[Skip to Content]
Access to paid content on this site is currently suspended due to excessive activity being detected from your IP address 18.207.249.15. Please contact the publisher to request reinstatement.
[Skip to Content Landing]
Original Article
July 2002

Downsizing of Voice Prosthesis Diameter in Patients With Laryngectomy: An In Vitro Study

Author Affiliations

From the Department of Otolaryngology–Head and Neck Surgery, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands.

Arch Otolaryngol Head Neck Surg. 2002;128(7):838-841. doi:10.1001/archotol.128.7.838
Abstract

In patients with laryngectomy, voice prostheses inserted into a tracheoesophageal fistula (TEF) are widely used for vocal rehabilitation. Gradual dilation of the TEF may cause bothersome leakage around voice prostheses. Prosthesis-related weight and mechanical trauma possibly exacerbate TEF dilation. If prosthesis size were to be decreased, with a concomitant decrease in prosthesis weight and diameter, dilation of the TEF would probably lessen. We performed in vitro tests to study the effects on aerodynamic prosthesis function when the prosthesis size—in particular, the inner diameter—was decreased. The effects on airflow and pressure were specifically studied in the airflow range of patients with laryngectomy. A 1-mm decrease of the regular inner prosthesis diameter from 5 mm to 4 mm showed no significant aerodynamic consequences at the average laryngectomized airflow point. Also, such a 1-mm decrease in diameter involved a prosthesis weight reduction of 18%. In view of these findings, downsizing the standard prosthetic diameter should be considered in future voice prosthesis development.

×