Key PointsQuestion
How accurate are Goldmann applanation tonometry (GAT) correction equations, and is there an association between glaucoma stage and discordance of GAT and dynamic contour tonometry intraocular pressure readings?
Findings
In this cross-sectional case series, GAT measurements were discordant from dynamic contour tonometry measurements in eyes with thin corneas and advanced glaucoma.
Meaning
In the management of patients with glaucoma, measurement inaccuracy associated with GAT must be taken into account, but GAT correction equations still involve the risk of unpredictable measurement errors.
Importance
Accurate determination of intraocular pressure (IOP) is crucial for the diagnosis and management of glaucoma. Objective clinical evaluation of the correction equations for Goldmann applanation tonometry (GAT) is lacking.
Objectives
To investigate the difference between corrected and conventional GAT and Pascal dynamic contour tonometry (DCT) measurements, as well as the correlation between discordant IOP values and stage of glaucoma.
Design, Setting, and Participants
This prospective cross-sectional case series was conducted at the Department of Ophthalmology, University Hospital Zurich, and Talacker Eye Center between July 1, 2011, and May 31, 2016, among 112 white patients with glaucoma.
Interventions
Intraocular pressure measurements were performed with GAT and DCT in a randomized order. Goldmann applanation tonometry measurements were modified with 5 correction equations.
Main Outcomes and Measures
The primary end point was degree of concordance between corrected or uncorrected GAT and DCT measurements. The secondary end point was association between discordant IOP measurements and the stage of glaucoma, as assessed by the Glaucoma Severity Score.
Results
Among the 112 patients (67 women and 45 men; mean [SD] age, 66.3 [13.1] years), 63 of the eyes in the study (56.3%) were left eyes and 85 patients (75.9%) were taking ocular antihypertensive medications. Mean (SD) IOP was 20.3 (4.5) mm Hg (95% CI, 19.4-21.1) as measured by DCT and 17.0 [4.1] mm Hg (95% CI, 16.3-17.8) as measured by GAT. The mean (SD) discordance between DCT and GAT measurements was –3.3 (2.0) mm Hg (95% CI, 2.9-3.6). The 5 corrected GAT values ranged from –2.7 to –5.4 mm Hg compared with DCT. The mean (SD) result of the Dresdner correction formula (17.6 [4.1] mm Hg) was closer to the DCT measurement than the original GAT measurement. The mean (SD) Glaucoma Severity Score was 4.7 (3.4) (95% CI, 4.1-5.4). The uncorrected discordance IOPDCT – IOPGAT showed a positive correlation with the Glaucoma Severity Score (rs = 0.33; P < .001) and a negative correlation with central corneal thickness (rs = –0.22; P = .02).
Conclusions and Relevance
In comparison with DCT measurements, these data suggest that GAT values are significantly discordant in eyes with thin corneas and advanced glaucoma. Application of GAT-based correction formulas involves a possible risk of creating an even greater number of unpredictable measurement errors. Hence, we advise with caution, especially pertaining to eyes with thin corneas, to not place reliance on GAT readings, and abandon any correction formula.
Trial Registration
clinicaltrials.gov Identifier: NCT01474070.
Since 1954, Goldmann applanation tonometry (GAT) has become the criterion standard for measurement of intraocular pressure (IOP). However, the accuracy of GAT is limited owing to its dependency on corneal properties.1-11 Although most correlations are largely unknown, the effect of central corneal thickness (CCT) on GAT measurements has been investigated.4,12-18 Goldmann calibrated his tonometer on 500 µm, which he assumed to be normal.11,12,19,20 In fact, CCT has a significantly large range, resulting in inaccurate GAT readings.18,21-25 The overestimation of IOP on thick corneas and underestimation on thin corneas has been demonstrated.11,12,18,26-28 Whereas slight overestimation is clinically acceptable, underestimation of true IOP must be avoided, as underestimation may delay the diagnosis of glaucoma and inhibit appropriate therapy.
In contrast to GAT, Pascal dynamic contour tonometry (DCT) measures IOP directly and continuously, eliminating most systematic errors caused by individual corneal properties.18,29-32 The principle behind DCT is contour matching of the piezoresistive sensor tip with the cornea, allowing noninvasive and direct transcorneal IOP measurement. In vivo and vitro studies comparing DCT measurements with those of an intracameral manometric pressure have shown that DCT gives IOP readings highly corresponding to manometry and, thus, noninvasively best approaches the “true” IOP.1,18,31,33,34
Since the first prototype in 2001, we experienced that DCT readings lie at least 2.0 mm Hg above GAT measurements.18,29,31,35,36 This systematical bias emphasizes the necessity to reevaluate GAT or seek out solutions to eliminate GAT measurement errors.18 To improve the accuracy of GAT, several correction formulas have been proposed. Objective evaluation with a manometric or DCT reference pressure and validation in clinical settings is still lacking.
The aim of this clinical trial is to compare DCT with conventional (uncorrected) GAT and corrected GAT, and to determine their degree of concordance with DCT. The second aim is to investigate the hypothesis that patients with discordant IOP readings have a more advanced stage of glaucoma owing to inappropriate treatment based on inaccurate IOP measurements.
This prospective observational study included outpatient visits to the Department of Ophthalmology, University Hospital Zurich, or the Talacker Eye Center Zurich. All patients provided written informed consent before the first study intervention. The study was approved by the Cantonal Ethics Committee of Zurich (KEK-ZH-Nr. 2011-0311), is registered in the trials registry of the US National Institutes of Health (NCT01474070), and adhered to the Declaration of Helsinki.37
A total of 215 eyes of 112 patients were examined between July 1, 2011, and May 31, 2016. One eye per patient was randomly selected for statistical analysis. To be included in the study, patients had to be at least 18 years of age, with a diagnosis of open angle glaucoma or ocular hypertension. Exclusion criteria were histories of contact lens wear, acute or chronic corneal diseases, corneal astigmatism greater than 2.0 diopters, and a history of laser refractive surgery or other corneal interventions.
A thorough ophthalmologic examination of all patients included refraction, visual acuity, slitlamp bio microscopy, optical biometry and pachymetry (Lenstar LS 900; Haag-Streit), echographic pachymetry (Tomey AL-1000; Tomey Europe GmbH), IOP measurement by GAT and DCT (Pascal; Ziemer Ophthalmic Systems), optical coherence tomography (OCT; Cirrus HD-OCT 5000; Carl Zeiss AG), and Octopus Perimetry (Octopus 900; Haag-Streit). Echographic pachymetry was performed after IOP measurements. The mean of 5 echographic CCT measurements was taken for subsequent analysis. Intraocular pressure was measured by applanation (IOPGAT) and Pascal (IOPDCT) twice each in a randomized order. The mean of 2 measurements was used for analysis. Goldmann applanation tonometry values were corrected by applying 5 correction formulas.12,17,38-40 Only DCT values of best quality were used (quality score Q1, which indicates excellent measurements). Perimetry and optical coherence tomography were performed to graduate the stage of glaucoma. The 5 GAT correction formulas are in Table 113,17,38-40 and described in the eAppendix in the Supplement.
The primary end point of the study was the degree of concordance between IOPGAT, its corrected values, and IOPDCT as the reference pressure. A difference in IOP of 2 mm Hg or more was considered clinically relevant. The association between discordant IOP measurements and the stage of glaucoma was our secondary study end point.
To determine the stage of glaucoma, we developed a Glaucoma Severity Score (GSS) ranging from 0 to 10 points (where 0 indicates a low likelihood of a glaucoma diagnosis and 10 indicates a high likelihood of a glaucoma diagnosis), including the following criteria: (1) superior and inferior peripapillary retinal nerve fiber layer (RNFL) thickness, (2) perimetric mean defect, and (3) agreement of anatomical and perimetric defects, assessed by 2 glaucoma specialists (M.T.-H. and C.K.) (eTable 1 in the Supplement).41
A pilot study with 35 patients found an SD of 3.9 mm Hg for the discordance of DCT and GAT. Given the clinically relevant effect of 2 mm Hg, a significance level of P = .005 and power of 99%, a sample size of 105 patients (independent eyes) was calculated.
Data were coded in Excel 2016 (Microsoft Corp) and analyzed with SPSS, version 22 (IBM Inc). Descriptive statistics, such as mean, SD, median, and interquartile range, in addition to absolute and relative frequencies were computed. Moreover, 95% CIs for the mean were derived. An agreement between the 2 glaucoma specialists (M.T.-H. and C.K.) was investigated by using the κ statistic. According to Altman,42 κ>0.8 indicates very good agreement. A 1-way analysis of variance (ANOVA) together with the Bonferroni post hoc test evaluated differences of discordance between DCT and GAT in addition to corrected GAT values in different GSS groups. Association between discordance of DCT and GAT, or corrected GAT values and the GSS score, as well as CCT, was investigated by using a nonparametric Spearman correlation and a simple and bivariate linear regression. In addition, the Bland-Altman method was applied to derive the 95% limits of agreement. Results of statistical analysis with P < .05 were considered statistically significant. All P values were adjusted for multiple testing. Adjustment was provided by the Bonferroni post hoc test for the 1-way ANOVA and by the multivariable regression analysis adjusting for confounders.
All 112 patients were examined between July 1, 2011, and May 31, 2016. Mean (SD) age was 66.3 (13.1) years (95% CI, 63.8-68.7), all patients were white, and there were 67 women and 45 men. Demographic data are summarized in Table 2.
Mean (SD) IOP was 17.0 (4.1) mm Hg (95% CI, 16.3-17.8) for GAT and 20.3 (4.5) mm Hg (95% CI, 19.4-21.1) for DCT. Mean (SD) CCT was 537 (36) µm (95% CI, 530-544). eTable 2 in the Supplement summarizes biometric data and IOPGAT and IOPDCT values. The mean (SD) IOP difference between the 2 tonometry principles (IOPDCT – IOPGAT) was 3.3 (2.0) mm Hg (Table 3).13,17,38-40 A Bland-Altman plot of the agreement between IOPGAT and IOPDCT illustrates the discordance between the 2 methods against the mean of both, showing 95% limits of agreement between DCT and GAT of –0.5 to 7.1 mm Hg, with a mean discordance of 3.3 mm Hg (Figure 1A). The mean discordances of the 5 corrected GAT values (IOPcorrected) from the DCT reading are listed in Table 3.13,17,38-40 All mean (SD) discordances were statistically highly significant (Ehlers et al,12 4.5 [2.9] mm Hg; Kohlhaas et al17 [Dresdner correction formula], 2.7 [2.3] mm Hg; Elsheikh et al [2009],39 5.4 [2.8] mm Hg; Elsheikh et al [2011],38 4.8 [2.0] mm Hg; and Spoerl et al,40 3.8 [2.1] mm Hg; all P < .001) and positive throughout owing to the higher DCT readings. The mean (SD) results of the Dresdner correction formula (17.6 [4.1] mm Hg) were closer to the DCT measurement than the original GAT reading (Figure 1B and Table 313,17,38-40).
Analysis of IOP Discordances and Glaucoma Severity Score
Our study population reached a mean (SD) GSS of 4.7 (3.4) points (eTable 3 in the Supplement). The Cohen κ coefficient (0.829) was determined and showed very good interrater agreement between M.T.-H. and C.K. for the parity of structural and functional defects. Thus, for statistical analysis, 1 specialist’s (C.K.) rating was applied to avoid half-point scores.
The Spearman rank-order correlation test was applied for the analysis of the GSS. A negative correlation was found between both the OCT superior (rs = –0.79; P < .001) and inferior (rs = –0.78; P < .001) RNFL thickness (criterion 1A and 1B) and the GSS score. For criterion 2 of the GSS, a strong positive correlation was found between functional defect and increasing GSS score (rs = 0.78; P < .001) (eFigure 1 in the Supplement). These findings are in accordance with the expected decrease of RNFL thickness with progressing glaucoma and increase in perimetric mean defect with higher GSS.
Discordant GAT values (IOPDCT – IOPGAT) and discordant modified GAT values (IOPDCT – IOPcorrected) were analyzed separately for each of the 3 GSS criteria using 1-way ANOVA. Superior RNFL thickness (criterion 1A) showed a statistically significant correlation with the discordances of Spoerl et al40 (F2,109, 3.83; P = .03), Elsheikh et al38 (F2,109, 5.10; P = .008), and uncorrected GAT (F2,109, 7.27; P = .001). Bonferroni post hoc testing revealed that the discordance between DCT and Spoerl et al40 (IOPDCT – IOPSPOERL) was significantly higher for the superior OCT score of 2 compared with score 0 (1.1; P = .01). The same applies to the discordance of DCT and GAT (1.3; P = .001). This finding corresponds to a higher discordance for IOPDCT – IOPSPOERL and IOPDCT – IOPGAT with thinner superior RNFL. Accordingly, there was a statistically significant difference between inferior OCT scores (criterion 1B) for the discordance of IOPDCT – IOPGAT as determined by 1-way ANOVA (F2,109, 5.48; P = .005). Again, Bonferroni post hoc testing showed that the discordance was significantly higher for GSS inferior OCT score 2 compared with score 0 (F2,109, 1.2; P = .006). No significant result was found regarding inferior OCT score for any of the discordant values from the correction formulas. Thus, the analysis demonstrates a significantly higher discordance of IOPDCT – IOPGAT in patients with structurally progressed glaucoma compared with patients with normal RNFL thickness, which was partly true for IOP values by Spoerl et al40 concerning superior RNFL thickness. Testing the second criterion, there was no statistically significant association between the discordances of the correction formulas and functional defects. For criterion 3, the discordance of IOPDCT – IOPGAT showed a statistically significant difference between the different score values as determined by 1-way ANOVA (F3,108, 4.1; P = .009). Results of the Bonferroni post hoc test were significant for scores 0 and 3 (F2,109, 1.4; P = .02).
The Spearman correlation coefficient was rs = 0.33 between the GSS and the discordance of DCT and GAT (P < .001). Thus, the Spearman correlation coefficient indicates increasing discordance of GAT and DCT glaucoma severity based on our score augments (Figure 2A). With the exception of the formula by Elsheikh et al38 (2011) (rs = 0.23; P = .01), none of the other correction equations indicated a significant correlation between the equations’ discordances from DCT and the GSS.
Univariate regression analysis indicated a significant dependence of the degree of discordance of IOPDCT – IOPGAT on the GSS (F1,110, 12.7; P = .001). The R2 value was 0.1, which signifies that only 10% of variance in the dependent variable IOPDCT – IOPGAT can be explained by the independent variable GSS. Multiple regression analysis was performed to analyze the influence of age, CCT, axial length, corneal radius, and GSS on the degree of discordance between DCT and GAT, as well as the corrected values. However, only GSS and CCT had a statistically significant effect on the discordances of DCT and GAT (eTable 4 in the Supplement).
Positive correlations between the discordances of all correction equations and CCT were highly statistically significant, except for the Elsheikh et al38 (2011) equation (eTable 4 in the Supplement). With increasing CCT, the discordances became larger and more scattered (eFigure 2A in the Supplement [the Dresdner correction table]). Thus, any formula provides even poorer corrections for thicker vs thinner corneas. The Elsheikh et al38 2011 formula showed the weakest positive correlation between its discordance from IOPDCT and CCT (eFigure 2B in the Supplement). On the contrary, a negative correlation was found between the discordance of IOPDCT – IOPGAT and CCT (rs = –0.22), indicating that with increasing CCT, the discordance between DCT and GAT becomes smaller (P = .02). Thus, GAT readings are closer to the “true” IOP in patients with thicker corneas.
Spearman rank-order correlation showed a negative correlation between CCT and GSS (rs = –0.38; P < .001). Therefore, our study population tended to have thinner corneas with advanced glaucoma as indicated by the GSS (Figure 2B). Analyzing CCT by glaucoma diagnosis resulted in a mean (SD) CCT of 531.3 (35.2) µm for eyes with primary open-angle glaucoma, 548.6 (32.7) µm for eyes with secondary glaucoma, and 541.2 (42.2) µm for patients with ocular hypertension. There was no significant difference in CCT among these diagnosis groups as determined by 1-way ANOVA (P = .11). A scatterplot showed equal distribution of all 3 diagnosis groups from thin to thick corneas.
Mixed-model analyses were performed for all 215 eyes (112 patients), adjusting for multiple observations (both eyes) within each patient. The results provided by the analyses at the patient level and the eye level were in strong agreement.
Accurate determination of intraocular pressure is crucial for the diagnosis and management of glaucoma. Goldmann applanation tonometry retains the reference and standard for IOP measurements despite its well-known limitations. The Pascal DCT was designed to overcome these limitations, providing an IOP measuring technique largely independent of corneal properties.3,27,30,33,36
The poor agreement between GAT and DCT may influence therapeutic decisions. With a mean (SD) difference of 3.3 (1.9) mm Hg between GAT and DCT, our discordance is within the previously reported range.3,15,35 In their systematic review, Cook et al35 reported a mean difference of 1.8 mm Hg between GAT and DCT, with a 95% level of agreement between –2.9 and 6.5 mm Hg. When using the Pascal DCT, Kaufmann et al36 suggested an addition of 2 mm Hg to the target pressure, which is usually based on GAT readings.
In our study, discordances of up to 9.2 mm Hg between GAT and DCT measurements were found. The correction equations resulted in a wider scatter and a larger deviation of DCT values compared with uncorrected GAT readings. The extensive scatter of the IOP values resulting from the correction equations is in accordance with Ang et al,43 who reported no improvement of agreement with DCT after adjustments of their GAT readings.
Regarding the influence of IOP discordances on the stage of glaucoma, the discordance of DCT and uncorrected GAT (IOPDCT – IOPGAT), as well as the discordance of DCT and the GAT as determined by Elsheikh et al38 (IOPDCT – IOPElsheikh), increased with advanced GSS. Only the equation by Elsheikh et al38 was independent of CCT, but nevertheless was influenced by severity of glaucoma. For all other formulas, the association was the reverse. Accordingly, the discordances of the remaining 4 correction formulas showed a positive correlation with CCT. Park et al44 additionally described an increasing discordance between DCT and adjusted GAT readings with increasing CCT.
Concerning all equations, our analyses clarify that the correction of IOP in patients with thinner CCT gives a pressure value closer to the DCT reading as compared with patients with thicker CCTs, because they presumably were developed for thinner corneas. However, discordances from DCT are still clinically relevant even in the lower CCT ranges. The approximately linear Dresdner correction table shows the smallest discordance overall and corresponds to the DCT value at a CCT of 450 µm. However, owing to our small sample size at this CCT level and discordances of more than 10 mm Hg, this result should be interpreted with caution. Contrarily, the difference between DCT and GAT decreased with increasing CCT, which is in accordance with other studies and is well known.15,18,45
Boehm et al33 compared DCT measurements with an intracameral IOP reference pressure in vivo and found excellent agreement, which had been described by Kniestedt et al31 in 2005 for an in vitro setting. Of all corneal factors examined by Boehm et al33 (CCT, corneal curvature, astigmatism, axial length, age), a statistically significant correlation between the discordance of both measurements was found only for CCT. Boehm et al,33 however, acknowledged the effect of CCT as clinically irrelevant, owing to a very small and very weak correlation between DCT and CCT (R2 = 0.00012; P = .03). With this study, we confirm their data with similar values of R2 = 0.04 and P = .04.
The dependence of IOPDCT – IOPGAT on both CCT and the GSS was judged to be clinically relevant (eTable 4 in the Supplement). Therefore, the discordance between DCT and GAT increased with an augmenting GSS score and decreased with growing CCT. In our study group, patients with thinner corneas had more advanced glaucoma. We deduce a causative association between discordant IOP readings and the stage of glaucoma. Patients with thinner corneas may be underdiagnosed owing to measurement inaccuracy of GAT, leading to increased risk of disease progression.
Our study shows that all 5 correction formulas calculate IOP values, which differ to an even greater extent from the “true” IOP than the GAT value itself. It seems clear that simple linear correction equations, as proposed by Ehlers et al12 or the Dresdner correction table, oversimplify the real association between CCT and IOP. The oversimplification may result in a corrected IOP, which may be much less accurate than the initial GAT measurement. In addition, other biomechanical properties of the eye are neglected. Contrarily, the more complex formulas, such as the equation by Elsheikh et al,38 may better reflect the complexity of the association, as additional cofactors are considered. However, they are still not able to provide a sufficient IOP correction. Our results suggest that no correction equation is suitable for the approximation of “true” IOP, and therefore, the risk of creating a significant error is present. Conversely, DCT is known to provide precise measurements of good interobserver reproducibility.
Despite the commonly known limitations of GAT, it is currently the most accessible method for IOP measurement and will likely remain the preferred reference technique. Therefore, it is important to be aware of the limitations and its potential for inaccuracy.35
This study has some limitations. We did not investigate the influence of previous cataract or glaucoma surgery and topical medication on the elasticity of the cornea and, thus, its possible influence on the discordance between corrected and uncorrected GAT and DCT. Furthermore, we did not investigate whether discordances remain unchanged in the same eye on follow-up visits and at different pressure levels. A follow-up study is warranted to address this issue and should be enhanced by corneal hysteresis as a measure for corneal elasticity.
We found a significant increase in discordance between GAT and DCT in patients with thin corneas and advanced glaucoma. Thus, we believe that patients with thin corneas may have a higher risk for glaucoma progression owing to measurement inaccuracy associated with GAT. It is advisable to investigate the discordance from the DCT value at the time of setting the treatment strategy. Furthermore, rather than correcting the GAT value with any correction equation, the discordance should be reevaluated when the glaucoma is uncontrolled and under progression.
Accepted for Publication: March 18, 2017.
Corresponding Author: Christoph Kniestedt, MD, FEBO, Department of Ophthalmology, University Hospital Zurich, Frauenklinikstrasse 24, 8091 Zurich, Switzerland (kniestedt@tazz.ch).
Published Online: May 11, 2017. doi:10.1001/jamaophthalmol.2017.1012
Author Contributions: Drs Wachtl and Kniestedt had full access to all the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis.
Study concept and design: Wachtl, Töteberg-Harms, Frimmel, Kniestedt.
Acquisition, analysis, or interpretation of data: Wachtl, Töteberg-Harms, Roos, Kniestedt.
Drafting of the manuscript: Wachtl, Töteberg-Harms, Roos, Kniestedt.
Critical revision of the manuscript for important intellectual content: All authors.
Statistical analysis: Wachtl, Roos, Kniestedt.
Obtained funding: Kniestedt.
Administrative, technical, or material support: Wachtl, Töteberg-Harms, Kniestedt.
Study supervision: Wachtl, Töteberg-Harms, Roos, Kniestedt.
Conflict of Interest Disclosures: All authors have completed and submitted the ICMJE Form for Disclosure of Potential Conflicts of Interest and none were reported.
Funding/Support: This study was supported by the Talacker Eye Center Zurich (TAZZ) foundation.
Role of the Funder/Sponsor: The funding source had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and the decision to submit the manuscript for publication.
1.Ceruti
P, Morbio
R, Marraffa
M, Marchini
G. Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes.
Eye (Lond). 2009;23(2):262-269.
PubMedGoogle ScholarCrossref 2.Davey
PG, Elsheikh
A, Garway-Heath
DF. Clinical evaluation of multiparameter correction equations for Goldmann applanation tonometry.
Eye (Lond). 2013;27(5):621-629.
PubMedGoogle ScholarCrossref 3.Kotecha
A, White
E, Schlottmann
PG, Garway-Heath
DF. Intraocular pressure measurement precision with the Goldmann applanation, dynamic contour, and ocular response analyzer tonometers.
Ophthalmology. 2010;117(4):730-737.
PubMedGoogle ScholarCrossref 4.Liu
J, Roberts
CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis.
J Cataract Refract Surg. 2005;31(1):146-155.
PubMedGoogle ScholarCrossref 6.Shimmyo
M. Central corneal thickness and measured IOP response to topical ocular hypotensive medication in the Ocular Hypertension Treatment Study.
Am J Ophthalmol. 2005;139(6):1148.
PubMedGoogle ScholarCrossref 7.Stodtmeister
R. Applanation tonometry and correction according to corneal thickness.
Acta Ophthalmol Scand. 1998;76(3):319-324.
PubMedGoogle ScholarCrossref 8.Velten
IM, Bergua
A, Horn
FK, Jünemann
A, Korth
M. Central corneal thickness in normal eyes, patients with ocular hypertension, normal-pressure and open-angle glaucomas—a clinical study [in German].
Klin Monbl Augenheilkd. 2000;217(4):219-224.
PubMedGoogle ScholarCrossref 9.Ventura
AC, Böhnke
M, Mojon
DS. Central corneal thickness measurements in patients with normal tension glaucoma, primary open angle glaucoma, pseudoexfoliation glaucoma, or ocular hypertension.
Br J Ophthalmol. 2001;85(7):792-795.
PubMedGoogle ScholarCrossref 11.Whitacre
MM, Stein
RA, Hassanein
K. The effect of corneal thickness on applanation tonometry.
Am J Ophthalmol. 1993;115(5):592-596.
PubMedGoogle ScholarCrossref 12.Ehlers
N, Bramsen
T, Sperling
S. Applanation tonometry and central corneal thickness.
Acta Ophthalmol (Copenh). 1975;53(1):34-43.
PubMedGoogle ScholarCrossref 13.Francis
BA, Hsieh
A, Lai
MY,
et al; Los Angeles Latino Eye Study Group. Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry.
Ophthalmology. 2007;114(1):20-26.
PubMedGoogle ScholarCrossref 14.Grieshaber
MC, Schoetzau
A, Zawinka
C, Flammer
J, Orgul
S. Effect of central corneal thickness on dynamic contour tonometry and Goldmann applanation tonometry in primary open-angle glaucoma.
Arch Ophthalmol. 2007;125(6):740-744.
PubMedGoogle ScholarCrossref 15.Kniestedt
C, Lin
S, Choe
J, Bostrom
A, Nee
M, Stamper
RL. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry.
Arch Ophthalmol. 2005;123(11):1532-1537.
PubMedGoogle ScholarCrossref 16.Kniestedt
C, Lin
S, Choe
J,
et al. Correlation between intraocular pressure, central corneal thickness, stage of glaucoma, and demographic patient data: prospective analysis of biophysical parameters in tertiary glaucoma practice populations.
J Glaucoma. 2006;15(2):91-97.
PubMedGoogle ScholarCrossref 17.Kohlhaas
M, Boehm
AG, Spoerl
E, Pürsten
A, Grein
HJ, Pillunat
LE. Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.
Arch Ophthalmol. 2006;124(4):471-476.
PubMedGoogle ScholarCrossref 18.Ku
JY, Danesh-Meyer
HV, Craig
JP, Gamble
GD, McGhee
CN. Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry.
Eye (Lond). 2006;20(2):191-198.
PubMedGoogle ScholarCrossref 19.Brandt
JD, Beiser
JA, Kass
MA, Gordon
MO. Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS).
Ophthalmology. 2001;108(10):1779-1788.
PubMedGoogle ScholarCrossref 21.Copt
RP, Thomas
R, Mermoud
A. Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma.
Arch Ophthalmol. 1999;117(1):14-16.
PubMedGoogle ScholarCrossref 22.Doughty
MJ, Zaman
ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.
Surv Ophthalmol. 2000;44(5):367-408.
PubMedGoogle ScholarCrossref 23.Hansen
FK. A clinical study of the normal human central corneal thickness.
Acta Ophthalmol (Copenh). 1971;49(1):82-89.
PubMedGoogle Scholar 24.Herman
DC, Hodge
DO, Bourne
WM. Increased corneal thickness in patients with ocular hypertension.
Arch Ophthalmol. 2001;119(3):334-336.
PubMedGoogle ScholarCrossref 25.Wolfs
RC, Klaver
CC, Vingerling
JR, Grobbee
DE, Hofman
A, de Jong
PT. Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study.
Am J Ophthalmol. 1997;123(6):767-772.
PubMedGoogle ScholarCrossref 26.Barleon
L, Hoffmann
EM, Berres
M, Pfeiffer
N, Grus
FH. Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects.
Am J Ophthalmol. 2006;142(4):583-590.
PubMedGoogle ScholarCrossref 27.Kotecha
A, White
ET, Shewry
JM, Garway-Heath
DF. The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry.
Br J Ophthalmol. 2005;89(12):1572-1575.
PubMedGoogle ScholarCrossref 28.Tonnu
PA, Ho
T, Newson
T,
et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.
Br J Ophthalmol. 2005;89(7):851-854.
PubMedGoogle ScholarCrossref 29.Kanngiesser
HE, Kniestedt
C, Robert
YC. Dynamic contour tonometry: presentation of a new tonometer.
J Glaucoma. 2005;14(5):344-350.
PubMedGoogle ScholarCrossref 30.Kaufmann
C, Bachmann
LM, Thiel
MA. Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis.
Invest Ophthalmol Vis Sci. 2003;44(9):3790-3794.
PubMedGoogle ScholarCrossref 31.Kniestedt
C, Nee
M, Stamper
RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states.
Graefes Arch Clin Exp Ophthalmol. 2005;243(4):359-366.
PubMedGoogle ScholarCrossref 32.Schneider
E, Grehn
F. Intraocular pressure measurement—comparison of dynamic contour tonometry and Goldmann applanation tonometry.
J Glaucoma. 2006;15(1):2-6.
PubMedGoogle ScholarCrossref 33.Boehm
AG, Weber
A, Pillunat
LE, Koch
R, Spoerl
E. Dynamic contour tonometry in comparison to intracameral IOP measurements.
Invest Ophthalmol Vis Sci. 2008;49(6):2472-2477.
PubMedGoogle ScholarCrossref 34.Kniestedt
C, Nee
M, Stamper
RL. Dynamic contour tonometry: a comparative study on human cadaver eyes.
Arch Ophthalmol. 2004;122(9):1287-1293.
PubMedGoogle ScholarCrossref 35.Cook
JA, Botello
AP, Elders
A,
et al; Surveillance of Ocular Hypertension Study Group. Systematic review of the agreement of tonometers with Goldmann applanation tonometry.
Ophthalmology. 2012;119(8):1552-1557.
PubMedGoogle ScholarCrossref 36.Kaufmann
C, Bachmann
LM, Thiel
MA. Comparison of dynamic contour tonometry with Goldmann applanation tonometry.
Invest Ophthalmol Vis Sci. 2004;45(9):3118-3121.
PubMedGoogle ScholarCrossref 37.World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects.
JAMA. 2013;310(20):2191-2194.
PubMedGoogle ScholarCrossref 38.Elsheikh
A, Alhasso
D, Gunvant
P, Garway-Heath
D. Multiparameter correction equation for Goldmann applanation tonometry.
Optom Vis Sci. 2011;88(1):E102-E112.
PubMedGoogle ScholarCrossref 39.Elsheikh
A, Alhasso
D, Pye
D. Goldmann tonometry correction factors based on numerical analysis.
J Biomech Eng. 2009;131(11):111013.
PubMedGoogle ScholarCrossref 40.Spoerl
E, Terai
N, Pillunat
LE. Age-dependent correction factors for Goldmann tonometry.
J Glaucoma. 2012;21(4):276-277.
PubMedGoogle Scholar 41.Wachtl
J, Töteberg-Harms
M, Frimmel
S, Kniestedt
C. A new glaucoma severity score combining structural and functional defects [published online January 31, 2017].
Klin Monbl Augenheilkd. 2017. doi:
10.1055/s-0042-123725PubMedGoogle Scholar 42.Altman
DG. Practical Statistics for Medical Research. London, UK: Chapman and Hall; 1991.
43.Ang
GS, Nicholas
S, Wells
AP. Poor utility of intraocular pressure correction formulae in individual glaucoma and glaucoma suspect patients.
Clin Exp Ophthalmol. 2011;39(2):111-118.
PubMedGoogle ScholarCrossref 44.Park
SJ, Ang
GS, Nicholas
S, Wells
AP. The effect of thin, thick, and normal corneas on Goldmann intraocular pressure measurements and correction formulae in individual eyes.
Ophthalmology. 2012;119(3):443-449.
PubMedGoogle ScholarCrossref 45.Umurhan Akkan
JC, Akkan
F, Sezgin Akcay
BI, Ayintap
E, Tuncer
K. Dynamic contour tonometry and Goldmann applanation tonometry: difference of intraocular pressure values between eyes with and without glaucomatous damage in thin corneas [in German].
Klin Monbl Augenheilkd. 2015;232(10):1190-1197.
PubMedGoogle ScholarCrossref