Ablation profile display on LADARWave (Alcon Laboratories, Inc, Fort Worth, Texas). Cartesian coordinates are in the top left corner. The cursor is currently at the 90° meridian at the 6.0-mm zone diameter (Cartesian coordinates: 0.0, 3.0).

Ablation profile display on LADARWave (Alcon Laboratories, Inc, Fort Worth, Texas). Cartesian coordinates are in the top left corner. The cursor is currently at the 90° meridian at the 6.0-mm zone diameter (Cartesian coordinates: 0.0, 3.0).

Figure 2. 
Ablation profile of myopic astigmatism (M3) and myopic sphere (A7) algorithms. An overlay of 2 ablation profiles shows that additional ablation in the M3 algorithm occurs at peak and 4-mm, 6-mm, and 7-mm zones. The mean ablation difference (in micrometers) at each zone is shown.

Ablation profile of myopic astigmatism (M3) and myopic sphere (A7) algorithms. An overlay of 2 ablation profiles shows that additional ablation in the M3 algorithm occurs at peak and 4-mm, 6-mm, and 7-mm zones. The mean ablation difference (in micrometers) at each zone is shown.

Table 1. 
Wavefront and Manifest Refractions
Wavefront and Manifest Refractions
Table 2. 
Ablation Depths and Depth Differences for M3 and A7 Algorithms
Ablation Depths and Depth Differences for M3 and A7 Algorithms
1.
Boxer Wachler  BSHuynh  VNEl-Shiaty  AFGoldberg  D Evaluation of corneal functional optical zone after laser in situ keratomileusis.  J Cataract Refract Surg 2002;28 (6) 948- 953PubMedGoogle ScholarCrossref
2.
Holladay  JTDudeja  DRChang  J Functional vision and corneal changes after laser in situ keratomileusis determined by contrast sensitivity, glare testing and corneal topography.  J Cataract Refract Surg 1999;25 (5) 663- 669PubMedGoogle ScholarCrossref
3.
Marcos  S Aberrations and visual performance following standard laser vision correction.  J Refract Surg 2001;17 (5) S596- S601PubMedGoogle Scholar
4.
Nepomuceno  RLBoxer Wachler  BSScruggs  R Functional optical zone after myopic LASIK as a function of ablation diameter.  J Cataract Refract Surg 2005;31 (2) 379- 384PubMedGoogle ScholarCrossref
5.
McGhee  CNCraig  JPSachdev  NWeed  KHBrown  AD Functional, psychological, and satisfaction outcomes of laser in situ keratomileusis for high myopia.  J Cataract Refract Surg 2000;26 (4) 497- 509PubMedGoogle ScholarCrossref
6.
Bailey  MDMitchell  GLDhaliwal  DKBoxer Wachler  BSZadnik  K Patient satisfaction and visual symptoms after laser in situ keratomileusis.  Ophthalmology 2003;110 (7) 1371- 1378PubMedGoogle ScholarCrossref
7.
Jabbur  NSSakatani  KO’Brien  TP Survey of complications and recommendations for management in dissatisfied patients seeking a consultation after refractive surgery.  J Cataract Refract Surg 2004;30 (9) 1867- 1874PubMedGoogle ScholarCrossref
8.
Holladay  JTJanes  JA Topographic changes in corneal asphericity and effective optical zone after laser in situ keratomileusis.  J Cataract Refract Surg 2002;28 (6) 942- 947PubMedGoogle ScholarCrossref
9.
Mrochen  MSeiler  T Influence of corneal curvature on calculation of ablation patterns used in photorefractive laser surgery.  J Refract Surg 2001;17 (5) S584- S587PubMedGoogle Scholar
10.
Gatinel  DMalet  JHoang-Xuan  TAzar  DT Analysis of customized corneal ablations: theoretical limitations of increasing negative asphericity.  Invest Ophthalmol Vis Sci 2002;43 (4) 941- 948PubMedGoogle Scholar
11.
Hersh  PSFry  KBlaker  W Spherical aberration after laser in situ keratomileusis and photorefractive keratectomy: clinical results and theoretical models of etiology.  J Cataract Refract Surg 2003;29 (11) 2096- 2104PubMedGoogle ScholarCrossref
12.
Roberts  C Biomechanical forces may also induce an oblate shape: the cornea is not a piece of plastic.  J Refract Surg 2000;16 (4) 407- 413PubMedGoogle Scholar
13.
Stiles  WS The directional sensitivity of the retina and the spectral sensitivities of the rods and cones.  Proc R Soc Lond B Biol Sci 1939;127 (846) 64- 105Google ScholarCrossref
14.
Crawford  BH The luminous efficiency of light entering the eye pupil at different points and its relation to brightness threshold measurements.  Proc R Soc Lond B Biol Sci 1937;124 (834) 81- 96Google ScholarCrossref
15.
Moon  PSpencer  DE On the Stiles-Crawford effect.  J Opt Soc Am 1944;34 (6) 319- 329Google ScholarCrossref
Clinical Sciences
August 11, 2008

A Comparison of CustomCornea Myopia Algorithms for Wavefront-Guided Laser In Situ Keratomileusis

Author Affiliations

Author Affiliations: Boxer Wachler Vision Institute, Beverly Hills, California.

Arch Ophthalmol. 2008;126(8):1067-1070. doi:10.1001/archopht.126.8.1067
Abstract

Objective  To determine whether the new CustomCornea M3 (myopic astigmatism) algorithm places additional pulses in the periphery to counter excimer laser peripheral ablation inefficiency.

Methods  Analysis of 25 myopic eyes of 18 patients that were within the approved indications for both the myopic astigmatism (M3) and myopic sphere (A7) algorithms. We analyzed ablation depth at peak and 4-mm, 6-mm, and 7-mm zone diameters for both the A7 and M3 algorithms.

Results  The M3 algorithm programmed for more of an ablation at the peak and peripheral zones than the A7 algorithm (P < .001). Even accounting for the additional peak ablation, there was significantly greater ablation in the periphery with the M3 algorithm compared with the A7 algorithm (P < .001). The mean (SD) manifest sphere was −4.09 (1.90) diopters (D), the mean (SD) manifest cylinder was −0.60 (0.52) D, and the mean (SD) manifest spherical equivalent was −4.39 (1.92) D. The manifest sphere ranged from −1.00 to −7.50 D, the manifest cylinder from 0.00 to −1.25 D, and the manifest spherical equivalent from −1.50 to −8.25 D.

Conclusion  The new CustomCornea M3 algorithm is programmed to perform additional ablation in the periphery to counter decreased pulse efficacy and the potential for induced spherical aberration.

Increased spherical aberration after myopic laser in situ keratomileusis (LASIK)1-4 can lead to symptoms of halos and glare as well as patient dissatisfaction despite excellent uncorrected visual acuity.5-7 The change in corneal shape from prolate to oblate is one of the main causes of spherical aberration. This may be due to the decreased effectiveness of laser pulses in the periphery as the angle of incidence of the laser to the cornea increases toward the periphery.2,8,9 It has also been found through mathematical modeling that a greater depth of central ablation than predicted by the Munnerlyn formula is required to maintain the prolate shape.2,10 The first-generation CustomCornea algorithm, known as myopic sphere or A7 (Alcon Laboratories, Inc, Fort Worth, Texas), was the initial software used for CustomCornea LASIK. The second-generation CustomCornea myopic astigmatism algorithm, or M3 (Alcon Laboratories, Inc), is meant to minimize induction of spherical aberration by placing greater pulses in the periphery of the ablation profile. The aim of our study was to evaluate the ablation profile of the M3 algorithm to determine whether additional pulses are placed in the periphery as compared with the A7 algorithm.

Methods

We conducted a technical analysis of 25 myopic consecutive eyes of 18 patients. Wavefront data were captured on the CustomCornea LADARWave (Alcon Laboratories, Inc). Inclusion criteria were no previous ocular surgery, phoropter spherical refraction of up to −7.90 diopters (D), and phoropter cylinder refraction of up to −1.49 D. We chose these criteria to remain within the allowable limits of both the A7 and M3 algorithms. This makes the data applicable to current US approved ranges for both algorithms. The M3 approved indications for use are phoropter refractions of up to −8.00 D sphere, −0.50 D to −4.00 D cylinder, and up to −8.00 D spherical equivalent. This is an expanded range compared with the original CustomCornea A7 algorithm (up to −7.00 D sphere and −0.50 D cylinder with an allowable difference of 1.00 D between the phoropter and wavefront refractions).

Using the LADARWave platform, we collected data on the proposed ablation profile for each eye using both the M3 and A7 algorithms. Measurements of the peak ablation along with the ablation depth at optical zones of 4-, 6-, and 7-mm diameters were taken. The LADARWave platform allows the user to move a cursor over various points of the proposed ablation profile to determine the ablation depth at that point. The position of the cursor is displayed as x and y Cartesian coordinates so that any spot of the ablation profile can be pinpointed (Figure 1).

Data points were taken in the horizontal (180°) and vertical (90°) meridians of all 4 quadrants (nasal, temporal, superior, and inferior) for each 4-, 6-, and 7-mm zone diameter. For each diameter group, the points were averaged to give a mean ablation depth at the 4-, 6-, and 7-mm zones. The peak ablation amounts and the ablation amounts at the 3 zones were compared across and within the 2 algorithms.

StatView analysis software (SAS Institute Inc, Cary, North Carolina) was used for data analysis. A paired t test was used for significance testing and a Spearman test was used for correlations. P < .05 was set for statistical significance. Institutional board review was not obtained as the study was retrospective and data collection and analysis did not affect clinical decisions. The primary surgeon (B.S.B.W.) followed the usual clinical practice in determining which algorithm was ultimately used for each eye. For α = .05 and a power of 0.80, a sample size of 18 was required to detect a mean difference of 5 μm.

Results

Wavefront and manifest refractions for the 25 eyes are outlined in Table 1. Table 2 shows that the mean (SD) peak ablation was 84.9 (33.5) μm in the M3 algorithm and 74.8 (29.9) μm in the A7 algorithm. This was a mean (SD) difference of 10.1 (3.7) μm, or 13.5% (P < .001).

The mean ablation at all of the outer zones was statistically greater in the M3 algorithm (P < .001). The mean (SD) ablation depths for the M3 algorithm were 69.7 (27.3) μm, 30.0 (10.8) μm, and 8.8 (4.9) μm at the 4-, 6-, and 7-mm zones, respectively. The mean (SD) ablation depths for the spherical algorithm were 58.9 (23.4) μm, 25.2 (9.5) μm, and 8.5 (4.8) μm at the 4-, 6-, and 7-mm zones, respectively (Table 2). At the 4-, 6-, and 7-mm zones, the mean additional amounts of ablation were therefore 10.8, 4.8, and 0.2 μm, respectively; in percentage terms, they were 18.8%, 20.3%, and 4.5%, respectively. The difference between the 2 algorithms at all of the zones and at the peak ablation correlated with the magnitude of the spherical and cylindrical corrections (P < .05) (Figure 2).

Table 2 also shows that additional peripheral ablation was performed by the M3 algorithm as compared with the A7 algorithm relative to the peak ablation (P < .001), ie, even accounting for the increased central ablation depth of the M3 algorithm, the peripheral zones had a greater percentage increase in ablation depth. The amount of additional ablation at the 4- and 6-mm zones with the M3 algorithm compared with its peak ablation correlated with attempted spherical equivalent correction (both P = .001). Even accounting for the increased peak ablation that occurs with the myopic astigmatism algorithm, significant additional ablation occurs in the peripheral zones. This is more evident at the 6- and 7-mm zones.

Comment

Induced spherical aberration after LASIK may occur owing to 3 primary interconnected factors: (1) pupil and optical zone mismatch; (2) lack of a transition zone or blend zone leading to an abrupt change in refractive power at the edge of the optical zone; and (3) change from a prolate corneal shape to an oblate shape. Peripheral rays undergo greater bending relative to the central rays in an oblate cornea. This results in symptoms including glare and halos, especially under scotopic conditions when the pupil dilates and encompasses these peripheral rays.

Most studies propose that the change in shape takes place because change in fluence or laser energy occurs when the laser beam strikes the cornea at an increasingly more oblique angle as it moves from the center to the periphery.2,8,9 This results in a decrease of the actual energy reaching the surface as well as increased reflectance of the incident beam. This has also been called the cosine effect because the perpendicular or effective component of the incident beam is given by cos2 φ.2,10 Other theories suggested to account for the corneal shape change after LASIK are either biological (wound healing or epithelial hyperplasia) or biomechanical (increased pull by peripheral corneal lamellae).8,10-12

It has been proposed that additional laser pulses in the periphery would help prevent spherical aberration. It is not known how much extra ablation should take place. Holladay et al2 have stated that the laser's effective energy decreases by 16% at 1 mm, 20% at 2 mm, and 28% at 3 mm. Increasing attempted correction would necessitate an increasing amount of ablation to maintain a prolate shape as well as a large effective optical zone.8,10,11 The Stiles-Crawford effect needs to be considered when formulating compensatory software. Photoreceptors are maximally stimulated by rays of light incident perpendicularly; as a result, peripheral rays do not contribute as much to the image perceived by the patient. This effect compensates for spherical aberration. A ray at the 6-mm margin has an effectivity of 35% to 40% of a central ray, and a ray at the 8-mm margin has a 20% effectivity.13-15 Gatinel et al10 have also shown through a mathematical model that the depth of the central ablation also must be greater to maintain a prolate shape. The more prolate the cornea is initially, the less ablation that is required.

The new CustomCornea M3 algorithm does appear to incorporate these factors when creating an ablation profile. Our results show that the new CustomCornea M3 software does result in more pulses centrally and in the midperiphery compared with the A7 algorithm. It also places additional pulses at the 4- and 6-mm zones relative to the central ablation. Even though the peak ablation depth is slightly increased, the peripheral ablation depth is significantly greater to counteract the decreased peripheral ablation efficiency that occurs with the excimer laser. The additional ablation performed by the M3 algorithm correlated very strongly with the degree of attempted correction. The degree of additional peripheral pulses also correlated with the degree of attempted correction. The increased central ablation is a limiting factor for greater corrections to avoid an increase in the risk of ectasia.

This study has limitations. We cannot comment on the spherical aberration outcomes as they were not clinically assessed. The purpose of this study was a technical assessment to verify how the manufacturer had incorporated spherical aberration compensation into the algorithm. This information is not released to the surgeon or technician and is proprietary. We also cannot comment on how or whether such software is incorporated into other laser platforms—although some manufacturers state that compensatory software is present. For example, the Wavelight platform (Wavelight Laser Technologie AG, Erlangen, Germany) has wavefront-optimized software that is meant to minimize induced spherical aberration based on higher-order aberration data derived from a normal population. The VISX laser software (VISX USA, Inc, Santa Clara, California) is also meant to have spherical aberration software compensation that uses K values to judge how to compensate for the cosine effect. We did not analyze our data to determine whether this is the case with the CustomCornea algorithms. However, owing to the proprietary nature of the software, no details on these other systems are available, nor are there any peer-reviewed articles to our knowledge that have analyzed these systems like we did in this study.

Our findings confirm that the new CustomCornea M3 algorithm is designed to add pulses in the periphery to counter UV reflectance and minimize induction of spherical aberration. This technique can be a benefit to all laser platforms by increasing ablation efficiency in the corneal periphery.

Correspondence: Brian S. Boxer Wachler, MD, Boxer Wachler Vision Institute, 465 N Roxbury Dr, Ste 902, Beverly Hills, CA 90210 (bbw@boxerwachler.com).

Submitted for Publication: May 26, 2006; final revision received November 16, 2006; accepted February 1, 2008.

Author Contributions: Dr Chan had full access to all of the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

Financial Disclosure: Dr Boxer Wachler is a consultant for Alcon Laboratories, Inc, and Addition Technology, Inc. Dr Chan is a consultant for VistaResearch Inc and Blackmores.

References
1.
Boxer Wachler  BSHuynh  VNEl-Shiaty  AFGoldberg  D Evaluation of corneal functional optical zone after laser in situ keratomileusis.  J Cataract Refract Surg 2002;28 (6) 948- 953PubMedGoogle ScholarCrossref
2.
Holladay  JTDudeja  DRChang  J Functional vision and corneal changes after laser in situ keratomileusis determined by contrast sensitivity, glare testing and corneal topography.  J Cataract Refract Surg 1999;25 (5) 663- 669PubMedGoogle ScholarCrossref
3.
Marcos  S Aberrations and visual performance following standard laser vision correction.  J Refract Surg 2001;17 (5) S596- S601PubMedGoogle Scholar
4.
Nepomuceno  RLBoxer Wachler  BSScruggs  R Functional optical zone after myopic LASIK as a function of ablation diameter.  J Cataract Refract Surg 2005;31 (2) 379- 384PubMedGoogle ScholarCrossref
5.
McGhee  CNCraig  JPSachdev  NWeed  KHBrown  AD Functional, psychological, and satisfaction outcomes of laser in situ keratomileusis for high myopia.  J Cataract Refract Surg 2000;26 (4) 497- 509PubMedGoogle ScholarCrossref
6.
Bailey  MDMitchell  GLDhaliwal  DKBoxer Wachler  BSZadnik  K Patient satisfaction and visual symptoms after laser in situ keratomileusis.  Ophthalmology 2003;110 (7) 1371- 1378PubMedGoogle ScholarCrossref
7.
Jabbur  NSSakatani  KO’Brien  TP Survey of complications and recommendations for management in dissatisfied patients seeking a consultation after refractive surgery.  J Cataract Refract Surg 2004;30 (9) 1867- 1874PubMedGoogle ScholarCrossref
8.
Holladay  JTJanes  JA Topographic changes in corneal asphericity and effective optical zone after laser in situ keratomileusis.  J Cataract Refract Surg 2002;28 (6) 942- 947PubMedGoogle ScholarCrossref
9.
Mrochen  MSeiler  T Influence of corneal curvature on calculation of ablation patterns used in photorefractive laser surgery.  J Refract Surg 2001;17 (5) S584- S587PubMedGoogle Scholar
10.
Gatinel  DMalet  JHoang-Xuan  TAzar  DT Analysis of customized corneal ablations: theoretical limitations of increasing negative asphericity.  Invest Ophthalmol Vis Sci 2002;43 (4) 941- 948PubMedGoogle Scholar
11.
Hersh  PSFry  KBlaker  W Spherical aberration after laser in situ keratomileusis and photorefractive keratectomy: clinical results and theoretical models of etiology.  J Cataract Refract Surg 2003;29 (11) 2096- 2104PubMedGoogle ScholarCrossref
12.
Roberts  C Biomechanical forces may also induce an oblate shape: the cornea is not a piece of plastic.  J Refract Surg 2000;16 (4) 407- 413PubMedGoogle Scholar
13.
Stiles  WS The directional sensitivity of the retina and the spectral sensitivities of the rods and cones.  Proc R Soc Lond B Biol Sci 1939;127 (846) 64- 105Google ScholarCrossref
14.
Crawford  BH The luminous efficiency of light entering the eye pupil at different points and its relation to brightness threshold measurements.  Proc R Soc Lond B Biol Sci 1937;124 (834) 81- 96Google ScholarCrossref
15.
Moon  PSpencer  DE On the Stiles-Crawford effect.  J Opt Soc Am 1944;34 (6) 319- 329Google ScholarCrossref
×