Objective
To correlate macular thickness and retinal nerve fiber layer (RNFL) thickness in normal and glaucomatous eyes using optical coherence tomography.
Methods
Complete examination, automated achromatic perimetry, and optical coherence tomography of the peripapillary RNFL and macula were performed. Exclusion criteria were visual acuity of less than 20/40, diseases other than glaucoma, and unreliable automated achromatic perimetry. Macular thickness measurements were generated using 6 radial optical coherence tomographic scans (5.9 mm) centered on the fovea, and mean and quadrantic macular thickness values were calculated.
Results
Fifty-nine eyes of 59 patients (29 normal and 30 glaucomatous) were enrolled (mean ± SD age, 56.7 ± 20.3 years; range, 20-91 years). All eyes with glaucoma had associated visual field loss (mean ± SD mean defect, −8.4 ± 5.8 d B). Mean macular thickness was significantly associated with visual field mean defect (R2 = 0.47; P<.001), pattern standard deviation(R2 = 0.32; P<.001), and mean RNFL thickness (R2 = 0.38; P<.001). In glaucomatous eyes with visual field loss localized to 1 hemifield (n = 11), mean ± SD macular thickness in the quadrant associated with the field defect (277 ± 28 µm) was significantly less (P = .005) than in the unaffected quadrant (286± 27 µm). Mean RNFL thickness in the affected quadrant (89 ± 53 µm) was significantly thinner (P = .009) than in the unaffected quadrant (121 ± 39 µm).
Main Outcome Measures
Mean total and quadrantic macular and RNFL thickness measurements.
Conclusions
Macular thickness changes are well correlated with changes in visual function and RNFL structure in glaucoma and may be a surrogate indicator of retinal ganglion cell loss.
IN GLAUCOMA, the essential pathologic process is the loss of retinal ganglion cells and their axons.1 Numerous studies have shown that glaucomatous damage to the retinal nerve fiber layer (RNFL) precedes functional loss by as much as 5 years.2-7 Loss of the retinal ganglion cells and their axons is also known to occur in the posterior pole, where these cells may constitute 30% to 35% of the retinal thickness in the macular region.8 Studies of chronic experimental glaucoma in monkey eyes have shown a substantial loss of retinal ganglion cells in the zone surrounding the fovea.9,10 Furthermore, a significant association exists between loss of retinal thickness in the posterior pole and visual function.8
The size and anatomical distribution of retinal ganglion cells varies throughout the posterior pole.9,11 Approximately 50% of retinal ganglion cells are located in the macular region 4 to 5 mm from the center of the fovea12 with the peak density occurring 750 to 1100 µm from the foveal center where the cell density may be 4 to 6 cell bodies thick.13 Although cell diameter distribution is variable, a skewed distribution toward larger cell diameters (eg, 14-16 µm) exists in the normal retina, and such cells have been shown to be selectively lost in human and experimental models of glaucoma.1,9,11 Postmortem analysis of enucleated human glaucomatous eyes has demonstrated that a 20% loss of retinal ganglion cells throughout the central 30° of the retina was associated with a 5-d B sensitivity loss with automated perimetry; a 40% loss corresponded to a 10-d B loss in sensitivity.1
Optical coherence tomography (OCT) is a noninvasive, noncontact, transpupillary imaging technology that can image retinal structures in vivo with an axial resolution of 10 to 15 µm.14,15 It provides quantitative, objective, and reproducible assessment of retinal and RNFL thickness.16-19 Good correlation has been reported between peripapillary RNFL thickness measured by OCT and visual function16,20-23 as well as histological measurements of the nerve fiber layer thickness.24
Zeimer et al8 reported a significant correlation between glaucomatous visual field defects and reductions in macular thickness using a retinal topographer (Retinal Thickness Analyzer; Talia Technology Ltd, Neve Ilan, Israel) based on the principles of slitlamp biomicroscopy. Significant losses in retinal thickness at the posterior pole of up to 34% were reported to occur in patients with early glaucoma (mean corrected pattern standard deviation, approximately 5.5 d B). The purpose of this investigation was to evaluate the correlation between macular thickness and RNFL thickness as measured by OCT in normal eyes and eyes with moderate glaucomatous optic neuropathy.
Normal (control) and glaucomatous eyes meeting the eligibility criteria were enrolled in this prospective study. Informed consent was obtained from all subjects by means of a consent form approved by the Institutional Review Board for Human Research of the University of Miami School of Medicine, Miami, Fla. All patients underwent complete ophthalmic examination, including slitlamp biomicroscopy, gonioscopy, Goldmann applanation tonometry, dilated stereoscopic examination of the optic disc and fundus, and achromatic automated perimetry.
Control subjects had no history of ocular disease. All had intraocular pressure of 21 mmHg or less by Goldmann applanation tonometry, normal optic disc appearance based on clinical stereoscopic examination and review of stereo disc photography, and normal perimetry. Absence of glaucomatous optic neuropathy was defined as vertical cup-disc asymmetry of less than 0.2, a cup-disc ratio of less than 0.6, and an intact neuroretinal rim without peripapillary hemorrhages, notches, localized pallor, or RNFL defect. Normal visual field indices were defined as a mean defect and corrected pattern standard deviation within 95% confidence limits and a glaucoma hemifield test result within normal limits.
Glaucomatous optic neuropathy was defined as either cup-disc asymmetry between fellow eyes of greater than 0.2, rim thinning, notching, excavation, or RNFL defect. Patients with glaucoma had glaucomatous optic nerve damage and associated achromatic visual field loss in the corresponding hemifield location. Patients with achromatic visual field abnormalities had at least 1 confirmatory visual field examination. Eyes with visual acuity of less than 20/40, retinal disease, or unreliable perimetry (greater than 25% fixation losses or false-positive and false-negative rates) were excluded from this investigation.
Optical coherence tomographic imaging (OCT 1; Zeiss-Humphrey Systems, Dublin, Calif) of the macular and peripapillary RNFL was performed in all patients within 6 months of clinical examination; OCT was performed using near-infrared, low-coherence illumination (840 nm) with a tissue resolution of approximately 10 to 17 µm.14,17,18 Image acquisition was performed by one of us (H.B.) and analyzed with version A6.1 software. After pupillary dilation to a minimum diameter of 5 mm, three 360° circular scans with a diameter of 3.4 mm centered on the optic disc were performed. Scan acquisition time was 1.0 second. Each scan consisted of 100 individual A-scan samples evenly distributed along the circle circumference. Mean RNFL thickness was calculated from the values of the 3 scans. Macular thickness measurements were generated using 6 radial scans (each 5.9 mm) centered on the fovea (Figure 1). These scans are processed to produce a topographic map of the macula (Figure 2B). Mean quadrantic measurements were generated from the retinal map consisting of sectoral measurements located 0.5 mm to 1.7 mm outside the center of the fovea.
Statistical analysis was performed using JMP software (SAS Institute Inc, Cary, NC). Analysis of variance was used to compare different measures among the groups. Statistical associations among macular thickness values, peripapillary RNFL thickness, and visual function were evaluated using the Pearson correlation coefficient. No adjustments were made for multiple comparisons. P≤.05 was considered statistically significant.
Fifty-nine eyes of 59 patients (29 normal and 30 glaucomatous) were enrolled (mean ± SD age, 56.7 ± 20.3 years; age range, 20-91 years). All eyes with glaucoma had associated visual field loss (mean ± SD mean defect, –8.4 ± 5.8 d B). Clinical characteristics of the study population are described in Table 1. As illustrated in Table 2, mean macular thickness and mean peripapillary RNFL thickness in glaucomatous eyes were significantly less than mean macular thickness and mean peripapillary RNFL thickness in control subjects. The relationships among visual field defects and RNFL and macular thickness for one case are illustrated in Figure 2.
Correlations between macular and RNFL thickness values were evaluated among a subgroup of 11 similar patients with glaucoma and visual field loss localized to a single hemifield (Table 3). As illustrated in Figure 3, mean macular thickness in the hemifield associated with the field defect (277± 28 µm) was significantly less compared with the unaffected hemifield (286 ± 27 µm) (P = .005; paired t test). Mean RNFL thickness was significantly less in the affected hemifield (89 ± 53 µm) compared with the unaffected hemifield (121 ± 39 µm) (P = .009; paired t test). Macular thickness was significantly correlated with RNFL thickness in the posterior segment quadrants associated with the field defect (P<.001; paired t test).
As illustrated in Figure 4, significant correlations were observed between OCT-generated mean macular thickness and visual field mean defect (R2 =0.47; P<.001) and pattern standard deviation (R2 = 0.32; P<.001). Significant correlations were similarly observed between mean peripapillary RNFL thickness and visual field mean defect (R2 = 0.45; P<.001) and pattern standard deviation(R2 = 0.35; P<.001). Mean macular thickness was significantly associated with mean RNFL thickness(R2 = 0.38; P<.001)(Figure 5).
Glaucoma is a complex multifactorial disorder characterized by a typical pattern of optic nerve damage and visual field loss that is usually but not always associated with elevated intraocular pressure. Accepted parameters for monitoring glaucoma include descriptions and photography of the optic disc appearance (eg, cup-disc ratio), measurement of intraocular pressure, and periodic threshold perimetry. Advances in posterior segment imaging technology14,15,25-28 provide a means for generating structural data useful in monitoring eyes with glaucomatous optic nerve damage. Objective, quantitative measurements of the optic nerve and surrounding RNFL generated with these technologies correlate with known characteristics of optic disc structure and visual function.
The results of this report suggest that macular thickness measurements generated with OCT represent a neglected structural end point for glaucoma. Although glaucoma is an optic nerve disorder, the fundamental defining abnormality is localized at the level of the retinal ganglion cell. Glaucoma is known to cause loss of ganglion cells and their axons, leading to a reduction in the thickness of the RNFL.1,4,29,30 Macular thickness measurements represent a surrogate indicator of retinal ganglion cell thickness and could prove to have clinical value for glaucoma diagnosis and detection of change. Our results support this hypothesis and illustrate a significant correlation between macular thickness and 2 established indicators of glaucomatous damage: RNFL loss and loss of visual field.
We found significant differences in mean macular thickness between control subjects and patients with moderately advanced glaucoma using direct measurements of retinal thickness generated with OCT. As illustrated in Figure 4, both macular and RNFL thickness assessments were strongly correlated with visual field global indices. The parallel slopes of the regression lines suggest that both parameters should be equally robust discriminators for disease detection. Furthermore, macular thickness and RNFL thickness assessments were significantly associated with each other, suggesting concordance between loss of retinal ganglion cells and their axons. These observations were emphasized in patients with visual field loss confined to a single hemifield who illustrated regional reductions in macular thickness that corresponded topographically to regional reductions in RNFL. Longitudinal studies are necessary to determine whether macular thickness reductions precede RNFL loss or vice versa. For macular thickness measurements to be clinically useful, collection of age-corrected normative data with 95% confidence limits is necessary. Furthermore, to detect glaucomatous progression, statistical criteria are necessary to differentiate test-retest variability from true biological change.
Eyes with visual field loss confined to a single hemifield had a mean difference between the affected and unaffected macular quadrant thickness of only 9 µm, suggesting that regional macular thickness data may have limitations. A 3.4-mm diameter macular map corresponds to a visual angle of 6° from fixation.31 Eyes with early visual field defects located far from fixation may have had undetectable ganglion cell loss using a macular map of this diameter. Furthermore, the topographic location of retinal ganglion cell death may have been temporal to the macular quadrant suspected to be associated with the field defect. As illustrated in Figure 2, the inferior macular quadrant was suspected to be associated with the superior field defect; however, perhaps the most significant reduction in macular thickness occurred in the inferior temporal fovea. Finally, although the foveola (diameter approximately 300 to 400 µm)32 is devoid of ganglion cells, relevant data may have been inadvertently eliminated by neglecting the central 1.0-mm diameter of the macular map thought to consist largely of photoreceptor cells. Nevertheless, a statistically significant reduction in thickness was identified in the macular quadrant predicted to be associated with the field defect. Advances in axial resolution of this technology, and software designed to extract data from more pertinent topographic areas, would be expected to improve detection of retinal ganglion cell loss in glaucoma.
Based on the principle of low-coherence interferometry, OCT provides high-resolution, cross-sectional imaging of the retina and the RNFL.14-18 A high level of correlation between OCT-generated RNFL thickness and visual function has been reported by several authors.16,21,22,33 As presently configured, OCT employs 100 A-scans with an axial resolution of 10 to 20 µm, which limits the ability to visualize and measure the retinal ganglion cell layer. A modification of this technology that was recently approved by the Food and Drug Administration employs 512 A-scans and has an axial resolution of approximately 8 µm with no need for pupillary dilation. This modification may provide a potential means to directly visualize and measure the retinal ganglion cell layer. Emerging therapeutic strategies such as neuroprotection emphasize the need to identify and measure such cells for glaucoma diagnosis and monitoring.
There is considerable evidence to support the relationship between RNFL loss and reductions in visual function using various posterior segment imaging technologies.20,22,23,28,34,35 The discriminating power of these instruments is limited by the wide distribution of normative RNFL data among the general population as well as technological assumptions.36,37 It remains unclear whether macular thickness loss may be a more robust indicator of early glaucomatous damage. Furthermore, as with RNFL thickness assessments, longitudinal studies are necessary to validate their ability to detect glaucomatous progression. It is important to emphasize that macular thickness measurements have limited use for monitoring glaucoma in eyes with macular comorbidity. Thus, eyes with diabetic or age-related maculopathy are not candidates for monitoring macular thickness changes as a strategy for glaucoma diagnosis or detection of glaucomatous progression.
In conclusion, macular thickness changes are well correlated with changes in visual function and RNFL structure in glaucoma and may represent a surrogate indicator of retinal ganglion cell loss. Macular thickness measurements with OCT may provide a new approach for the detection and monitoring of glaucomatous damage.
Corresponding author and reprints: David S. Greenfield, MD, Bascom Palmer Eye Institute, 7108 Fairway Dr, Suite 340, Palm Beach Gardens, FL 33418(e-mail: dgreenfield@med.miami.edu).
Submitted for publication March 4, 2002; final revision received June 25, 2002; accepted August 29, 2002.
This study was supported in part by the New York Community Trust, New York; the Kessel Foundation, Bergenfield, NJ; a grant from Barney Donnelley, Palm Beach, Fla; and grant R01-EY08684 from the National Institutes of Health, Bethesda, Md.
1.Quigley
HADunkelberger
GRGreen
WR Retinal ganglion cell atrophy correlated with automated perimetry in human eyes with glaucoma.
Am J Ophthalmol. 1989;107453- 464
Google Scholar 2.Hoyt
WFFrisén
LNewman
NM Funduscopy of nerve fiber layer defects in glaucoma.
Invest Ophthalmol. 1973;12814- 829
Google Scholar 3.Quigley
HAMiller
NRGeorge
T Clinical evaluation of nerve fiber layer atrophy as an indicator of glaucomatous optic nerve damage.
Arch Ophthalmol. 1980;981564- 1571
Google ScholarCrossref 4.Sommer
AKatz
JQuigley
HA
et al. Clinically detectable nerve fiber layer atrophy precedes the onset of glaucomatous field loss.
Arch Ophthalmol. 1991;10977- 83
Google ScholarCrossref 5.Quigley
HAKatz
JDerick
RJGilbert
DSommer
A An evaluation of optic disc and nerve fiber layer examinations in monitoring progression of early glaucoma damage.
Ophthalmology. 1992;9919- 28
Google ScholarCrossref 6.Airaksinen
PJDrance
SMDouglas
GR
et al. Diffuse and localized nerve fiber loss in glaucoma.
Am J Ophthalmol. 1984;98566- 571
Google Scholar 7.Harwerth
RSCarter-Dawson
LShen
F
et al. Ganglion cell losses underlying visual field defects from experimental glaucoma.
Invest Ophthalmol Vis Sci. 1999;402242- 2250
Google Scholar 8.Zeimer
RAsrani
SZou
SQuigley
HAJampel
H Quantitative detection of glaucomatous damage at the posterior pole by retinal thickness mapping.
Ophthalmology. 1998;105224- 231
Google ScholarCrossref 9.Glovinsky
YQuigley
HAPease
ME Foveal ganglion cell loss in size dependent in experimental glaucoma.
Invest Ophthalmol Vis Sci. 1993;34395- 400
Google Scholar 10.Frishman
LJShen
FFDu
L
et al. The scotopic electroretinogram of macaque after retinal ganglion cell loss from experimental glaucoma.
Invest Ophthalmol Vis Sci. 1996;37125- 141
Google Scholar 11.Glovinsky
YQuigley
HADunkelberger
GR Retinal ganglion cell loss is size dependent in experimental glaucoma.
Invest Ophthalmol Vis Sci. 1991;32484- 491
Google Scholar 13.Wässle
HGrunert
URoherbbeck
JBoycott
BB Cortical magnification factor and the ganglion cell density of the primate retina.
Nature. 1989;341643- 646
Google ScholarCrossref 15.Izatt
JAHee
MRSwanson
EA
et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography.
Arch Ophthalmol. 1994;1121584- 1589
Google ScholarCrossref 16.Schuman
JSHee
MRPuliafito
CA
et al. Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography.
Arch Ophthalmol. 1995;113586- 596
Google ScholarCrossref 17.Hee
MRIzatt
JASwanson
EA
et al. Optical coherence tomography of the human retina.
Arch Ophthalmol. 1995;113325- 332
Google ScholarCrossref 18.Schuman
JSPedut-Kloizman
THertzmark
E
et al. Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography.
Ophthalmology. 1996;1031889- 1898
Google ScholarCrossref 19.Baumann
MGentile
RCLiebmann
JMRitch
R Reproducibility of retinal thickness measurements in normal eyes using optical coherence tomography.
Ophthalmic Surg Lasers. 1998;29280- 285
Google Scholar 20.Bowd
CAZangwill
LMBerry
CC
et al. Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function.
Invest Ophthalmol Vis Sci. 2001;421993- 2003
Google Scholar 21.Zangwill
LMWilliams
JBerry
CCKnauer
SWeinreb
RN A comparison of optical coherence tomography and retinal nerve fiber layer photography for detection of nerve fiber layer damage in glaucoma.
Ophthalmology. 2000;1071309- 1315
Google ScholarCrossref 22.Hoh
STGreenfield
DSMistlberger
ALiebmann
JMIshikawa
HRitch
R Optical coherence tomography and scanning laser polarimetry in normal, ocular hypertensive, and glaucomatous eyes.
Am J Ophthalmol. 2000;129129- 135
Google ScholarCrossref 23.Mistlberger
ALiebmann
JMGreenfield
DS
et al. Heidelberg retina tomography and optical coherence tomography in normal, ocular hypertensive, and glaucomatous eyes.
Ophthalmology. 1999;1062027- 2032
Google ScholarCrossref 24.Huang
YCideciyan
AVPapastergiou
GI
et al. Relation of optical coherence tomography to microanatomy in normal and rd chickens.
Invest Ophthalmol Vis Sci. 1998;392405- 2416
Google Scholar 25.Weinreb
RNDreher
AWBille
JF Quantitative assessment of the optic nerve head with the laser tomographic scanner.
Int Ophthalmol. 1989;1325- 29
Google ScholarCrossref 26.Zangwill
LSchakiba
SCaprioli
JWeinreb
RN Agreement between clinicians and a confocal scanning laser ophthalmoscope in estimating cup-to-disc ratios.
Am J Ophthalmol. 1995;119415- 421
Google Scholar 27.Dreher
AWReiter
KWeinreb
RN Spatially resolved birefringence of the retinal nerve fiber layer assessed with a retinal ellipsometer.
Appl Opt. 1992;313730- 3735
Google ScholarCrossref 28.Weinreb
RNShakiba
SZangwill
L Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes.
Am J Ophthalmol. 1995;119627- 636
Google Scholar 29.Sommer
AMiller
NRPollack
IMaumenee
AEGeorge
T The nerve fiber layer in the diagnosis of glaucoma.
Arch Ophthalmol. 1977;952149- 2156
Google ScholarCrossref 31.Wyszecki
GStiles
WS The eye. In:
Color Science: Concepts and Methods, Quantitative Data and Formulae 2nd New York, NY John Wiley & Sons1982;98- 101
Google Scholar 32.Sjostrand
JPopovic
ZConradi
NMarshall
J Morphometric study of the displacement of retinal ganglion cells subserving cones within the human fovea.
Graefes Arch Clin Exp Ophthalmol. 1999;2371014- 1023
Google ScholarCrossref 33.Teesalu
PTuulonen
AAiraksinen
PJ Optical coherence tomography and localized defects of the retinal nerve fiber layer.
Acta Ophthalmol Scand. 2000;7849- 52
Google ScholarCrossref 34.Weinreb
RNShakiba
SSample
PA
et al. Association between quantitative nerve fiber layer measurement and visual field loss in glaucoma.
Am J Ophthalmol. 1995;120732- 738
Google Scholar 35.Zangwill
LMBowd
CBerry
CC
et al. Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph.
Arch Ophthalmol. 2001;119985- 993
Google ScholarCrossref 36.Greenfield
DSKnighton
RW Stability of corneal polarization axis measurements for scanning laser polarimetry.
Ophthalmology. 2001;1081065- 1069
Google ScholarCrossref 37.Greenfield
DSHuang
X-RKnighton
RW Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry.
Am J Ophthalmol. 2000;129715- 722
Google ScholarCrossref