Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry | Cornea | JAMA Ophthalmology | JAMA Network
[Skip to Navigation]
Access to paid content on this site is currently suspended due to excessive activity being detected from your IP address 18.206.238.77. Please contact the publisher to request reinstatement.
1.
Goldmann  HSchmidt  T Über Applanationstonometrie.  Ophthalmologica 1957;134221- 242PubMedGoogle ScholarCrossref
2.
Goldmann  HSchmidt  T Weiterer Beitrag zur Applanationstonometrie.  Ophthalmologica 1961;141441- 456PubMedGoogle ScholarCrossref
3.
Whitacre  MMStein  R Sources of error with use of Goldmann-type tonometers.  Surv Ophthalmol 1993;381- 30PubMedGoogle ScholarCrossref
4.
Whitacre  MMStein  RAHassanein  K The effect of corneal thickness on applanation tonometry.  Am J Ophthalmol 1993;115592- 596PubMedGoogle Scholar
5.
Bron  AMCreuzot-Garcher  CGoudeau-Boutillon  Sd’Athis  P Falsely elevated intraocular pressure due to increased central corneal thickness.  Graefes Arch Clin Exp Ophthalmol 1999;237220- 224PubMedGoogle ScholarCrossref
6.
Damji  KFMunger  R Influence of central corneal thickness on applanation intraocular pressure.  J Glaucoma 2000;9205- 207PubMedGoogle ScholarCrossref
7.
Doughty  MJZaman  ML Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.  Surv Ophthalmol 2000;44367- 408PubMedGoogle ScholarCrossref
8.
Ehlers  NHansen  FK Central corneal thickness in low-tension glaucoma.  Acta Ophthalmol (Copenh) 1974;52740- 746PubMedGoogle ScholarCrossref
9.
Shah  SChatterjee  AMathai  M  et al.  Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic.  Ophthalmology 1999;1062154- 2160PubMedGoogle ScholarCrossref
10.
Shah  S Accurate intraocular pressure measurement—the myth of modern ophthalmology [editorial]?  Ophthalmology 2000;1071805- 1807PubMedGoogle ScholarCrossref
11.
Stodtmeister  R Applanation tonometry and correction according to corneal thickness.  Acta Ophthalmol Scand 1998;76319- 324PubMedGoogle ScholarCrossref
12.
Ehlers  NBramsen  TSperling  S Applanation tonometry and central corneal thickness.  Acta Ophthalmol (Copenh) 1975;5334- 43PubMedGoogle ScholarCrossref
13.
Orssengo  GJPye  DC Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo.  Bull Math Biol 1999;61551- 572Google ScholarCrossref
14.
Feltgen  NLeifert  DFunk  J Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings.  Br J Ophthalmol 2001;8585- 87PubMedGoogle ScholarCrossref
15.
Wolfs  RCWKlaver  CCVVingerling  JRGrobbee  DEHofman  Ade Jong  PT Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study.  Am J Ophthalmol 1997;123767- 772PubMedGoogle Scholar
16.
Mark  HHRobbins  KPMark  TL Axial length in applanation tonometry.  J Cataract Refract Surg 2002;28504- 506PubMedGoogle ScholarCrossref
17.
Mark  HH Corneal curvature in applanation tonometry.  Am J Ophthalmol 1973;76223- 224PubMedGoogle Scholar
Clinical Sciences
April 2006

Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry

Author Affiliations

Author Affiliations: Department of Ophthalmology, University Carl Gustav Carus, Dresden (Drs Kohlhaas, Boehm, Spoerl, and Pillunat), and Department of Optometry, University of Applied Sciences, Jena (Ms Pürsten and Dr Grein), Germany. Dr Kohlhass is now with the Augenklinik, St. Johannes-Hospital, Dortmund, Germany.

Arch Ophthalmol. 2006;124(4):471-476. doi:10.1001/archopht.124.4.471
Abstract

Objective  To evaluate the effect of central corneal thickness (CCT), corneal curvature, and axial length on applanation tonometry in an in vivo study.

Methods  In a masked, prospective clinical trial, we examined 125 eyes of 125 patients scheduled for cataract surgery. Corneal curvature was measured by means of keratometry and axial length by A-scan ultrasonography. By cannulating the anterior chamber before surgery, intraocular pressure (IOP) was set to 20, 35, and 50 mm Hg in a closed system by means of a water column. After measuring thickness, the IOP was measured with an applanation tonometer. Pearson product moment correlations and multiple linear regression analyses were performed, and significance levels were evaluated by the paired, 2-tailed t test.

Results  The difference between measured and real IOP was significantly dependent (P<.001) on CCT. The associations between IOP and corneal curvature or IOP and axial length were not statistically significant (P = .31). The association between IOP reading and CCT is shown in the “Dresdner correction table,” which illustrates an approximately 1-mm Hg correction for every 25-μm deviation from a CCT of 550 μm. The correction values were positive as thickness decreased and negative as thickness increased.

Conclusions  Central corneal thickness significantly affects IOP readings obtained by applanation tonometry according to the Goldmann principle. A correction of IOP readings by considering CCT according to the Dresdner correction table might be helpful for determining an accurate IOP value.

×