Steps for measuring choroidal melanoma using 3-dimensional ultrasound.(1) Vertical diameter (white line). (2) Horizontal plane, cut perpendicular to the vertical diameter. (3) Clockwise image rotation. (4) Horizontal diameter(white line). (5) Brightness profile (line between square and triangle) perpendicular to the horizontal diameter, and measurement of maximum height with the help of the brightness profile spikes. (6) Volume measurement (white outline).

Steps for measuring choroidal melanoma using 3-dimensional ultrasound.(1) Vertical diameter (white line). (2) Horizontal plane, cut perpendicular to the vertical diameter. (3) Clockwise image rotation. (4) Horizontal diameter(white line). (5) Brightness profile (line between square and triangle) perpendicular to the horizontal diameter, and measurement of maximum height with the help of the brightness profile spikes. (6) Volume measurement (white outline).

Table 1. 
Fifty 3-Dimensional Ultrasound Images of 42 Consecutive Cases of Choroidal Melanoma*
Fifty 3-Dimensional Ultrasound Images of 42 Consecutive Cases of Choroidal Melanoma*
Table 2. 
Reproducibility of 3-Dimensional Ultrasound Choroidal Melanoma Measurement*
Reproducibility of 3-Dimensional Ultrasound Choroidal Melanoma Measurement*
Table 3. 
Measurement of Choroidal Melanoma: Comparison Between Clinical Examination, Standardized Ultrasound, 3-Dimensional Ultrasound, and Pathological Evaluation*
Measurement of Choroidal Melanoma: Comparison Between Clinical Examination, Standardized Ultrasound, 3-Dimensional Ultrasound, and Pathological Evaluation*
Table 4. 
Linear Regression Analysis to Predict Choroidal Melanoma Volume From Area and Height*
Linear Regression Analysis to Predict Choroidal Melanoma Volume From Area and Height*
Table 5. 
Comparison of Average Reproducibility of Indirect Ophthalmoscopy, A-scan Ultrasound, and 3-Dimensional Ultrasound Choroidal Melanoma Measurements*
Comparison of Average Reproducibility of Indirect Ophthalmoscopy, A-scan Ultrasound, and 3-Dimensional Ultrasound Choroidal Melanoma Measurements*
Table 6. 
Average Accuracy and Reproducibility of 3-Dimensional Ultrasound Measurements*
Average Accuracy and Reproducibility of 3-Dimensional Ultrasound Measurements*
1.
Shammas  HFBlodi  FC Prognostic factors in choroidal and ciliary body melanomas.  Arch Ophthalmol. 1977;9563- 69Google ScholarCrossref
2.
Karlsson  ULAugsburger  JJShields  JAMarkoe  AMBrady  LWWoodleigh  R Recurrence of posterior uveal melanoma after 60Co episcleral plaque therapy.  Ophthalmology. 1989;96382- 388Google ScholarCrossref
3.
Martin  JARobertson  DM Extrascleral extension of choroidal melanoma diagnosed by ultrasound.  Ophthalmology. 1983;901554- 1559Google ScholarCrossref
4.
Char  DHStone  RDIrvine  AR  et al.  Diagnostic modalities in choroidal melanoma.  Am J Ophthalmol. 1980;89223- 230Google Scholar
5.
Servodidio  CAAbramson  DHMcCormick  BFass  D Choroidal melanoma responses after brachytherapy treatment.  Acta Ophthalmol (Copenh). 1992;70687- 689Google ScholarCrossref
6.
Shields  CLShields  JAKarlsson  UMarkoe  AMBrady  LW Reasons for enucleation after plaque radiotherapy for posterior uveal melanoma: clinical findings.  Ophthalmology. 1989;96919- 924Google ScholarCrossref
7.
The Collaborative Ocular Melanoma Study Group, Design and methods of a clinical trial for a rare condition: the Collaborative Ocular Melanoma Study (COMS Report No.3).  Control Clin Trials. 1993;14362- 391Google ScholarCrossref
8.
Finger  PT Radiation therapy for choroidal melanoma.  Surv Ophthalmol. 1997;42215- 232Google ScholarCrossref
9.
Packer  SRotman  M Radiotherapy of choroidal melanoma with 125I.  Ophthalmology. 1980;87582- 590Google ScholarCrossref
10.
Coleman  DJRondeau  MJSilverman  RHLizzi  FL Computerized ultrasonic biometry and imaging of intraocular tumors for monitoring of therapy.  Trans Am Ophthalmol Soc. 1987;8549- 77Google Scholar
11.
Frazier-Byrne  SGreen  RL Ultrasound of the Eye and Orbit.  St Louis, Mo Mosby Year-Book1992;140- 150
12.
Char  DHKroll  SStone  RDHarrie  RKerman  B Ultrasonographic measurement of uveal melanoma thickness: inter-observer variability.  Br J Ophthalmol. 1990;74183- 185Google ScholarCrossref
13.
Snyder  WBFuller  DGFish  GE An inexpensive fiberoptic light pipe to aid in placement of episcleral radioactive plaques.  Ophthalmic Surg. 1988;1962- 63Google Scholar
14.
Robertson  DMFuller  DGAnderson  RE A technique for accurate placement of episcleral iodine-125 plaques.  Am J Ophthalmol. 1987;10363- 65Google Scholar
15.
Reminick  LRFinger  PTRitch  RWeiss  SIshikawa  H Ultrasound biomicroscopy in the diagnosis and management of anterior segment tumors.  J Am Optom Assoc. 1998;69575- 581Google Scholar
16.
Kidd  MNLyness  RWPatterson  CCJohnston  PBArcher  DB Prognostic factors in malignant melanoma of the choroid: a retrospective survey of cases occurring in Northern Ireland between 1965 and 1980.  Trans Ophthalmol Soc U K. 1986;105114- 121Google Scholar
17.
Coleman  DJSilverman  RHRondeau  MJLizzi  FLMcLean  IWJakobiec  FA Correlations of acoustic tissue typing of malignant melanoma and histopathologic features as a predictor of death.  Am J Ophthalmol. 1990;110380- 388Google Scholar
18.
Byrne  SFBoldt  HCDiBernardo  CL  et al. for the Collaborative Ocular Melanoma Study, Echography (ultrasound) procedures for the Collaborative Ocular Melanoma Study (COMS).  J Ophthalmic Nurs Technol. 1999;18143- 149Google Scholar
19.
Byrne  SFBoldt  HCDiBernardo  CL  et al. for the Collaborative Ocular Melanoma Study, Echography (ultrasound) procedures for the Collaborative Ocular Melanoma Study (COMS).  J Ophthalmic Nurs Technol. 1999;18219- 232Google Scholar
20.
Foldberg  RVerdick  RWeingeist  TAMontaque  PR The gross examination of eyes removed for choroidal and ciliary body melanomas.  Ophthalmology. 1986;931643- 1647Google ScholarCrossref
21.
Nicholson  DHFrazier-Byrne  SChiu  MTSchiffman  JHughes  JRNovinski  EK Echographic and histologic tumor height measurements in uveal melanoma.  Am J Ophthalmol. 1985;100454- 457Google Scholar
22.
Rask  RJensen  PK Precision of ultrasonic estimates of choroidal melanoma regression.  Graefes Arch Clin Exp Ophthalmol. 1995;233777- 782Google ScholarCrossref
23.
Silverman  RHColeman  JRondeau  MJWoods  SMLizzi  FL Measurement of ocular tumor volumes from serial, cross-sectional ultrasound scans.  Retina. 1993;1369- 74Google ScholarCrossref
24.
Jensen  PKHansen  MK Ultrasonographic three-dimensional scanning for determination of intraocular tumour volume.  Acta Ophthalmol (Copenh). 1991;69178- 186Google ScholarCrossref
25.
Fisher  YHanustsaha  PTong  SFenster  AMazarin  G Three-dimensional ophthalmic contact B-scan ultrasonography of the posterior segment.  Retina. 1998;18251- 256Google ScholarCrossref
26.
Augsburger  JJGamel  JWBailey  RSDonoso  LAGonder  JRShields  JA Accuracy of clinical estimates of tumor dimensions: a clinical-pathologic correlation study of posterior uveal melanomas.  Retina. 1985;526- 29Google ScholarCrossref
27.
Nicholson  DHFrazier-Byrne  SChiu  MTSchiffman  JHughes  JRNovinski  EK Echographic and histologic tumor height measurements in uveal melanoma[letter].  Am J Ophthalmol. 1986;101125- 128Google Scholar
28.
Favilla  IBarry  WMcNeil  JMcKenzie  JGosbell  ADEllims  P Measurement of choroidal melanomas: a comparison of methods.  Aust N Z J Ophthalmol. 1988;1659- 66Google ScholarCrossref
29.
Finger  PTRomero  JMRosen  RBIezzi  REmery  RBerson  A Three-dimensional ultrasonography of choroidal melanoma: localization of radioactive eye plaques.  Arch Ophthalmol. 1998;116305- 312Google ScholarCrossref
30.
Romero  JMFinger  PTIezzi  RRosen  RBCocker  RS Three-dimensional ultrasonography of choroidal melanoma: extrascleral extension.  Am J Ophthalmol. 1998;126842- 844Google ScholarCrossref
Clinical Sciences
September 2001

Three-Dimensional Ultrasound for the Measurement of Choroidal Melanomas

Author Affiliations

From the New York Eye and Ear Infirmary (Drs Romero, Finger, Rosen, and Iezzi), and the New York Eye Cancer Center (Drs Romero and Finger), New York, NY. The authors have no proprietary or financial interest related to this article, including the equipment used in this project.

Arch Ophthalmol. 2001;119(9):1275-1282. doi:10.1001/archopht.119.9.1275
Abstract

Objective  To evaluate the reliability of 3-dimensional ultrasound (3D-US) for the measurement of choroidal melanomas.

Design  Retrospective case series.

Participants  Forty-two consecutive cases of choroidal melanoma imaged with 3D-US.

Methods  Tumor measurements obtained with ophthalmoscopy, transillumination, standard ultrasound techniques, 3D-US, and pathological studies. Tumor diameters, heights, and volumes were compared. Our 3D-US tumor measurement techniques were tested for intraobserver and interscan reproducibility.

Results  Fifty 3D-US images were studied. The 3D-US tumor measurements were found to be reproducible (height coefficient of variation [CV] ≤3%; diameter CV ≤9.7%; volume CV ≤13.2%). There was significant correlation with the usual methods of tumor measurement (diameter r= 0.76; height r≥0.98). Significant differences were found between measurements at pathological examination, as compared with both 2-dimensional and 3D-US height measurements (range, 0.73-0.83 mm). This finding was thought to be due to specimen shrinkage. Three-dimensional ultrasound was found to be at least as reproducible as clinical examination and standard ultrasound techniques used for measurement of diameter and height of choroidal melanomas. It was our impression that the 3D-US volume measurements accounted for the geometry of the tumor better than volume estimates calculated from basal area and tumor height.

Conclusions  Three-dimensional ultrasound measurements of choroidal melanoma were reproducible, correlated well with other tumor measurement techniques, and can be used for measurement of choroidal melanomas.

CHOROIDAL melanoma size is the most important factor in determining both treatment options and prognosis (ocular and for metastasis).1-9 For example, apical tumor height and largest tumor dimension (typically the largest diameter in contact with sclera) have been reported to be risk factors for extrascleral extension, posttreatment recurrence, and metastasis.1-3 Accurate measurements of tumor height and basal diameter are also critical for monitoring tumor growth and are used to define treatment alternatives.4-9 For example, accurate measurements of tumor apical height and basal diameters are critical for choosing plaque size and radiation dose.8,9

In clinical practice, tumor height and basal diameters are typically measured using combinations of ultrasonography (2-dimensional [2D] B-scan and A-scan), ophthalmoscopy, fluorescein angiography, and transillumination.4,7,10,11 The accuracy and reproducibility of these methods have been described.4,12 Ultrasonography is currently one of the most reliable techniques for measurement of choroidal melanomas (except in cases of anteriorly located tumors). This is because tumor diameters obtained by ophthalmoscopy are based on relative proportional measurements compared with normal ocular structures and the ophthalmoscopic field of view (20 diopters[D] = 12 mm; 28 D = 16 mm), and can be affected by the axial length of the eye. However, ophthalmoscopy can reveal flat regions of the tumor margin that can be missed with ultrasound. Measurements using photographs and measuring grids can be more accurate, but many tumors do not fit into the conventional fundus camera field or are located anterior to the equator, rendering them difficult to photograph.

Transillumination and high-frequency ultrasound are used for measurement of anterior uveal melanomas.13-15 Measurement of tumor diameters in cases of large collar-button choroidal melanomas can be difficult with both transillumination and funduscopic examination owing to the tumor's shadow, which can be cast onto the sclera.

We believe that tumor volume measurements will become more important in the evaluation of intraocular tumors. Though most melanomas are shaped like a dome or mushroom, each dome or mushroom is typically different. Though apical height and basal diameter are standards for monitoring growth or regression, tumor volume also reflects the tumor size, which is dependent on tumor shape.16,17

The primary focus of this study was to determine if the clinical use of 3-dimensional (3D) ultrasound (3D-US) was comparable to standard 2D ultrasound measurements of choroidal melanomas. Three-dimensional ultrasound measurements were also evaluated for intraobserver and interscan reproducibility. Three-dimensional ultrasound measurements were compared with measurements obtained by ophthalmoscopy, transillumination, and pathological evaluation. Lastly, 3D-US tumor volumes were compared with tumor volumes calculated from basal area and apical height. This study examines the current capabilities of 3D-US for the measurement of choroidal melanomas.

Materials and methods
Study design and clinical assessments

We present a clinical case series of 42 patients whose choroidal melanomas were measured with 3D-US. The tumor's minimum and maximum diameters were measured by ophthalmoscopic examination by a trained specialist in ophthalmic oncology(P.T.F.). Transillumination was used for measurement of anterior tumor diameters. B-scan and A-scan examinations were performed using the Ophthascan S system(Alcon, Fort Worth, Tex). The ultrasound protocol used by the Collaborative Ocular Melanoma Study as well as interpolative A- and /or B-scan techniques were used as applicable.10,11,18,19 Ultrasound techniques were used to achieve a "best clinical assessment" of tumor size based on ultrasonography. Ultrasound measurements were documented with Polaroid pictures (Polaroid Corp, Cambridge, Mass).

3d-us studies

Three-dimensional ultrasound was requested after the clinical assessment. Three-dimensional ultrasound studies were performed using the 3D i-scan (Ophthalmic Technologies Inc, Downsview, Ontario). Three-dimensional ultrasound uses a conventional brightness mode (B-mode) transducer, combined with a motorized, rotating holder and computerized image processing. During data acquisition, the transducer is rotated while multiple 2D images are collected and processed by the computer to form a 3D block. After acquisition, it is possible to view and manipulate the 3D block interactively. Since the 3D block can be rotated and sliced, it allows for viewing 2D images derived from unique perspectives.

The 3D B-scan probe operates at a frequency of 10 MHz, with an axial resolution of 0.15 mm and a lateral resolution of 0.27 mm. The focal point is 25 mm, with a total image depth of 50 mm. For image processing, the proprietary software 3D i-scan was run on a Macintosh computer (Apple Computer Corp, Cupertino, Calif). The acquisition and reconstruction times were 7.5 and 6 seconds, respectively. All 3D-US measurements were made at an almost equivalent sound velocity of 1532 m/s (speed of sound in vitreous), as compared with 1550 m/s used for standardized A-scan. This difference would make 3D-US height measurements 1.16% smaller than those derived by A-scan. This difference was not considered to be a significant factor in our measurements of heights or volumes.

Three-dimensional ultrasound was done within 1 week after the clinical assessment. These evaluations were performed by one of us (J.M.R.), who spent 2 years in training as a retinal imaging fellow with clinical experience in ultrasonography, and under the supervision of the other authors. He had no prior knowledge of the clinical measurements (performed by P.T.F.). During the examination, the patient lay in the supine position, and the ultrasound probe (with coupling media) was in contact with the eyelid. The patient was asked to look at a fixation target while the ultrasound probe was oriented in the direction of the tumor. The area of interest was centered in the ultrasound screen. The ultrasound gain was adjusted to obtain the best image possible(typically 70 dB). During acquisition time, the patient was asked to be still. Ordinarily, 3 sets of images were saved for later measurement.

This study was complied in accordance with the Declaration of Helsinki recommendations on biomedical research involving human subjects. Before scanning, all patients were informed that 3D-US would be performed, that the 3D-US transducer emissions were essentially equivalent to a 2D unit, and that there was no additional risk.

3d-us tumor measurements

The 3D-US tumor measurement technique was performed as follows (Figure 1): (1) Consecutive vertical planes were examined to obtain the vertical diameter. (2) A horizontal plane was cut perpendicular to the vertical diameter. (3) The image was rotated clockwise.(4) Consecutive horizontal planes were examined to obtain the horizontal diameter.(5) A brightness profile was placed perpendicular to the horizontal diameter, and maximum height was obtained from consecutive horizontal planes. (6) Area outlines were done in consecutive horizontal planes to obtain volume.

The tumor diameters help to determine planes and lines perpendicular to the tumor base. Therefore, tumor height measurements are perpendicular to the tumor base. Interactive analysis of the 3D block allowed the observer to interactively check the height measurement in consecutive parallel planes, in order to always obtain the maximum tumor height measurement. Our 3D-US tumor measurement technique took approximately 20 minutes, of which 15 minutes were spent on volume measurement. Images and measurements were saved in magneto-optical media.

To obtain the tumor volume, the 3D block is sliced in parallel planes, and the observer must outline the tumor in each of them. We used a slice interval of 0.3 mm, which is close to the system's lateral resolution. The larger the tumor diameters, the longer it takes the observer to outline the whole tumor. The system calculates the tumor volume based on the outlined areas.

3d-us measurement reproducibility

To measure the intraobserver reproducibility, the same observer (J.M.R.) measured images of a small, a medium, and a large tumor 10 times at a 24-hour interval between measurements. Collaborative Ocular Melanoma Study criteria were used to classify the tumor size.7 To measure the interscan reproducibility, another 4 cases were scanned 3 consecutive times with the same ultrasound settings (same gain and dynamic range) and the same transducer orientation. The observer measured each image once on the same day. The measurement display was covered, and the measurements were saved in the computer memory for later analysis. Interobserver reproducibility was not evaluated.

Data recollection

Medical records of the cases were reviewed. Information from the date of 3D-US imaging was obtained: status (ie, prebrachytherapy), affected eye, tumor location, tumor shape, clinical diameters (transillumination or fundoscopy), A-scan ultrasound tumor apical height, 3D-US tumor measurements, and pathological study measurements (as available).

Gross and histopathological studies were performed at the Department of Pathology of the New York Eye and Ear Infirmary (New York, NY). The specimens were typically preserved in formaldehyde for 1 to 2 weeks prior to processing. Eye sectioning and tumor measuring techniques have been described.20 Our data were retrieved from the patient's pathology records.

Analysis

Coefficient of variation (CV) (standard deviation–mean ratio) was used to measure the 3D-US intraobserver and interscan measurement variability. Pearson correlation and a 2-tailed paired t test(α = .05) were performed between the different measurements of heights and diameters. Since vertical and horizontal diameters are measured in 3D-US images, minimum and maximum diameters are measured during clinical and pathological examination; the mean diameter was used for comparison. Basal area was calculated from basal diameters. Linear regression analysis was used to estimate volume from basal area and apical height. For statistical analysis, JMP software(SAS Institute Inc, Cary, NC) was used.

Results
Tumor characteristics

Fifty 3D-US images of 42 eyes were analyzed in this study (Table 1). The cases were collected from October 1996 to May 1998. The status of the cases when imaged was as follows: 6 cases were under observation (1 dormant, 3 small, and 2 treatment refused), 19 were prebrachytherapy (17 palladium 103, and 2 iodine 125), 12 were postbrachytherapy, and 13 were pre-enucleation (1 tumor regrowth, 2 neovascular glaucoma, 10 as primary treatment). Three cases had imaging studies performed on more than one date. The tumor apex locations were as follows: 3 posterior, 16 posterior-equator, 18 equator-posterior, 8 equatorial, 5 equator-anterior. Thirty-nine of the imaged tumors were dome-shaped, and 11 were mushroom-shaped. Five of the mushroom-shaped tumors were treated by enucleation.

Clinical diameters were not documented in 18 cases. In 8 of these cases, the tumor was too large for the base to be completely visualized by ultrasound. The other 10 were treated with plaque radiation therapy, with a good clinical response, and no posttreatment measurements of tumor base diameters were recorded. A-scan ultrasound height measurements were not obtained in 3 cases. In these 3 cases, 2 large and 1 medium-sized melanomas were evaluated with 3D-US only. Tumor heights obtained at pathological evaluation were not included in 2 cases because the eye section did not include the maximum tumor elevation.

3d-us measurements

Based on 3D-US measurements, the tumor dimensions were as follows: mean diameters ranged from 5.7 to 18.5 mm (mean, 10.3 mm; SD, 3.1 mm), heights ranged from 1.3 to 13.9 mm (mean, 4.5 mm; SD, 3.2 mm), and volumes from 17.8 to 1569.1 mm3 (mean, 277.5 mm3; SD, 375.5 mm3)(Table 1).

The intraobserver and interscan reproducibility for 3D-US measurements of the choroidal melanomas were also tested (Table 2). When the observer performed repeated measurements of an image, smaller coefficients of variation were noted with increasing tumor size. From the measured dimensions, height was less variable (CV, 0.7%-3.0%), diameter was intermediate (CV, 0.0%-9.7%), and volume was more variable (3.6%-13.2%). The volume CV was inversely proportional to the size of the tumor. Diameters were more variable than heights. The explanation for that could be that the system has less lateral resolution (0.27 mm) than axial resolution (0.15 mm), and that, as any experienced ocular ultrasonographer knows, tumor margins are less distinct with ultrasonography.

3d-us vs standard clinical evaluations

Results of Pearson correlation and paired t tests comparing clinical examination, 3D-US, and pathological examination measurements are presented in Table 3. Three-dimensional ultrasound was found to be equivalent to 2D-US for measurement of tumor height. In fact, statistically significant correlations were shown between all methods of diameter and height measurement. Measurements of diameter on clinical examination were slightly larger than those done with 3D-US by a mean difference of 0.4 mm. These differences were not statistically significant. Similarly, measurements with 3D-US were slightly larger than those made with pathological examination. The standardized ultrasound and 3D-US height measurements were 0.8 mm larger, on average, than heights obtained at pathological evaluation, which could have be due to cut of perfusion, fixation, and pathology processing. Clearly, there was no significant difference between 3D-US and 2D-US height measurements.

Tumor volume

Linear regression analysis was used to predict choroidal melanoma volume(V) (measured by 3D-US) from basal area and apical height (A × H) (Table 4). Tumors were divided into groups by volume and shape. All groups, except those with an A × H greater than 1600 mm3, resulted in significant correlations. For all 50 images, volume could be estimated using the prediction formula:

plus or minus 38.7 mm3 (95% confidence interval). Prediction equations with smaller confidence intervals were obtained in small low-volume and dome-shaped tumors, which had more predictable geometry. Equations with larger confidence intervals were obtained in large collar-button–shaped tumors. It was difficult to predict the volume of large tumors and collar-button–shaped tumors. Volumes determined by 3D-US were less variable (lower SD in repeated measurements) than volumes calculated using area and height (large confidence interval).

Comment
3d-us reproducibility

Investigators have studied the reproducibility of indirect ophthalmoscopy,4 standardized A-scan ultrasound,4,12,21 and 3D-US22-24 for measurement of choroidal melanomas. We compared the reported reproducibility of those methods with the reproducibility of 3D-US (Table 5). Three-dimensional ultrasound was found to offer better intraobserver and interscan reproducibility than those conventional methods. Three-dimensional ultrasound in vivo interobserver reproducibility is not available.

Fisher et al25 studied the accuracy and reproducibility of measurements in vitro using the 3D i-scan system, and we studied the reproducibility of measurement in vivo using the same system (Table 6). Fisher et al found that the in vitro measurements were accurate, with an error of 0.2% for linear measurements and 2.9% for volume measurements. Based on the data of Fisher et al and our own data, we have found that 3D-US is accurate and reproducible for measuring choroidal melanomas.

Comparisons with pathological studies

There have been studies correlating pre-enucleation tumor measurements with pathological studies.21,26 Those investigators reported correlation of height measurements (r = 0.92) and ophthalmoscopic diameters (r= 0.81). It is well known that there is tumor shrinkage after cutting the blood supply and formaldehyde fixation.21,27 In our study, we found that both A-scan and 3D-US height measurements had excellent correlation with measurements obtained at pathological examination(r = 0.99 and r = 0.98, respectively). Also, in our pathological studies, tumor height measurements were significantly smaller than preoperative measurements. Nicholson et al27 also noted shrinkage of the tumor dimensions after enucleation.

Linear regression analysis of tumor volume

Favilla et al28 used linear regression analysis to find the functional relationship between B-scan volume and the volume calculated from A × H. The B-scan volume was measured using ultrasonographic tomography and computer-assisted calculation of area of each tomographic slice, the sum of which constituted the volume. They analyzed 51 follow-up measurements of 15 cases with tumor volumes on B-scan that were less than 500 mm3. For the total of tumor measurements, the obtained prediction equation to estimate tumor volume was:

This formula was slightly greater than the volume formula of a cone(0.33 A × H). In our series, the obtained prediction equation for tumors with an A × H less than 600 mm3 (V <300 mm3)was:

This formula was in between the volume formula of a cone (0.33 A × H) and a hemisphere (0.67 A × H).

3d-us measurements of tumor volume

In comparison with the linear regression analysis, in which a dome-shaped tumor with an A × H of 167.57 mm3 has an estimated volume of 81.41 ± 21.00 mm3 using equation 3 (the prediction equation with the smallest confidence intervals), the same tumor repeatedly scanned with 3D-US had a volume measurement of 72.40 ± 5.18 mm3. Note the smaller confidence interval of the 3D-US–measured volume. This finding was the same for all the other tumors we measured. It was our impression that the 3D-US volume measurements were closer to the true tumor volume than volume estimates from A × H because 3D-US measurements account for the shape or geometry of the whole tumor. Our study showed that 3D-US volume measurements were more reproducible than volume estimates from A × H.

3d-us vs conventional ultrasound

Conventional ultrasound and 3D-US share the same ophthalmic ultrasound principles and limitations. Both techniques can have limitations in visualizing the tumor margins and anteriorly located tumors. While both 2D ultrasound and 3D-US studies are initially dynamic, they differ in that the collected images from conventional ultrasound studies are visualized in a 2D format. Typically the user takes 5 to 10 minutes imaging the tumor with A- and B-scans, while performing a mental 3D representation. The observer records a limited number of 2D images. Future review of the entire tumor volume is not possible. Static 2D images may demonstrate the largest tumor diameter or may not include the tumor's apex. Clearly, most of the tumor's surface characteristics are lost for comparison to future examinations.

In contrast, 3D-US involves acquisition of multiple 2D B-scan images, which are stored with each transducer rotation.29,30 Software uses this information to reconstruct a 3D block. Then, without further contact with the patient, the 3D image, which is a block of data, can be rotated, sliced, and viewed in planes oriented in any direction. After an image acquisition, there can be no missing points of the lesion. Any tumor dimension can be measured in consecutive planes until obtaining the most appropriate measurement. A plane perpendicular to the tumor diameters can be displayed to obtain a more reproducible tumor height measurement. It is also possible to measure volume by outlining areas of tumor on consecutive planes.

In this study, we are reporting our experience with 3D-US, a newly available technology. This case series demonstrates that it can be as reliable as conventional methods for measuring choroidal melanomas. Because the image of the whole tumor is stored and analyzed, it has the potential of providing more reproducible measurements. In addition, 3D-US provides the advantage of the ability to measure tumor volume, no matter the shape of the tumor. As with conventional ultrasound, it is important to combine ultrasound evaluation with clinical examination and indirect ophthalmoscopy to identify all tumor margins, including flat areas that can be missed by ultrasonography.

Accepted for publication March 30, 2001.

This study was supported in part by the EyeCare Foundation, New York, NY (http://www.eyecarefoundation.org).

Corresponding author and reprints: Paul T. Finger, the New York Eye Cancer Center, 115 E 61st St, New York, NY 10021 (e-mail: pfinger@eyecancer.com).

References
1.
Shammas  HFBlodi  FC Prognostic factors in choroidal and ciliary body melanomas.  Arch Ophthalmol. 1977;9563- 69Google ScholarCrossref
2.
Karlsson  ULAugsburger  JJShields  JAMarkoe  AMBrady  LWWoodleigh  R Recurrence of posterior uveal melanoma after 60Co episcleral plaque therapy.  Ophthalmology. 1989;96382- 388Google ScholarCrossref
3.
Martin  JARobertson  DM Extrascleral extension of choroidal melanoma diagnosed by ultrasound.  Ophthalmology. 1983;901554- 1559Google ScholarCrossref
4.
Char  DHStone  RDIrvine  AR  et al.  Diagnostic modalities in choroidal melanoma.  Am J Ophthalmol. 1980;89223- 230Google Scholar
5.
Servodidio  CAAbramson  DHMcCormick  BFass  D Choroidal melanoma responses after brachytherapy treatment.  Acta Ophthalmol (Copenh). 1992;70687- 689Google ScholarCrossref
6.
Shields  CLShields  JAKarlsson  UMarkoe  AMBrady  LW Reasons for enucleation after plaque radiotherapy for posterior uveal melanoma: clinical findings.  Ophthalmology. 1989;96919- 924Google ScholarCrossref
7.
The Collaborative Ocular Melanoma Study Group, Design and methods of a clinical trial for a rare condition: the Collaborative Ocular Melanoma Study (COMS Report No.3).  Control Clin Trials. 1993;14362- 391Google ScholarCrossref
8.
Finger  PT Radiation therapy for choroidal melanoma.  Surv Ophthalmol. 1997;42215- 232Google ScholarCrossref
9.
Packer  SRotman  M Radiotherapy of choroidal melanoma with 125I.  Ophthalmology. 1980;87582- 590Google ScholarCrossref
10.
Coleman  DJRondeau  MJSilverman  RHLizzi  FL Computerized ultrasonic biometry and imaging of intraocular tumors for monitoring of therapy.  Trans Am Ophthalmol Soc. 1987;8549- 77Google Scholar
11.
Frazier-Byrne  SGreen  RL Ultrasound of the Eye and Orbit.  St Louis, Mo Mosby Year-Book1992;140- 150
12.
Char  DHKroll  SStone  RDHarrie  RKerman  B Ultrasonographic measurement of uveal melanoma thickness: inter-observer variability.  Br J Ophthalmol. 1990;74183- 185Google ScholarCrossref
13.
Snyder  WBFuller  DGFish  GE An inexpensive fiberoptic light pipe to aid in placement of episcleral radioactive plaques.  Ophthalmic Surg. 1988;1962- 63Google Scholar
14.
Robertson  DMFuller  DGAnderson  RE A technique for accurate placement of episcleral iodine-125 plaques.  Am J Ophthalmol. 1987;10363- 65Google Scholar
15.
Reminick  LRFinger  PTRitch  RWeiss  SIshikawa  H Ultrasound biomicroscopy in the diagnosis and management of anterior segment tumors.  J Am Optom Assoc. 1998;69575- 581Google Scholar
16.
Kidd  MNLyness  RWPatterson  CCJohnston  PBArcher  DB Prognostic factors in malignant melanoma of the choroid: a retrospective survey of cases occurring in Northern Ireland between 1965 and 1980.  Trans Ophthalmol Soc U K. 1986;105114- 121Google Scholar
17.
Coleman  DJSilverman  RHRondeau  MJLizzi  FLMcLean  IWJakobiec  FA Correlations of acoustic tissue typing of malignant melanoma and histopathologic features as a predictor of death.  Am J Ophthalmol. 1990;110380- 388Google Scholar
18.
Byrne  SFBoldt  HCDiBernardo  CL  et al. for the Collaborative Ocular Melanoma Study, Echography (ultrasound) procedures for the Collaborative Ocular Melanoma Study (COMS).  J Ophthalmic Nurs Technol. 1999;18143- 149Google Scholar
19.
Byrne  SFBoldt  HCDiBernardo  CL  et al. for the Collaborative Ocular Melanoma Study, Echography (ultrasound) procedures for the Collaborative Ocular Melanoma Study (COMS).  J Ophthalmic Nurs Technol. 1999;18219- 232Google Scholar
20.
Foldberg  RVerdick  RWeingeist  TAMontaque  PR The gross examination of eyes removed for choroidal and ciliary body melanomas.  Ophthalmology. 1986;931643- 1647Google ScholarCrossref
21.
Nicholson  DHFrazier-Byrne  SChiu  MTSchiffman  JHughes  JRNovinski  EK Echographic and histologic tumor height measurements in uveal melanoma.  Am J Ophthalmol. 1985;100454- 457Google Scholar
22.
Rask  RJensen  PK Precision of ultrasonic estimates of choroidal melanoma regression.  Graefes Arch Clin Exp Ophthalmol. 1995;233777- 782Google ScholarCrossref
23.
Silverman  RHColeman  JRondeau  MJWoods  SMLizzi  FL Measurement of ocular tumor volumes from serial, cross-sectional ultrasound scans.  Retina. 1993;1369- 74Google ScholarCrossref
24.
Jensen  PKHansen  MK Ultrasonographic three-dimensional scanning for determination of intraocular tumour volume.  Acta Ophthalmol (Copenh). 1991;69178- 186Google ScholarCrossref
25.
Fisher  YHanustsaha  PTong  SFenster  AMazarin  G Three-dimensional ophthalmic contact B-scan ultrasonography of the posterior segment.  Retina. 1998;18251- 256Google ScholarCrossref
26.
Augsburger  JJGamel  JWBailey  RSDonoso  LAGonder  JRShields  JA Accuracy of clinical estimates of tumor dimensions: a clinical-pathologic correlation study of posterior uveal melanomas.  Retina. 1985;526- 29Google ScholarCrossref
27.
Nicholson  DHFrazier-Byrne  SChiu  MTSchiffman  JHughes  JRNovinski  EK Echographic and histologic tumor height measurements in uveal melanoma[letter].  Am J Ophthalmol. 1986;101125- 128Google Scholar
28.
Favilla  IBarry  WMcNeil  JMcKenzie  JGosbell  ADEllims  P Measurement of choroidal melanomas: a comparison of methods.  Aust N Z J Ophthalmol. 1988;1659- 66Google ScholarCrossref
29.
Finger  PTRomero  JMRosen  RBIezzi  REmery  RBerson  A Three-dimensional ultrasonography of choroidal melanoma: localization of radioactive eye plaques.  Arch Ophthalmol. 1998;116305- 312Google ScholarCrossref
30.
Romero  JMFinger  PTIezzi  RRosen  RBCocker  RS Three-dimensional ultrasonography of choroidal melanoma: extrascleral extension.  Am J Ophthalmol. 1998;126842- 844Google ScholarCrossref
×