Background
Multidetector computed tomography has been proposed as a tool for routine screening for coronary artery calcification in asymptomatic individuals. As proposed, such screening could involve tens of millions of individuals, but detailed estimates of radiation doses and potential risk of radiation-induced cancer are not currently available. We estimated organ-specific radiation doses and associated cancer risks from coronary artery calcification screening with multidetector computed tomography according to patient age, frequency of screening, and scan protocol.
Methods
Radiation doses delivered to adult patients were calculated from a range of available protocols using Monte Carlo radiation transport. Radiation risk models, derived using data from Japanese atomic bomb survivors and medically exposed cohorts, were used to estimate the excess lifetime risk of radiation-induced cancer.
Results
The radiation dose from a single coronary artery calcification computed tomographic scan varied more than 10-fold (effective dose range, 0.8-10.5 mSv) depending on the protocol. In general, higher radiation doses were associated with higher x-ray tube current, higher tube potential, spiral scanning with low pitch, and retrospective gating. The wide dose variation also resulted in wide variation in estimated radiation-induced cancer risk. Assuming screening every 5 years from the age of 45 to 75 years for men and 55 to 75 years for women, the estimated excess lifetime cancer risk using the median dose of 2.3 mSv was 42 cases per 100 000 men (range, 14-200 cases) and 62 cases per 100 000 women (range, 21-300 cases).
Conclusions
These radiation risk estimates can be compared with potential benefits from screening, when such estimates are available. Doses and therefore risks can be minimized by the use of optimized protocols.
Computed tomography (CT) has been proposed as a tool for routine screening for coronary artery calcification (CAC) in asymptomatic individuals as part of a comprehensive risk assessment. A national survey in the United States reported that approximately 27% of diagnostic radiologists already read CAC CT screening scans regularly, making it the most common type of CT screening currently performed in the United States.1The Screening for Heart Attack Prevention and Education (SHAPE) guidelines recommend screening of all asymptomatic men 45 to 75 years of age and asymptomatic women 55 to 75 years of age except those defined as very low risk.2 Such screening in the United States could involve tens of millions of individuals. However, the benefits from this type of screening have not yet been demonstrated directly in randomized trials with cardiovascular events or mortality as an end point, but it has been suggested that the use of CAC scoring can detect disease in asymptomatic persons who would be at low risk when assessed by traditional risk factors.3 However, the potential risks of screening, including the risk of radiation-induced cancer, have to be considered along with the potential benefits.
It is impractical to estimate the risk of radiation-induced cancer from CT scans directly through an observational study, because such a study would require the follow-up of hundreds of thousands of patients for their entire lifetime.4 The difficulty was also underscored in a recent study.5 The magnitude of the risks can be estimated indirectly by extrapolating risk models from existing long-term studies of the effects of radiation exposure, such as the Life Span Study of Japanese atomic bomb survivors.6-8
In the present study, we reviewed a number of protocols for CAC screening by multidetector CT (MDCT) to estimate organ-specific radiation doses. We then used the radiation risk models from the National Research Council's Biological Effects of Ionizing Radiation (BEIR) VII committee to estimate the potential radiation-induced cancer risk from screening according to age and sex.
We reviewed the literature for recent protocols for CAC measurement using MDCT. Unlike other cardiac imaging modalities, such as nuclear medicine and echocardiography, where protocols are more standardized,9,10 cardiac CT protocols still vary widely depending on institutions and scanners. There have been no agreed-on standard scan protocols for CAC measurement by MDCT, and various protocols have been used in previous studies.11-18 In the present study, we considered CT scan protocols that were established by some national cardiac studies19-23 and by one international study on standardization in cardiac CT.24 In practice, most CT technologists are not believed to adjust the calcium scoring protocols much. They use a default protocol by just adjusting scan area. Also, protocols currently used clinically at 3 different university hospitals (Columbia University/New York–Presbyterian Hospital, New York, NY; Cleveland Clinic, Cleveland, Ohio; and Penn State University, Hershey, Pennsylvania) were included. The protocols, including technical parameters, are summarized in Table 1.
Radiation dose calculation
Radiation doses were calculated using the CT scan protocols and CT dosimetry programs (CTDosimetry version 0.99x and CT-Expo version 1.6).25,26 The programs use organ dose databases generated based on Monte Carlo radiation transport modeling by the National Radiological Protection Board in the United Kingdom and the National Research Center for Environment and Health in Germany.27-30
According to the CT operational manuals for studies 1 through 3, each participant receives 2 scans for the purposes of increased reliability and quality control.19-21,23 The double scan is performed for research purposes and is not typical clinical practice. Therefore, radiation dose calculation and risk estimation in this study was performed assuming only a single CT scan per participant. The studies and scanner-specific CT setting parameters presented in Table 1 were used for dose calculation, with the additional assumption that the typical scan length is 12 cm.17 Some protocols provide a range of x-ray tube current-time products (milliampere seconds [mAs]) for patients of different sizes. For these protocols, tube current-time products for medium-size patients were used for dose calculation. Organ-specific radiation doses and the effective dose for each protocol were calculated. The effective dose is the sum of weighted absorbed doses in all irradiated tissues and organs in the body and describes nonuniform dose in order of equivalent whole-body dose.31 The effective dose is one of the most frequently reported dosimetric quantities from CT scans. International Commission on Radiological Protection publication 103 tissue-weighting factors were used to calculate the effective dose.32
We used the radiation risk models for sex- and organ-specific cancer incidence that were developed by the BEIR VII committee33 combined with our organ-specific dose estimates to estimate the risk of radiation-induced cancer. Site-specific models were not available for some cancer sites, such as the esophagus, pancreas, skin, and kidney; therefore, cancer risks at these sites were estimated using the radiation coefficients from the excess relative risk model for “other solid cancers” in the BEIR VII report. For most cancer sites, the committee's risk models were estimated using data from the Life Span Study of the Japanese atomic bomb survivors33; the major exceptions were the risk models for breast cancer and thyroid cancer. The breast cancer model was based on an excess relative risk model from a pooled analysis of 8 cohort studies, including both atomic bomb survivors and medically exposed individuals, by Preston et al.34 Thyroid cancer was based on a pooled analysis of 7 studies by Ron et al.35 Background cancer incidence rates for the general US population were estimated using site-specific cancer incidence rates for all races from the Surveillance, Epidemiology and End Results Program cancer registries for 2000 through 2005.36
After an initial lag period (assumed to be 5 years for solid cancers and 2 years for leukemia), the risk of radiation-induced cancer remains elevated for the rest of the person's lifetime.37 Therefore, the total risk of radiation-induced cancer was calculated using life table methods as a cumulative lifetime risk with adjustment for competing causes of death made using all-cause mortality rates for the US population.38 Potential radiation-induced cancer risks from CAC screening by MDCT were estimated according to age at screening and sex. The SHAPE guidelines recommend screening of all asymptomatic men 45 to 75 years of age and asymptomatic women 55 to 75 years of age except those defined as being at very low risk.2 While SHAPE suggests screening with either CAC or carotid intima-media thickness measurement, the former is more reproducible and predictive of future events,39 and it would be expected that most patients would therefore undergo CAC screening. There are approximately 50 million persons in the United States who are within the age range recommended for CAC screening by the SHAPE guidelines. The SHAPE guidelines recommended reassessment within 5 years for those with a positive test result for atherosclerosis and every 5 to 10 years for those with a negative test result. In the present study, we estimated cancer risks under 2 different scenarios. First, we estimated radiation-induced cancer risk from a single CAC screening by MDCT at any age between 40 and 80 years for each of the different protocols. Second, we estimated the total radiation-induced cancer risk in accordance with the SHAPE guidelines: repeated CT screening every 5 years from the age of 45 to 75 years for men and 55 to 75 years for women.
The estimated effective dose from a single CAC screening varied widely from 0.8 to 10.5 mSv across the CT protocols (Table 2). The median and mean values were 2.3 and 3.1 mSv, respectively. The organs or tissues that were estimated to receive measurable radiation doses, in approximate order of magnitude, were breast, lung, thymus, esophagus, bone surface, adrenal glands, bone marrow, liver, pancreas, skin, spleen, muscle, stomach, and kidney. The wide variation in radiation doses can be attributed to many factors, including different CT scanner models and different CT scan techniques. In general, higher radiation doses were associated with higher x-ray tube current, higher tube potential, and spiral scanning, with low pitch and retrospective gating.
The Figure shows the estimates of the excess lifetime risk of radiation-induced cancer from a single CAC CT scan according to age at screening. The estimated risks decreased with increasing age at screening for all protocols primarily because of the reduced life expectancy. For the median dose of 2.3 mSv (range, 0.8-10.5 mSv), a single screening at the age of 40 years was estimated to result in a lifetime excess cancer risk of 9 (range, 3-42) and 28 (range, 9-130) cancers per 100 000 persons for men and women, respectively. This risk decreased to 3 (range, 1-13) (for men) and 6 (range, 2-26) (for women) per 100 000 persons from screening at the age of 80 years.
The largest proportion of the total radiation-induced cancer incidence, 72% and 71% for men and women, respectively, was attributable to lung cancer, followed by breast cancer for women (20%), and then leukemia (12% and 4% for men and women, respectively) (Table 3). The estimated risk from all cancers combined was higher for women than for men because of the contribution from radiation-induced breast cancer and also because the risk of radiation-induced lung cancer was estimated to be about 2-fold higher for women than for men. Assuming screening every 5 years from age 45 to 75 years for men and 55 to 75 years for women, the cumulative radiation-induced cancer risk from the median dose of 2.3 mSv (range, 0.8-10.5 mSv) was estimated to be 42 per 100 000 for men (range, 14-200) and 62 per 100 000 for women (range, 21-300).
We observed a more than 10-fold variation in radiation doses from CAC screening with MDCT and therefore wide variation in the estimated radiation-induced cancer risk. The range of estimated effective doses is similar to the range from a recent literature review (range, 1.0-12 mSv; mean, 3.0 mSv).40 Many factors influence the radiation dose from medical radiation sources. For CAC CT, these include the CT scanner model, scan mode, electrocardiographic triggering or gating, x-ray tube potential (kilovolt peak [kVp]), tube current-time product (mAs), pitch,41,42 and scan length. Even for the same CT scanner model, CT scan technique or parameter settings may vary between hospitals. For example, there are 3 different protocols in Table 1 for the Volume Zoom CT scanner. Despite the fact that the current-time products were similar for these 3 protocols (range, 50-55 mAs), the estimated effective radiation doses varied from 1.3 to 3.1 mSv. The higher estimates were primarily attributable to the use of higher tube potential or spiral scanners with a lower pitch.
Slower gantry speed and retrospective gating were the main explanations for the higher radiation exposures in some of the protocols (eg, study 1, LightSpeed QX/i scanner). Because of the intrinsic slower gantry speed, 2 full gantry rotations are necessary to create an image during the desired phase of the cardiac cycle.22 Also, retrospective gating results in the x-ray beam being turned on during the whole rotation time. Both features result in an extended exposure time of 1.6-second per imaging level and thus a higher radiation dose, whereas exposure times for the other protocols ranged from 0.2 to 0.5 seconds. Low pitch and retrospective gating also yielded relatively high radiation doses (eg, study 4, Sensation 64 and Volume Zoom scanners). Radiation dose is also inversely proportional to pitch, and a low pitch factor, typically between 0.2 and 0.375, is used for spiral scanning for cardiac CT.
The radiation dose increases rapidly with kVp (as a rule of thumb, radiation dose is proportional to kVp with power of 2.543). The study 1 protocol for the Volume Zoom scanner uses 140 kVp, while most other protocols we reviewed recommend 120 kVp (Table 1). The higher kVp resulted in an estimated effective dose for this protocol that was about 70% higher than the otherwise similar protocol for the same scanner (study 4, axial scan). The radiation dose is linearly proportional to the tube current time product (mAs). Two of the university hospital protocols in our study use the same Siemens Definition scanner but differ considerably in radiation dose owing to the different mAs used for imaging (protocols from hospitals 2 and 3). The same mAs will result in a higher radiation dose44 and less image noise for smaller patients owing to less radiation attenuation by less tissue. Some protocols suggest reducing the mAs for small patients (Table 1), although, in practice, most CT technologists use a default protocol for calcium scoring, adjusting just the scan area to cover from the carina to below the cardiac apex.
The differences in dose observed between scanners and protocols highlights the importance of using scanner equipment that enables a low-dose scan to be performed and of optimizing protocols for a specific scanner model. These considerations are even more important when the CT scan is being used for screening a population that will mostly involve healthy individuals. Moreover, the broader the population that is to be scanned (SHAPE proposes screening approximately 50 million Americans), the greater the potential impact in terms of attributable cancer from a small increase in radiation dose. Therefore, it is essential to optimize calcium scoring protocols to minimize dose while maintaining adequate image quality to yield a reliable calcium score. The wide variation in the protocols reviewed herein also highlights the fact that there are still no agreed-on standard protocols for CAC quantification by MDCT. While the International Consortium has recently recommended that protocols should be standardized, to date it has only published protocols for a limited number of scanners, including few current-generation scanners.24 Further efforts by professional societies are necessary to standardize protocols.
Diamond and Kaul45 have performed cost-effectiveness analyses of the SHAPE paradigm in comparison with other cardiovascular prevention approaches. They estimated that 1-time screening of 50 million individuals could, assuming perfect statin adherence in patients with high CAC, prevent 24 000 deaths and 96 000 nonfatal cardiovascular events, at a net cost of $17 billion, equivalent to $32 000 per life-year equivalent saved in comparison to a standard prevention strategy based on National Cholesterol Education Program guidelines.46 Assuming that 50 million individuals in the United States in the age group had a single screening with 1 of the protocols with the median radiation dose of 2.3 mSv (hospital 1, second protocol), our estimates suggest that the single screening could result in about 5600 individuals (range, 2700-37 000 individuals depending on CT protocol) developing a radiation-induced cancer in the future. Estimates of the radiation-induced cancers were not included in the cost-effectiveness calculations described above. These calculations also provide an informal indication of how the radiation risks might compare with the potential benefits under the best case scenario of 100% treatment compliance.
There are a number of sources of uncertainty in radiation risk estimates owing to the lack of precision in the parameter estimates and uncertain assumptions, such as the form of the dose-response relationship at low doses. In the present study, the extrapolation was performed under the linear no-threshold assumption, with a dose and dose-rate effectiveness factor of 1.5. However, there is also epidemiological and radiobiological evidence that supports both downwardly and upwardly curving slopes, and these alternatives to the BEIR VII risk models would result in higher or lower risk estimates, respectively.5
Another uncertainty involved in radiation risk assessment is the transfer of risk models estimated from the Japanese to other populations with different background cancer rates. The BEIR VII committee's approach to this uncertainty was to use a weighted average of 2 risk models with different underlying assumptions: the excess relative risk model, which is based on the assumption that the risk from radiation exposure multiplies the background cancer risk in the population, and the absolute excess risk model, which is based on the assumption that the risk from radiation exposure adds to the background cancer risk in the population. Although we have not calculated formal confidence bounds for our radiation risk estimates, the BEIR VII committee did conduct such calculations for the risk of all cancers combined, and their results suggest that combination of the uncertainty in the dose and dose-rate effectiveness factor, the transport of the risk models from the Japanese to the US population, and the radiation risk coefficients could mean that the projected cancer risks for each protocol could be higher or lower than our point estimates by a factor of 2.33
We reviewed protocols and estimated radiation doses and associated cancer risks from CAC screening with MDCT. These risks can be compared with estimates of the benefits from such screening once they are available so as to design appropriate screening and prevention strategies for coronary artery disease. There have been no widely agreed-on standard protocols for CAC screening by MDCT. Radiation doses and relevant radiation-induced cancer risks vary depending on protocols up to an order of magnitude. Many technical factors influence radiation dose from CAC measurement with MDCT. Careful optimization of these factors may reduce radiation exposure without detriment to the clinical purpose of the screening examination. Further efforts by professional societies are necessary to standardize protocols in order to decrease unnecessary radiation exposure and to minimize cancer risk.
Correspondence: Kwang Pyo Kim, PhD, Department of Nuclear Engineering, Kyung Hee University, 1 Seocheon-dong, Giheung-gu, Yongin-si, Gyeonggi-do, Republic of Korea (kpkim@khu.ac.kr).
Accepted for Publication: March 20, 2009.
Author Contributions:Study concept and design: Kim, Einstein, and Berrington de González. Acquisition of data: Kim, Einstein, and Berrington de González. Analysis and interpretation of data: Kim, Einstein, and Berrington de González. Drafting of the manuscript: Kim, Einstein, and Berrington de González. Critical revision of the manuscript for important intellectual content: Kim, Einstein, and Berrington de González. Statistical analysis: Kim and Berrington de González. Obtained funding: Einstein. Administrative, technical, and material support: Kim. Study supervision: Kim, Einstein, and Berrington de González.
Financial Disclosure: Dr Einstein has received a research grant from the Nuclear Cardiology Foundation that was sponsored by Covidien, has served as a consultant to GE Healthcare, and has received travel expenses from GE Healthcare, Philips Medical Systems, and Toshiba America Medical Systems.
Funding/Support: Dr Einstein is supported in part by a National Institutes of Health K12 Institutional Career Development Award (5 KL2 RR024157-03).
Additional Contributions: Sandra Halliburton (Cleveland Clinic), PhD, and Kevin Moser (Penn State Milton S. Hershey Medical Center), PhD, provided the protocol for the CT scans that were used at the hospitals.
1.Burger
IMKass
NESunshine
JHSiegelman
SS The use of CT for screening: a national survey of radiologists' activities and attitudes.
Radiology 2008;248
(1)
160- 168
PubMedGoogle ScholarCrossref 2.Naghavi
MFalk
EHecht
HS
et al. SHAPE Task Force, From vulnerable plaque to vulnerable patient, III: executive summary of the Screening for Heart Attack Prevention and Education (SHAPE) Task Force report.
Am J Cardiol 2006;98
(2A)
2H- 15H
PubMedGoogle ScholarCrossref 3.Waugh
NBlack
CWalker
S McIntyre
LCummins
EHillis
G The effectiveness and cost-effectiveness of computed tomography screening for coronary artery disease: systematic review.
Health Technol Assess 2006;10
(39)
iii-iv, ix-x, 1- 41
PubMedGoogle Scholar 5.Brenner
DJDoll
RGoodhead
DT
et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know.
Proc Natl Acad Sci U S A 2003;100
(24)
13761- 13766
PubMedGoogle ScholarCrossref 6.Berrington de González
ABKim
KPSamet
JM Radiation-induced cancer risk from annual computed tomography for patients with cystic fibrosis.
Am J Respir Crit Care Med 2007;176
(10)
970- 973
PubMedGoogle ScholarCrossref 7.Einstein
AJHenzlova
MJRajagopalan
S Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography.
JAMA 2007;298
(3)
317- 323
PubMedGoogle ScholarCrossref 8.Berrington de González
ADarby
S Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries.
Lancet 2004;363
(9406)
345- 351
PubMedGoogle ScholarCrossref 9.Henzlova
MJCerqueira
MDMahmarian
JJYao
SSQuality Assurance Committee of the American Society of Nuclear Cardiology, Stress protocols and tracers.
J Nucl Cardiol 2006;13
(6)
e80- e90
PubMedGoogle ScholarCrossref 10.Pellikka
PANagueh
SFElhendy
AAKuehl
CASawada
SGAmerican Society of Echocardiography, American Society of Echocardiography recommendations for performance, interpretation, and application of stress echocardiography.
J Am Soc Echocardiogr 2007;20
(9)
1021- 1041
PubMedGoogle ScholarCrossref 11.Trabold
TBuchgeister
MKuttner
A
et al. Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating.
Rofo 2003;175
(8)
1051- 1055
PubMedGoogle ScholarCrossref 12.Poll
LWCohnen
MBrachten
SEwen
KModder
U Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation (“ECG pulsing”): phantom measurements.
Rofo 2002;174
(12)
1500- 1505
PubMedGoogle ScholarCrossref 14.Mollet
NRCademartiri
Fvan Mieghem
CAG
et al. High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography.
Circulation 2005;112
(15)
2318- 2323
PubMedGoogle ScholarCrossref 15.Hunold
PVogt
FMSchmermund
A
et al. Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT.
Radiology 2003;226
(1)
145- 152
PubMedGoogle ScholarCrossref 16.Flohr
TGSchoepf
UJKuettner
A
et al. Advances in cardiac imaging with 16-section CT systems.
Acad Radiol 2003;10
(4)
386- 401
PubMedGoogle ScholarCrossref 18.Coles
DRSmail
MANegus
IS
et al. Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography.
J Am Coll Cardiol 2006;47
(9)
1840- 1845
PubMedGoogle ScholarCrossref 20.MESA Coordinating Center,
Multiethnic Study of Atherosclerosis: Computed Tomography Manual of Operations Exam II.
Seattle University of Washington November14 2003;
21.MESA Coordinating Center,
Multiethnic Study of Atherosclerosis: Computed Tomography Manual of Operations Exam III.
Seattle University of Washington April27 2005;
22.Carr
JJNelson
JCWong
ND
et al. Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study.
Radiology 2005;234
(1)
35- 43
PubMedGoogle ScholarCrossref 24. McCollough
CHUlzheimer
SHalliburton
SSShanneik
KWhite
RDKalender
WA Coronary artery calcium: a multi-institutional, multimanufacturer international standard for quantification at cardiac CT.
Radiology 2007;243
(2)
527- 538
PubMedGoogle ScholarCrossref 26.Imaging Performance Assessment of Computed Tomography Group,
ImPACT CT Patient Dosimetry Calculator—Version 0.99x.
London, England Imaging Performance Assessment of Computed Tomography2006;
27.Zankl
MPanzer
WDrexler
G
The Calculation of Dose from External Photon Exposures Using Reference Human Phantoms and Monte Carlo Methods, Part VI: Organ Doses From Computed Tomographic Examinations.
Neuherberg, Germany Gesellschaft für Strahlen und Umweltforschung mbH1991;
28.Zankl
MPanzer
WDrexler
G
Tomographic Anthropomorphic Models, Part II: Organ Doses From Computed Tomographic Examinations in Paediatric Radiology.
Neuherberg, Germany Gesellschaft für Strahlen und Umweltforschung mbH1993;
29.Jones
DGShrimpton
PC
Survey of CT Practice in the UK, Part 3: Normalized Organ Doses Calculated Using Monte Carlo Techniques. Chilton, England National Radiological Protection Board1991;
30.Jones
DGShrimpton
PC
Normalized Organ Doses for X-Ray Computed Tomography Calculated Using Monte Carlo Techniques. Chilton, England National Radiological Protection Board1993;
31.Smith
H
1990 Recommendations of the International Commission on Radiological Protection.
New York, NY International Commission on Radiological Protection1991;
32.Valentin
J
The 2007 Recommendations of the International Commission on Radiological Protection.
New York, NY International Commission on Radiological Protection2007;
33.
Health Risks From Exposure to Low Levels of Ionizing Radiation: BEIR VII, Phase 2.
Washington, DC National Academy of Sciences2005;
34.Preston
DLMattsson
AHolmberg
EShore
RHildreth
NGBoice
JD
Jr Radiation effects on breast cancer risk: a pooled analysis of eight cohorts.
Radiat Res 2002;158
(2)
220- 235
PubMedGoogle ScholarCrossref 35.Ron
ELubin
JHShore
RE
et al. Thyroid cancer after exposure to external radiation: a pooled analysis of seven studies.
Radiat Res 1995;141
(3)
259- 277
PubMedGoogle ScholarCrossref 36.National Cancer Institute,
Surveillance, Epidemiology, and End Results.
Bethesdal, MD National Cancer Institute2008;
38.Centers for Disease Control and Prevention,
National Vital Statistics Reports: United States Life Tables 1999.
Hyattsville, MD Centers for Disease Control and Prevention2002;
39.Folsom
ARKronmal
RADetrano
RC
et al. Coronary artery calcification compared with carotid intima-media thickness in the prediction of cardiovascular disease incidence: the Multi-Ethnic Study of Atherosclerosis (MESA).
Arch Intern Med 2008;168
(12)
1333- 1339
PubMedGoogle ScholarCrossref 40.Mettler
FA
JrHuda
WYoshizumi
TTMahesh
M Effective doses in radiology and diagnostic nuclear medicine: A catalog.
Radiology 2008;248
(1)
254- 263
PubMedGoogle ScholarCrossref 41.Conference of Radiation Control Program Directors,
Nationwide Evaluation of X-Ray Trends (NEXT): Tabulation and Graphical Summary of 2000 Survey of Computed Tomography.
Frankfort, KY Conference of Radiation Control Program Directors2007;
42.International Electrotechnical Commission,
Medical Electrical Equipment, Part 2-44: Particular Requirements for the Safety of X-Ray Equipment for Computed Tomography.
Geneva, Switzerland International Electrotechnical Commission Central Office2002;
43.Brix
GNagel
HDStamm
G
et al. Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey.
Eur Radiol 2003;13
(8)
1979- 1991
PubMedGoogle ScholarCrossref 44.Lee
CLodwick
DWilliams
JLBolch
WE Hybrid computational phantoms of the 15-year male and female adolescent: applications to CT organ dosimetry for patients of variable morphometry.
Med Phys 2008;35
(6)
2366- 2382
PubMedGoogle ScholarCrossref 45.Diamond
GAKaul
S The things to come of SHAPE: cost and effectiveness of cardiovascular prevention.
Am J Cardiol 2007;99
(7)
1013- 1015
PubMedGoogle ScholarCrossref 46.National Institutes of Health,
Third Report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III) Final Report.
Bethesda, MD National Institutes of Health2002;