena cava filter at our institution between 2004 and 2009, usually as treatment for deep-vein thrombus or pulmonary embolus (Table 1). Because the Bard Recovery filter was first on the market and was later replaced by the Bard G2, the duration of time between filter implantation and fluoroscopy for the Bard G2 filter is shorter than that of the Bard Recovery filter.
Thirty-five of the 189 patients with Bard filters had died by the time of the study; 10 others had undergone routine scheduled prophylactic recovery of their filter; and 1 was pregnant and could not undergo fluoroscopy. Attempts were made to contact the remaining 143 patients by letter and telephone, and 84 patients were contacted, of whom 80 agreed to participate in the trial. After informed consent, all 80 patients underwent fluoroscopy of the device. The patient enrollment flowchart is shown in Figure 2. Thirteen of the 80 patients who underwent fluoroscopy were found to have at least 1 nitinol arm or leg fracture (16%). The average age for this group was 48.9 years, and the device was evaluated, on average, 37.8 months after implantation (Table 2).
The subset of patients who had received the Bard Recovery filter had a 25% prevalence of device fragmentation and embolization (7 of 28). All Bard Recovery filter fractured fragments embolized via the venous system from the inferior vena cava (IVC) to an end organ. Five of the 7 fractured devices had at least 1 fragment embolize and come to rest in the right ventricle of the heart (71%). Two of these patients experienced symptoms of pleuritic chest pain with documented nonsustained ventricular tachycardia. Three months after completion of this study, 1 of these 2 patients experienced sudden death at home. Included in this group is the patient who experienced hemorrhagic pericardial effusion with subsequent cardiac tamponade requiring emergency open-heart surgery with drainage and fragment removal. In total, 6 of 7 patients who experienced fracture of a Bard Recovery filter (86%) had a fragment either embolize and come to rest in the right ventricle of the heart or continue moving within the venous system through the heart and into the pulmonary arteries. The 7 fractured Bard Recovery filters found on follow-up had been implanted by 3 different physicians, including an interventional radiologist and 2 different general surgeons.
The Bard G2 filter had a lower, but still high, fracture prevalence (12%; 6 of 52). Two of these 6 patients had a fragment embolize cephalad to the IVC, with 1 fragment coming to rest in the hepatic vein and 1 in the lung. Four of the 6 Bard G2 fragmented filters had the fractured strut(s) remain local to the original device implantation location in the IVC (67%). Three different physicians implanted the 6 filters that fragmented, including a vascular surgeon, an interventional radiologist, and a general surgeon.
Complications subsequent to the initial placement of vena cava filters are multiple, and some occur with significant frequency. It is reported that up to 5% of all filters migrate further than 1 cm, with almost all occurrences being asymptomatic.2 Perforation or erosion of the filter through the vena cava at the site of implantation is more common3 but is usually without clinical consequence. Fracture of a filter strut has been reported rarely in the literature but has resulted in life-threatening events.4,5 In 2 of the cases we found in the literature, the fractured filter fragment embolized to the heart and perforated the right ventricle. In 1 case, the patient had chest pain, and in the second case, the patient had life-threatening cardiac tamponade. Both patients underwent successful emergency open-heart surgery for removal of the fractured filter fragment. The Bard Recovery filter was the implanted device implicated in both of these cases. Hull and Robertson6 reported a third patient who had chest pain and nonsustained ventricular tachycardia necessitating removal of fragments from the heart and recovery of the remaining Bard Recovery filter from the vena cava.
Our retrospective review of patients receiving either the Bard Recovery filter or the Bard G2 filter demonstrates a high prevalence of fracture of these devices. The Bard Recovery filter had an overall fracture prevalence of 25% (7 of 28). Six of these 7 patients had at least 1 fragment embolize to the heart or beyond to the lungs (86%) (Figure 1), and 3 of the 7 patients experienced life-threatening symptoms. While the Bard G2 filter incorporated engineering modifications to reduce these occurrences, 12% of the implanted Bard G2 filters also fractured (6 of 52). The modifications might have reduced the ability of the fragments to distally embolize—two-thirds of the fragments remained locally within the IVC near the original implantation site—but as demonstrated in patients 10 and 13, the potential for embolization out of the IVC to other vital organs still exists.
The observed prevalence of filter fracture was 25% with the Bard Recovery filter (7 of 28) and 12% in the Bard G2 filter (6 of 52). These data initially suggest that the fracture rate for the Bard G2 filter is approximately half that of the Bard Recovery filter. However, on further analysis, this conclusion may not be accurate. The average time between filter implantation and assessment of filter integrity for the Bard Recovery filter was 1498 days, or approximately 50 months. For the Bard G2 filter, the average time interval was 717 days, or approximately 24 months. The average time intervals in patients where fracture was observed in the Bard Recovery and Bard G2 groups were nearly identical to those of all patients in those respective groups.
Because nitinol metal fatigue may play a role in the filter fracture, it is reasonable to assume that the incidence of filter fracture would be directly proportional to the time that the filter is allowed to dwell in the patient after implantation. Lynch and Kekulawela7 reported their experience in removing Bard G2 filters before and after 180 days following implantation. A total of 3.4% of the filters removed had fractured prior to removal, and all fractures were observed in patients who had had the filter implanted for more than 180 days. Cantwell et al8 reported a lower observed prevalence of fracture than we observed of the Bard Recovery and Bard G2 filters on removal of the devices, but 95% of the filters in that study were recovered within 15.5 months of implantation (mean, 7.5 months). There was a high prevalence of fracture in this short period with the Bard Recovery filter (9%) but no fractures of the Bard G2 filter. Our follow-up periods of 50 months for the Bard Recovery Filter and 24 months for the Bard G2 filter were much longer. These devices are therefore experiencing prolonged exposure to repetitive movement and subsequent metal fatigue, potentially predisposing them to fail.
We would contend that the fracture prevalence seen by Cantwell et al8 might have been significantly higher if the filter (which has a permanent implant indication) was allowed to remain in place for a longer period. If one were to extrapolate our observed prevalence of Bard G2 filter fractures to 50 months (essentially double the observed period), the prevalence of fracture would be identical to that observed for the Bard Recovery filter, thus challenging the hypothesis that the Bard G2 filter represents an improvement in fracture resistance.
Of the filters that went on to fracture in our study, 6 different physicians from 3 different disciplines had implanted the devices. While it is possible that our findings may represent a local phenomenon, it is difficult to assert that fracture and fragmentation is an operator-dependent event.
The results of this study point to a difficult clinical challenge to the treating physician advising patients with and without filter fragmentation. Many of the implanted devices are left in place, owing to potential long-term indications for pulmonary embolism prophylaxis. Although the recoverable nature of the filter allows for recovery at a later date, the low rate of recovery in our study is similar to that reported in other series.9 As such, the ability to safely remove the device after it experiences local fibrosis into the vena cava may be compromised. It is encouraging that Hull and Robertson6 reported 100% successful retrieval of the Bard Recovery filter without complications an average of 899 days after initial implantation in patients requesting Bard Recovery filter removal (12 of 12). In addition, Lynch and Kekulawela7 reported a similarly large-volume experience with safe late removal (>180 days after implantation) of the Bard G2 filter.
The fact that fragment removal from the heart obviously requires open-heart surgery also poses a difficult clinical decision for a physician caring for a patient who has had embolization of fragments to the heart. While echocardiography is useful to determine if a pericardial effusion is present in these patients, absence of effusion at evaluation does not exclude the possibility of its later development or the occurrence of ventricular tachycardia.
Thirty-five of the 189 patients who underwent vena cava filter implantation at our center had died by the time of this study (19%). Indication for filter implantation obviously selects a patient population with a high mortality rate. Twenty-four of the 35 dead patients in our overall population (69%) died either during the hospital stay when the filter was implanted or within 6 months after the implantation.
Reported filter fractures and migration are rare, and it is not a well-known potential clinical complication. Therefore, the clinical suspicion of a filter fracture and migration being the cause of a patient's complaints or symptoms will be low. Patients with perforation of the right heart and tamponade may have shortness of breath, pleuritic chest discomfort, dizziness, and syncope. Death might also result. These are the same symptoms found in patients with pulmonary emboli. One could hypothesize that some patients who might subsequently be seen with presumed symptoms of recurrent pulmonary emboli may in fact have had filter fracture, migration, and perforation.
It is essential that patients and their treating physicians be educated about this previously underrecognized and potentially life-threatening complication of these devices. Armed with this knowledge, educated patients can be alert to the presence of pleuritic chest pain and other symptoms that should prompt immediate evaluation. Such early awareness and evaluation could certainly be life saving. In addition, the propensity for filter fragmentation may be directly related to the duration of implantation. Patients and their physicians should be educated about this fact so that they have the opportunity to consider having the filter removed.
It is important to recognize that our data represent only a single-center experience. However, 6 different physicians from 3 different specialties implanted the devices that went on to fracture. While our medical center is representative of most large tertiary care hospitals, it is important that our data be corroborated with independent evaluation of implanted devices from other centers. In addition, other brands of IVC filters must be evaluated to determine if our findings are specific to the Bard devices or are a flaw of all such filters.
In conclusion, the prevalence of fragmentation and embolization was found to be high in both the Bard Recovery (25%, 7 of 28) and the Bard G2 (12%, 6 of 52) vena cava filters. Dissemination of these results has a clear benefit for the medical community. Further implantation of these particular devices has been halted at our institution. Educating the medical community and patients on the possibility of device fragmentation and embolization will lead to future patient protection. The optimal approach to the treatment of patients who have already received one of these devices is unclear. Filter fragmentation and embolization might be due to metal fatigue, and therefore its incidence might be directly related to the time that it remains implanted. The decision whether to remove an implanted device will have to be tailored to each individual's particular clinical scenario. It is essential that other medical centers evaluate patients who have received a Bard retrievable filter or any other IVC filter, both for patient safety and to corroborate our single-center findings.
Correspondence: William Nicholson, MD, 25 Monument Rd, Ste 200, York, PA 17403 (wjnichmd2@aol.com).
Accepted for Publication: May 26, 2010.
Published Online: August 9, 2010. doi:10.1001 /archinternmed.2010.316
Author Contributions: Dr W. Nicholson had full access to all of the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis. Study concept and design: W. Nicholson, Goldberg, and Schuler. Acquisition of data: W. Nicholson, W. J. Nicholson, Tolerico, Taylor, Solomon, Schryver, McCullum, Mills, Shears, Agarwal, and Tuohy. Analysis and interpretation of data: W. Nicholson, Solomon, McCullum, Siddoway, and Tuohy. Drafting of the manuscript: W. Nicholson, Schuler, and Shears. Critical revision of the manuscript for important intellectual content: W. Nicholson, W. J. Nicholson, Tolerico, Taylor, Solomon, Schryver, McCullum, Goldberg, Mills, Siddoway, Agarwal, and Tuohy. Statistical analysis: W. Nicholson and Tuohy. Obtained funding: W. Nicholson. Administrative, technical, and material support: W. Nicholson, Tolerico, Solomon, McCullum, Schuler, Shears, Siddoway, and Tuohy. Study supervision: W. Nicholson.
Financial Disclosure: None reported.
Additional Contributions: We acknowledge the efforts of the research coordinators at the York Hospital Heart Center: Barbara DelioCox, RN, Kathy Hutcheson, RN, Elayne Grim, Kathaleen Patierno, RN, and Linda Gerberick, RN, for their diligent efforts in coordinating this study.
This article was corrected for errors on May 29, 2012.
1.Silverstein MD, Heit JA, Mohr DN, Petterson TM, O’Fallon WM, Melton LJ III. Trends in the incidence of deep vein thrombosis and pulmonary embolism: a 25-year population-based study.
Arch Intern Med. 1998;158(6):585-5939521222
PubMedGoogle ScholarCrossref 2.Ferris EJ, McCowan TC, Carver DK, McFarland DR. Percutaneous inferior vena caval filters: follow-up of seven designs in 320 patients.
Radiology. 1993;188(3):851-8568351361
PubMedGoogle ScholarCrossref 3.Poletti PA, Becker CD, Prina L,
et al. Long-term results of the Simon nitinol inferior vena cava filter.
Eur Radiol. 1998;8(2):289-2949477285
PubMedGoogle ScholarCrossref 4.Kumar SP, Mahtabifard A, Young JN. Fractured inferior vena cava filter strut presenting as a penetrating foreign body in the right ventricle: report of a case.
J Card Surg. 2008;23(4):378-38118384575
PubMedGoogle ScholarCrossref 5.Saeed I, Garcia M, McNicholas K. Right ventricular migration of a recovery IVC filter's fractured wire with subsequent pericardial tamponade.
Cardiovasc Intervent Radiol. 2006;29(4):685-68616604409
PubMedGoogle ScholarCrossref 6.Hull JE, Robertson SW. Bard Recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration.
J Vasc Interv Radiol. 2009;20(1):52-6019026564
PubMedGoogle ScholarCrossref 7.Lynch FC, Kekulawela S. Removal of the G2 filter: differences between implantation times greater and less than 180 days.
J Vasc Interv Radiol. 2009;20(9):1200-120919640738
PubMedGoogle ScholarCrossref 8.Cantwell CP, Pennypacker J, Singh H, Scorza LB, Waybill PN, Lynch FC. Comparison of the recovery and G2 filter as retrievable inferior vena cava filters.
J Vasc Interv Radiol. 2009;20(9):1193-119919640733
PubMedGoogle ScholarCrossref 9.Grande WJ, Trerotola SO, Reilly PM,
et al. Experience with the recovery filter as a retrievable inferior vena cava filter.
J Vasc Interv Radiol. 2005;16(9):1189-119316151059
PubMedGoogle ScholarCrossref