April 1976

A Plasma Scalpel: Comparison of Tissue Damage and Wound Healing With Electrosurgical and Steel Scalpels

Author Affiliations

From the Department of Surgery (Drs Link and Glover), Wishard Memorial Hospital, Indiana University School of Medicine, Indianapolis, and the School of Mechanical Engineering (Dr Incropera), Purdue University, West Lafayette, Ind.

Arch Surg. 1976;111(4):392-397. doi:10.1001/archsurg.1976.01360220088014
Abstract

• We have been evaluating a plasma scalpel (the term has no relation to blood plasma) for four years that, with the generation of a small, hot gas jet (3,000 C), can cut tissue and simultaneously cauterize blood vessels of 3 mm in diameter.

Twenty-nine plasma scalpel hepatectomies in dogs and comparative skin wound healing (with steel and electrosurgical scalpels) in 90 mice showed that bleeding was reduced, and the thermal insult in liver tissue was limited to 2 mm from the incision; both liver and mouse skin incisions healed without complication. The completion of epithelization of mouse skin wounds occurred at 2 to 6, 6 to 14, and 6 to 18 days, and the average scar width was 0.8, 1.4, and 2.1 mm for the steel, electrosurgical, and plasma scalpels, respectively.

The plasma scalpel effectively cauterizes blood vessels as it cuts, leaving limited damaged tissue. We have recently begun human trials, and the device shows promise as a clinical tool.

(Arch Surg 111:392-397, 1976)

References
1.
Fidler JP, Law E, Rockwell RJ, et al:  Carbon dioxide laser excision of acute burns with immediate autografting .  J Surg Res 17:1-11, 1974.Crossref
2.
Stellar S, Levine N, Ger R, et al:  Laser excision of acute third-degree burns followed by immediate autograft replacement: An experimental study in the pig .  J Trauma 13:45-53, 1973.Crossref
3.
Rockwell RJ:  Current status of laser applications in medicine and biology .  CRC Crit Rev Bioeng 1:49-67, 1971.
4.
Roberts TS, Brayshaw FG: Experimental use of the plasma tissue-cutting device. Read before the 22nd Annual Conference on Engineering in Medicine and Biology, Chicago, 1969.
5.
Hishimoto K, Rockwell RJ, Drefer R, et al: Some technical problems in plasma scalpel hepatectomy. Read before the 22nd Annual Conference on Engineering in Medicine and Biology, Chicago, 1969.
6.
Henriques FC, Moritz AR:  Studies of thermal injury: I. The conduction of heat to and through skin and the temperature attained therein. A theoretical and experimental investigation .  Am J Pathol 23:531-549, 1947.
7.
Henriques FC:  Studies of thermal injury: V. The predictability and significance of thermally induced rate process leading to epidermal injury .  Arch Pathol 43:489-502, 1947.
8.
Link WJ, Incropera FP, Glover JL: The thermal response of tissue subjected to plasma scalpel heating. Read before the winter annual meeting of the American Society of Mechanical Engineers, Detroit, 1973.
9.
Henderson MR, Link WJ, Incropera FP, et al:  Gas transport resulting from plasma scalpel surgery .  Med Biol Eng 12:208-213, 1974.Crossref
10.
Link WJ, Zook EG, Glover JL:  Plasma scalpel excision of burns: An experimental study .  Plast Reconstr Surg 55:657-663, 1975.Crossref
×