Background
Toxic endothelial cell destruction (TECD) syndrome after intraocular ophthalmic surgery is rare and can result from exposure to a variety of toxins. During January 8 to 14, 1998, 6 patients developed TECD with corneal edema associated with unreactive or dilated pupils at Hospital A.
Methods
A case patient was any Hospital A patient with TECD within 24 hours after surgery during January 5 to 14, 1998 (epidemic period). A control was any hospital A ophthalmic surgery patient without TECD during the epidemic period. The medical records of hospital A ophthalmology surgery patients during the pre-epidemic (ie, September 1, 1997-January 4, 1998) and epidemic periods were reviewed. Inductively coupled plasma atomic emission spectrometry was used to detect trace inorganic elements on sterilized surgical instruments. Cannulated surgical instruments and laboratory rinsates were perfused directly to the corneal endothelium of isolated rabbit and human corneas. Corneal endothelial ultrastructure and swelling were assessed.
Results
The rate of TECD at hospital A was higher during the epidemic than pre-epidemic period (6/12 vs 0/118, P<.001). The only change during the periods was the introduction, on November 5, 1997, of a new sterilization method, AbTox Plazlyte, for sterilization of ophthalmic surgery instruments. Findings from spectrometry revealed that copper and zinc residues were higher in instruments sterilized with Plazlyte than in those sterilized with ethylene oxide (median copper value, 7.64 mg/L vs 0.14 mg/L, respectively, P = .02; median zinc value, 5.90 mg/L vs 1.35 mg/L, respectively, P = .2). Corneal endothelial perfusion of Plazlyte sterilized–instrument rinsates or laboratory solution with copper and zinc produced irreversible damage, similar to toxic corneal endothelial destruction, to rabbit and human corneas.
Conclusion
A new sterilization method degraded brass to copper and zinc on cannulated surgical instruments resulting in TECD of the cornea.
TOXIC ENDOTHELIAL cell destruction (TECD) syndrome is a rare complication of intraocular surgery.1-6 It is clinically manifest by unexpected profound corneal edema and opacification within 24 hours after surgery. Toxins implicated in TECD include detergent residues on ophthalmic instruments,3 topical antiseptic solutions,4,5 or preservatives in intraocular medications.6 Corneal endothelial cell toxic effects have been demonstrated experimentally with presurgical topical antiseptic solutions,5 intraocular irrigating solutions,7 high concentrations of intraocular medications,8 antibiotics in corneal storage media,9,10 preservatives in medications,3,6,8,11-13 detergent residues on instruments,14 hydrogen peroxide,15 or intraocular air.16 Corneal edema following ophthalmic surgery also can result from mechanical trauma, high intraocular pressure, or inflammation.17 Severe corneal endothelial decompensation can require corneal transplantation. Of 1.4 million cataract surgeries performed in the United States annually,18 it is estimated that 0.62% are complicated by corneal edema or corneal transplantation requiring rehospitalization.19
Because instruments routinely used in ophthalmic surgery often have small lumens, careful cleaning and sterilization are essential. Steam autoclaving and ethylene oxide (ETO) are two commonly used methods for sterilization of ophthalmic instruments. Each technique has its drawbacks. Deposition of rust on instruments and a decrease in sharpness can be seen with steam autoclaving. The National Institute for Occupational Safety and Health, Centers for Disease Control and Prevention (CDC), Atlanta, Ga, consider ETO to be an occupational carcinogen and reproductive toxin.20,21 Because of this and the environmentally harmful effects of ETO, the Environmental Protection Agency, Washington, DC, is encouraging health care facility personnel to reduce use of this form of sterilization.
Recently, a new method of sterilization using plasma gas technology such as the AbTox Plazlyte sterilization system (AbTox Inc, Mundelein, Ill) has been introduced.22,23 This method sterilizes instruments by vaporization of a mixture of peracetic acid, acetic acid, and hydrogen peroxide in combination with low temperature (40°C) and an electromagnetic field; the vapor is removed with argon, oxygen, and hydrogen gas.22,23
During January 8 to 14, 1998, 6 of 8 patients receiving elective intraocular surgery at one hospital (hospital A) developed endothelial cell destruction within 24 hours after surgery. When no source for the outbreak could be identified, CDC was invited to assist in the investigation.
Case and control definition
We defined a case patient as any patient undergoing intraocular surgery from January 5 to 14, 1998 (epidemic period) at hospital A who had cornea edema and opacification (TECD) within 24 hours after surgery. We defined a control patient as any hospital A patient undergoing intraocular surgery during the epidemic period who did not develop TECD. To identify case patients and determine the background rate, we reviewed medical records of patients undergoing intraocular surgery at hospital A during the pre-epidemic (September 1, 1997-January 4, 1998) and epidemic periods.
Analytical epidemiology and statistical methods
A case-control study was conducted. Control patients were all patients who had an intraocular surgical procedure on the same day as the case patients. Data were collected on standardized forms, entered into a computer, and analyzed using Epi Info version 6.04 software (CDC) or SAS for personal computers (SAS Institute Inc, Cary, NC). Data collected included race; age; sex; dates of admission, surgery, and discharge; preoperative and postoperative vision assessment (≤24 hours and at 1-year follow-up); reason for eye surgery; type of surgery performed; type and model of intraocular lens (IOL) implant; type of anesthesia; all operative personnel including surgeons, anesthesiologists, and scrub or circulating nurses; medications used before, during, or after surgery; underlying disease; and instrument sterilization methods.
An ophthalmic surgical procedure (cataract removal) was observed. Infection control policies and procedures before, during, and after ophthalmic surgery were reviewed and observed, including the cleaning, rinsing, and packaging of surgical instruments.
To determine whether the method of sterilization may have been associated with adverse outcomes, we randomly selected 16 ophthalmic surgery instruments (ie, scissors, forceps, cannulas, hooks) from a hospital A ophthalmic surgical kit for chemical evaluation. All 16 instruments were placed in Micropaks (Riley Medical Inc, Auburn, Me), a plastic microsurgical instrument tray, and sterilized (Figure 1). Then, 8 instruments were sterilized with the Plazlyte system and 8 with ETO at hospital A. Following sterilization, these instruments in trays were forwarded to CDC for chemical analysis. At CDC, the 16 instruments were removed from the trays and rinsed with water polished to 18 MΩ-cm resistance with a Milli-Q system (Millipore Corp, Bedford, Mass). The rinsates were then evaluated by inductively coupled argon plasma atomic emission spectrometry24,25 (JY 70 plus; Instrument SA Inc, Edison, NJ) to determine the presence of any trace elements. To determine the effect of instrument rinsing on residual trace element contaminants, 9 of 16 instruments were washed an additional 3 times with Milli-Q water and evaluated after each rinse by atomic emission spectrometry for residual trace elements.
Next, to determine whether the type of packaging and the method of sterilization may have been associated with the adverse outcome, we resterilized 15 instruments, 8 of the initial 16 and 7 additional instruments (Figure 1). To determine if the adverse reactions could be associated specifically with hospital A sterilization (Plazlyte or ETO) machines, we sent instruments to be sterilized to hospital B in Dallas, Tex. Eight instruments were individually packaged in a Tyvek Mylar (DuPont, Wilmington, Del) paper and plastic pouch; 4 were sterilized in Plazlyte, (2 at hospital A; 2, hospital B) and 4 in ETO (2 at hospital A; 2, hospital B). The remaining 7 instruments were sterilized at hospital A in Micropak trays—5 in Plazlyte and 2 in ETO. To determine if any trace elements were present, these 15 instruments were rinsed in balanced salt solution (BSS) (Cytosol Laboratories Inc, Braintree, Mass), and the rinsates were evaluated by atomic emission spectrometry.24,25
To detect whether residual peracetic (peroxyacetic) acid and hydrogen peroxide were present on sterilized ophthalmic instruments, 10 mL of the rinsate was collected from the instruments. Then, an analytical titration method was performed in which hydrogen peroxide was titrated with ceric sulfate and the presence of peracetic acid was back titrated using starch indicator and sodium thiosulfate.26
An in vitro rabbit corneal endothelial perfusion model was used to evaluate the effects of the poststerilized instrument rinsates and various test solutions.15,27,28 Additionally, a human corneal endothelial perfusion model using Optisol GS–stored eye bank corneas (Chiron, Claremont, Calif) was used for evaluating laboratory-prepared solutions of 10-mg/L copper and 10-mg/L zinc (n = 5).
New Zealand white rabbits weighing 2 to 3 kg were anesthetized with an intramuscular mixture of ketamine hydrochloride, 0.5 mL (30 mg/kg of body weight), and xylocaine hydrochloride, 0.5 mL (4 mg/kg of body weight). The rabbits were euthanized with an intracardiac injection of pentobarbital sodium, 324 mg/mL (1.0 Euthanasia-5 solution [1 mg/2.5-kg rabbit]; Henry Schein Inc, Fort Washington, NY), and the eyes were enucleated with the conjunctiva and eyelids intact. All experiments were conducted under the Association for Research in Vision and Ophthalmology, Bethesda, Md, guidelines for animal research.
The rabbit corneas were mounted in an in vitro specular microscope. The test solutions were perfused directly to the corneal endothelium (rabbit and human) after a 1-hour stabilization perfusion period with a control solution (BSS Plus; Alcon Laboratories Inc, Ft Worth, Tex). Temperature was maintained at 37°C, perfusion pressure at 15 mm Hg, and perfusion rate at 30 µL per minute. There were 31 solutions used; 9 were rinsates from Plazlyte-sterilized instruments; 5, rinsates from ETO-sterilized instruments; 6, laboratory-prepared solutions of 10-mg/L copper or 10-mg/L zinc; 4, 10-mg/L copper only; and 7, 10-mg/L zinc only. Each of 31 rabbits had 1 cornea perfused with a test solution and the other cornea perfused with a BSS Plus control solution. We measured the change in corneal thickness (micrometers) over time (hours) and calculated a swelling rate (micrometers per hour) by linear regression analysis. At the end of perfusion, the corneas were fixed in 2.5% glutaraldehyde (Electron Microscopy Sciences, Fort Washington, Penn) in 0.1 mol/L cacodylate buffer and examined by scanning electron microscopy (SEM) and transmission electron microscopy (TEM) to determine if TECD had occurred.27 The tissue parameters for human corneas stored in Optisol GS were as follows: age 60.0 ± 11.4 years; time from death to enucleation, 3.2 ± 0.7 hours; and storage time, 5.5 ± 0.5 days.
Descriptive and analytical epidemiology
Six patients met the case definition. Toxic endothelial cell destruction after intraocular ophthalmic surgery was more frequent during the epidemic than pre-epidemic period (6/12 vs 0/118; P<.001). All case patients were men and ranged in age from 43 to 85 years (median age, 67 years). All had long-term systemic diseases, such as coronary artery disease, diabetes, or hypertension (Table 1).
Two case patients had extracapsular cataract extraction and a posterior chamber IOL implant; 2, cataract removal by phacoemulsification and a posterior chamber IOL implant; 1, repositioning of a previously implanted anterior chamber IOL; and 1, a trabeculectomy for glaucoma. Case patients had corneal edema associated with visual loss within 24 hours of postoperative examination. At 1-year follow-up, 3 of 6 case patients had received a corneal transplant for persistent corneal edema. Of 3 remaining case patients, corneal edema has resolved in 2, and 1 has slowly resolving corneal edema (Table 2).
When we compared case with control patients, there were no significant differences in the following preoperative, intraoperative, or postoperative conditions: medications, including local or general anesthesia; operating room; operative personnel, including surgeons, anesthesiologists, and scrub or circulating nurses; preoperative vision; length of surgery or time under anesthesia; type of surgery; or brand of lens implant. In contrast, case patients had poorer vision and greater postoperative corneal thickness than controls (Table 1).
After introduction of the AbTox Plazlyte system, some ophthalmology personnel noticed blue-green residues on some of the instruments and commented that they smelled a pungent odor similar to acetic acid when opening instrument packs. Therefore, we next reviewed intraocular surgery procedures, including instrument cleaning and reprocessing. There had been no changes in surgical procedures. In contrast, surgery instrument sterilization practices had changed with the introduction of the AbTox Plazlyte sterilization system in November 1997. Postoperatively, instruments were washed with sterile water, placed in a solution of Klenzyme (Steris Co, St Louis, Mo) and tap water for 10 minutes, rinsed with tap water, washed by hand in a solution of Klenzyme and tap water, and rinsed again with tap water. Next, they were hand-rinsed sequentially with tap water, deionized water, and distilled water. Then, all instruments were hand-dried; the smaller instruments and those with lumens were blown dry with compressed air. All ophthalmic surgery instruments to be sterilized were placed in Micropaks and sterilized in the Plazlyte machine.
Next, we evaluated the relationship between sterilization method and residual inorganic metals (copper and zinc), including the influences of packaging materials on metal residuals and the importance of rinsing. First, we assessed the relationship between sterilization method (Plazlyte vs ETO) and presence of residual trace elements. Eight of 16 instruments sterilized in Micropaks by using either ETO or Plazlyte had measurable amounts of copper and zinc by atomic emission spectrometry; all 8 were cannulas with brass hubs and very small lumens, which are placed directly into the eye during surgery (Figure 1). Of 8 cannulated instruments sterilized, the 4 sterilized in the Plazlyte machine had higher levels of copper and zinc than the 4 sterilized in ETO, although the latter did not reach statistical significance (median copper values: 7.64 vs 0.14 mg/L, respectively, P = .02; median zinc values: 5.90 vs 1.35 mg/L, respectively, P = .20).
Second, we assessed the influence of type of packaging (Micropaks vs Tyvek Mylar) and method of sterilization (Plazlyte vs ETO) on instrument copper or zinc residuals. Also, we evaluated whether copper or zinc would be found on cannulas after sterilization in hospital B's machines (Plazlyte vs ETO). The 8 cannulas placed in Micropaks and sterilized by using the Plazlyte method had rinsates with significantly higher levels of copper and zinc than rinsates from instruments sterilized in Micropaks and ETO (Table 3). All rinsates from Plazlyte-sterilized cannulas individually packaged in Tyvek Mylar had lower levels of copper and zinc than the rinsates from cannulas sterilized in plastic Micropaks. The copper and zinc residuals were significantly lower in ETO-sterilized instruments whether in Tyvek Mylar or Micropaks. Ophthalmic surgical instruments sterilized with Plazlyte at hospital A and at hospital B had similar copper and zinc levels.
Third, we determined if sequential rinsing of sterilized ophthalmic surgical instruments would affect the residual amount of copper or zinc. Nine instruments, 2 of which were cannulas, sterilized in the Plazlyte machine, rinsed once, were rinsed 3 additional times. Findings from atomic emission spectrometric analysis of these 2 cannula rinsates showed a rapid decrease in the trace elements in the first 5 mL wash, with very low levels of copper and zinc in the subsequent washes (Table 4). The other noncannulated instruments had little to no initial measurable amounts of copper and zinc. When we examined the Plazlyte-sterilized cannulas for residual sterilant (ie, peracetic acid or hydrogen peroxide), neither was detected.
Next, we performed in vitro rabbit corneal endothelial perfusions to assess whether the rinsates from the instruments would cause TECD. Nine rabbit corneas were perfused with rinsates from Plazlyte-sterilized cannulas (ie, BSS Plus rinsed through the sterilized cannulas); 5 were infused with rinsates from ETO-sterilized cannulas. Additionally, 6 were perfused with a laboratory-made solution of 10-mg/L copper, 10-mg/L zinc, and BSS Plus; 4 with 10-mg/L copper only; and 7 with 10-mg/L zinc only.
The rabbit corneas perfused with ETO cannula rinsates or BSS Plus alone (controls) showed minimal corneal swelling (7.0 ± 1.1 µm/h vs 3.8 ± 1.7 µm/h, respectively) (Figure 2) with no significant differences between the two. Findings from SEMs and TEMs revealed no structural damage with a normal hexagonal mosaic (SEM), intact borders, and normal intracellular organization (TEM) with no evidence of endothelial cell edema (Figure 3).
In contrast, rabbit corneas perfused with Plazlyte cannula rinsates swelled at 35.5, 38.0, and 16.6 µm/h, vs BSS Plus, 11.2 ± 1.5 µm/h, and corneas exposed to the laboratory-made solutions of copper and zinc swelled at 34.0 ± 3.0 µm/h vs BSS Plus, 8.5 ± 1.8 µm/h, P<.001 (Figure 4 and Figure 5). Rabbit corneas perfused with either 10-mg/L zinc only (24 ± 2.7 µm/h vs BSS Plus, 13.1 ± 2.0 µm/h, P<.01) or 10-mg/L copper only (21 ± 4.2 µm/h vs BSS Plus, 5.9 ± 2.9 µm/h, P<.03) showed less corneal swelling than those perfused with both (Figure 5). Electron micrographs of Plazlyte rinsate–perfused endothelial cells demonstrated marked endothelial cell destruction (SEM) with cell borders pulling apart and endothelial cell edema (TEM) compared with normal-appearing controls (Figure 6). Findings from SEM and TEM of endothelial cells perfused with laboratory-made solutions of copper and zinc showed endothelial cell damage (SEM) with intracellular vacuolization and disorganization (TEM) compared with normal-appearing controls (Figure 7).
We performed in vitro corneal endothelial perfusions with Optisol GS–stored human corneas to assess how human corneas perfused with laboratory-prepared samples of 10-mg/L copper and 10-mg/L zinc would compare with results of the perfused rabbit corneas of 10 mg/L-copper and 10 mg/L-zinc. Similar to results seen in the rabbit, the human corneas swelled significantly (28.2 ± 0.9 µm/hr vs BSS Plus, 14.2 ± 0.9 µm/hr, P<.001). Electron micrographs reveal endothelial cell destruction (SEM) with large vacuolization and endothelial cell edema (TEM) (Figure 8).
We investigated an outbreak of profound corneal endothelial decompensation observed within 24 hours after surgery. Our initial epidemiologic investigation failed to identify any risk factors for adverse outcome. The recent changes that occurred in ophthalmic surgery instrument sterilization methods led us to evaluate the relationship between these changes and the adverse outcomes. We found that rinsates of cannulated sterilized ophthalmic surgery instruments with small lumens had measurable copper and zinc levels. Regardless of the packaging, our laboratory experiments showed that rinsates from cannulas sterilized with the Plazlyte system had elevated levels of both copper or zinc compared with instruments sterilized in ETO; these levels were elevated whether the instruments were sterilized at Hospital A or Hospital B, suggesting that the method of sterilization rather than specific hospital practices was responsible. Rinsate from cannulas sterilized with Plazlyte in surgical kits had the highest levels of copper and zinc.
Since the Plazlyte sterilization machine was introduced for instrument sterilization in November 1997, it is unclear why more patients have not developed TECD. We hypothesize that cannulated instruments that were manufactured with chrome-covered brass hubs may have been oxidized by the acetic and peracetic acids from the Plazlyte sterilization machine. The chrome's degradation, through repeated reprocessing, exposed the brass; then, brass decomposed into its primary elements, copper and zinc. It also is possible that the acetic and peracetic acids, used during the Plazlyte sterilization process, did not quickly dissipate, especially in those cannulas sterilized in Micropaks. The increase in contact time of the acetic and peracetic acids on the instruments may have allowed oxidation of the brass to occur, each time exposing more and more of the brass. This process may have been repeated each time the cannulas were sterilized. Following the sterilization, copper and zinc would have remained in the cannula channel and then flushed into the patient's eye during the surgical procedure, producing TECD.
Since the oxidation process would have occurred over time, the breakdown of the brass by the acetic and peracetic acids may have been related to the number of times the cannulas were sterilized in the Plazlyte machine or the age of the cannulas. Hospital personnel did not always record which kit instrument was used or how many times it was previously used or sterilized. Hospital personnel did not perform many (<5 per day) intraocular ophthalmic surgery procedures, so it is possible that some surgical kits were used rarely and not repeatedly resterilized; thus, little oxidation of the cannula brass hubs of these instruments may have occurred. These later instruments may have been used during the eye procedures of those patients who did not develop TECD. Alternatively, the copper and zinc residuals were water soluble. Sufficient rinsing before the procedure could reduce or eliminate the copper and zinc residuals, reducing or diminishing the risk of their infusion into the eye. Hospital A personnel noted that it was their policy to rinse routinely the instruments and irrigate cannulas before their use in the eye.
Contamination of instruments with the trace metals copper and zinc would allow the formation of free hydroxyl and peroxyl radicals within the eye. Any of these reactive elements would indiscriminately combine with intraocular tissues, resulting in the unexpected TECD. Additionally, the smell of acetic acid on some instruments suggests that residual hydrogen peroxide was present, although we were not able to detect it.
The AbTox Plazlyte machine had not been approved by the Food and Drug Administration (FDA), Washington, DC, for instrument sterilization. The FDA had approved an earlier design of the Plazlyte machine for use on stainless steel items but not instruments with small lumens or hinges, such as many intraocular surgery instruments. Furthermore, instruments sterilized in the AbTox Plazlyte machine require careful cleaning and drying before sterilization to avoid residual chemicals or water, which can react with the Plazlyte chemicals or gas to produce by-products.
The results of this investigation led to a CDC Morbidity Mortality Weekly Report29 and an FDA safety alert about the use of AbTox Plazlyte Sterilization System in sterilizing eye instruments.29-31 Until the FDA approves its use, cannulated instruments should not be sterilized in the Plazlyte machine.
Corneal endothelial cells are a sensitive indicator of potential toxins inadvertently introduced during ophthalmic surgery. Our case patients demonstrated the potentially catastrophic effect that copper and zinc contamination can have on intraocular tissues. This investigation documented the importance of active postmarket surveillance for complications associated with newly introduced medical devices or sterilization methods. Furthermore, it emphasizes that medical personnel should ensure that the FDA has approved a medical device or sterilization method before it is used on patients or patient equipment. As new sterilization methods and devices are developed and implemented, health care personnel must be aware of and determine whether changes in practice are safe to avoid adverse patient outcomes.
Accepted for publication February 8, 2000.
This study was supported in part by the National Institutes of Health, National Eye Institute, Bethesda, Md, grant EY00933 (Dr Edelhauser).
Use of trade names and commercial sources is for identification purposes only and does not imply endorsement by the Public Health Service or the US Department of Health and Human Services.
We thank Douglas R. Dobson, BS, and H. Denny Donnell, MD, MPH, at the Missouri Department of Health, Jefferson City, and Phillip Boston, BS, and Jeffery A. Hangartner, BS, at the Food and Drug Administration, St Louis, Mo, for their assistance in this investigation.
Toxic Endothelial Cell Destruction Syndrome Investigative Team: Sandra Sides, RN, Chaille Fisher, RN, BSN, John Cochran Division, Veteran Affairs Medical Center, St Louis, Mo; and Johnny M. Khoury, MD, Department of Ophthalmology and Visual Sciences, Washington University, St Louis, and the Ophthalmology Section, Surgical Service of the St Louis Veteran Affairs Medical Center.
Reprints: William R. Jarvis, MD, CDC, Hospital Infections Program, 1600 Clifton Rd NE, MS-E69, Atlanta, GA 30333.
1.Glasser
DBSchultz
ROHyndiuk
RA The role of viscoelastics, cannulas, and irrigating solution additives in post-cataract surgery corneal edema: a brief review.
Lens Eye Toxic Res. 1992;9351- 359
Google Scholar 2.Glasser
DB Pathophysiology of corneal endothelial dysfunction. Chandler
JWSugar
JEdelhauser
HFeds
External Disease Cornea, Conjunctiva, Sclera, Eyelids, Lacrimal System. St Louis, Mo Mosby–Year Book Inc1994;81- 819
Google Scholar 3.Breebaart
ACNuyts
RPels
EEdelhauser
HFVerbraak
FD Toxic endothelial cell destruction of the cornea after routine extracapsular cataract surgery.
Arch Ophthalmol. 1990;1081121- 1125
Google ScholarCrossref 4.Phinney
RBMondino
BJHofbauer
JD
et al. Corneal edema related to accidental Hibiclens exposure.
Am J Ophthalmol. 1988;106210- 215
Google Scholar 5.Mac Rae
SMBrown
BEdelhauser
HF The corneal toxicity of presurgical skin antiseptics.
Am J Ophthalmol. 1984;97221- 232
Google Scholar 6.Edelhauser
HFHyndiuk
RAZeeb
ASchultz
RO Corneal edema and the intraocular use of epinephrine.
Am J Ophthalmol. 1982;93327- 333
Google Scholar 7.Edelhauser
HFGonnering
RVan Horn
DL Intraocular irrigating solutions: a comparative study of BSS Plus and lactated Ringer's solution.
Arch Ophthalmol. 1978;96516- 520
Google ScholarCrossref 8.Kadonosono
KIto
NYazama
F
et al. Effects of intracameral anesthesia on the corneal endothelium.
J Cataract Refract Surg. 1998;241377- 1381
Google ScholarCrossref 9.Garcia-Ferrer
FJPepose
JSMurray
PRGlaser
SRLass
JHGreen
WR Antimicrobial efficacy and corneal endothelial toxicity of DexSol corneal storage medium supplemented with vancomycin.
Ophthalmology. 1991;98863- 869
Google ScholarCrossref 10.Hyndiuk
RASchultz
RO Overview of the corneal toxicity of surgical solutions and drugs: and clinical concepts in corneal edema.
Lens Eye Toxic Res. 1992;9331- 350
Google Scholar 11.Green
KCheeks
LSteward
DATrask
D Role of toxic ingredients in silicone oils in the induction of increased corneal endothelial permeability.
Lens Eye Toxic Res. 1992;9377- 384
Google Scholar 12.Ohguro
NMatsuda
MKinoshita
S The effects of denatured sodium hyaluronate on the corneal endothelium in cats.
Am J Ophthalmol. 1991;112424- 430
Google Scholar 13.Chang
CHLin
CPWang
HZ Cytotoxicity of intracameral injection drugs to corneal endothelium as evaluated by corneal endothelial cell culture.
Cornea. 1995;1471- 76
Google ScholarCrossref 14.Nuyts
RMEdelhauser
HFPels
EBreebaart
AC Toxic effects of detergents on the corneal endothelium.
Arch Ophthalmol. 1990;1081158- 1162
Google ScholarCrossref 15.Artola
AAlio
JLBellot
JLRuiz
JM Protective properties of viscoelastic substances (sodium hyaluronate and 2% hydroxymethylcellouse) against experimental free radical damage to cornea endothelium.
Cornea. 1993;12109
Google ScholarCrossref 16.Olson
RJ Air and the corneal endothelium: an in vivo specular microscopy study in cats.
Arch Ophthalmol. 1980;981283- 1284
Google ScholarCrossref 17.Edelhauser
HFGlasser
DB Surgical pharmacology intraocular solutions and drugs for cataract surgery. Buratto
Led
Phacoemulsification Principles and Techniques Thorofare, NJ Stark Inc1998;275- 292
Google Scholar 18.Agency for Health Care Policy and Research, Cataracts in Adults: Management of Functional Impairment. Rockville, Md US Dept of Health and Human Services, Agency for Health Care Policy and Research1993;AHCPR publication 93-0542.
19.Canner
JKJavitt
JCMcBean
AM National outcomes of cataract extraction, III: corneal edema and transplant following inpatient surgery.
Arch Ophthalmol. 1992;1101137- 1142
Google ScholarCrossref 20.National Institute for Occupational Safety and Health, Current Intelligence 52: Ethylene Oxide Sterilizers in Health Care Facilities: Engineering Controls and Work Practices. Cincinnati, Ohio US Dept of Health and Human Services, Public Health Service, Centers for Disease Control, NIOSH1989;DHHS Pub No. (NIOSH) 89-115.
21.National Institute for Occupational Safety and Health, Current Intelligence 35: Ethylene Oxide (ETO): Evidence of Carcinogenicity. Cincinnati, Ohio US Dept of Health and Human Services, Public Health Service, Centers for Disease Control, NIOSH1981;DHHS publication (NIOSH) 81-130.
22.Lynch
M Gas plasma low-temperature sterilization technology.
Infect Control Today. 1998;253- 56
Google Scholar 23.Bryce
EAChia
ELogelin
GSmith
JA An evaluation of the AbTox Plazlyte Sterilization System.
Infect Control Hosp Epidemiol. 1997;18646- 653
Google ScholarCrossref 24.Kimberly
MPaschal
D Screening for selected toxic elements in urine by sequential-scanning inductively-coupled plasma atomic emission spectrometry.
Anal Chim Acta. 1985;174203- 210
Google ScholarCrossref 25.EnvironmentalProtection Agency, Methods for the Determination of Metals in Environmental Samples. Washington, DC Environmental Protection Agency1994;600/R-94-111, PB94-184942.
26.Not Available, European Pharmacopoeia, Vol 3-5. 2nd ed. Sainte-Ruffine, France Council of Europe1980;European Treaty Series No. 50.
27.McCarey
BEEdelhauser
HFVan Horn
DL Functional and structural changes in the corneal endothelium during in vitro perfusion.
Invest Ophthalmol. 1973;12410- 417
Google Scholar 28.Kim
THolley
GPLee
JHBroocker
GEdelhauser
HF The effects of intraocular lidocaine on the corneal endothelium.
Ophthalmology. 1998;105125- 130
Google ScholarCrossref 29.Not Available, Corneal decompensation after Intraocular Ophthalmic Surgery—Missouri, 1998.
MMWR Morb Mortal Wkly Rep. 1998;47306- 309
Google Scholar 30.Food and Drug Administration, A Nationwide Warning Against the Use of "AbTox" Plazlyte Sterilization System. Rockville, Md US Dept of Health and Human Services, Food and Drug Administration1998;
31.Food and Drug Administration, FDA Safety Alert: Warning Regarding the Use of the AbTox Plazlyte Sterilization System. Rockville, Md US Dept of Health and Human Services, Food and Drug Administration April13 1998;