Peripheral Visual Field Testing in Glaucoma by Automated Kinetic Perimetry With the Humphrey Field Analyzer

Author Affiliations

From the Duke University Eye Center, Durham, NC.

Arch Ophthalmol. 1992;110(12):1730-1732. doi:10.1001/archopht.1992.01080240070033
Abstract

• A Humphrey automated perimeter was used to measure the central 24° of vision with static threshold targets and the peripheral field with two automatic kinetic stimuli in 100 eyes of 100 patients with glaucoma or a suspicion of glaucoma and to compare the additional information gained with the peripheral tests. The peripheral visual field supported the diagnosis made with central field testing in approximately one third of the eyes and added additional diagnostic information in another fourth of the cases. In 4% of patients a normal central field was associated with a glaucomatous peripheral defect. Virtually all peripheral defects were in the nasal quadrant, and the more sensitive isopter uncovered the vast majority of the defects.